A02. PWS Attach 3 - Analytical Methods v2.docx
DOCX document 17 KB Posted
- Attached to
- Trace Metal Sample Analysis
- Solicitation number
- HT942726RE010
- Issued by
- Not on record
About this file
This is a Performance Work Statement Attachment specifying the analytical methods and technologies required for trace metals analysis across multiple sample matrices. The attachment establishes EPA-approved methodologies for analyzing trace metals in fresh water, salt water, soil, sediment, and tissue samples using Inductively Coupled Plasma Mass Spectrometry (ICP/MS) as the primary method, with ICP/OES as an alternative for freshwater analysis under EPA 200.7. For specialized analyses, the document requires Cold Vapor Atomic Fluorescence (CVAF) for mercury detection across all matrices per EPA 1631/1631E, Hydride Generation-Continuous Thermospray Gas Chromatography-Atomic Absorption Spectroscopy (HG-CT-GC-AAS) for arsenic speciation per EPA 1632, and Ion Chromatography with post-column reaction and UV-Vis spectrophotometry for hexavalent chromium speciation per EPA 1636.
The attachment also addresses sample preparation requirements, including digestion procedures and preconcentration methods, with vendors required to submit Standard Operating Procedures (SOPs) for approval. Media and supplies cleaning protocols must comply with EPA 1669 standards for equipment such as Teflon or silicone tubing, in-line sampling filters, and strainers. This specification supports the Defense Health Agency Contracting Activity's solicitation for trace metals analysis services under solicitation number HT942726RE010, classified as NAICS 541380 (Testing Laboratories) with a $19M size standard, issued under full and open competition.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| A02 PWS FINAL v2 COR.docx | DOCX document | |
| Attachment 6-Pricing Sheet v2.xlsx | XLSX spreadsheet | |
| Solicitation - HT942726RE010-0001.docx | DOCX document | |
| A02. PWS Attach 1 - Final Analytical Report & EDD Requirements v2.docx | DOCX document | |
| A02. PWS Attach 4 - Limit of Quantitation (LOQ) v2.docx | DOCX document | |
| A02. PWS Attach 3 - Analytical Methods.pdf | ||
| A02. PWS Attach 4 - Limit of Quantitation (LOQ).pdf | ||
| A02. PWS Attach 1 - Final Analytical Report & EDD Requirements.pdf | ||
| Attachment 6-Pricing Sheet.xlsx | XLSX spreadsheet | |
| Solicitation - HT942726RE010_final.docx | DOCX document | |
| A02. PWS Attach 2 - EDD Specifications.pdf |
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 3 – Analytical methods
Parameter
Technology
Method Reference
| Trace Metals (fresh water) |
| ICP/MS |
ICP/OES (if necessary)
EPA 1638
EPA 200.7
| Trace Metals (salt water) |
| ICP/MS |
EPA 1640
| Trace Metals (soil/sediment) |
| ICP/MS |
EPA 1638
| Trace Metals (tissue) |
| ICP/MS |
EPA 1638
Trace Mercury (all water, soil/ sediment, and tissue)
| CVAF |
| EPA 1631 / 1631E |
| Arsenic Speciation III (trivalent, 7778-43-0) and V (pentavalent, 7778-43-0) |
| HG-CT-GC-AAS |
| EPA 1632 |
| Chromium Speciation VI (Hexavalent, 18540-29-9) |
| Ion Chromatography (IC) with post-column reaction and UV-Vis spectrophotometry |
| EPA 1636 |
| Methyl Mercury (water) |
| Purge & Trap CVAF |
| EPA 1630 |
| Sample Preparation (digestion for all matrices) |
| Various |
| Submit SOP’s for approval |
| Sample Preparation (preconcentration) |
| Various |
| EPA 1640 (or equivalent) |
| Media/Supplies Cleaning (for example - Teflon or silicone tubing, in-line sampling filters, strainers, etcetera) |
| Various |
| EPA 1669 |
Attachment 3 – ACRONYMS CVAF – Cold Vapor Atomic Fluorescence EPA – U.S. Environmental Protection Agency ICP/MS – Inductively Coupled Plasma Mass Spectrometry ICP/OES – Inductively Coupled Plasma Optical Emission spectroscopy SOP – Standard Operating Procedure