This is a Statement of Objectives (SOO) for the Department of War's requirement for an on-site, fully automated Liquid Nitrogen (LN2) Generator system. The solicitation references the Rutherford & Titan - Titan 10 L/Day Liquid Nitrogen Generator as the brand name specification, with provisions for "or equal" alternatives that must meet or exceed all stated requirements. The system must produce up to 10 liters per day with integrated storage capacity between 10 and 55 liters, and deliver nitrogen purity greater than 99%. Physical constraints limit the maximum footprint to 67.3" H x 35" D x 26.7" W and system weight to 595 lbs, with integrated heavy-duty rollers/casters for on-site mobility.
The generator must operate as an air-cooled system requiring 230VAC 50/60Hz power with maximum consumption of 2.2 kW/H during active production, and function continuously in ambient temperatures from +77°F to +82.4°F with acoustic output not exceeding 60–65 dBA at 2 feet. Key functional requirements include plug-and-play automated operation with fully automated startup/shutdown cycles, an integrated intelligent control system equivalent to FlowSync™ Technology that monitors and automatically adjusts production based on internal LN2 levels and usage, and built-in remote monitoring and diagnostics capability via Wi-Fi or Ethernet equivalent to SmartConnect™ Monitoring. All wireless/wired connectivity must comply with federal network security standards or be capable of complete disabling based on local security protocols.
View the file
Other files
Show all 13
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SECTION 1: Scope of Requirement The Department of War has a requirement for an on-site, fully automated Liquid Nitrogen (LN2) Generator. The system must meet or exceed the technical and functional capabilities of the Rutherford & Titan - Titan 10 L/Day Liquid Nitrogen Generator (Brand Name or Equal).
SECTION 2: Salient Characteristics (Brand Name or Equal) To be considered technically acceptable, any proposed "or equal" system must meet or exceed the following objective performance and physical parameters:
Characteristic Requirement / Minimum Specification
Production Capacity Not to exceed 10 Liters per Day (L/Day).
Integrated Storage Capacity
Must be a minimum of 10 Liters and must not exceed 55 Liters.
Nitrogen Purity Greater than 99% pure liquid nitrogen.
Physical Dimensions Maximum footprint of 67.3" H x 35" D x 26.7" W to fit existing spatial constraints.
System Weight Maximum of 595 lbs.
Mobility Must be integrated with heavy-duty rollers/casters for on-site mobility.
Cooling Method Air-cooled system (no external water-cooling infrastructure required).
Power Requirements Must operate standard electrical configurations: Requires (230VAC 50/60Hz) 220V+.
Power Consumption Energy efficient; maximum of 2.2 kW/H during active production.
Acoustic Signature Maximum noise level of 60–65 dBA at a distance of 2 feet.
Environmental Tolerance
Must be capable of continuous, automated operation in high ambient temperatures +77°F (+25°C) to Max +82.4°F (+28°C)
SECTION 3: Key Functional & Smart Features Any proposed "or equal" system must possess equivalent automated capabilities as described below:
Feature Performance and Functional Standards
Automated Operation
"Plug-and-play" capability; fully automated startup/shutdown cycles requiring only power connection and a single manual/digital "start" input.
Intelligent Production Control
Must feature an integrated, intelligent control system (equivalent to FlowSync™ Technology) that continuously monitors internal LN₂ levels to automatically adjust, start, and stop production based on usage and capacity.
Remote Monitoring & Diagnostics
Must include built-in remote access capability (via Wi-Fi or Ethernet) to allow real-time system monitoring, troubleshooting, and diagnostics (equivalent to SmartConnect™ Monitoring).
Security Standards
Any wireless/wired connectivity must be compliant with standard federal networks or capable of being completely disabled/isolated based on local security protocols.