36E77618R0094-005.pdf
PDF 1 MB Posted
- Attached to
- Prepare Digital Radiography Rooms Site Prep - South Dakota Sites (Sioux Falls, Fort Meade and Hot Springs)
- Solicitation number
- 36E77618R0094
About this file
36E77618R0094 ATTACHMENT 3 - SOW Appendix 2 - Fort Meade - Samsung Equip Dwgs - 6-14-17.pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36E77619C0014-000.docx | DOCX document | |
| 36E77618R0094-0002002.docx | DOCX document | |
| 36E77618R0094-0002000.docx | DOCX document | |
| 36E77618R0094-0002001.docx | DOCX document | |
| 36E77618R0094-0001001.docx | DOCX document | |
| 36E77618R0094-0001000.docx | DOCX document | |
| 36E77618R0094-008.zip | ZIP file | |
| 36E77618R0094-010.zip | ZIP file | |
| 36E77618R0094-009.zip | ZIP file | |
| 36E77618R0094-011.docx | DOCX document | |
| 36E77618R0094-004.pdf | ||
| 36E77618R0094-007.pdf | ||
| 36E77618R0094-001.docx | DOCX document | |
| 36E77618R0094-002.docx | DOCX document | |
| 36E77618R0094-003.docx | DOCX document | |
| 36E77618R0094-012.docx | DOCX document | |
| 36E77618R0094-013.docx | DOCX document | |
| 36E77618R0094-006.pdf | ||
| 36E77618R0094-000.docx | DOCX document |
Show all 19
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
- SITE PROGRESS CHECKLIST -
REVIEW EQUIPMENT ORDER FOR EXACT ITEMS PURCHASED. OPTIONAL/-
FUTURE ITEMS NOT ON ORDER MAY BE INDICATED ON THESE PLANS.
ALL ROOM DIMENSIONS ARE CRITICAL! IMMEDIATELY CONTACT GENE NOWACZYK
AT RADSOURCE IF CHANGES OCCUR OR DIMENSIONS ARE NOT CORRECT.
CONTACT A RADIATION PHYSICIST OR CONSULTANT TO SPECIFY
REQUIREMENTS FOR RADIATION CONTAINMENT.
PROVIDE A LOCKABLE EQUIPMENT HOLDING AREA CLOSE TO THE
INSTALLATION FOR STORING TOOLS AND TEST EQUIPMENT.
MAKE SURE A DUST FREE, TEMPERATURE AND HUMIDITY CONTROLLED
ENVIRONMENT IS AVAILABLE FOR STORING THE EQUIPMENT IF YOUR SITE
IS NOT READY FOR INSTALLATION AT THE TIME OF DELIVERY. ONCE THE
SITE IS PREPARED, YOU ARE THEN RESPONSIBLE FOR DELIVERING THE
MAKE ARRANGEMENTS FOR ANY RIGGING, SPECIAL HANDLING, OR FACILITY
MODIFICATIONS THAT MUST BE MADE IN ORDER FOR THE EQUIPMENT TO BE
DELIVERED TO THE INSTALLATION SITE. IF DESIRED, YOUR LOCAL RADSOURCE
TEAM REPRESENTATIVE CAN SUPPLY A REFERENCE LIST OF RIGGERS.
EQUIPMENT TO THE SITE.
CUSTOMER
ALL REQUIREMENTS FOR THIS
EQUIPMENT ARE NOT NOTED ON THIS
SHEET. IT IS SUGGESTED THAT THESE
DRAWINGS BE REVIEWED BY QUALIFIED
PROFESSIONALS WHO CAN ASSIST
WITH MAKING DECISIONS REGARDING
RADIATION CONTAINMENT, MAGNETIC
FIELD CONTAINMENT, ELECTRICAL,
STRUCTURAL AND MECHANICAL
REQUIREMENTS. ALTHOUGH THE
EQUIPMENT MAY BE INSTALLED IN AN
EXISTING ROOM OF SIMILAR FUNCTION,
REQUIREMENTS STILL NEED TO BE
CHECKED.
SITE VISIT
NORMAL PROCEDURE IS TO ATTEND AN
INITIAL CONSTRUCTION MEETING AND
THEN PERFORM AN INSPECTION OF
THE ELECTRICAL BUILD OUT AT OR
ABOUT THE SAME TIME AS THE
ELECTRICAL ROUGH IN (PRIOR TO
SHEETROCK OR LEAD BEING
INSTALLED), ADDITIONAL VISITS MAY BE
REQUIRED
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT MANAGER
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
RADIOLOGY MANAGER
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE:
DRAWING SHEET #
NOTED
A.W.
COVER
6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
INSTALL LOCATION MAP
SAMSUNG
GC85
~ DRAWING INDEX ~
A1.1 EQUIPMENT LAYOUT ROOM 146
A1.2 EQUIPMENT LAYOUT ROOM 164
AN NOTES
E1 ELECTRICAL LAYOUT GC85
ROOM# 164
E2 ELECTRICAL LAYOUT GC85
ROOM# 146
E1.1 ELECTRICAL DETAILS
E1.3 MAIN PANEL DETAILS
S1 OVERHEAD STRUCTURAL
LAYOUT ROOM# 164
S2 OVERHEAD STRUCTURAL
LAYOUT ROOM# 146
S1.1 STRUCTURAL NOTES GENERAL
CONNECTION DETAILS
Q1 EQUIPMENT DETAILS GC85
QE1 EQUIPMENT CABLE LENGTH GC85
PROJECT MANAGER: SHAWN SALANDER
5257 27TH ST S. SUITE 101
FARGO,NORTH DAKOTA
58104
OFFICE PHONE 701-297-0305
CELL PHONE 701-371-0505
E-MAIL SSALANDER@IMAGINGSOLUTIONSINC.COM
RECEIPT OF DRAWING FINAL
THIS SIGNATURE PRINTED OR OTHERWISE
REPRESENTS RECEIPT OF THIS SET OF
PLANS, IT IS UNDERSTOOD THAT ANY
DEVIATION FROM THESE DRAWINGS,
DETAILS AND SPECIFICATIONS MAY
ENCROACH UPON THE EQUIPMENT
OPERATION, SERVICEABILITY, OR SAFETY
GUIDELINES. ALSO I AM AWARE THAT ANY
CHANGES MADE AFTER 16 June 2017
COULD RESULT IN ADDITIONAL EXPENSES
BEING INCURRED IF RADSOURCE IS NOT
NOTIFIED IN WRITING i.e. E-MAIL ETC... NEW
SHEETS WILL BE ISSUED AS ADDITIONS
AND REVISIONS WILL HAVE CURRENT
DATES ADDED.
