Attachment C - Datasheet_THORNTON_M300_Transmitters.pdf
PDF 611 KB Posted
- Solicitation number
- 485281
- Issued by
-
Department of Energy
About this file
This document is a product brochure for the METTLER TOLEDO THORNTON M300 series of analytical transmitters used for comprehensive liquid analysis in pure water treatment applications.
The M300 transmitter line offers single-channel, multi-channel, and multi-parameter models measuring conductivity/resistivity, pH/ORP, dissolved oxygen, dissolved ozone, and flow rates. Available in both ¼ DIN panel-mount and ½ DIN wall/panel/pipe-mount configurations, the transmitters feature a large backlit 4-line display, dual-level password protection, on-line sensor diagnostics, internal PID control, and USB connectivity for configuration and data acquisition. Conductivity/resistivity models achieve ±0.5% accuracy with temperature ranges from -40 to 200°C; pH/ORP models deliver ±0.03 pH accuracy with ranges from -1.00 to 15.00 pH; multiparameter models combine any two parameters with 4 analog output signals and 6 relay setpoints; and flow models accommodate 1 or 4 channels with 0-9999 GPM capability. All models include universal 100-240 VAC or 20-30 VDC power supply, multiple language support, CE compliance, and pending UL listing. Pricing and ordering information lists part numbers for various configurations and accessories including sensor cables, calibration modules, and standard solutions. Primary applications target semiconductor rinsing, pharmaceutical water systems, power plant cycle chemistry monitoring, and wastewater treatment requiring precise analytical measurement and control capabilities.
View the file
Other files
Show all 14
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
An al yt ic al T ra ns m itt er s
M300 Transmitters for Comprehensive Liquid Analysis
THORNTON
Leading Pure Water Analytics
M300 Transmitters Conductivity / Resistivity pH / ORP
Dissolved Oxygen
Dissolved Ozone
Flow / Total Flow
M
Tr an sm itt er s
The METTLER TOLEDO THORNTON M300 series of process analytical instruments provide single-channel, multi-channel and multi-parameter models to measure conductivity/ resis-tivity, pH/ORP, dissolved oxygen, ozone and flowrate. Both ¼ DIN and ½ DIN enclosures are available for especially convenient panel, wall and pipe mounting.
The standard for measurement in pure water treatment
Parameter Channels Outputs Specs on page*
Conductivity / Resistivity 1 2 analog; 4 relays 3
Conductivity / Resistivity 2 4 analog; 4 relays 3 pH ORP 1 2 analog; 4 relays 4
Conductivity / Resistivity, pH / ORP
Dissolved Oxygen, Ozone 2 4 analog; 6 relays 5 - 6
Flow 1 2 analog; 4 relays 4
Flow 4 4 analog; 4 relays 4
*physical and electrical specifications on page 7, order numbers on page 10
M od el s el ec tio n
Features � Large backlit 4-line display � Compact ¼ DIN panel-mounting with NEMA 4X, IP65 panel seal � Robust ½ DIN wall-mounting with NEMA 4X, IP65 back cover � User-selectable dual-level password protection � On-line sensor diagnostics for pH � Internal PID control with relay or analog output � USB port for configuration and data acquisition � Direct sensor inputs for all parameters � Compatible with existing Thornton sensors � Plug-in terminal connectors for easy wiring � Universal AC/DC power supply � Multiple languages: English, French, German, Italian, Spanish � CE compliant, UL Listing pending for US and Canadian standards
Applications Pure and ultrapure water treatment for semiconductor rinsing, critical power/steam makeup and pharmaceutical waters.
Semiconductor processing in rinsers and wet benches with precise resistivity measurement and temperature compensation.
Power plant cycle chemistry and stator cooling monitoring with exceptionally accurate temperature compensation for specific and cation conductivity and pH and very low maintenance dissolved oxygen measurement.
Pharmaceutical water monitoring to meet USP, EP, and JP conductivity requirements including built in tables of alarm limits.
