This Statement of Work (SOW) details the procurement of a Hematology Analyzer for the 49th Medical Group at Holloman Air Force Base, New Mexico. The analyzer must perform at least 50 complete blood counts per hour, utilize optical/laser technology for white blood cell differential counts, have barcode capabilities, and interface with Composite Health Care System and Military Health System Genesis. Key requirements include FDA approval, ability to operate at 0-30% humidity and 4,500 feet altitude, and direct printer capabilities.

The five-year contract includes comprehensive installation, maintenance, and training services. The contractor must provide a certified installation specialist, perform system validation, offer 24/7 technical phone support, conduct preventative maintenance visits every six months, and provide a replacement system if needed. Training includes one initial off-site training slot and annual on-site training for technicians. The facility will be billed on a cost-per-test basis, estimated at 500 monthly tests, with the contractor responsible for providing reagents, controls, and all necessary parts. Performance objectives include maintaining equipment in 98% serviceable condition and 100% compliance with service and training requirements.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK (SOW)

FOR

Hematology Analyzer

1. SCOPE OF WORK: The contactor shall provide all personnel, equipment, tools, materials, startup reagents, supervision, transportation, one service/operator’s manual and other items and services necessary for the installation of the Hematology Analyzer and ancillary equipment in the

Clinical Laboratory located at Holloman AFB, NM. Contractor shall comply with current

National Electric Code (NEC), National Fire Protection Association (NFPA) standards, and all

Environmental and Safety codes, laws and regulations.

1.1 The specifications of the hematology analyzer must be as follows: The analyzer shall perform at least 50 complete blood counts per hour. The analyzer should also use an optical and/or laser technology to provide accurate readings for white blood cell (WBC) counts with a 5 part WBC differential which also enumerates bands, immature granulocytes, blast, variant lymphocytes and nucleated red blood cells that is FDA approved for laboratory use.

Additionally, the white cell counts should provide the correct WBC count in most cases in the presence of nucleated red blood cells, platelet clumps, giant platelets; lyse resistant red blood cells or fragile lymphocytes. The system must have barcode capabilities and must be fully interface-able to the Composite Health Care System (CHCS) and upcoming the Military Health

System Genesis (MHS Genesis). Analyzer must be able to barcode patient samples without technician direct intervention to prevent patient identification errors. Analyzer must be able to work at 0-30% humidity and an altitude of 4500 feet. Analyzer must have direct printer capabilities and automatic printing for samples requiring further review. Due to systems firewalls, analyzer must have capability to download quality data to a disk for upload to a peer group comparison, USB ports are not acceptable.

2. OPERATIONAL REQUIREMENTS AND SPECIFIC TASKS

2.1. The contractor shall provide the government a minimum of (14) calendar days advance notice for commencement of work. Contractor shall identify in writing to the laboratory officer or designee the names and identification required by security forces for background check and the number of individuals that will be working during the installation annually between 14

December and 1 January of each calendar year.

2.2. The contractor shall provide a manufacturer’s certified installation specialist. A Clinical

Application Specialist must perform system validation procedures to include calibrations, accuracy, quality control and mode to mode studies.

The requirements must include:

Solicitation # FA480125Q0005

Attachment 1 - SOW Page 1 of 5

1. Setting up QC files and entering QC limits.

2. Performing correlation studies with current method.

3. Linearity verification utilizing commercial linearity material to validate all measured parameters. If commercial linearity material is not available for any parameter to validate linear range, appropriate patient samples may be used.

4. To obtain acceptable performance on the analyzer, all studies must be presented to the laboratory officer and/or medical technologist for acceptance and approval prior to the contractor’s manufacturer’s (installation specialist, clinical application specialist and/or service engineer) departure.

2.3 LIS INTERFACE: The contractor shall furnish interface assistance during installation. If interface cannot be completed prior to installation teams’ departure, the contractor agrees to provide telephone support to allow for interfacing of system. This will occur initially and during any information system change. Instrument must be compatible to the Composite Health Care

System as well as the Military Health Care System Genesis. If it is found the instrument is not compatible to either system, the contract will be null and void and the contracting company will not receive any monetary compensation from the government.

2.4 MAINTENANCE AND REPAIR/SERVICE: The contractor shall provide the customer with service required to maintain the instrument, contractor provided computer, and software at its full capabilities for the term of this agreement at the business hours agreement stated below. If a replacement system is needed before the end of the term, the contractor will provide a replacement at no cost to the government. The system replacement will take place after the manufacturer has exhausted its troubleshooting options through its service escalation process.

2.5 All contractor employees must check in with the 49th

Medical Group’s Biomedical

Equipment Repair Office (BMET) prior to performing any repairs or maintenance. When all necessary repairs or maintenance are, the contractor’s representative will supply the laboratory with a written report explaining the repairs/maintenance performed prior to departing the facility.

Additionally, the contractor’s representative must check out with the BMET office and provide a documentation of all work accomplished. The contractor shall provide a maintenance and service agreement to include:

Service must include:

1. Years 1 – 5: Must provide at least a routine service IAW manufacturer’s standards (Mon –

Fri, 730 am to 430 pm)

2. If on-site repair is needed based on vendor service escalation process, response shall be within 24 hours

3. Technical Phone Support 24 hours a day/ 7 days a week (including holidays)

4. Receive preventative maintenance visits every six months with 72 hour notice prior to arrival.

Attachment 1 - SOW Page 2 of 5

Parts: The contractor shall be responsible for all parts required to maintain a serviceable system.

2.6 SHIPPING AND HANDLING: The contractor shall provide all manufacturer brand reagent and controls necessary to meet the average monthly test volume as stated in the General Services

Administration (GSA)/Federal Supply Schedule (FSS) Price List assuming reasonably efficient utilization. The facility will be billed on a cost per test basis for patient tests and reruns based on an estimated 500 tests performed monthly. Controls must be provided on a mutually-agreed shipping schedule for the assigned monthly sample volume at no additional cost to the government. A schedule of routine shipments of reagents shall be provided along with a shipping mode based on geographical location. The mode of shipment used shall be appropriate for the product being shipped to arrive within manufacturer’s specifications. The shipment of temperature controlled products that are restricted from shipping over weekends and holidays shall be sent to arrive during normal duty hours.

2.7 TRAINING: Prior to installation, the contractor should provide one training slot for one individual, off-site at their training facility; the training must be completed within the first 12 months. The contractor should provide one additional slot annually over the course of the 5-year agreement. During the installation on-site, training must be performed for all other testing personnel by a certified system specialist. If the company does not provide off-site training for a key operator, the installation specialist will train all current employees and return annually to train all new technicians on all items to include maintenance, troubleshooting, and quality control procedures.

2.8 The contractor must provide Material Safety Data Sheets (MSDS) for all required supply items that are utilized on the system at no cost to the government.

2.9 The contractor must provide all instrument freight charges incurred to have the instrument delivered for installation.

3 HOURS OF OPERATION: Normal work hours are from 0730 to 1630 hours, Monday thru

Friday. Request for work other than normal work hours must be coordinated and approved by the

Contracting Office and Laboratory Management.

Normal business hours are from 0730 to 1630 hours, Monday through Friday, excluding holidays. No preventative maintenance or consumable deliveries shall be made on

Saturday, Sunday or federal holidays as shown:

New Year’s Day – 1 January

Martin Luther King’s Day – Third Monday of the month in January

President’s Day – Third Monday of month in February

Memorial Day – Last Monday in May

Juneteenth

Independence Day

Labor Day

Columbus Day

Attachment 1 - SOW Page 3 of 5

Veteran’s Day

Thanksgiving

Christmas Day

4 The laboratory POC shall be the Officer in Charge, 575-572-3587.

4.1 SUPERVISION: Laboratory leadership will ensure a minimum of one (1) technician remains on-site during length of installation project to safeguard/secure laboratory supplies.

5 ENVIRONMENTAL COMPLIANCE: The contractor shall be responsible for the removal and disposal of any hazardous waste products generated by the contractor. The contractor shall prepare all disposal manifests for hazardous wastes or chemicals. Contractor shall obtain signatures from the Base Environmental office of any hazardous waste disposal prior to removal of items off the base.

5.1 HAZMAT PHARMACY REGISTRATION: The contractor shall register with the base

HAZMAT Pharmacy any chemicals, etc., that are brought onto the base. Contractor shall comply with all Hazmat Pharmacy requirements and procedures.

6 GENERAL INFORMATION: When the contractor representative is working in a potential infectious area of the laboratory, they must wear a lab jacket/coat and gloves which will be provided.

6.1 CLEAN- UP. The contractor shall maintain the work area in a clean condition by the end of the contract performance period. Contractor shall dispose of all waste at an off-base permitted disposal site.

6.2 APPEARANCE: Contractor’s personnel shall present a neat appearance and be easily recognized. This may be accomplished by wearing distinctive clothing bearing the name of the company or by wearing appropriate badges, which contain the company and employee’s name. The wear of open-toed shoes is prohibited for infection control purposes.

6.3 CONTROLLED UNCLASSIFIED INFORMATION (CUI). The contractor shall comply with DoDI 5200.48 DAFI 16-1403, Controlled Unclassified Information (CUI), requirements. This publication establishes policy, assigns responsibilities, and prescribes procedures for CUI throughout the DoD and establishes the official DoD CUI Registry.

7. PERFORMANCE OBJECTIVES

Performance Objectives PWS paragraph Performance Threshold

Equipment maintained in serviceable condition

- Efficient processing of blood samples with few errors

2.4

1.1

98%

98%

Attachment 1 - SOW Page 4 of 5

Contractor Compliance:

- Compliance with calls for emergency service

- Compliance with timely supply of reagents

- Continuity in training of users, 1 trained per year

2.4

2.5

2.6

100%

Attachment 1 - SOW Page 5 of 5

File details come from the government source that posted it. Updated .