This document outlines test slates and specifications for lube oil quality management testing support services. The contractor shall perform routine testing of various oil samples according to test slates that specify analyses such as viscosity, water content, oxidation levels, wear metals, and particle counting. Tests will follow standards from ASTM, SAE, and ISO. The contractor must also conduct spectrochemical analysis of wear metals in oil samples and analytical ferrography as directed. Reporting requirements include submitting test results and analytical ferrography reports to the government client with relevant details and analysis. This RFP from the Department of the Navy seeks these lube oil quality management testing support services with a performance work statement and data transfer requirements outlined in the attachments.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Reference A – Test Slates and Specifications

I. Test Slates by Oil Type: The routine test slates for each type of oil samples are defined in the table below. The

Contractor shall perform these tests by using the test methods outlined in Paragraphs II through IV below.

Test Slate 1 Viscosity, Viscosity Index, Water, Total Base Number (TBN), Spectrochemical Analysis, Fuel Dilution, Oxidation, Insolubles, Direct Reading Ferrography or Particle Quantification

Test Slate 2 Viscosity, Viscosity Index, Water, Spectrochemical Analysis, Total Acid Number (TAN), Particle Count

Test Slate 3 Viscosity, Viscosity Index, Water, Particle Count, Spectrochemical Analysis TAN

Test Slate 4 Water, Particle Count

Test Slate 5 Viscosity, Viscosity Index, Water, Particle Count

Test Slate 6 Water, Particle Count, TAN

Test Slate 7 Viscosity, Viscosity Index, Water, Spectrochemical Analysis, TAN, Direct Reading Ferrography or Particle Quantification, Particle Count, Flashpoint, Oxidation

Test Slate 8 Viscosity, Viscosity Index, Water, Spectrochemical Analysis, TAN, Particle Count, and Flashpoint, Oxidation

Test Slate 9 Viscosity, Viscosity Index, Water, TBN, and Spectrochemical Analysis Test Slate 10 Viscosity, Viscosity Index, Water, TAN, and Spectrochemical Analysis

Test Slate 11 Water, Spectrochemical Analysis, Particulate Count

Test Slate 12 Analytical Ferrography

Test Slate 13 Viscosity and Viscosity Index

Test Slate 14 Water

Test Slate 15 Total Base Number

Test Slate 16 Total Acid Number

Test Slate 17 Insolubles

Test Slate 18 Oxidation

Test Slate 19 Direct Reading Ferrography

Test Slate 20 Particle Quantification

Test Slate 21 Fuel Oil Dilution

Test Slate 22 Particle Count

Test Slate 23 Flash Point

Test Slate 24 Spectrochemical Analysis

Test Slate 25 RPVOT

Definitions of Test Slates:

Test Slate 1: This test slate will be used for testing used diesel engine oil samples.

Test Slate 2: This test slate will be used for testing used hydraulic oil samples with Particle Count.

Test Slate 3: This test slate shall be used for testing used refrigeration, circulating, compressed air, fire retardant and non-PC hydraulic oil samples.

Test Slate 4: This test slate shall be used for designated hydraulic samples.

Test Slate 5: This test slate shall be used for designated hydraulic samples.

Test Slate 6: This test slate shall be used for designated hydraulic samples.

Test Slate 7: This test slate shall be used for testing used gear oil samples.

Test Slate 8: This test slate shall be used for testing used controllable pitch propeller oil samples.

Test Slate 9: This test slate shall be used for testing new diesel engine oil samples.

Test Slate 10: This test slate shall be used for testing new non-diesel engine oil samples and gas turbine samples.

Test Slate 11: This test slate shall be used for testing steering fluid.

All other test slates are self-defined and shall be run when instructed by the COR or designated representative.

II. Physical Test Properties

Physical Test

Property Description Unit of Measure

Viscosity Test in accordance with ASTM D-445. Report Viscosity at 40°C and 100°C. Ability to test ISO 3200 oil, as needed.

cSt

Viscosity Index Calculated in accordance with ASTM D-2270 unitless

Water

“Crackle” test is an acceptable screening method for test slate 1 and 6. For all “crackle” failures and all other test slates water content in accordance with the Karl Fischer water test (ASTM D-6304) shall be performed. Report any Free Water observed as percent by volume.