NAME: NICHOLAS A. CHRISTENSEN
DATE:_______
PRINTED:________________________________
COMPANY:______________________________
NAME: KEVIN FRAVEL
DATE:_______
PRINTED:________________________________
COMPANY:______________________________
NAME:ANTHONY KOESSL
DATE:_______
PRINTED:________________________________
COMPANY:______________________________
NAME:______________________DATE:_______
PRINTED:________________________________
COMPANY:______________________________
VA BLACK HILLS HEALTH CARE SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
LAST PLOT DATE: 16 June 2017
LAST REVIEW: 6/16/2017 10:36:27 AM
3X4 14'-0"3" 5'-0"
1'-8"
3'-4"
2' -1
01 2"
5'-5"
3' -1
01 2"
7'-51 4"
1' -1
1 2"
1'-7"
4"3'-53
4"3'-03 4"
'-8
4' -8
4'
-8
4"
ROOM LAYOUT - RADIOLOGY 164
SCALE: 1/2" = 1'-0"
MINIMUM CEILING HEIGHT: 9'-2" - 9'-5"ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT
PATIENT TABLE
WALL / CHEST STAND
440LBS
330 LBS
2,046 BTU
511 BTU
3 1 GENERATOR CABINET 220 LBS 11,798 BTU
4 1 CEILING RAIL 140 LBS 3400 BTU
5 1 CEILING MOUNTED TUBE HEAD
6 1 WORK STATION 35 LBS 1,364 BTU
ENVIRONMENT
AMBIENT OPERATION TEMPERATURE: 55-75° (F)
ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR
HUMIDITY: 20-80 PERCENT NON CONDENSING
ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR
EQUIPMENT LEGEND SAMSUNG XGEO-GC85
PROVIDED BY INSTALLED BY
OWNER
OWNER
NOTE:
SPECIAL DELIVERY TO BUILDING NOT COVERED BY SAMSUNG , TO BE
COORDINATED BETWEEN "VA BLACK HILLS HEALTH CARE SYSTEM
AND/OR N/A" AND SAMSUNG
OWNER
OWNER
OWNER
OWNER
616 LBS 1,909 BTU
7 1 STITCHING STAND 235 LBS N/A BTU OWNER
OWNER: VA BLACK HILLS HEALTH CARE SYSTEM
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
3'1' 2' 4'
1m
5' 7'
2m
6' 8' 10'
3m
9' DRAWING SHEET #
A1.2
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
NOTE:
ARCHITECT / CONTRACTOR IS TO VERIFY ALL
DIMENSIONS ON THIS DRAWING AGAINST AS BUILT.
DIMENSIONS ILLUSTRATED WITH IN HAVE BEEN
USED FOR PROPER EQUIPMENT PLACEMENT AND
MAY BE USED IN THE FUTURE FOR RADIATION
SHIELDING PROTECTION BY THE CUSTOMERS
SELECTED PHYSICIST.
EACH RESPECTIVE ROOM WILL HAVE AN ENLARGED
VIEW ILLUSTRATING THE PLACEMENT OF THE
RADIOLOGY EQUIPMENT FOR THE ASSOCIATED
ROOM IDENTIFICATION AS SHOWN ON SHEETS
LABELED WITH THE LETTER "AQ-?".
SUPPLEMENTAL NOTES:
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED /
DESIGNED BY THE ARCHITECT N/A. THE ARCHITECT
N/A SHALL CONSULT WITH VA BLACK HILLS HEALTH
CARE SYSTEM FOR ANY ADDITIONAL INFORMATION.
· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,
MAMMO, R/F, CT, ETC., SHALL HAVE " LED" CAN
LIGHTING AND/OR FLORESCENT LIGHT
FIXTURES. THE LED FIXTURES ARE TO BE
DISTRIBUTED AROUND THE ROOM AND SHOULD
BE PROVIDED WITH DIMMER CONTROLS. THE
DIMMER CONTROLS SHOULD BE LOCATED NEAR
THE ENTRANCES AND IN THE CONTROL AREAS.
· IT IS RECOMMENDED THAT ALL DIAGNOSTIC
ROOMS SHOULD HAVE TILE OR VINYL
FLOORING.
· A TILE DROP CEILING IS RECOMMENDED FOR
ANY DIAGNOSTIC ROOMS. MINIMUM CEILING
HEIGHT 9'-2" - 9'-5"
· ALL CABINETRY WORK IS TO BE DESIGNED BY
THE ARCHITECT AND PROVIDED BY VA BLACK
HILLS HEALTH CARE SYSTEM AND OR N/A .
RECOMMEND ALL COUNTERS SHALL HAVE
ROUNDED CORNERS.
· PROVIDE FOR REFUSE REMOVAL AND DISPOSAL
(E.G. CRATES, CARTONS, PACKING)
· PROVIDE A WORKING RESTROOM WITH
SUPPLIES WITH IN THE FACILITY
NOTE:
THIS SAMSUNG GC85 ROOM CONFIGURATION
UTILIZES A STANDARD THREE (3) METER TRANSVERSE
BRIDGE AND STANDARD FOUR (4) METER LONGITUDE
RAILS OTHER VARIATIONS ARE AVAILABLE AT AN
ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT
YOUR SALES PERSONAL FOR INFORMATION.
COUNTER TOP IN OPERATOR CONTROL AREA SHOULD
BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
RADIOLOGY MANAGER
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
A.W.
