This document is an appendix that outlines the device accuracy and calibration requirements for the CSD Governor Test Stand. It provides the minimum required accuracy for various devices such as pressure gauges, pressure transducers, temperature measurement devices, and RPM sensors, along with the instruments used by PMEL to calibrate the existing test stand configuration. The document is in support of a federal contract opportunity for the CSD Governor Test Stands Upgrade, which is a 100% Small Business Set-aside solicitation issued by the Department of the Air Force Materiel Command Air Force Sustainment Center. The purpose of this draft solicitation is to allow potential contractors to submit questions, and a Rough Order of Magnitude (ROM) quote is not required at this time.

View the file

Other files for this federal contract opportunity

Show all 18

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

CSD Governor Test Stand Device Accuracy and Calibration Requirements

The minimum required accuracy is shown for each device along with the instrument that PMEL uses to calibrate the existing test stand configuration.

Device Range Minimum Required Accuracy

Existing PMEL Calibration Instrument

Underdrive Analog Pressure Gauge

0-400 PSI ± 2% full scale XP2i Crystal Gauge

Overdrive Analog Pressure Gauge

0-400 PSI ± 2% full scale XP2i Crystal Gauge

Underdrive Pressure Transducer

0-400 PSI ± 2% full scale XP2i Crystal Gauge

Overdrive Pressure Transducer

0-400 PSI ± 2% full scale XP2i Crystal Gauge

Charge Pressure Transducer

0-600 PSI ± 2% full scale XP2i Crystal Gauge

Lube Pressure Transducer

0-30 PSI ± 2% full scale XP2i Crystal Gauge

Oil Temperature Measurement Devices

50-300F ± 2% full scale Fluke 5520A

AC Voltage transducers

0-120VAC ± 2% full scale Fluke 5520A

DC current transducers

0-100mA ± 0.5% full scale Fluke 5520A

RPM 0-8000RPM ± 7 RPM Non-contact tachometer