PPM or Percent by volume

Total Base Number

(TBN)

Test in accordance with ASTM D-2896/D-4739. Testing Range 1-70

Total Acid Number

(TAN)

Test in accordance with ASTM D-664 or ASTM D974.

Testing Range up to 5.0

Insolubles Test in accordance with ASTM D-893 or using Fourier Transform

Infrared (FTIR) spectroscopy.

Testing Range between 0.01 to 5.0

Oxidation Test in accordance with ASTM D7214 or D7414 using Fourier Transform Infrared (FTIR) spectroscopy.

Testing Range 1-50 A/Cm

Direct Reading Ferrography

Particle size index shall be reported in microns using industry accepted test methods.

Testing Range 0-1000 microns

Particle Quantifier

(PQ)

Particle size index shall be reported in microns using industry accepted test methods for Particle Quantifier (Ferrous Density)

Testing Range 0- 1000 microns

Analytical Ferrography

Perform test as directed by DR/PQ Wear Particle Content (WPC) and/or trend analysis and requested by the Government.

Fuel Oil Dilution Fuel dilution shall be tested in accordance with ASTM

D7593 or ASTM D8004.

Percentage by volume (0.1-15%)

Particle Counting

Test in accordance with SAE AS4059.

Test results provided shall be the actual particle count readings in

NAS 1638 and ISO 4406 4/6/14 micron size formats.

NAS 1638 and

ISO 4406

Flash Point Test in accordance with ASTM D93 or ASTM D6450. Instrument shall be capable of completing a 350-450F sweep for lube oil testing. A separate instrument shall be used for any fuel testing completed at the lab.

Testing Range:

350-450 Degree

Fahrenheit

RPVOT Test in accordance with ASTM D2722. minutes

III.Spectrochemical Analysis:

The Contractor shall conduct oil elemental analysis in accordance with ASTM D5185 or D6595) and report concentrations of the following elements:

Wear Metal Symbol Unit of measurement Aluminum (Al) PPM Antimony (Sb) PPM Boron (B) PPM Barium (Ba) PPM Calcium (Ca) PPM Cadmium (Cd) PPM Chromium (Cr) PPM Copper (Cu) PPM Iron (Fe) PPM Lead (Pb) PPM Magnesium (Mg) PPM Manganese (Mn) PPM Molybdenum (Mo) PPM Nickel (Ni) PPM Phosphorous (P) PPM Potassium (K) PPM Tin (Sn) PPM Silicon (Si) PPM Silver (Ag) PPM Sodium (Na) PPM Titanium (Ti) PPM

Vanadium (V) PPM

Zinc (ZN) PPM

IV. Analytical Ferrography Testing Standards:

The Contractor shall conduct ferrographic analysis as dictated by the results of DR/PQ Wear Particle Content (WPC) or as otherwise requested by Government. The following elements shall be included on each analytic ferrography report submitted to the Government:

A representative photograph of sample from 50x-1000x in magnification Wear particle count Normal wear concentration Sliding wear concentration Severe wear concentration Wear particle size determination Dark oxide concentration Red oxide concentration Corrosive wear particle concentration Non-Metallic Particle Concentration Non-Ferrous Metallic Particle Concentration

a. A narrative with a statistical list of suspected problems and the recommended actions

b. Sample analysis data

Report analytical ferrographic results via NSWCPD site as a PDF document or Adobe Acrobat with file linking it to the original oil sample data.

The data analyst shall provide remarks to indicate the test results have been reviewed and verified. Additional information with regards to sample condition such as abnormal visual appearance, presence of contaminants or inability to perform test results due to insufficient sample, shall also be provided in the laboratory comments of the data format (Attachment B).

The data analyst shall employ analysis techniques using trend analysis or USN threshold limits to determine oil sample condition (normal, monitor, abnormal or critical) and provide alert indications.