NOTED
5X4
1'-8"3'-4"
3' -0
4" 8"
18'-8"
3"
'-8
1 2"
4"
4' -8
4'
-8
6' -1
4'
-7 1 2"
4"
1' -2
1 2"
3'-73 4" 4'-93
4" 3'
-5 1 2"
10'-51 2"
1' -2
1 2"
ROOM LAYOUT - RADIOLOGY 146
SCALE: 1/2" = 1'-0"
MINIMUM CEILING HEIGHT: 9'-2" - 9'-5"ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT
PATIENT TABLE
WALL / CHEST STAND
440LBS
330 LBS
2,046 BTU
511 BTU
3 1 GENERATOR CABINET 220 LBS 11,798 BTU
4 1 CEILING RAIL 140 LBS 3400 BTU
5 1 CEILING MOUNTED TUBE HEAD
6 1 WORK STATION 35 LBS 1,364 BTU
ENVIRONMENT
AMBIENT OPERATION TEMPERATURE: 55-75° (F)
ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR
HUMIDITY: 20-80 PERCENT NON CONDENSING
ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR
EQUIPMENT LEGEND SAMSUNG XGEO-GC85
PROVIDED BY INSTALLED BY
OWNER
OWNER
NOTE:
SPECIAL DELIVERY TO BUILDING NOT COVERED BY SAMSUNG , TO BE
COORDINATED BETWEEN "VA BLACK HILLS HEALTH CARE SYSTEM
AND/OR N/A" AND SAMSUNG
OWNER
OWNER
OWNER
OWNER
616 LBS 1,909 BTU
7 1 STITCHING STAND 235 LBS N/A BTU OWNER
OWNER: VA BLACK HILLS HEALTH CARE SYSTEM
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
SAMSUNG
3'1' 2' 4'
1m
5' 7'
2m
6' 8' 10'
3m
9' DRAWING SHEET #
A1.1
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
NOTE:
ARCHITECT / CONTRACTOR IS TO VERIFY ALL
DIMENSIONS ON THIS DRAWING AGAINST AS BUILT.
DIMENSIONS ILLUSTRATED WITH IN HAVE BEEN
USED FOR PROPER EQUIPMENT PLACEMENT AND
MAY BE USED IN THE FUTURE FOR RADIATION
SHIELDING PROTECTION BY THE CUSTOMERS
SELECTED PHYSICIST.
EACH RESPECTIVE ROOM WILL HAVE AN ENLARGED
VIEW ILLUSTRATING THE PLACEMENT OF THE
RADIOLOGY EQUIPMENT FOR THE ASSOCIATED
ROOM IDENTIFICATION AS SHOWN ON SHEETS
LABELED WITH THE LETTER "AQ-?".
SUPPLEMENTAL NOTES:
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED /
DESIGNED BY THE ARCHITECT N/A. THE ARCHITECT
N/A SHALL CONSULT WITH VA BLACK HILLS HEALTH
CARE SYSTEM FOR ANY ADDITIONAL INFORMATION.
· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,
MAMMO, R/F, CT, ETC., SHALL HAVE " LED" CAN
LIGHTING AND/OR FLORESCENT LIGHT
FIXTURES. THE LED FIXTURES ARE TO BE
DISTRIBUTED AROUND THE ROOM AND SHOULD
BE PROVIDED WITH DIMMER CONTROLS. THE
DIMMER CONTROLS SHOULD BE LOCATED NEAR
THE ENTRANCES AND IN THE CONTROL AREAS.
· IT IS RECOMMENDED THAT ALL DIAGNOSTIC
ROOMS SHOULD HAVE TILE OR VINYL
FLOORING.
· A TILE DROP CEILING IS RECOMMENDED FOR
ANY DIAGNOSTIC ROOMS. MINIMUM CEILING
HEIGHT 9'-2" - 9'-5"
· ALL CABINETRY WORK IS TO BE DESIGNED BY
THE ARCHITECT AND PROVIDED BY VA BLACK
HILLS HEALTH CARE SYSTEM AND OR N/A .
RECOMMEND ALL COUNTERS SHALL HAVE
ROUNDED CORNERS.
· PROVIDE FOR REFUSE REMOVAL AND DISPOSAL
(E.G. CRATES, CARTONS, PACKING)
· PROVIDE A WORKING RESTROOM WITH
SUPPLIES WITH IN THE FACILITY
NOTE:
THIS SAMSUNG GC85 ROOM CONFIGURATION
UTILIZES A STANDARD THREE (4) METER TRANSVERSE
BRIDGE AND STANDARD FOUR (5) METER LONGITUDE
RAILS OTHER VARIATIONS ARE AVAILABLE AT AN
ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT
YOUR SALES PERSONAL FOR INFORMATION.
COUNTER TOP IN OPERATOR CONTROL AREA SHOULD
BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
RADIOLOGY MANAGER
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
MINIMUM SITE PREPARATION REQUIREMENTS
WALLS TO BE PAINTED OR COVERED, BASEBOARDS INSTALLED, FLOORS TO BE
TITLED AND OR COVERED, CEILING SHALL HAVE GRID TILES AND LIGHTING
FIXTURES INSTALLED.
DOORS AND WINDOWS, ESPECIALLY RADIATION PROTECTION BARRIERS,
INSTALLED AND FINISHED WITH LOCK SETS OPERATIONAL.
ALL ELECTRICAL CONVENIENCE, CONDUIT, RACEWAY AND JUNCTION BOXES
INSTALLED.
INCOMING MAINS POWER OPERATIONAL AND CONNECTED TO X-RAY ROOM
BREAKER.
115 VOLTS CONVENIENCE OUTLETS OPERATIONAL.
ALL SUPPORT STRUCTURES CORRECTLY INSTALLED.
ALL CHANNELS, PIPES, BEAM'S AND OR OTHER SUPPORTING DEVICES SHOULD
BE LEVEL, PARALLEL AND FREE OF LATERAL OR LONGITUDINAL
MOVEMENTS.
ALL CONTRACTOR SUPPLIED CABLES PULLED AND TERMINATED.
A DUST FREE ENVIRONMENT IN AND AROUND THE PROCEDURE ROOM.
ALL HEATING AND VENTILATION / AIR-CONDITIONING INSTALLED AND
OPERATIONAL.
