The Department of Energy Office of Science seeks quotation for a high voltage power supply with proposals due October 2, 2026.

Place of performance is Upton, New York. The supply must deliver dual output voltages of 80 kV at 0.3 mA output current with 24 W power capacity, weigh approximately 35 kg, and meet strict performance specifications including 10 ppm RMS ripple, 50 ppm/hour stability, 50 ppm/°C temperature coefficient, and 10 ppm line and load regulation. The unit operates on 115 VAC 10%, single-phase, 50/60 Hz input power within a temperature range of 0 to +45°C and 20–80% relative humidity (non-condensing). The power supply must incorporate overvoltage, overcurrent, short-circuit, and arc protection; remote voltage control via 0–10 VDC; and remote digital control and monitoring via USB, LAN, RS-232C, RS-485, or GPIB.

Suppliers must comply with ISO 9001 quality standards or maintain a documented quality program subject to Brookhaven National Laboratory assessment. Key requirements include change management with prior written buyer approval, M&TE calibration traceable to NIST standards, counterfeit parts prevention, ESD protection for susceptible items, NRTL certification for electrical and fire protection equipment, OSHA compliance for specialized equipment, and chemical hazard compliance per OSHA's Global Harmonized System. Age/shelf life control shall maintain three-quarters shelf life remaining at receipt. Suppliers must implement immediate notification of product recalls to ProductRecallAlert@BNL.gov and retain records for a minimum of three years.

Quotations must include pricing, country of origin, lead time, UEI number, and factory testing data with required factory test documentation. The NAICS code is 335999, All Other Miscellaneous Electrical Equipment and Component Manufacturing; PSC code is 6120, Transformers: Distribution and Power Station.

Notice text

Please see attached for specifications.

When you submit your quote please make sure the following are on the quote:

Pricing

Country of Origin

Lead Time

UEI number

Files

File Type Posted
QA-101 482087.docx DOCX document
Specs for WS482087.xlsx XLSX spreadsheet

Similar open opportunities

Opportunity Agency Set-aside Responses due
LT-SPM Upton, New York 11973, United States · NAICS 334516 Office of Science No set-aside
Spectrum Analyzer, No Equivalents Or Substitutions New York 11973, United States · NAICS 334515 Office of Science No set-aside
Power supply NAICS 335931 Naval Supply Systems Command No set-aside
Power supply NAICS 335312 Land and Maritime No set-aside
Power supply NAICS 335999 Land and Maritime No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity