About this opportunity

The Department of Veterans Affairs Strategic Acquisition Center is conducting market research through this Sources Sought Notice to identify qualified vendors capable of supplying Pachymeters: Corneal Ultrasound equipment (DGH Pachmate 2 brand name or equal) on an agency-wide basis for delivery to VA medical centers throughout the United States. The procurement seeks handheld pachymeters meeting specific technical specifications, including a measurement range of 300-999 microns, accuracy of ±5 microns, battery operation, minimum 9 readings per test capability, and weight under 3.65 ounces. Vendors must be registered in SAM with appropriate NAICS code 339112 certification, and service-disabled veteran-owned small businesses must be verified in the SBA database. While this is a planning document and does not constitute a formal solicitation for bids or proposals, the VA intends to award a single Requirements Contract with Firm-Fixed Price orders to a qualified source. Responses to this Sources Sought Notice are due by December 30, 2025, and must include detailed company information, technical literature, pricing, service plans, and potential subcontracting opportunities.

This procurement carries no set-aside designation and represents an open competition opportunity. The anticipated contract will consist of one 12-month base period followed by four optional 12-month extension periods, with estimated annual quantity requirements ranging from 15 to 40 units for both the Pachmate 2 device and its detachable probe. The contract structure utilizes Firm-Fixed Price orders under FAR 16.503, with contract continuation dependent upon successful performance during each preceding year and demonstrated continued need by the government. The size standard for this NAICS code 339112 procurement is 1,000 employees, establishing the small business threshold for potential offerors. Specific award values are not disclosed in this pre-solicitation phase, though the government's decision to proceed will be based on identified qualified sources and their capability to meet the Veterans Health Administration's equipment requirements on a national scale.

Notice text

2 versions

Update #2 · Latest ·

PLEASE SEE THE ATTACHED DOCUMENT FOR ADDITONAL DETAILS REGARDING QUANTITY REQUIREMENTS.

Sources Sought Notice Sources Sought Notice Page 1 of 4 1.0 DESCRIPTION THIS IS A SOURCES SOUGHT NOTICE ONLY. This is not a solicitation for bids, proposals, proposal abstracts, or quotations. The Department of Veterans Affairs (VA) Strategic Acquisition Center (SAC) is issuing this Sources Sought Notice in accordance with the FAR 10.002(b)(2) as a means of conducting market research to identify qualified sources with the capability and the resources to meet the Veterans Health Administration s (VHA) Non-Expendable Equipment National Program requirement for Pachymeters: Corneal Ultrasound. The SAC in conjunction with VHA is seeking a qualified source to supply brand name or equal DGH Pachmate 2 on an agency-wide basis. VA intends to award a single Requirements Contract with Firm-Fixed Price (FFP) orders IAW FAR 16.503, Requirements Contract to a supplier of this equipment. Vendors will be required to deliver Pachymeters: Corneal Ultrasound to VA medical centers and facilities throughout the United States. The anticipated period of performance is for one 12-month base period with four 12-month option periods from the date of award. However, the Government's decision as to whether to continue with the contract, upon annual review, will be based upon successful performance during each previous year and continued need. The associated North American Industrial Classification System (NAICS) code for this procurement is 339112 Surgical and Medical Instrument Manufacturing and the associated size standard is 1,000 employees. The FAR Provision found at 52.204-7(b)(1) requires that All interested Offerors must be registered in SAM prior to submitting an offer or quotation. You may access the SAM website at SAM.gov | Home. All offerors must be certified with the appropriate NAICS code on the SAM website. All SDVOSB/VOSB offerors must be verified in the SBA database at: https://veterans.certify.sba.gov/#search at the time of submission. Submissions for this sources sought shall be submitted by email no later than December 30, 2025, 11:59am Eastern to all the following: Cyiamaudia.Jones@va.gov Kimberly.LeMieux@va.gov This notice is for planning purposes only and does not constitute a commitment on the part of the Government to award a contract, nor does the Government intend to pay for any information submitted as a result of this notice. This Sources Sought Notice is part of a continuous process for obtaining the latest information on the commercial status of the industry with respect to their current and near-term abilities. The information provided herein is subject to change and in no way binds the Government to solicit for, or award, a competitive contract. It is emphasized that this is a notice solely for planning and informational purposes. Respondents are advised that the Government is under no obligation to acknowledge receipt of the information received or provide feedback to respondents with respect to any information submitted. The Government may or may not issue a solicitation as a result of this notice. There is no solicitation available at this time. 2.0 SCOPE OF WORK Pachymeters are ophthalmic measuring instruments designed to determine the thickness of the cornea. The instruments may consist of contact devices using ultrasonic technology that generate pulses and measure the time differences of the echoes. Pachymeters are used mainly to detect glaucoma or for planning refractive surgery. The following Contract Line Items are being considered: Contract Line Items Manufacture Part Number Description 0001 DGH DGH 55B Pachmate 2 Handheld Pachymeter 0002 DGH DET 2006 Detachable Probe for Pachmate 2 The following Salient Characteristics (SCs) are being considered: SC # SALIENT CHARACTERISTICS METHOD OF EVALUATION SC 1 Must be handheld for easy portability Literature Review SC 2 Data can be reviewed on the instrument s LCD screen Literature Review SC 3 Must be battery operated or rechargeable Literature Review SC4 Must have a measurement range of at least 300-999 microns Literature Review SC 5 Accuracy must be at least +/- 5 microns Literature Review SC 6 Must have resolution of at least 1 micron Literature Review SC 7 Must take a minimum of 9 readings per test Literature Review SC 8 Must be able to clean/disinfect the probe/tip without the use of probe/tip covers Literature Review SC 9 Must weight less than 3.65 oz Literature Review SC 10 Must have audible indication when measurement acquired Literature Review SC 11 Must have at least one year warranty Literature Review SC 12 Any accessories must function with CLIN 0001 For each product (brand name and part number), the response must include descriptive literature demonstrating the product meets or exceeds the salient characteristics (SC) specified above. The Government's intent is to include all ancillary items to ensure the equipment can function as designed by the OEM and as clinically required. Vendors are encouraged to provide any product solution or configuration so long as they meet the salient characteristics. Additionally, responses can include any additional product that may be beneficial. These items must be clearly identified by brand name and part number. Responses to this Sources Sought Notice shall include the following: Full name and address of company. SAM Unique Entity Identifier (UEI). Business Size. Manufacturer or Distributor: If distributor, provide full name, business size and address of manufacturer. If distributor, provide evidence that your firm is an authorized distributor for the manufacturer of the products proposed. Country of Origin designation for all products proposed. Ability to provide uninterrupted supply of products on a national scale. Technical Literature that clearly shows product(s) meets the identified salient characteristics and page numbers where each salient characteristic is met. Does vendor offer an annual service plan, Yes or No? If Yes: What does the plan cover? Does plan cover software updates only? Does the plan cover parts repair or replacement only? Does the plan cover both software and parts? Vendor is requested to provide documentation that shows information on their service plan. Is the service provided by the OEM or a subcontractor? If distributor, provide full name, business size and address of manufacturer. Any additional product solution or configuration that would be beneficial to the clinical functionality of the product line identified. Item pricing information of products, inclusive of flat rate shipping. Include any quantity discounts, noting at what level of quantity ordered should the Government expect tiered price reductions, if any. NOTE: In order for the Contracting Officer to determine whether a set-aside can be pursued, it is essential to provide pricing information. Without this information, the Contracting Officer will be unable to assess whether there is a reasonable expectation of receiving fair and reasonable pricing from socio-economic concerns. Does your firm have any subcontracting opportunities for this equipment? Identify any federal contracts vehicles held by your firm that provides this equipment. Are there any further standards or criteria VA should consider in a procurement action for this equipment? Are the Government s specifications unduly restrictive? Any additional information the Government should know prior to issuing a solicitation. Although not required, vendors responding to this Sources Sought may also submit a Capabilities Statement.

Update #1 ·

Sources Sought Notice

Sources Sought Notice

Page 1 of 4

1.0 DESCRIPTION

THIS IS A SOURCES SOUGHT NOTICE ONLY. This is not a solicitation for bids, proposals, proposal abstracts, or quotations. The Department of Veterans Affairs (VA) Strategic Acquisition Center (SAC) is issuing this Sources Sought Notice in accordance with the FAR 10.002(b)(2) as a means of conducting market research to identify qualified sources with the capability and the resources to meet the Veterans Health Administration s (VHA) Non-Expendable Equipment National Program requirement for Pachymeters: Corneal Ultrasound.

The SAC in conjunction with VHA is seeking a qualified source to supply brand name or equal DGH Pachmate 2 on an agency-wide basis. VA intends to award a single Requirements Contract with Firm-Fixed Price (FFP) orders IAW FAR 16.503, Requirements Contract to a supplier of this equipment. Vendors will be required to deliver Pachymeters: Corneal Ultrasound to VA medical centers and facilities throughout the United States.

The anticipated period of performance is for one 12-month base period with four 12-month option periods from the date of award. However, the Government's decision as to whether to continue with the contract, upon annual review, will be based upon successful performance during each previous year and continued need.

The associated North American Industrial Classification System (NAICS) code for this procurement is 339112 Surgical and Medical Instrument Manufacturing and the associated size standard is 1,000 employees.

The FAR Provision found at 52.204-7(b)(1) requires that All interested Offerors must be registered in SAM prior to submitting an offer or quotation. You may access the SAM website at SAM.gov | Home. All offerors must be certified with the appropriate NAICS code on the SAM website. All SDVOSB/VOSB offerors must be verified in the SBA database at: https://veterans.certify.sba.gov/#search at the time of submission.