Reclaim, recycle and wastewater treatment for the above industry applications for contaminant detection, diversion and neutralization.
michael.hughett Rectangle
� One & Two channel models � Highest accuracy measurement & temperature compensation � Wide range of measurement allows verification in ASTM standards even for pure water measurements � Two- & four-electrode conductivity/resistivity measurements � Reverse osmosis % rejection computation on two-channel models � Direct acid & caustic concentration readout � USP ‹645› & EP alarm setpoints
Measurement Specifications
M300 Conductivity / Resistivity Models
Functional
Conductivity/Resistivity Ranges
0.01 Constant Sensor 0.002 to 200 μS/cm (5000 ohm-cm to 500 Mohm-cm)
0.1 Constant Sensor 0.02 to 2000 μS/cm (500 ohm-cm to 50 Mohm-cm)
10 Constant Sensor 50 to 40,000 μS/cm (25 to 20 Kohm-cm)
4-electrode Sensor 0.01 mS/cm to 650 mS/cm (1.54 ohm-cm to 100 Kohm-cm) readout in equivalent S/m ranges is selectable
Concentration Ranges of HCl, NaOH, H2SO4 0-20%, 0-15%, 0-20%
TDS Ranges (CaCO3 and NaCl) cover equivalent conductivity ranges
Calculated Paraameters (2-channel) % Rejection, power plant calculations of pH based on specific and cation conductivity, and C02 based on cation and degassed conductivity
Resolution four significant digits, auto-ranged
Update rate display and outputs, once per second
Temperature Measurement Range -40 to 200°C (-40 to 392 °F), resolution 0.1°
Temperature Sensor Input RTD, Pt1000 (Pt100 with adapter)
Temperature Compensation selectable as: Std (standard high purity Thornton/Light), Light 84, Std referenced to 75°C, linear %/°C, 50% glycol, 100% glycol, cation, ammonia, isopropyl alcohol, none
Sensor maximum distance 61 m (200 feet); 15 m (50 ft) with 4-electrode sensors
Performance
Conductivity/Resistivity Accuracy ± 0.5% of reading or 0.5 ohm, whichever is greater, up to 18 Mohm-cm
Repeatability ± 0.1% of reading
Temperature Relative Accuracy ± 0.25 °C (± 0.45 °F)
Temperature Resolution 0.01 °
Temperature Repeatability ± 0.13 °C (± 0.23 °F)
Outputs – see p.7 for ratings Single channel Two channel
Setpoints/Alarms 2 - high, low, outside, between, USP, EP 6 - high, low, outside, between, USP or EP
Relays 2 SPDT, 2 SPST reed 2 SPDT, 1 SPST NO, 1 SPST NC, 2 SPST reed
Analog Output Signals 2 4
Discrete Inputs 1 2
Tr an sm itt er s
M300 pH/ORP Models
Functional pH, ORP Ranges -1.00 to 15.00 pH, -1500 to 1500 mV
Temperature Range - 30 to + 130°C (-22 to 266 °F)
Temperature Sensor RTD, Pt1000 (Pt100 with adapter)
Temperature Compensation automatic / manual for electrode output, plus adjustable solution temperature coefficient for solution ionization effects
Update rate display and outputs, once per second
Calibration 1- or 2- point, with auto buffer recognition
Diagnostics selectable continuous checking of membrane resistance and reference diaphragm/junction resistance (with solution ground sensors)
Sensor Maximum Distance 10 m (33 ft)
Performance pH, ORP Relative Accuracy ± 0.03 pH, ± 2 mV pH, ORP Resolution 0.01 pH, 1 mV
Temperature Accuracy ± 0.25 °C (± 0.45 °F)
Temperature Resolution 0.1 °C
Outputs – see p.7 for ratings
Setpoints/Alarms 4 - high, low, outside, or between
Relays 2 SPDT, 2 SPST reed
Analog Output Signals 2
Discrete Inputs 1
� Direct electrode input � On-line sensor diagnostics for measuring and reference electrode � Auto buffer recognition with extensive buffer library � Internal PID control � Combined high/low setpoint for alarming outside a band � Solution temperature compensation for pure water ionization effects as well as conventional electrode temperature compensation