A CLEAR DOOR OPENING FOR MOVING EQUIPMENT INTO THE BUILDING MUST
BE 42 IN. X 82 IN. OR LARGER CONTINGENT ON AN 8' CORRIDOR WIDTH.
NOTICE:
THIS DRAWING SET IS THE SOLE PROPERTY OF
NOWACZYK / PEMBER & ASSOCIATES
MEP / ARCHITECTURAL ENGINEERS
ITS USE IS AUTHORIZED ONLY FOR THE CUSTOMER VA
BLACK HILLS HEALTH CARE SYSTEM/CONTRACTOR N/A TO
DESIGN AND INCORPORATE OUR CONCEPT INTO
CONSTRUCTION AND PREPARATION FOR IMAGING
EQUIPMENT INSTALLATION. THESE DRAWINGS SHALL NOT
BE USED FOR ANY PURPOSE OTHER THAN THE AGREED
UPON DESIGN BETWEEN NEUROLOGICA AND THEIR
CUSTOMER.
THE CUSTOMER VA BLACK HILLS HEALTH CARE SYSTEM IS
RESPONSIBLE FOR ALL ROOM PREPARATION COSTS, FEES,
PERMITS AND INSPECTIONS UNLESS OTHERWISE
SPECIFIED IN THE GENERAL ORDER FOR THE EQUIPMENT
PURCHASE.
RADIATION SHIELDING IS NOT SHOWN ON THESE PLANS.
HOWEVER, THE CUSTOMER VA BLACK HILLS HEALTH CARE
SYSTEM SHALL, AT THEIR OWN EXPENSE, HAVE SHIELDING
CALCULATIONS FOR THE ROOM PREPARED BY A LICENSED
RADIATION PHYSICIST.
PREFACE
THESE DRAWINGS HAVE BEEN PREPARED BY
NEUROLOGICA , THEIR PURPOSE IS TO PROVIDE THE
SITING, ELECTRICAL, MECHANICAL AND ENVIRONMENTAL
SPECIFICATIONS REQUIRED TO ACCOMMODATE THE
INSTALLATION AND OPERATION OF THE DIAGNOSTIC
IMAGING EQUIPMENT AND SUB-COMPONENTS
ILLUSTRATED.
THE LAYOUT(S) PROVIDED FOR ALL COMPONENTS
PURCHASED FROM AND/OR PROVIDED BY THE EQUIPMENT
MANUFACTURER SHALL SERVE AS A GUIDE FOR
INSTALLATION BY THE LOCAL SERVICE/INSTALLATION
REPRESENTATIVES.
INFORMATION IN THESE DRAWINGS RELATING TO
BUILDING/FACILITY SPECIFICATIONS THAT WILL SUPPORT
IMAGING EQUIPMENT COMPONENTS SUCH AS ELECTRICAL,
STRUCTURAL, MECHANICAL AND ENVIRONMENTAL
REQUIREMENTS SHALL BE UTILIZED AS A GUIDE BY THE
CUSTOMER'S ARCHITECT N/A AND/OR CONTRACTOR N/A.
FACILITY CONDITIONS MAY DICTATE ACTUAL
CONSTRUCTION METHODS AND MATERIALS APPLIED.
HOWEVER, ALL METHODS AND MATERIALS MUST COMPLY
WITH EQUIPMENT MANUFACTURER SPECIFICATIONS, AS
WELL AS LOCAL AND/OR NATIONAL BUILDING CODES.
THE REQUIRED CEILING HEIGHT OF 9'-2" - 9'-5" INDICATED
ON THESE PLANS IS TO INSURE EQUIPMENT FUNCTION IS
NOT INHIBITED. CONSULT WITH YOUR EQUIPMENT
INSTALLATION SPECIALIST REGARDING ACCEPTABILITY OF
THE OTHER CEILING HEIGHTS. CHECK ALL DOOR
OPENINGS FROM DELIVERY LOCATION TO WHERE
EQUIPMENT IS TO BE INSTALLED.
THE QUALITY OF CONSTRUCTION METHODS, MATERIALS
AND CONFORMANCE TO EQUIPMENT MANUFACTURER
SPECIFICATIONS AND TOLERANCES, WILL AFFECT
EQUIPMENT PERFORMANCE.
MAGNETIC INTERFERENCE SPECIFICATIONS
DIGITAL FLAT PANEL MUST BE LOCATED IN AMBIENT STATIC
MAGNETIC FIELDS OF LESS THAN ONE GAUSS TO
GUARANTEE SPECIFIED IMAGING PERFORMANCE.
X-RAY TUBES AND EQUIPMENT MUST BE LOCATED IN
AMBIENT STATIC MAGNETIC FIELDS OF LESS THAN TEN
GAUSS TO GUARANTEE SPECIFIED PERFORMANCE DATA
INTEGRITY
SYSTEM ELECTRONICS / EQUIPMENT MUST BE LOCATED IN
AMBIENT STATIC MAGNETIC FIELDS OF LESS THAN TEN
GAUSS TO GUARANTEE DATA INTEGRITY
CONSOLE EQUIPMENT MUST BE LOCATED IN AMBIENT
STATIC MAGNETIC FIELDS OF LESS THAN ONE GAUSS TO
OBTAIN SPECIFIED GEOMETRIC LINEARITY.
SITE ENVIRONMENT SPECIFICATIONS
AMBIENT OPERATING TEMPERATURE: 55⁰ TO 75⁰ F (16⁰ TO
24⁰ C) MAXIMUM
ALLOWABLE TEMPERATURE CHANGE OF 15⁰ F, (5⁰ C) /
HOUR MAXIMUM ROOM
TEMPERATURE GRADIENT 5⁰ F, (3⁰ C)
HUMIDITY: 20 TO 80 PERCENT NON-CONDENSING, MAXIMUM
ALLOWABLE CHANGE OF 5 PERCENT / HOUR
ALTITUDE: NOT TO EXCEED 10,000 FT (3050M) ABOVE SEA
LEVEL
THE ENVIRONMENT FOR THE ELECTRONICS CABINET MUST
BE CONTROLLED SO THE ABOVE RESTRICTIONS ARE NOT
EXCEEDED.
