Statement of Work Summary: Sterilizer Purchase DHN-Europe
This Statement of Work describes the Defense Health Agency's (DHA) requirement to purchase, ship, and provide operator training for commercial steam tabletop sterilizers across multiple European military medical and dental treatment facilities (MTFs and DTFs) in Germany, Italy, Belgium, the United Kingdom, Spain, and Turkey. The requirement is to replace existing sterilization equipment that has reached or exceeded its operational life expectancy. The contractor must supply Large-Capacity Tabletop Sterilizers (approximately 85L) and Class B Pre-Vacuum Tabletop Sterilizers (approximately 27L) capable of processing a wide range of surgical instruments with multiple rack configurations. All equipment must comply with ANSI/AAMI ST79, ISO 17665-1, EN 13060, and EN 285 standards for moist heat steam sterilization, and shall have FDA approval and CE Mark certification per Medical Devices Directive 93/42/EEC.
The sterilizers must utilize standardized user interfaces, control systems, and parts inventories across all facilities to facilitate operator transition and maintenance consistency. Key technical requirements include a minimum 4.5-liter water reservoir with removable filter, operating temperature range of 250°F to 274°F (121°C to 134°C), automated closed-door drying systems with 0.2 µm HEPA filters, double safety locking mechanisms preventing door opening during pressurization, automatic water level detection and shutdown, and 230v, 50/60 Hz electrical compatibility without requiring step-up/step-down transformers. The contractor must provide all transportation and delivery costs (all-inclusive), coordinate installation at four specified European locations, deliver at least two copies of English-language operator and technical manuals to each site, provide Monday through Friday telephonic support (0800-1700 CET) during the warranty period, and furnish basic operational and user-level maintenance training with formal training certificates authorizing military Biomedical Equipment Technicians (BMETs) and clinical staff to perform maintenance without voiding the commercial warranty. Equipment shall include a minimum 12-month parts-only warranty with an optional additional 12-month supplemental warranty available under separate agreement, and the complete manufacturer's commercial warranty terms must be provided in English with quote submission and equipment delivery.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
Sterilizer Purchase DHN-Europe
1. Background
The Defense Health Agency (DHA) European Network has a requirement for the purchase, shipment, and operator training via telephonic support on all standardized Tabletop Steam Sterilizer platforms. The DHA Europe Network operates Medical and Dental Treatment Facilities ((MTFs and DTFs), across multiple European host nations which includes Germany, Italy, Belgium, the United Kingdom, Spain, and Turkey). Standardizing medical sterilizers across these facilities is critical to ensuring a uniform Sterility Assurance Level. The equipment is required to aid in preventing the spread of infection and cross-contamination among both patients and staff by thoroughly disinfecting dental/surgical instruments prior to use.
2.0. Description of supplies/services
2.1. The Government requires sterilization equipment to replace the existing units that are at or beyond their life expectancy.
2.2. Select equipment shall be capable of processing a wide range of surgical instruments with multiple rack configurations.
2.3. Equipment Class: The requirement is for commercial steam tabletop sterilizers (autoclaves), comprising Large-Capacity Tabletop Sterilizers (~85L) and Class B Pre-Vacuum Tabletop Sterilizers (~27L) for medical and dental instrument reprocessing.
2.4. The sterilizers shall be compliant with ANSI/AAMI ST79 and ISO 17665-1 standards for moist heat steam sterilization, as well as applicable EN 13060 and EN 285 standard.
2.5. Door Safety Systems: All units shall incorporate automated safety locking mechanisms that prevent door opening while the chamber is pressurized. In the event of a power failure, the chamber shall allow safe manual opening once depressurized.
2.6. Power Failure: In the event of a power failure, the door shall be capable of being opened manually. A door interlock feature is provided to prevent cross-contamination.
3.0. Scope
3.1. The contractors shall provide all transportation required to ensure onsite delivery.
3.2. Identification. All sterilizers shall be compliant with the regulations and standards of the Food and Drug Administration (FDA) and The Joint Commission (TJC). However, when available, the unit shall have CE Mark per Governing Medical Devices Directive 93/42/EEC as amended by 2007/47/EC. This does not preclude the requirement for FDA Approval.
4.0. System Requirements for Equipment Standardization
4.1. Standardization. All procured sterilizers must utilize a standardized user interface, control system, and parts inventory to ensure ease of operator transition and simplify the equipment maintenance program across the DHA Europe Network.