� Automatic sensor cleaning
Measurement Specifications
M300 Multiparameter Models
Functional Conductivity/Resistivity Ranges
0.01 Constant Sensor 0.002 to 200 μS/cm (5000 ohm-cm to 500 Mohm-cm)
0.1 Constant Sensor 0.02 to 2000 μS/cm (500 ohm-cm to 50 Mohm-cm)
10 Constant Sensor 50 to 40,000 μS/cm (25 to 20 Kohm-cm)
4-electrode Sensor 0.01 mS/cm to 650 mS/cm (1.54 ohm-cm to 100 Kohm-cm) readout in equivalent S/m ranges is selectable
Concentration Ranges of HCl, NaOH, H2SO4 0-20%, 0-15%, 0-20%
TDS Ranges (CaCO3 and NaCl) cover equivalent conductivity ranges
RO % Rejection 0 -100% (computed from two conductivity measurements)
Conductivity Resolution four significant digits, auto-ranged pH, ORP Ranges -1.00 to 15.00 pH, -1500 to 1500 mV; resolution 0.01 pH, 1 mV
Dissolved Oxygen Ranges 0-20,000 ppb or μg/L; 0-20 ppm or mg/L, 0-200% saturation; resolution 0.1 ppb or μg/L
Dissolved Ozone Ranges 0-5,000 ppb or μg/L, 0-5 ppm or mg/L; resolution 0.1 ppb or μg/L
Temperature Measurement Range -40 to 200 °C (-40 to 392 °F); resolution 0.1 °
Temperature Sensor Input RTD, Pt1000 (Pt100 with adapter)
Update rate display and outputs, once per second
Temperature Compensation
Conductivity/Resistivity selectable as: Std (standard high purity Thornton/Light), Light 84, Std referenced to 75°C, linear %/°C, 50% glycol, 100% glycol, cation, ammonia, isopropyl alcohol, none pH automatic / manual for electrode output, plus adjustable solution temperature coefficient for solution ionization effects
Dissolved Oxygen automatic, for membrane permeability and oxygen solubility
Dissolved Ozone automatic, for membrane permeability and ozone solubility
Conductivity/Resistivity Calibration 1-point; 2-point for 4-E sensors pH Calibration 1- or 2-point, with auto buffer recognition pH Diagnostics selectable continuous checking of membrane resistance and reference diaphragm/junction resistance (with solution ground sensors)
Dissolved Oxygen Calibration air/comparison, zero
Dissolved Ozone Calibration comparison, zero
Sensor maximum distance 61 m (200 ft); 15 m (50 ft) with 4-electrode sensors
Performance Accuracy
Conductivity/Resistivity ± 0.5% of reading or 0.5 ohm, whichever is greater, up to 18 Mohm-cm pH, ORP ± 0.03 pH, ± 2 mV
Dissolved Oxygen ± 1% (± 2% with long life probe) of reading or ± 1 ppb, system accuracy
Dissolved Ozone ± 2 % of reading or ± 3 ppb, system accuracy
Outputs – see p. 7 for ratings
Setpoints/Alarms 6 - high, low, outside, between, USP or EP
Relays 2 SPDT, 1 SPST NO, 1 SPST NC, 2 SPST reed
Analog Output Signals 4
Discrete Inputs 2
� Two Field-configurable channels for any pair of parameters:
� Conductivity/resistivity, pH/ORP, dissolved oxygen and/or dissolved ozone � Reduces number of instruments and amount of panel space � Includes all features of single parameter instruments
Measurement Specifications
Tr an sm itt er s
� Available in one and four channel models � Accept input from most pulse flow sensors � Compute total flow with high resolution � Configurable for batch control � Compute reverse osmosis % recovery on four channel model
Measurement Specifications
M300 Flow Models
Functional
Flowrate Range 0 to 9999 GPM, L/min, m3/hr
Total Flow Range 0 to 9,999,999 Gallons, 37,850,000 Liters, 37,850 m3
RO % Recovery Range 0 to 100%
Flow Velocity Range equivalent ft/s, m/s
Frequency Range 1 to 4000 Hz
Calculated Parameters (4-channel) ratio, sum and difference of two flowrates
Resolution 4 significant digits, auto-ranged; up to 8 digits for total flow