DO NOT RESTRICT THE AIR INTAKE OR AIR EXHAUST OF THE
SYSTEM COMPONENTS.
ENVIRONMENTAL CONDITIONS LISTED ABOVE MUST BE
MAINTAINED AT ALL TIMES, INCLUDING OVERNIGHT AND
HOLIDAYS.
DRAWING SHEET #
AN
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
RADIOLOGY MANAGER
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
UB
FB
WL
OC
ST1
ACR
XG XT
E1.1
E1.1
6'-41 4"
8' -3
3 4"
3' -8
1 4"
9'-73 4"
2' -2
1 2"
4' -5
1 4"
1'-43 4"
4'-11"
ELECTRICAL LEGEND
MAIN DISCONNECT CONTRACTOR SUPPLIED, PART #
( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE
OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE
NOWACZYK AT 816-223-6378
480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC
CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM.
FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1)
NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM
BREAKER PER N.E.C.
INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'
OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL
CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT
LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,
TERMINATE FLEX WITH A 90DEG BOTH ENDS.
SEE SHEET E-3 FOR DETAILS
1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING,
PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING
ABOVE CEILING OUTLET.
ACR
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM
OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE
BASE
1 @ COVER
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
UB
SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX,
SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP
BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT
MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY
ELECTRICAL CONTRACTOR
ST1
WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND
INSTALLED BY CONTRACTORWL
XG
CBE1
1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT
COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"
PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)
INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER
LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL
CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND
CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.
1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT
MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET
(E1.3)XT
1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER
OF BOX TO 18.0" A.F.F.
1 @ COVER
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
OC
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF
BOX FLUSH TO FINISHED FLOOR
1 @ COVER
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
FB
110/115VAC 20 AMP
PHONE / DATA
PHONE
DATA
WALL MOUNTED PHONE 48"A.F.F.W
GROUND FAULT INTERRUPT RECEPTACLE
G.F.I.
SYMBOLS LEGEND
ABOVE CEILING CONDUIT
BELOW GRAD CONDUIT
3 x 10 HORIZONTAL WALL DUCT
3 x 10 FLUSH TO FINISHED FLOOR DUCT
ELECTRICAL NOTES
ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,
FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR
STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER,
SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR
DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING
OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS
MUST BE CONTINUOUS COPPER STRANDED AND FREE
FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN.
WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER
SIZES MAY BE REQUIRED BY LOCAL CODES.
IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,
AS REQUIRED IN ACCORDANCE WITH NATIONAL AND
LOCAL ELECTRICAL CODES.
LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO
THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT
AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE
HOSPITAL APPROVED OUTLET OR EQUIVALENT.
GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.
CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT
FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO
CEILING MOUNTING COMPONENTS AND WIRING IF HIGH
WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,
USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT
LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING
MOUNTED ACCESSORIES WILL BE PARKED.
ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST
RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE
NEED FOR GREATER THAN STANDARD CABLE LENGTHS
(REFER TO THE INTERCONNECTION DIAGRAM FOR
MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1
CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH
MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND
LOCAL ELECTRICAL CODES.
PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG
EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL
CONTRACTOR XXX WITH THE SUPERVISION OF A
RADSOURCE IMAGING REPRESENTATIVE. THE
RADSOURCE IMAGING REPRESENTATIVE WOULD BE
REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION
LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG
EQUIPMENT.
ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE
OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED
FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL
PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW
CONNECTIONS.
LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT
ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED
INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT
OR CONDUCTORS WILL BE ALLOWED.
ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL
CONDUIT (FMC)
ELECTRICAL LAYOUT SAMSUNG GC85
SCALE: 3/8" = 1'-0"
MINIMUM CEILING HEIGHT: 9'-2" - 9'-5"
REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR
1 CBE1 TO XG
CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @
#3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS
AT "XG". LABEL "FOR X-RAY POWER"
2 CBE1 T0 ST1
CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__
CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE
3 XG TO OC
2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW
GRAD IS ACCEPTABLE
4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND
5 N/A TO N/A
1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,
LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR
WARNING LIGHT"
6 XG TO FB 2.0" CONDUIT BELOW GRAD
7 XG TO UB 2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW
GRAD IS ACCEPTABLE
8 INCOMING
MAINS
TO CBE1 CONDUIT AS REQUIRED
SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED
NEA-43080-5W-ST24DC-SPD-WLC-FC
MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT
TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA
SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH
COVER 15.4"w x 32.4"h
NEA-ESTOP22 EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS
NEA-ESTOPCV EMERGENCY STOP COVER PVC
NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY
XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT
W1035-KIT-GC
WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES
10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X
12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED
JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL
JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL
DRAWING SHEET #
E1
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
FB
WL
OC
ST1
ACR
XG XT
E1.1
E1.1
8'-10"
8' -9
1'-4"
5'-23 4"1'
-1 13 4"
3' -8
1 2"
13'-21 2"
9' -8
1 4"
ELECTRICAL LEGEND
MAIN DISCONNECT CONTRACTOR SUPPLIED, PART #
( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE
OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE
NOWACZYK AT 816-223-6378
480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC
CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM.
FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1)
NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM
BREAKER PER N.E.C.
INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'
OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL
CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT
LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,
TERMINATE FLEX WITH A 90DEG BOTH ENDS.
SEE SHEET E-3 FOR DETAILS
1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING,
PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING
ABOVE CEILING OUTLET.
ACR
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM
OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE
BASE
1 @ COVER
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
UB
SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX,
SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP
BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT
MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY
ELECTRICAL CONTRACTOR
ST1
WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND
INSTALLED BY CONTRACTORWL
XG
CBE1
1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT
COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"
PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)
INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER
LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL
CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND
CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.
1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT
MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET
(E1.3)XT
1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER
OF BOX TO 18.0" A.F.F.
1 @ COVER
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
OC
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF
BOX FLUSH TO FINISHED FLOOR
1 @ COVER
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
FB
110/115VAC 20 AMP
PHONE / DATA
PHONE
DATA
WALL MOUNTED PHONE 48"A.F.F.W
GROUND FAULT INTERRUPT RECEPTACLE
G.F.I.
