SPECIFICATIONS - VEA CT System - Dated 5-30-2025.pdf
PDF 361 KB Posted
- Attached to
- DLA Aviation Computed Tomography System
- Solicitation number
- SP4703-25-R-0007
- Issued by
-
Defense Logistics Agency
About this file
These specifications detail a federal contract opportunity for the Defense Logistics Agency (DLA) Aviation to purchase a turn-key, state-of-the-art 7-axis digital radiography (DR) computed tomography (CT) system for the Richmond facility. The comprehensive system requirements include a fully shielded cabinet-style unit capable of scanning parts up to 32" diameter and 48" in length, with advanced technical specifications for the X-ray tube, detector, manipulator, software, and associated components.
The solicitation (SP4703-25-R-0007) includes extensive training and support provisions, such as 7 days of on-site DR/CT system operation training, 3 seats for a 40-hour digital radiography and intro to CT training course, and 2 seats for an advanced 40-hour CT training course. Delivery will be to Building 47 at 6090 Strathmore Road, Richmond, VA, with installation in Room 150 (Metrology Lab). The contract includes a 12-month hardware warranty, one preventative maintenance visit, and requires all on-site personnel to meet Department of Defense facility access requirements. Proposals are due on May 16, 2025, at 2:00 PM EST, with the contract being a firm fixed-price acquisition.
View the file
Other files
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Source Selection Information -- See FAR 2.101 and 3.104 1
Defense Logistics Agency
DLA Installation Operations
DM-FRIF
SPECIFICATIONS
DLA Aviation Richmond, Sourcing and Reliability Program Branch (VEA) Bldg. 47, Purchase and Install One (1) New Computed Tomography System
Source Selection Information -- See FAR 2.101 and 3.104 2
Technical Requirements:
This procurement is for a turn-key, state-of-the-art, 7-axis, digital radiography (DR) computed tomography (CT) system. The system must be able to scan parts up to a 32” diameter and 48” in length. The system must be a fully shielded and interlocked cabinet-style unit. It must comply with all applicable industry regulations. The system must include a flat panel detector, 225Kv Microfocus Tube, 450 HP Tube, DR Image processing acquisition workstation and software, and a separate CT reconstruction workstation including software and meet Technical Requirements listed below.
Technical Requirements System Capabilities
· Advanced 2D X-ray inspection
· 2D CT Slice reconstruction
· CT volume reconstruction for 3D inspection
· 3D internal and external surface scanning
· Measurement and Analysis tools
System Specifics
X-ray tube: The x-ray tube shall have the following characteristics:
Microfocus 225kV Directional
· Tube Voltage: 25 - 225kV
· mA Range: 0 -3 mA
· Power: 320 W max
· Focal Spot Size: < 6 microns
· Detail Recognition: < 2 microns
· 10 Meter, High Tension Cable
· Closed Loop Water Cooler
· X-ray Control Software:
o Windows based o Automatic tube warm-up o Automatic filament current adjustment o Automatic focusing o Automatic target loading control o Intensity control o Spare parts and service kit
Source Selection Information -- See FAR 2.101 and 3.104 3
225 Microfocus Automated/Motorized Filter Wheel:
· Integration with DR motion programs
· Selection of several different thicknesses of copper
· Tungsten beam blocker
· Open slot for use without filter
· Fast technique optimization
450Kv Mini focus X-ray Source
· 450/11 metal ceramic tube head
· ASTM E2903 FS 7 focal spot class (EN 12543 Spec 0.4mm and 1.0mm dual focal spots)
· Continuous rating: 700 watts/1.5mA; 1500 watts/3.3mA
· 5mm Be window
· Target angle: 11 degrees
· Radiation coverage: 30 degrees x 40 degrees
· CF-450 high stability generators or equivalent, 1800 watt
· Digital controller with enclosure
· Remote Control via RS232
· I/O power distribution panel
· 2-10 meter high-tension cable
· Oil to air cooler with flow and temperature interlocks
· All associated cables, cooling lines and documentation
450HP X-ray Tube Shutter
· Four leaf, two axis tube shutter shielded for 450kV
· Independent axis control
· Laser alignment system
Spare Pack of 225kV Filaments
· (1) Spare Pack of 225kV Filaments will be provided
Spare 225kV Target
· (1) Spare 225kV Target will be provided
Source Selection Information -- See FAR 2.101 and 3.104 4
X-ray detector: The x-ray detector shall have the minimum following characteristics:
4343N DRZ +
· 17” x 17” (432mm x 432mm) active area
· DRZ+ (Gadox) Scintillator
· 150-micron pixel pitch
· 2880 x 2880 pixel matrix
· 16-bit A/D Conversion
· 15 fps (1x1); 30fps (2x2 pixel binning)
· Premium CT Grade
· Tungsten shielded to 450kV (increased lifetime)
· Power & Ethernet cables
Manipulator: The manipulator shall have the minimum following characteristics:
Movement of the inspection object is limited to x, z, tilt and rotation, along with detector and tube vertical movement.
· All drives are motorized with variable speed joystick operation.
· Tethered desktop control console
· Detector Manipulator:
o Focal Distance: 0” to 48” (0cm to 122cm) o Detector and Tube Vertical Axis – 48” (122cm)
· Part Manipulator:
o X Axis / front to back translation: 33” (84cm) o Z Axis / focal distance tube to detector translation: 48” (122cm) o Tilt: +20 /- 20 Degrees
· CT precision nickel plated 9” (22cm) rotational stage - 360 Degrees Continuous.
· Resolution 0.001 degree
· 500lb (227kg) static capacity (not including tilt)
· Integrated electronics.
· Heavy duty welded and precision machined frame construction for reliability
· Precision linear bearing rail
Radiation Enclosure 450kV
· Radiation shielding to energies of 450kV constant potential.
· Enclosure will have the ability to pass through an 8’ x 8’ (2438.4 mm2) door frame for installation. Modular assembly is optional to aid in entry to the lab space for installation.
· Steel/lead/steel construction.
· Cable access port with cover.
· Large, motorized sliding access door.
· Primary and secondary X-ray safety interlocks.
· Interior lighting.
· X-ray on warning light.
· Designed to meet all state and federal X-ray safety regulations.
Motion Control Module
Source Selection Information -- See FAR 2.101 and 3.104 5
The Motion Control Module will provide an operator-friendly teach-and-learn interface to create complex motion control programs easily and quickly. Will include additional 4k monitor.
4 Camera Color Surveillance System Internal Camera surveillance system w/split screen monitor and minimum 4 cameras.
Computer hardware and Software:
The system shall include a computer workstation with the following minimum requirements to operate the system. The workstations shall include:
DR Workstation (Minimum Requirements)
· Windows® 11 Operating System
· Ten-Core Intel® Xeon® Processor
· 48 GB RAM
· 512 GB Operating System Drive
· 4TB of Data Storage in RAID 0 Configuration
· DVD+/-RW Drive
· 10/100/1000 Network Interface
· NVidia High Performance Display Adaptor
· 31.5” (800mm) 4K Color Display
· Wired Keyboard and Mouse
· SSD Data Drive
DR software (Minimum Requirements) High performance image analysis system with a comprehensive set of image analysis tools.
· Windows® based (Compatible with Windows 10 or higher)
· Non-proprietary image storage format
· DICONDE compliant
· Tools to help ensure ASTM compliance
· High performance image processing and measurement functions utilizing GPU technology
· Automatic parallelization for systems with multiple CPU’s and GPU’s
· Live Imaging up to 30 frames per second
· Live Averaging, Live Histogram, Live line Profile, Live Rotation between portrait and landscape modes
· Live Measurements with up to 1-pixel resolution
· Live image gain/offset calibration and defective pixel correction
· Automatic creation of Technique sheet
· Capture video into industry standard video format, i.e., AVI, MPEG, etc.
Source Selection Information -- See FAR 2.101 and 3.104 6
· Supports digital flat panel detectors, LDA’s and digital/analog cameras at 8, 10, 12 and 16 bits
· Read and store images in standard image formats, i.e., TIFF, PNG, JPEG, etc.