Update rate display and outputs, once per 2 seconds
Input pulses low <1.0 volt; high >1.4 volts (36 volts max)
Performance
Accuracy ± 0.5 Hz
Repeatability ± 0.2 Hz
Outputs – see p. 7 for ratings Single channel Four channel
Setpoints/Alarms 4 - high, low, outside, or between 8 - high, low, outside, or between
Relays 2 SPDT, 1 SPST NO, 1 SPST NC 2 SPDT, 1 SPST NO, 1 SPST NC
Analog Output Signals 2 4
Discrete Inputs, for external totalizer reset 1 2
Enclosure/Power
Operator interface 4 line back-lit LCD; 5-tactile keys
1/4 DIN Model Dimensions (H x W x D) 4.01 x 4.01 x 5.51 in (102 x 102 x 140 mm)
Max. Depth,- panel mounted 4.96 in (126 mm), excluding plug-in connectors
1/2 DIN Model Dimensions (H x W x D) 5.90 x 5.90 x 4.57 in (150 x 150 x 116 mm)
Max. Depth,- panel mounted 3.46 in (88 mm)
Material ABS/polycarbonate
Weight, 1/4 DIN Models 1.5 lb (0.7 kg)
Weight, 1/2 DIN Models 2 lb (1 kg)
UL Electrical Environment Installation (overvoltage) Category II
Ratings/Approvals UL (pending, US & Canada), CE compliant; NEMA 4X, IP 65 - 1/2 DIN wall mount and 1/4 DIN panel seal
EMC Emissions according to EN55011 Class A
Power universal 100-240 VAC, 50-60 Hz or 20-30 VDC; 5 W
(on power loss, all settings are retained in non-volatile memory without batteries)
Environmental
Storage temperature -40 to 70 °C (-40 to 158 °F)
Ambient temperature operating range -10 to 50 °C (14 to 122 °F)
Relative humidity 0 to 95%
Outputs
Analog outputs (as specified for individual models) powered 0/4-20 mA, 22 mA alarm, 500 ohms maximum load; not for use with externally powered circuits
Analog output accuracy ± 0.05 mA
Analog output scaling linear, bi-linear, logarithmic (1,2,3 or 4 decades), auto-ranging
Relays (as specified for individual models) all contacts are potential free, with adjustable hysteresis and time delay
SPDT, SPST NO, SPST NC 250 VAC/30 VDC, 3 A, resistive
SPST reed 300 VDC, 0.5 A, 10 W
Digital communications USB, type B connector, for remote configuration and data acquisition
Discrete input (as specified for individual models) accepts dry contact closure for remote flow totalizer reset or remote PID control auto/manual selection
PID Control
Display auto/manual status and % output on bottom line of display
Settings auto/manual, setpoint, deadband, non-linear corner points, control limits, proportional gain, integral reset time
(min.), derivative rate time (min.)
Manual Station controlled by up/down arrow keys in manual mode; remote auto/manual selection by discrete input
Control Output Types one or two analog signals, relays—pulse frequency, or relays—pulse length
Physical & Electrical Specifications - all models
Tr an sm itt er s
Dimensional information
Dimensions for 1/4 DIN panel-mount models
Mounting Brackets 2 Places
Dimensions for 1/2 DIN wall/panel-mount models
Dimensions for 1/2 DIN pipe mount kit
Tr an sm itt er s
M300 Transmitters Description Part Number
1/4 DIN enclosure – panel mounting kit included
M300 1-channel Conductivity/Resistivity 58 002 301
M300 2-channel Conductivity/Resistivity 58 001 304
M300 1-channel pH/ORP 58 001 303
M300 2-channel Multiparameter 58 001 306
M300 1-channel Flow 58 004 302
M300 4-channel Flow 58 001 305
1/2 DIN enclosure – panel/wall/pipe mount*
M300 1-channel Conductivity/Resistivity 58 002 311
M300 2-channel Conductivity/Resistivity 58 001 314
M300 1-channel pH/ORP 58 001 313
M300 2-channel Multiparameter 58 001 316
M300 1-channel Flow 58 004 312
M300 4-channel Flow 58 001 315
* Panel and pipe mounting require kits, listed below, ordered separately.