SYMBOLS LEGEND
ABOVE CEILING CONDUIT
BELOW GRAD CONDUIT
3 x 10 HORIZONTAL WALL DUCT
3 x 10 FLUSH TO FINISHED FLOOR DUCT
ELECTRICAL NOTES
ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,
FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR
STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER,
SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR
DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING
OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS
MUST BE CONTINUOUS COPPER STRANDED AND FREE
FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN.
WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER
SIZES MAY BE REQUIRED BY LOCAL CODES.
IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,
AS REQUIRED IN ACCORDANCE WITH NATIONAL AND
LOCAL ELECTRICAL CODES.
LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO
THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT
AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE
HOSPITAL APPROVED OUTLET OR EQUIVALENT.
GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.
CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT
FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO
CEILING MOUNTING COMPONENTS AND WIRING IF HIGH
WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,
USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT
LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING
MOUNTED ACCESSORIES WILL BE PARKED.
ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST
RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE
NEED FOR GREATER THAN STANDARD CABLE LENGTHS
(REFER TO THE INTERCONNECTION DIAGRAM FOR
MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1
CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH
MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND
LOCAL ELECTRICAL CODES.
PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG
EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL
CONTRACTOR XXX WITH THE SUPERVISION OF A
RADSOURCE IMAGING REPRESENTATIVE. THE
RADSOURCE IMAGING REPRESENTATIVE WOULD BE
REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION
LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG
EQUIPMENT.
ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE
OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED
FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL
PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW
CONNECTIONS.
LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT
ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED
INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT
OR CONDUCTORS WILL BE ALLOWED.
ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL
CONDUIT (FMC)
ELECTRICAL LAYOUT SAMSUNG GC85
SCALE: 3/8" = 1'-0"
MINIMUM CEILING HEIGHT: 9'-2" - 9'-5"
REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR
1 CBE1 TO XG
CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @
#3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS
AT "XG". LABEL "FOR X-RAY POWER"
2 CBE1 T0 ST1
CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__
CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE
3 XG TO OC
2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW
GRAD IS ACCEPTABLE
4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND
5 N/A TO N/A
1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,
LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR
WARNING LIGHT"
6 XG TO FB 2.0" CONDUIT BELOW GRAD
7 XG TO UB 2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW
GRAD IS ACCEPTABLE
8 INCOMING
MAINS
TO CBE1 CONDUIT AS REQUIRED
SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED
NEA-43080-5W-ST24DC-SPD-WLC-FC
MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT
TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA
SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH
COVER 15.4"w x 32.4"h
NEA-ESTOP22 EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS
NEA-ESTOPCV EMERGENCY STOP COVER PVC
NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY
XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT
W1035-KIT-GC
WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES
10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X
12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED
JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL
JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL
DRAWING SHEET #
E2
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
DAT. 0'-0"
FINISHED FLOOR
DAT. 9'-4"
FINISHED CEILING
1'-117 8"
2 UPRIGHT ELEVATION
SCALE: 3/8" = 1'-0"
1 JUNCTION BOX ELEVATION
SCALE: 3/8" = 1'-0"
XG
XT
CBE1
DAT. 0'-0"
FINISHED FLOOR
DAT. 9'-4"
FINISHED CEILING
1' -0
"1'-4"
FIT INTO NEAREST
STUD POCKET
6"
14.5" 3.5"
1' -0
SUPPLIED
DIVIDER
TO
A
B O
V E
F
IN
IS
H
E D
C E
IL
IG
SUPPLIED
DIVIDER
10."3.5"
XG
XT
CONNECT 2.0" CND
FOR INCOMING LINE
FEED TO CBE1
JUNCTION BOX/DUCT DETAILS
SCALE: NTS
DRAWING SHEET #
E1.1
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
AC
In Use Sign
X-RAY
Generator Digital
X-RAY
PDU
Imaging Interface CabinetInternal
Controls Local Indication
Emergency Power Off(EPO)
DC
Ethernet
Main Circuit
Breaker
A
B
C
GND
A
B
C
GND
MAIN DISCONNECT ( NEA-43080-4W-ST24DC-SPD-WLC-FC)
Class 2
24VDC
Class 2
24VDC
Contractor Supplied 480V ac, 3 Phase +
GND.
50/60Hz
Phase Type
SPD
ST24DC
(Dry Contact)
24VDC
INCHESDUAL DIMENSIONS:
MILLIMETERS
H150 F OR S
J250 F OR S AND CJ250NS
NEMA/EEMAC TYPE 1
CONDUCTORS
PER
TERMINAL
CIRCUIT
BREAKER
CATALOG NO
PREFIX
HD
JD
WIRING
DIA
A,B A,B
AMPERE
RATING
15-150 150-250
FIELD INSTALLABLE CIRCUIT BREAKER DATA
UL LISTED INTERRUPTING RATING
RMS SYMMETRICAL AMPERES
AC VOLTAGE DC VOLTAGE
240 480Y/277
25K 25K
480 600Y/
600 125 250 500
HG
JG
HJ
JJ
HL
JL
A,B A,B A,B A,B A,B A,B
15-150 150-250 15-150
150-250 15-150
150-250 125K 125K
65K 65K
100K 100K
18K 18K 35K 35K 65K 65K 65K 65K
65K 65K 65K 65K
35K 35K
18K 18K
14K 14K 18K 18K 25K 25K 25K 25K
25K 25K 25K 25K
18K 18K
14K 14K
20K 20K
20K 20K 20K 20K
20K 20K
20K 20K
20K 20K
20K 20K
20K 20K
WIRING DIAGRAM
A B S N N
S
FIELD INSTALLABLE SOLID NEUTRAL ASSEMBLIES AND GROUND KIT
CATALOG
NUMBER
SN100FA 1
MAXIMUM
AMPERE
RATING
NUMBER OF
TERMINALS
100A
PK0GTJ250
WIRE SIZE
(AWG/KCMIL)
MAX MIN
SN400LA 250A
250A
2 1
OR 2
#4
#1/0
#1/0 #14
#12
#1 #1
#600 #250
#6
#300
#300
MM
IN MM
DIAMETER
1.38 35
2.00 51
2.50
3.63 92
3.00 76
IN
SYMBOL
1.00C
1.50
2.00
2.50
3.00
E
F
G
H
CONDUIT SIZE
KNOCKOUTS
J250S 14.35
J250F 32.50
J250S 31.43
27.00
11.90
J250F 15.42
5.96
1.85
31.00
2.85
25.30
2.85
14.00
9.25.25
1.82
2.37
C,E,F,G,H
C,E,F,G,H
C,E,F,G,H
DASHED LINES
INDICATE
SURFACE COVER
(2 HOLES)
.38
[10].25 DIA
[6]
.50 DIA
[12]
.28 DIA
[7]
(2 HOLES)
AA
SEISMIC NOTES:
USE (4) 1/4" DIA GRADE 5 STEEL MOUNTING BOLTS @ HOLES AA
MAX CONFIGURED WEIGHT 41 LB FOR THE PURPOSE OF DETERMINING
SEISMIC ANCHORAGE REQUIREMENTS. FOR ALL OTHER APPLICATIONS,
CONTACT SCHNEIDER ELECTRIC.
2.98
FROM
MTG
SURFACE
5.95
13.50
CG
CG
2.18
13.50
1.23
CIRCUIT BREAKER TERMINAL LUG DATA
15-150
150-175
RATING
AMPERE
JD, JG
CATALOG NO.
HD, HG
PREFIX
CIRCUIT
BREAKER
#3/01
#4/0 #4
MAX
(AWG/KCMIL)
WIRE SIZE
PER PHASE
CONDUCTORS
#14
MIN
TYPE
HJ, HL, HR
AL/CU
JJ, JL, JR AL/CU
200-250
350 #3/0
PK0GTH150 150A 2 1 #14#2
HR
JR
A,B A,B
15-150 150-250
125K 125K
65K 65K
65K 65K
25K 25K
25K 25K
20K 20K
20K 20K
CJ250NS AND H150 COVER OPENING
J250 COVER OPENING
WITH TRIP SETTING ACCESS
3.102.08
.25
.08
4.00 2.08
.25 .08
1.00
4.25
1.25
HAS
OFFSET
FOR 200% NEURALS USE SN400LA (AMP RANGE 100 TO 250,
CU WIRE ONLY).
SEE COVER
OPENING DETAILS
TYPE
CU
AL
AL OR CU
AL OR CU
AL OR CU
AL OR CU
AL OR CU
DRAWING SHEET #
E1.3
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
WALL STAND
TABLE
CABINET
CEILING
WORK STATION
POWER
CABINET WALLSTAND CABINET CABINETTABLE
WIRING : WALL or FLOOR WIRING : FLOOR WIRING : FLOOR
CEILINGCABINET
WIRING : WALL AND CEILING
WIRING : FLOOR
WORK-CABINET
STATION
3.9 m
& CIB
13.1 m
11.2 m
STANDTABLE
WIRING : FLOOR
13 m
WALL-
8.5 m
WALLSTAND CEILING
WIRING : FLOOR
13.65 m
7.85 m
CONTROL ROOM
EXAM ROOM
DISTRIBUTION
POWER
DISTRIBUTION
DRAWING SHEET #
QE1
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
TABLE DETAIL DRAWING
FRONT VIEW SIDE VIEW
MIN MAX
TOP VIEW
95.73
67.52
31.73
.4
6 35 .9
10.71
95.58 18.98
5.
.6
WALL STAND DETAIL DRAWING
FRONT VIEW SIDE VIEW
TOP VIEW
MIN
MAX
21.97
.5
.1
.4
20°
31.74
.9
CABINET DETAIL DRAWING
FRONT VIEW SIDE VIEW TOP VIEW
19.29
.7
29.53
.7
29.53
.7
THU DETAIL DRAWING
FRONT VIEW SIDE VIEW
TOP VIEW
Focal Spot
Rotation Conter
Focal Spot
118.50
.3
.3
DRAWING SHEET #
Q1
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
STRUCTURAL LAYOUT SAMSUNG GC85
SCALE: 3/8" = 1'-0"
RECOMMENDED CEILING HEIGHT 9'-4.0"
A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY
UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO
WHICH THE STATIONARY CEILING TRACKS ARE FASTENED.
THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO
ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH
INSTALLATION.
THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE
SPECIFIED ON SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE
OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND
MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS,
ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE
CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED TO MEET SPECIFIC SITE
REQUIREMENTS BY THE ENGINEER OF RECORD.
CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN.
STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS.
THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT
MOVEMENT IN ANY DIRECTION. THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE
BUILDING STRUCTURE TO HOLD THE DEAD WEIGHTS INDICATED. see EQUIPMENT LEGEND "SHEET A1"
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL
PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE
REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING
DRAWINGS IS UNDERTAKEN.
1. CUSTOMER VA BLACK HILLS HEALTH CARE SYSTEM CAN ALSO UTILIZE AN EXISTING SUPPORT
STRUCTURE, ONLY IF THE STRUCTURE IS CERTIFIED BY A
LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.
2. SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS FOR WIDTH,
DEPTH, STRENGTH,
ETC.
3. CUSTOMER VA BLACK HILLS HEALTH CARE SYSTEM IS RESPONSIBLE FOR MEETING SITE
CONDITIONS FOR STATIC LOADS, LEVELNESS, ETC.
GENERAL STRUCTURAL CEILING SUPPORT REQUIREMENTS
GENERAL STRUCTURAL NOTES
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
BE LEVEL TO 1MM IN 10.0'.
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
CUSTOMERS CONTRACTOR
PENETRATIONS IN POST TENSION FLOORS.
CUSTOMERS CONTRACTORS
ALL HARDWARE FOR "THROUGH THE FLOOR" ANCHORING
AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS FLOORS.
THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRILLING THAT
CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUCTION
ENCOUNTERED BY THE INSTALLER SUCH AS REBAR ETC...