· Acquisition of images on supported systems
· Cropping support for reduced file size and versatility
· Seamless integration with commercial CT software
DR CT Acquisition Module The CT acquisition module shall provide automated CT acquisition capabilities for:
· Capable of CT acquisition up to 10,000 angle steps
· Automatic creation of Technique sheet
· Monitors x-ray performance and will pause the CT scan if x-ray interruption is detected
· Acquisition of Spiral CT scans
· Ring artifact suppression capabilities
Fast CT Acquisition Module Module included that shall allow for continuous fast CT acquisition of projections for fast CT scans capturing approximately 600-1000 projections.
Reconstruction Workstation (Minimum Requirements)
· Windows® based (Compatible with Windows 10 or higher)
· Dual 12-Core Processors (Total 24 Cores)
· 1 TB RAM
· 1 TB SSD Operating System Drive
· 24 TB Hard Drive with high speed Raid 5 capabilities
· 4 NVidia High Performance Display Adapters for Display and Reconstruction
· 15,360 total processing cores
· 96 GB of GPU Memory
· 3D Rendering capabilities
· 31.5” (800mm) 4K Color Display
· Wired Keyboard and Mouse
CT Computed Tomography Software CT software shall include:
· CT calibration, Cone-Beam reconstruction, 3D real time visualization and analysis tools in the same interface.
· The software shall not limit the size of the reconstruction
· The software shall be capable of the following:
o Combining 2 or more datasets taken at different current settings o Significantly improving CT quality of widely varying densities.
o Increasing the dynamic range of the CT dataset.
Source Selection Information -- See FAR 2.101 and 3.104 7 o Reducing noise and photon starvation in the CT reconstruction o The software shall be able to perform local CT reconstructions of a small area of interest inside a larger object.
o Simplified processing - The software shall include a simplified, user-friendly mode that lets the user quickly navigate through the full CT scanning process, from the creation of the project to the final results. The simplified interface shall include features to guide the user through the calibration and reconstruction process.
o Automatic Calibration - This tool is needed to optimize cone-beam CT reconstruction process and to simplify CT and μCT equipment.
o Cone Beam Reconstruction –Reconstruction performed on GPU hardware.
o 3D Volume Rendering
• Real time volume rendering
• Export slices at any angle
• Create movie animations
• Perform localized reconstruction.
• Ring Reduction - Uses a combination of hardware and software to virtually eliminate ring artifacts caused by nonuniform pixel responses.
o 3D Surface Rendering - Surface extraction to CAD compatible format:
• Export to STL (Stereolithography), cloud points, OBJ, DXF, WRL, PLY. Visualization in wireframe mode. Map textures with reflection.
• Export an executable viewer module with a dataset to allow others to view, slice, rotate, window and manipulate the dataset on a traditional laptop or desktop computer. Ideally, the viewer module is fully capable and equivalent to the post-processing capabilities of the CT Workstation.
High Accuracy Geometry Tools ⋅ High accuracy geometry tools for use with metrology applications o Serialized small, medium & large tools o Certificate of Conformance o High accuracy CMM inspection o Precision engineered mounting fixtures that meet the scope of this document o Heavy duty carrying case
⋅ Spiral Acquisition o The system shall be capable of performing automated spiral acquisition with a digital flat- panel detector for the scanning of elongated objects that traditionally cannot fit into a single exposure. The scan should not require volume stitching and the object length should only be limited by the system scanning area. The scan should reduce or eliminate the cone angle artifacts caused by flat, planar features.
Source Selection Information -- See FAR 2.101 and 3.104 8
⋅ Image stitching o Larger parts may require multiple scans to completely scan the part when it is larger than the detector panel. The software shall be able to stitch multiple images into one seamless image of the part or assembly.
⋅ Sub-pixel resolution o The designed motion of the manipulator and the software shall be capable of displaying and analyzing scan data utilizing sub-pixel analysis routines that is typically double of what the detector alone can achieve.
Focal Spot Measurement & Correction for Microfocus X-Ray Source:
⋅ Hardware (Microfocus Automated/Motorized Filter Wheel): and Software for providing an automated method of measuring the focal spot of MicroFocusX-ray tubes in compliance with ASTM E2903 and EN12543-5. The Digital Radiography Acquisition Software must include the following capabilities.
⋅ A method for automatically measuring focal spot size in a minimum of two directions in Realtime.
