Questions and Answers Environmental Test Chamber 22 SEP 2026.pdf

PDF 114 KB Posted

Attached to
Reach-in Environmental Chamber
Solicitation number
PANRSA26P0000037084
Issued by

About this file

This is a Request for Quotation (RFQ) Questions and Answers document for a Reach-In Environmental Test Chamber procurement issued by the U.S. Government. The solicitation number is PANRSA-26-P-0000 037084, and this Q&A document was released on September 22, 2026. The Government seeks quotations for an environmental test chamber with specific performance capabilities and safety requirements. Delivery must occur no later than ten months from award, though vendors with longer lead times (approximately 90 weeks) are encouraged to submit proposals. Payment will be provided in full following Government acceptance of equipment, with acceptance criteria established in PWS paragraph 2.6; pre-delivery payments are not available. The original quote due date of 10:30 AM EDT on September 22, 2026 was extended to 2:00 PM CST on the same date due to fiscal year-end constraints.

The chamber must accommodate test specimens up to nearly the interior workspace dimensions while maintaining appropriate airflow. The chamber floor must be heavy-duty and capable of supporting heavy loads, with offerors required to specify floor load ratings in total pounds and pounds per square foot; however, the Government indicated maximum load weight is unknown and varies by application. Test liquids will typically consist of water or a 50/50 mix of water and glycol (specifically NEMCO's inhibited glycol with an auto-ignition temperature reference of >400°C), though other non-hazardous liquids may be used. The Government provided Safety Data Sheets for propylene glycol (flashpoint 98°C) and clarified that the 50/50 glycol/water mix will be used for testing. Vendors raised concerns regarding heater surface temperature limitations based on flammability properties and requested clarification on maximum anticipated loads, test specimen sizes, and required fluid volumes; the Government indicated these parameters vary significantly by application and are detailed in the PWS.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Reach-In Environmental Test Chamber Procurement

PANRSA-26-P-0000 037084

RFQ Questions/Answers September 22, 2026

1. The SDS attached before is for 100% glycol, while the other document shows requirement is for 50/50? Required pack size?

USG Response: Our test applications vary significantly. Unfortunately, I don’t have a solid answer on test specimen size. Interior workspace dimensions are identified within PWS. Test specimens may be sized up to nearly the dimensions of the interior workspace (while leaving appropriate space for airflow).

2. Required volumes?

USG Response: See the answer above.

3. Required ratio?

USG Response: PWS states:

4. Payment Terms: Thermotron’s standard payment terms for this equipment are:

50% due at time of order, Net 30 days / 50% due from shipment, Net 30 days. Our pricing is based on these standard terms. If you require different or special payment terms, please let us know so we can account for them in the quotation.

USG Response: Pre-delivery payments are not available. Payment provided in full following Government acceptance of equipment. Acceptance criteria are established in PWS paragraph 2.6.

5. Lead Time: Page 2, Section 1.5.1.1 states: “Delivery of equipment shall be no later than ten (10) months from award.” Our current estimated lead time for this chamber is approximately 90 weeks. Would this lead time be acceptable?

USG Response: A 90-week lead time is not optimal for this requirement; however vendor is encouraged to submit their proposal for consideration.

6. Heavy-Duty Shelves/Floor: Page 7, Section 2.2.1.6 states: “The chamber floor shall be heavy duty and capable of supporting heavy loads. Potential offerors shall indicate floor load rating in total pounds and pounds per square foot.” Could you please provide the maximum anticipated weight/load on the chamber floor and shelves so our engineering team can confirm the appropriate ratings?

USG Response: Our test applications can vary widely. Maximum load weight is unknown.

7. Glycol / Hazardous Liquid: Page 8, Section 2.2.2.5.3 states: “Liquid loadings will typically be water or a 50/50 mix of water and glycol but may include other non-hazardous liquids.” Thermotron considers certain types of glycol to be combustible, which may require additional safety provisions. If engineering determines that the glycol being used presents a hazard, a Minimal Spark package may be required. Could you please confirm what type of glycol will be used, its flash point, and whether it is considered combustible? Also, could you please complete and quickly return the attached Minimal Spark Checklist so our engineering team can review the application? Our concern is safety.

USG Response: Glyocol SDS is attached. For testing, we utilize a 50/50 Glycol/water mix.

8. Quote Due Date / Extension: The solicitation currently closes at 10:30 AM EDT on

Tuesday, September 22, 2026. Would it be possible to extend the due date to Friday, September 25, 2026, to allow us enough time to complete the engineering review and prepare our quotation?

USG Response: Given the rapid approach of the end of FY26, we can only extend the solicitation deadline to 2pm CST, September 22, 2026.

9. SDS for Propylene Glycol does not list an auto-ignition (AI) temperature; it only lists a Flashpoint (FP) of 98°C. When we are quoting a chamber where chemicals are being used in the chamber, we have to limit the surface temperature of the heaters as to minimize any issues with flammability in the chamber. We limit the heater surface temperature based on the AI temperature or the FP of the chemicals being used so that the heater never gets hot enough to be a source of spark or ignition.

We prefer to use the AI temperature as it is always much higher than the FP.

However, if there is no AI listed, we have to use the FP. The formulas we use to determine the heater surface temperature are: For the AI – AI temp x .80 – 20°C (For example, if the AI is 500°C, 500 x .80 = 400°C -20°C = 380°C). In this scenario, we would limit the heater surface temperature to 380°C. For the FP – FP temp – 20°C (For example, if the FP is 280°C, 280°C – 20°C = 260°C). In this scenario, we would limit the heater surface temperature to 260°C. Regardless of which formula is used to determine the heater surface temperature, the chamber workspace temperature will only reach approximately 30°C below the heater surface temperature and it could take a long time to get there. What that means is that with a flashpoint of 98°C, the workspace temp will only reach about 48°C (98°C - 20°C – 30°C = 48°C). Please ask for an SDS with the AI temp.

USG Response: NEMCO’s safety datasheets for inhibited glycol (concentrate or premixed) do not contain auto-ignition temperatures. Their inhibited concentrate lists an AI temp of >400C, as a reference point. All 3 sheets are attached for reference. Note that test ladings will be a 50%/50% mix of glycol and water.