4.1.1. Configurations. The Contractor shall offer various chamber sizes ranging from small tabletop sterilizers (for dental and health clinics) to include larger bulk configurations.
4.1.2. Water Reservoir. The Reservoir shall hold no less than 4.5 liters of water in order to ensure sufficient steam generation.
4.1.3. The unit shall contain a removable water filter (pump strainer) to ensure clean water is used during system operation.
4.1.4. Water Quality. The offered equipment's steam generator/condensate water supply must comply with EN 285 water quality requirements for steam sterilization.
4.1.5. Operating Temperature. The temperature range shall be from 250°F to 274°F (121°C to 134°C) to ensure effective sterilization.
4.1.6.Drying System: All sterilizers shall be equipped with an automated closed-door drying system incorporating a replaceable 0.2 µm HEPA air filter (or post-vacuum drying) to prevent recontamination of sterile loads.
5.0. System Safety
5.1. Chamber Lockout. The unit shall have a double safety locking device that prevents the door from opening while the chamber is pressurized and an automatic shut-off at the end of both sterilization and drying cycles.
5.2. Insufficient Water Detection. The sterilizer shall automatically shut off in the event the water level falls below the minimum required level for operation.
5.3. Safety Lockout Feature. The sterilizer shall employ a feature that prevents a cycle from starting if the door is not fully closed.
5.4. Door Interlock. The sterilizer shall have a door interlock feature that prevents the door from opening while the chamber is pressurized, ensuring user safety during operation.
6.0. Contractor Requirements
6.1. The sterilizer shall be compliant with all the regulatory guidelines identified in para 3.2.
6.2. Telephonic Support. Both Operator and Technical Support shall be provided as needed to the end user to ensure the equipment is properly placed into service and remain fully operational during the warranty period. Telephonic support shall be made available Monday through Friday, 0800 to
1700 Central European Time (CET) / Germany Local Time, excluding U.S. Federal and Host- Nation German holidays
6.3. Training and maintenance. The contractor shall provide basic operational and user-level maintenance training for each location. Training to be scheduled and coordinated upon delivery.
The contractor shall issue formal Training Certificates and written Manufacturer Authorization certifying military Biomedical Equipment Technicians (BMETs) and clinical staff to perform user-level maintenance and routine parts replacement without voiding the manufacturer’s commercial warranty
6.4. Onsite or remote training shall be determined and agreed upon between the medical facility and the Contractor.
6.5. Operator Manuals and Technical Literature. The Contractor shall provide no less than two copies of operator and technical manuals. All manuals or accompanying literature shall be in English and provided to each location.
7.0. Operating Voltage
7.1. All equipment and components requiring standard electrical power shall be 230v, 50/60 Hz compliant, and capable of operating on standard European electrical power.
7.2. Equipment requiring an external step-down or step-up transformer shall not be allowed.
8.0. Warranty
8.1. The equipment provided shall have a warranty period of no less than 12 months (parts only) and include an additional 12 supplemental warranty for up to a total of 24 months. If required, the optional 12 months shall be provided under a separate agreement after the initial warranty period.
A complete copy of the manufacturer's commercial warranty terms and conditions must be provided in English with quote submission and equipment delivery.
9.0. Delivery
9.1. Shipping and handling costs shall be all inclusive with this purchase.
9.2. The Contractor shall be responsible for coordinating, packing, shipping, and receipt of all the equipment to include all required hardware and software associated with the equipment.
9.3. Below are the desired delivery Locations and Points of Contact (POC) for this contract:
Site Property Management POC
Site Property Management POC
Landstuhl Regional Medical Center PMB-Warehouse, Attn: Joachim Bannes Bldg./Geb. # 3739 66849 Landstuhl-Kirchberg, Germany
DSN: 314.590. 4654
COM: +49 06371.9464.4654/6347
Email: pia.m.kessler.ln@health.mil
86th MDSS/MEMO Section Lincoln Blvd., Geb. # 2497 Attn: Ms. Rosemarie Joyner 66877 Ramstein-Flugplatz, Germany
DSN: 314.727.3393
COM: +49 06371.46.2430
Email: rosemarie.joyner.ln@health.mil
Medical Logistics Purchasing POC 52 Medical Group HCOS/SGGM Bolling Drive. # 550, Attn: SrA Noah Cardiel 54529 Spangdahlem-Air Base, Germany
DSN: 314.452.3316/3673
COM: +49 06565.61.3316/3325