⋅ A method for automatically calculating and displaying the geometric unsharpness and number of pixels of unsharpness based on the newly measured focal spot, the system identified geometry, and the pixel pitch for a given setup.
⋅ A method of providing suggestive, and adjustable unsharpness thresholds based on the measured focal spot to assure image quality and speed requirements are both optimized for a given application.
⋅ A method of measuring and recording focal spot size into ongoing Electronic Process Control Records.
⋅ A method of automatically updating and storing the measured focal spot to an electronically stored technique sheet.
⋅ A method of using the measured focal spot and calculated geometric unsharpness (Ug) to calculate the maximum unsharpness (Uim) of a given setup per ASTM E2698 and provide indication of whether the focal spot size is sufficient to meet the maximum unsharpness requirements of a given technique
X-ray System CT Simulation and Offline Programming SIMULATION Software Simulation Software - Windows based software that simulates the operation of the CT system which shall include the following items:
⋅ Some of the key features include:
⋅ Windows® based (Compatible with Windows 10 or higher) ⋅ Offline Technique Development
Source Selection Information -- See FAR 2.101 and 3.104 9 o Import/export configurations easily from the system in industry standard formats.
o Import/Export configurations directly from the CT system.
o Optimize system settings & positions offline o Import CAD models of parts/fixtures/etc. using industry standard file types, i.e., STL, OBJ, etc. Part files must be millimeter, centimeter, or inch scale.
Additionally, the STL or OBJ files used must define watertight, physical solids.
o Configure part alloy with alloy mixing o Simulated machine environment o Visualize system crash protection o Ability to define custom materials based on a pre-populated repository and the ability to generate custom materials based on custom inputs. Ability to export CAD model files of optimized part orientation for use in creation of support structure o Software shall have operation and functions that can be adjusted in the simulation software to mirror those on actual system to take the parameters used and match them on the actual system to provide a scan as consistent as possible with the model outputs o Ability to define custom materials based on a pre-populated repository and the ability to generate custom materials based on custom inputs. Ability to export CAD model files of optimized part orientation for use in creation of support structure
⋅ Offline Motion Programming o Program motion routines without tying up a system o Controls look/feel/act the same as the real system
⋅ Imaging Simulation o Simulate different results based on changing settings o Change beam hardening parameters o Realistic polychromatic beam results
⋅ Training o Remotely learn system operation o Train new employees o Reduce system damage by training in a virtual environment
⋅ Programmable CNC Motion Control o The system shall utilize programmable CNC motion control.
⋅ Detector Lateral Shift o Detector Lateral Shift maximum: 60”
⋅ Continuous CT scan: mode will be included for faster CT scans.
⋅ Local CT reconstruction capabilities: Will have the ability to reconstruct just a small area of interest inside a larger object
Source Selection Information -- See FAR 2.101 and 3.104 10
⋅ No limitation of reconstruction size: Provide example of reconstruction limits, such as 10000x10000x10000 voxels reconstruction
Detector Qualification Module Module allows for the qualification of pixel mapping and the verification of the detector quality, spatial resolution, and contrast sensitivity.
Also included in this module is an SMPTE test pattern used for checking the quality of the display monitor.
To include the following:
· Pixel Mapping Qualification
· Detector Quality Check
· Spatial Resolution
· Contrast Sensitivity Check
· SMPTE Display Test Pattern
· EN-462 D13 Duplex Wire Gauge
· Photo Light Meter
Secondary Hard Drive Set - DR & CT Hot swappable hard drives in both the DR and CT workstations to quickly and easily separate scan data from the CT system and transfer to an air-gapped analysis computer. The system shall be capable of complying with 32 CFR Part 2002 and NIST SP 800-53 requirements for Critical Unclassified Information to include Critical Technological information. A protective storage/carrying case will be included.
