Salient Characteristics LN2 Generator v6.pdf

PDF 70 KB Posted

Attached to
124th Liquid Nitrogen Generator
Solicitation number
W50S7326QA002
Issued by

About this file

This is a Statement of Objectives (SOO) for an on-site, fully automated Liquid Nitrogen (LN2) Generator procurement by the Department of War.

The requirement specifies a system equivalent to or exceeding the Rutherford & Titan - Titan 10 L/Day Liquid Nitrogen Generator. Technical specifications include a production capacity of 10 liters per day, integrated storage capacity not exceeding 55 liters, nitrogen purity greater than 98%, physical dimensions of maximum 63" H × 25" D × 36" W, system weight of 450 lbs maximum, and air-cooled operation requiring no external water infrastructure. The system must operate on standard 120V, 60Hz electrical service with maximum 20-amp draw and energy consumption not exceeding 2.2 kW/H during active production. The equipment must include heavy-duty rollers/casters for mobility and maintain acoustic signature of 60-65 dB(A) at two feet distance. Functional requirements include "plug-and-play" automated operation with single start input, an intelligent production control system (equivalent to FlowSync™ Technology) that automatically monitors and adjusts nitrogen production based on internal levels and usage, and remote monitoring and diagnostics capability via Wi-Fi or Ethernet (equivalent to SmartConnect™ Monitoring). Wireless or wired connectivity must comply with federal network security standards or be capable of complete disablement based on local security protocols.

View the file

Other files

Show all 13

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SECTION 1: Scope of Requirement The Department of War has a requirement for an on-site, fully automated Liquid Nitrogen (LN2) Generator. The system must meet or exceed the technical and functional capabilities of the Rutherford & Titan - Titan 10 L/Day Liquid Nitrogen Generator (Brand Name or Equal).

SECTION 2: Salient Characteristics (Brand Name or Equal) To be considered technically acceptable, any proposed "or equal" system must meet or exceed the following objective performance and physical parameters:

Characteristic Requirement / Minimum Specification

Production Capacity Not to exceed 10 Liters per Day (L/Day).

Integrated Storage Capacity

Not to exceed 55 Liters (55L) internal storage tank.

Nitrogen Purity Greater than 98% pure liquid nitrogen.

Physical Dimensions Maximum footprint of 63″ H × 25″ D × 36″ W to fit existing spatial constraints.

System Weight Maximum of 450 lbs.

Mobility Must be integrated with heavy-duty rollers/casters for on-site mobility.

Cooling Method Air-cooled system (no external water-cooling infrastructure required).

Power Requirements Must operate standard electrical configurations: 120V, 60Hz, maximum draw of 20 Amps.

Power Consumption Energy efficient; maximum of 2.2 kW/H during active production.

Acoustic Signature Maximum noise level of 60–65 dBA at a distance of 2 feet.

SECTION 3: Key Functional & Smart Features Any proposed "or equal" system must possess equivalent automated capabilities as described below:

Feature Performance and Functional Standards

Automated Operation

"Plug-and-play" capability; fully automated startup/shutdown cycles requiring only power connection and a single manual/digital "start" input.

Intelligent Production Control

Must feature an integrated, intelligent control system (equivalent to FlowSync™ Technology) that continuously monitors internal LN₂ levels to automatically adjust, start, and stop production based on usage and capacity.

Remote Monitoring & Diagnostics

Must include built-in remote access capability (via Wi-Fi or Ethernet) to allow real-time system monitoring, troubleshooting, and diagnostics (equivalent to SmartConnect™ Monitoring).

Security Standards

Any wireless/wired connectivity must be compliant with standard federal networks or capable of being completely disabled/isolated based on local security protocols.