Accessories Description Part Number
Panel Mount Kit for 1/2 DIN models 52 500 213
Pipe Mount Kit for 1/2 DIN models 52 500 212
Conductivity Calibration Module 58 082 300
Adapter, VP to Standard, for calibrating conductivity with VP Patch Cord 58 080 102
Configuration Software & Data Logger Kit 58 077 300
Adapter Panel - M300 to 200/2000 cutout 58 083 305
Conductivity Sensor Cables Description Part Number
Standard VP*
1 ft (0.3 m) 1001-67 -
5 ft (1.5 m) 1005-67 58 080 201
10 ft (3 m) 1010-67 58 080 202
15 ft (4.5 m) 1015-67 58 080 203
25 ft (7.6 m) 1025-67 58 080 204
50 ft (15.2 m) 1050-67 58 080 205
75 ft (23 m) - 58 080 206
100 ft (30.5 m) 1110-67 58 080 207
200 ft (61 m) 1120-67 58 080 208
* For VP Conductivity sensors only.
Ordering Information michael.hughett Rectangle michael.hughett Rectangle pH / ORP / Dissolved Oxygen / Ozone Sensor Cables Description Part Number pH, Dissolved Oxygen**, Ozone
3 ft (1 m) VP, -30 to 70 °C 52 300 107
10 ft (3 m) VP, -30 to 70 °C 52 300 108
16 ft (5 m) VP, -30 to 70 °C 52 300 109
33 ft (10 m) VP, -30 to 70 °C 52 300 110
High temp 3 ft (1 m) VP, -40 to 135 °C 52 300 111
High temp 10 ft (3 m) VP, -40 to 135 °C 52 300 112
High temp 16 ft (5 m) VP, -40 to 135 °C 52 300 113
High temp 33 ft (10 m) VP, -40 to 135 °C 52 300 114
ORP
3 ft (1 m) AS9 10 001 0102
10 ft (3 m) AS9 10 001 0302
16 ft (5 m) AS9 10 001 0502
33 ft (10 m) AS9 10 001 1002
** For long life dissolved oxygen sensor 58 037 220 only, use standard conductivity cables in previous table.
Calibration / Verification Solutions Description Part Number
Conductivity Standard Solutions
Conductivity standard, 25 μS/cm, 500 mL, HCl, ± 3% 58 078 001
Conductivity standard, 100 μS/cm, 500 mL, KCl, ± 1% 58 078 002
Conductivity standard, 1000 μS/cm, 500 mL, KCl, ± 1% 58 078 003
Conductivity standard, 10,000 μS/cm, 500 mL, KCl, ± 1% 58 078 004
Conductivity standard, 100,000 μS/cm, 500 mL, KCl, ± 1% 58 078 005 pH and ORP (Redox) Standard Buffer Solutions pH Buffer, 4.01, 250 mL 51 340 057 pH Buffer, 7.00, 250 mL 51 340 059 pH Buffer, 9.21, 250 mL 51 300 193 pH Buffer, 10.00, 250 mL 51 340 056
ORP Buffer, 468 mV, 6 x 30 mL 51 319 058
ORP Buffer, 220 mV, 6 x 250 mL 51 340 081
Mettler-Toledo Thornton, Inc.
36 Middlesex Turnpike Bedford, MA 01730 USA Tel. +1-781-301-8600 Fax +1-781-301-8701 Toll Free +1-800-510-PURE thornton.info@mt.com
Subject to technical changes © Mettler-Toledo Thornton, Inc.
ML0122 Rev.B 02/07
Visit for more information www.mt.com/thornton
Quality certificate.
Development, production and testing to ISO 9001.
A certified ServiceXXL provider.
M300 Conductivity Calibration Module 58 082 300
• Includes resistances for all ranges of conductivity/resistivity and temperature
• Used with 1- and 2- channel conductivity and multiparameter models
• With NIST-traceable certificate of calibration
• Enables efficient QA program
• Includes standard patch cord connection; 58 080 102 adapter available for use with VP patch cords
File details come from the government source that posted it. Updated .