Powered MicroPositioner
· 2 Axis
· 85mm travel (each axis)
· Jog movements controlled through DR
· Precision centering of samples
225Kv Power Microfocus Filter Wheel Some features of the MicroFocus Filter Wheel:
· Integration with DR motion programs
· Selection of several different thicknesses of copper
· Tungsten beam blocker
· Open slot for use without filter
· Fast technique optimization
Source Selection Information -- See FAR 2.101 and 3.104 11
Additional Components
Copper Filter Kit Set of copper filters for use with microfocus tubes
· Selection of 0.005", 0.010", 0.020", 0.040", 0.060", 0.080" thick filters
High Energy Copper Filter Kit Set of copper filters for use with minifocus tubes
· Selection of 0.093", 0.250", 0.375" thick filters
Turn Table Riser The turn table riser will raise the height of the rotary table. The turn table riser will be used to help reduce possible scatter caused by electrical components below the main rotate stage.
Maintenance 12 Month Hardware Warranty and one preventative maintenance visits
Local support in EST time zone. Support by main manufacturing facility located in the USA.
All personnel required to perform on-site must be able to meet the requirements for access to a DoD facility as an unescorted visitor. See https://www.dla.mil/Aviation/Installation/Visitors-and-ID- Office/ for more information.
https://www.dla.mil/Aviation/Installation/Visitors-and-ID-Office/ https://www.dla.mil/Aviation/Installation/Visitors-and-ID-Office/
Source Selection Information -- See FAR 2.101 and 3.104 12
All forementioned Software must be developed and Sold by the Mfg. of the Provided X- Ray System.
o All software included in the delivery must be applicable to the provided hardware.
o For example:
If the workstation is a Windows-based system, then the software will need to be developed for Windows.
If the workstation is a Linux-based system, then the software will need to be developed for Linux.
o The requirement focuses on the specific X-ray control, imaging, simulation, and analysis software.
o It is acceptable to have a manufacturer’s software for the detector or x-ray source not originate from the vendor if they are not the original manufacturer, but all software must be sold as a configured solution for the capability of the machine sold.
o Integration of developed software for an OS is permitted.
First year DR & CT Software Maintenance/Updates
Technical Manual: The system shall include a user’s guide technical manual that covers system operation, routine maintenance and basic troubleshooting information. The standard commercial user/technical manual(s) for the hardware and software should meet this requirement.
Demonstration of Salient Characteristics:
Failure to provide the following samples and successful demonstrations will prevent one from becoming technically acceptable.
· Provided a sample part that is up to 25” in length and under 6” in width: the seller must demonstrate the ability to use one continuous helical motion in which the sample is CT scanned using a single automated spiral acquisition. This shall not be separate CT scans stitched together. This will demonstrate the spiral acquisition capabilities
· Provided a sample part that is up to 20” in length and under 20” in width: the seller must demonstrate the ability to seamlessly stitch together the 2D images from multiple projection angles and reconstruct a single CT volume of the entire part. This will demonstrate the image stitching capabilities.
· Provided a sample: demonstrate the ability to automatically move the detector a sub pixel distance both horizontally and vertically, capturing 4 images within each pixel for each CT projection. The system shall be capable of performing sub-pixel resolutions scans in both step-scan and continuous-scan modes. The result shall produce a CT voxel size equal to half that of a standard cone beam CT scan. This will demonstrate the sub-pixel resolution capabilities.
· For all of the above samples the vendor will provide interior feature resolution of measurements to the closest .005”
X-Ray Simulation Software:
Source Selection Information -- See FAR 2.101 and 3.104 13
· Provide x-ray simulation software that is representative of the actual system configuration including the manipulator, x-ray tube, detector, and scan envelope geometry. The simulation software must provide a polychromatic x-ray beam representing the kV and current range of the actual system. It shall also provide for controls of the detector and motion system and complete image acquisition parameters for both 2D DR imaging a CT scanning. The software shall be capable of inserting a representative part file into the simulator for imaging and providing acquisition of full motion programs. The software shall provide for material alloy selection of the product sample as well as allow for inserting assemblies and fixtures of varying materials and x-ray beam filters.
Source Selection Information -- See FAR 2.101 and 3.104 14
· The simulation software controls shall be representative of the actual system software controls. The sample must demonstrate within the simulation software, the creation of a CT data acquisition and a reconstruction data set of a representative part as well as a multi position DR motion control program which records all imaging technique and geometry parameters to a file. This simulation generation demonstration shall be performed on a different system than the actual system. This sample must also demonstrate the placement of the simulation file onto the actual x ray system along with the insertion of the actual part into the actual system to generate DR images from the simulation recorded motion control program and a CT Data set of the actual part. This will demonstrate the x ray system CT simulation and offline programming capabilities.