Submissions for this sources sought shall be submitted by email no later than December 30, 2025, 11:59am Eastern to all the following:

Cyiamaudia.Jones@va.gov

Kimberly.LeMieux@va.gov

This notice is for planning purposes only and does not constitute a commitment on the part of the Government to award a contract, nor does the Government intend to pay for any information submitted as a result of this notice. This Sources Sought Notice is part of a continuous process for obtaining the latest information on the commercial status of the industry with respect to their current and near-term abilities. The information provided herein is subject to change and in no way binds the Government to solicit for, or award, a competitive contract. It is emphasized that this is a notice solely for planning and informational purposes. Respondents are advised that the Government is under no obligation to acknowledge receipt of the information received or provide feedback to respondents with respect to any information submitted. The Government may or may not issue a solicitation as a result of this notice. There is no solicitation available at this time.

2.0 SCOPE OF WORK

Pachymeters are ophthalmic measuring instruments designed to determine the thickness of the cornea. The instruments may consist of contact devices using ultrasonic technology that generate pulses and measure the time differences of the echoes. Pachymeters are used mainly to detect glaucoma or for planning refractive surgery.

The following Contract Line Items are being considered:

Contract Line Items

Manufacture

Part Number

Description

0001

DGH

DGH 55B

Pachmate 2 Handheld Pachymeter

0002

DGH

DET 2006

Detachable Probe for Pachmate 2

The following Salient Characteristics (SCs) are being considered:

SC #

SALIENT CHARACTERISTICS

METHOD OF EVALUATION

SC 1

Must be handheld for easy portability

Literature Review

SC 2

Data can be reviewed on the instrument s LCD screen

Literature Review

SC 3

Must be battery operated or rechargeable

Literature Review

SC4

Must have a measurement range of at least 300-999 microns

Literature Review

SC 5

Accuracy must be at least +/- 5 microns

Literature Review

SC 6

Must have resolution of at least 1 micron

Literature Review

SC 7

Must take a minimum of 9 readings per test

Literature Review

SC 8

Must be able to clean/disinfect the probe/tip without the use of probe/tip covers

Literature Review

SC 9

Must weight less than 3.65 oz

Literature Review

SC 10

Must have audible indication when measurement acquired

Literature Review

SC 11

Must have at least one year warranty

Literature Review

SC 12

Any accessories must function with CLIN 0001

For each product (brand name and part number), the response must include descriptive literature demonstrating the product meets or exceeds the salient characteristics (SC) specified above.

The Government's intent is to include all ancillary items to ensure the equipment can function as designed by the OEM and as clinically required. Vendors are encouraged to provide any product solution or configuration so long as they meet the salient characteristics. Additionally, responses can include any additional product that may be beneficial. These items must be clearly identified by brand name and part number.

Responses to this Sources Sought Notice shall include the following:

Full name and address of company.

SAM Unique Entity Identifier (UEI).

Business Size.

Manufacturer or Distributor:

If distributor, provide full name, business size and address of manufacturer.

If distributor, provide evidence that your firm is an authorized distributor for the manufacturer of the products proposed.

Country of Origin designation for all products proposed.

Ability to provide uninterrupted supply of products on a national scale.

Technical Literature that clearly shows product(s) meets the identified salient characteristics and page numbers where each salient characteristic is met.

Does vendor offer an annual service plan, Yes or No?

If Yes:

What does the plan cover?

Does plan cover software updates only?

Does the plan cover parts repair or replacement only?

Does the plan cover both software and parts?

Vendor is requested to provide documentation that shows information on their service plan.

Is the service provided by the OEM or a subcontractor?

If distributor, provide full name, business size and address of manufacturer.

Any additional product solution or configuration that would be beneficial to the clinical functionality of the product line identified.

Item pricing information of products, inclusive of flat rate shipping. Include any quantity discounts, noting at what level of quantity ordered should the Government expect tiered price reductions, if any.

NOTE: In order for the Contracting Officer to determine whether a set-aside can be pursued, it is essential to provide pricing information. Without this information, the Contracting Officer will be unable to assess whether there is a reasonable expectation of receiving fair and reasonable pricing from socio-economic concerns.

Does your firm have any subcontracting opportunities for this equipment?

Identify any federal contracts vehicles held by your firm that provides this equipment.

Are there any further standards or criteria VA should consider in a procurement action for this equipment?

Are the Government s specifications unduly restrictive?

Any additional information the Government should know prior to issuing a solicitation.

Although not required, vendors responding to this Sources Sought may also submit a Capabilities Statement.

Files

File Type Posted
36C10G26Q0022 Attachment 1.docx DOCX document
36C10G26Q0022.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
6515--BCMA Barcode Scanners Department of Veterans Affairs VA Phoenix Healthcare System Phoenix, Arizona 85012, United States · NAICS 334118 Veterans Integrated Service Network 22 No set-aside
6515--Wechsler Memory Scale Department of Veterans Affairs Network Contracting Office 8 (NCO 8) Orlando FL 32827, United States · NAICS 334515 Veterans Integrated Service Network 8 No set-aside
METRx Instruments OR EQUAL San Francisco, California 94121, United States · NAICS 339112 Veterans Integrated Service Network 21 Total Small Business
NX EQ Microscope Spine United States · NAICS 339113 VA Strategic Acquisition Center Frederick No set-aside
Neuro Micro Instruments Houston, Texas 77030, United States · NAICS 339112 Veterans Integrated Service Network 16 No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors