Solicitation Amendment W50S7326QA0020003 SF 30.pdf
PDF 4 MB Posted
- Attached to
- 124th Liquid Nitrogen Generator
- Solicitation number
- W50S7326QA002
- Issued by
-
Department of the Army National Guard
About this file
This is a Standard Form 30 Amendment of Solicitation (Amendment 0003) issued by the U.S. Air Force Space Operations Command (W7MS USPFO Activity IDANG 124) on August 24, 2026, for solicitation W50S7326QA002, originally dated July 10, 2026. The amendment revises technical salient characteristics for CLIN 0001 and does not extend the offer receipt date.
The primary change modifies mandatory salient characteristics for a liquid nitrogen generation system. The nitrogen purity requirement was reduced from greater than 99% to greater than 98%. All other specifications remain unchanged, including: production capacity not exceeding 10 liters per day; integrated storage capacity not exceeding 55 liters; physical dimensions limited to 63" H x 25" D x 36" W; system weight not exceeding 450 lbs; air-cooled operation requiring no external water infrastructure; standard 120V/60Hz electrical operation at maximum 20 Amps; power consumption capped at 2.2 kW/H; acoustic signature not exceeding 60-65 dBA at 2 feet; environmental tolerance to +125 degrees Fahrenheit; mobility via integrated heavy-duty casters; and key functional features including fully automated plug-and-play operation, intelligent production control (FlowSync Technology equivalent), remote monitoring and diagnostics (SmartConnect Monitoring equivalent), and military security standards compliance allowing wireless/wired connectivity to be disabled or air-gapped without degrading core automated production functions. Offerors must acknowledge receipt of this amendment prior to the solicitation deadline or risk offer rejection. The Contracting Officer signature is dated August 25, 2026.
View the file
Other files
Show all 13
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers is extended. is not extended.
Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:
(a) By completing items 8 and 15, and returning copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted;
or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY RESULT IN REJECTION OF YOUR OFFER.
If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
12. ACCOUNTING AND APPROPRIATION DATA (If required)
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS.
IT MODIFIES THE CONTRACT/ORDER NUMBER AS DESCRIBED IN ITEM 14.
CHECK ONE A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER
NUMBER IN ITEM 10A.
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES (such as changes in paying office, appropriation data, etc.) SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER (Specify type of modification and authority)
E. IMPORTANT: Contractor is not is required to sign this document and return copies to the issuing office.
14. DESCRIPTION OF AMENDMENT/MODIFICATION (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER (Type or print) 16A. NAME AND TITLE OF CONTRACTING OFFICER (Type or print)
15B. CONTRACTOR/OFFEROR
(Signature of person authorized to sign)
15C. DATE SIGNED 16B. UNITED STATES OF AMERICA
(Signature of Contracting Officer)
16C. DATE SIGNED
Previous edition unusable STANDARD FORM 30 (REV. 11/2016) Prescribed by GSA FAR (48 CFR) 53.243
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT 1. CONTRACT ID CODE PAGE OF PAGES
2. AMENDMENT/MODIFICATION NUMBER 3. EFFECTIVE DATE 4. REQUISITION/PURCHASE REQUISITION NUMBER 5. PROJECT NUMBER (If applicable)
6. ISSUED BY CODE 7. ADMINISTERED BY (If other than Item 6) CODE
8. NAME AND ADDRESS OF CONTRACTOR (Number, street, county, State and ZIP Code) (X) 9A. AMENDMENT OF SOLICITATION NUMBER
9B. DATED (SEE ITEM 11)
10A. MODIFICATION OF CONTRACT/ORDER NUMBER
10B. DATED (SEE ITEM 13)
CODE FACILITY CODE
SF 30 BLOCK 14 CONTINUATION PAGE
SUMMARY OF CHANGES
Continuation of Supplies or Services and Prices/Costs
The following CLIN(s) / SLIN(s) / ELIN(s) were modified:
INFORMATION FROM TO
Description MANDATORY SALIENT
CHARACTERISTICS (BRAND
NAME OR EQUAL)
To be deemed technically acceptable, any proposed "or equal" system must meet or exceed the following objective performance and physical parameters:
1. PRODUCTION CAPACITY: Not to exceed 10 Liters per Day (L/Day).
2. INTEGRATED STORAGE
CAPACITY: Not to exceed 55 Liters (55L) internal storage tank.
3. NITROGEN PURITY: The system must consistently produce liquid nitrogen with a purity greater than 99%
4. PHYSICAL DIMENSIONS: The physical footprint must not exceed 63 inches Height by 25 inches Depth by 36 inches Width (63" H x 25" D x 36" W) to comply with rigid, existing spatial limitations.
5. SYSTEM WEIGHT: The overall system weight must not exceed 450
MANDATORY SALIENT
CHARACTERISTICS (BRAND
NAME OR EQUAL)
To be deemed technically acceptable, any proposed "or equal" system must meet or exceed the following objective performance and physical parameters:
1. PRODUCTION CAPACITY: Not to exceed 10 Liters per Day (L/Day).
2. INTEGRATED STORAGE
CAPACITY: Not to exceed 55 Liters (55L) internal storage tank.
3. NITROGEN PURITY: The system must consistently produce liquid nitrogen with a purity greater than 98%
4. PHYSICAL DIMENSIONS: The physical footprint must not exceed 63 inches Height by 25 inches Depth by 36 inches Width (63" H x 25" D x 36" W) to comply with rigid, existing spatial limitations.