Delivery, Installation, and Training
System Delivery Delivery location will be Building 47, 6090 Strathmore Road, Richmond, VA 23237 with proper truck and material handling equipment to offload and transport system to staging area. A staging area is to be provided and located in Room 150 (Metrology Lab).
Access from Parking Lot or nearest pabed access point to Staging is flat andlimited to either a poured concrete path (8’ wide) or ground path (unobstructed). Approximate travel distance is less than 40 yards. Access to Room 150 is limited to a double hung door with a measurement of 96 x 96 inch. The ceiling height of Room 150 has a clearance of 11.5 feet.
NOTE: All concrete walkways are rated as 3000 PSI concrete, 4” thickness. This will add to the complexity of movement from the parking lot or nearest paved access point to the staging area in Room 150.
Shipping/Freight/Crating included, no touch freight for customer during delivery.
System Installation
· Pre-Install Site Visit will be conducted and coordinated with PM/COR/Installation Management Richmond Representatives to coordinate installation, delivery dates, delivery methods, and identify any issues with install, delivery or any changes.
· Installation will be performed with factory trained personnel.
· Installation will be in Room 150 (Metrology Lab) and to be placed in the corner of the room
(which corner TBD). 12” to 18” will need to be between the cabinet and walls to allow for maintenance.
· Disposal of all materials once system is assembled and clean-up of entire area. Customer will provide dumpsters for disposal.
Source Selection Information -- See FAR 2.101 and 3.104 15
DR / CT System Training:
7 Days On-Site DR/CT system operation training. The offeror shall provide 7 Days On-Site training for system operators covering: digital radiography, computed tomography, and operation/maintenance/safety of the delivered CT system operation. This training shall be provided as part of the system delivery and acceptance.
3 Seats to Digital Radiography & Intro to CT Training 40-hour ASTM/ASNT Certified DR/CT training course, held at vendors site.
This course shall follow the recommended curriculum of FWG-IDR, MAI, ASTM and ASME in a format for compliance with NAS 410 Qualification/Certification program.
2 Seats to Advanced CT Training 40-hour class at vendors’ site that dives into the advanced capabilities of Industrial CT technology. The course shall include CT Technique Development training and CT Reconstruction, the class shall focus on the advanced capabilities of the delivered system and Industrial CT technology. The course shall include CT Technique Development training and CT Reconstruction.
Combatting Trafficking in Persons (CTIP) The Contracting Officer Representative (COR) will monitor the vendor compliance with Combatting Trafficking in Persons (CTIP) clause in accordance with DFARS PGI 222.1703(4), such that noncompliance with 52.222-50 is documented to the Contracting Officer, utilizing and annually completing the CTIP checklist at: http://ctip.defense.gov/.
http://ctip.defense.gov/
| Microfocus 225kV Directional |
| 225 Microfocus Automated/Motorized Filter Wheel: |
| 450Kv Mini focus X-ray Source |
| 450HP X-ray Tube Shutter |
| Spare Pack of 225kV Filaments |
| Spare 225kV Target |
| Manipulator: The manipulator shall have the minimum following characteristics: |
| Radiation Enclosure 450kV |
| Motion Control Module |
| 4 Camera Color Surveillance System |
| Computer hardware and Software: |
| DR Workstation (Minimum Requirements) |
| DR software (Minimum Requirements) |
| DR CT Acquisition Module |
| Fast CT Acquisition Module |
| Reconstruction Workstation (Minimum Requirements) |
| CT Computed Tomography Software |
| High Accuracy Geometry Tools |
| Spiral Acquisition |
| Image stitching |
| Sub-pixel resolution |
| Programmable CNC Motion Control |
| Detector Lateral Shift |
| Detector Qualification Module |
| Secondary Hard Drive Set - DR & CT |
| Copper Filter Kit |
| High Energy Copper Filter Kit |
| Turn Table Riser |
| Maintenance |
| System Delivery |
| System Installation |
| DR / CT System Training: |
| 3 Seats to Digital Radiography & Intro to CT Training |
| 2 Seats to Advanced CT Training |