5. SYSTEM WEIGHT: The overall system weight must not exceed 450
W50S7326QA0020003
lbs.
6. MOBILITY: The system must feature integrated, heavy-duty rollers /casters to enable rapid and safe on-site transport and repositioning.
7. COOLING METHOD: The system must be entirely air-cooled, requiring no external water-cooling infrastructure or external chiller facilities
8. POWER REQUIREMENTS: The system must operate on standard electrical configurations of 120V, 60Hz, with a maximum current draw of 20 Amps.
9. POWER CONSUMPTION: The system must be energy efficient, consuming a maximum of 2.2 kW/H during active production.
10. ACOUSTIC SIGNATURE: The operating noise level must not exceed 60 to 65 dBA when measured at a distance of 2 feet.
11. ENVIRONMENTAL
TOLERANCE: The system must be engineered for continuous, automated operations in extreme, high-ambient temperatures reaching up to +125 degrees Fahrenheit (+51.
6 degrees Celsius).
KEY FUNCTIONAL & SMART
FEATURES
lbs.
6. MOBILITY: The system must feature integrated, heavy-duty rollers /casters to enable rapid and safe on-site transport and repositioning.
7. COOLING METHOD: The system must be entirely air-cooled, requiring no external water-cooling infrastructure or external chiller facilities
8. POWER REQUIREMENTS: The system must operate on standard electrical configurations of 120V, 60Hz, with a maximum current draw of 20 Amps.
9. POWER CONSUMPTION: The system must be energy efficient, consuming a maximum of 2.2 kW/H during active production.
10. ACOUSTIC SIGNATURE: The operating noise level must not exceed 60 to 65 dBA when measured at a distance of 2 feet.
KEY FUNCTIONAL & SMART
FEATURES
Any proposed "or equal" system must possess equivalent automated and integrated technology features as described below:
1. FULLY AUTOMATED
OPERATION: The system must
Any proposed "or equal" system must possess equivalent automated and integrated technology features as described below:
1. FULLY AUTOMATED
OPERATION: The system must support a "plug-and-play" installation capability. It must utilize fully automated startup and shutdown cycles that require only an active power connection and a single manual or digital "start" input to operate.
2. INTELLIGENT PRODUCTION
CONTROL: The system must feature an integrated, intelligent control system (equivalent to FlowSync Technology). This system must continuously monitor internal LN2 levels to automatically adjust, start, and stop production cycles based on real-time usage and capacity.
3. REMOTE MONITORING &
DIAGNOSTICS: The system must include built-in remote access capability (via Wi-Fi or Ethernet) to support real-time system monitoring, troubleshooting, and diagnostics (equivalent to proprietary SmartConnect Monitoring).
4. MILITARY SECURITY
STANDARDS: To comply with local military security protocols, any wireless or wired connectivity support a "plug-and-play" installation capability. It must utilize fully automated startup and shutdown cycles that require only an active power connection and a single manual or digital "start" input to operate.
2. INTELLIGENT PRODUCTION
CONTROL: The system must feature an integrated, intelligent control system (equivalent to FlowSync Technology). This system must continuously monitor internal LN2 levels to automatically adjust, start, and stop production cycles based on real-time usage and capacity.
3. REMOTE MONITORING &
DIAGNOSTICS: The system must include built-in remote access capability (via Wi-Fi or Ethernet) to support real-time system monitoring, troubleshooting, and diagnostics (equivalent to proprietary SmartConnect Monitoring).
4. MILITARY SECURITY
STANDARDS: To comply with local military security protocols, any wireless or wired connectivity components must be fully compliant with standard federal networks, OR possess the hardware/software capability to be completely disabled, isolated, or physically air-gapped components must be fully compliant with standard federal networks, OR possess the hardware/software capability to be completely disabled, isolated, or physically air-gapped without degrading the core automated production functions of the generator.
without degrading the core automated production functions of the generator.
| IDCode: J |
| Page: 1 |
| Pages: 5 |
| AmendNo: 0003 |
| EffDate: 24 AUG 2026 |
| ReqNo: |
| ProjNo: |
| IssCode: W50S73 |
| AdmCode: |
| IssuedBy: W7M5 USPFO ACTIVITY IDANG 124 |
KO FOR IDANG DO NOT DELETE, 4460 S BYRD ST BOISE AIR TERMINAL
BOISE, ID 83705-8048
UNITED STATES
TREYVLIN JONES, EMAIL: TREYVLIN.JONES@US.AF.MIL TELEPHONE: 208-422-5558
AdminBy:
SCD: PAS:
| ContractorNameAdd: |
| Code: |
| FacCode: |
| SolChg: 1 |
| SolNo: W50S7326QA002 |
| SolDate: 10 JUL 2026 |
| AwardChg: Off |
| AwardNo: |
| AwardDate: |
| Amended: 1 |
| OffrExt: Off |
| OffrNoEx: 1 |
| NoCopies: 0 |
| ApprData: SEE CONTINUATION |
| ChgeOrd: Off |
| ChgeOrder: |
| Modify: Off |
| SuppAgre: Off |
| SuppAuth: |
| ModOthr: Off |
| OthrSpec: |
| NoReq: Off |
| Require: Off |
| Copies: |
| Descript: The purpose of this amendment is to revise technical salient characteristics. |