IT IS THE RESPONSIBILITY OF THE CUSTOMER
FLOOR OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
ENSURING THAT NO SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR STRUCTURAL SUPPORTS E.G. POST TENSION
CABLES OR REBAR WILL INTERFERE OR COME IN CONTACT
WITH SUBSURFACE PENETRATION OPERATIONS, E.G. DRILLING
AND INSTALLATION OF ANCHORING PERFORMED DURING THE
INSTALLATION PROCESS. TO ENSURE WORKERS SAFETY
INSTALLERS WILL ONLY PERFORM SURFACE PENETRATION
OPERATIONS ONLY AFTER THE CUSTOMERS CONTRACTOR HAS
VALIDATED THE SURFACE TO BE PENETRATED.
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VENTS,
SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT TO EXTEND
MORE THEN 1/4" BELOW THE FINISHED CEILING.
ALL STEEL WORK AND PARTS NECESSARY TO SUPPORT CEILING
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CUSTOMER
VA BLACK HILLS HEALTH CARE SYSTEM
SPANNING MEMBERS:
STEEL CHANNEL STRUT FLUSH WITH FINISHED
CEILING. SPACED AS INDICATED ON SHEET "S1"
RECOMMENDED FABRICATION DETAILS ON SHEET
"S2" LOADS SHOULD BE CALCULATED FROM SHEET
G1-1 (EQUIPMENT LEGEND ITEMS 4 & 5)
AREA OF EXAM ROOM TO BE LEVELED FOR X-RAY
EQUIPMENT, CONTRACTOR TO VERIFY MINIMUM
SPECIFICATION OF FLOOR SLABS ON WHICH
EQUIPMENT IS TO BE INSTALLED MUST BE LEVEL TO
1MM IN 10.0'".
CHEST STAND MOUNTING PLATE / ANCHOR BOLT
PATTERN3
TABLE MOUNTING BOLT PATTERN AND LOCATION4
5"
9"
1'-01 2"
AA 6'
-0
11'-103 4"
4' -8
71 2"
2" 4'-11
4"
4'-8"
9' -4
2'-2" 2'-2" 2'-2" 2'-2" 2'-2"
3"1'-81 4"
GENERAL STRUCTURAL NOTES
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
BE LEVEL TO 1MM IN 10.0'.
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
CUSTOMERS CONTRACTOR N/A MUST PROVIDE ALL
PENETRATIONS IN POST TENSION FLOORS.
CUSTOMERS CONTRACTORS N/AMUST PROVIDE AND INSTALL
ALL HARDWARE FOR "THROUGH THE FLOOR" ANCHORING
AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS FLOORS.
THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRILLING THAT
CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUCTION
ENCOUNTERED BY THE INSTALLER SUCH AS REBAR ETC...
IT IS THE RESPONSIBILITY OF THE CUSTOMER VA BLACK HILLS
HEALTH CARE SYSTEM/CONTRACTOR N/A TO PERFORM ANY
FLOOR OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
CUSTOMER VA BLACK HILLS HEALTH CARE
SYSTEM/CONTRACTOR N/A IS ALSO RESPONSIBLE FOR
ENSURING THAT NO SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR STRUCTURAL SUPPORTS E.G. POST TENSION
CABLES OR REBAR WILL INTERFERE OR COME IN CONTACT
WITH SUBSURFACE PENETRATION OPERATIONS, E.G. DRILLING
AND INSTALLATION OF ANCHORING PERFORMED DURING THE
INSTALLATION PROCESS. TO ENSURE WORKERS SAFETY
INSTALLERS WILL ONLY PERFORM SURFACE PENETRATION
OPERATIONS ONLY AFTER THE CUSTOMERS CONTRACTOR HAS
VALIDATED THE SURFACE TO BE PENETRATED.
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VENTS,
SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT TO EXTEND
MORE THEN 1/4" BELOW THE FINISHED CEILING.
ALL STEEL WORK AND PARTS NECESSARY TO SUPPORT CEILING
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CUSTOMER
VA BLACK HILLS HEALTH CARE SYSTEM OR THEIR CONTRACTOR
N/A.
ALL UNITS THAT ARE WALL MOUNTED OR WALL SUPPORTED ARE
TO BE PROVIDED WITH SUPPORTS WHERE NECESSARY. WALL
SUPPORTS ARE TO BE SUPPLIED AND INSTALLED BY THE
CUSTOMER VA BLACK HILLS HEALTH CARE SYSTEM OR THEIR
CUSTOMER OR CONTRACTOR'S PREPARATION
SAMSUNG
(CEILING SUPPORT STRUT)
(X-RAY CEILING RAIL INSTALL)
CUSTOMER OR CONTRACTOR'S PREPARATION
(FINISHED CEILING)
CEILING SUPPORT UNISTRUT
UPPER SLAB
TABLE
X-RAY
FLOOR FINISH
CEILING FINISH
SECTION A-A
DRAWING SHEET #
S1
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
GENE NOWACZYK
MEP
SITE PLANNING &
PROJECT DEVELOPMENT
SAMSUNG GC85
VA BLACK HILLS HEALTH CARE
SYSTEM
113 COMANCHE ROAD, FT MEADE, SD, 57741
PROJECT
ARCHITECT:
N/A p.N/A N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
N/A p.XXX-XXX-XXXX m.XXX-XXX-XXXX N/A
PROJECT MANAGER FOR CUSTOMER:
KEVIN FRAVEL
p.XXX
KEVIN.FRAVEL@VA.GOV
IT
ANTHONY KOESSL
p. XXX
ANTHONY.KOESSL@VA.GOV
INFECTIOUS DISEASE
NICHOLAS A. CHRISTENSEN
p.XXX
NICHOLAS.CHRISTENSEN2@VA.GOV
GENE N.DRAWN BY:
CHK'D BY:
SCALE:
DATE: 6/16/2017 8:53:45 AM
FILE NAME:
Fort Meade VA F1 06-16-2017.dwg
REVISION: DATE:
CUSTOMER :
STRUCTURAL LAYOUT SAMSUNG GC85
SCALE: 3/8" = 1'-0"
RECOMMENDED CEILING HEIGHT 9'-4.0"
A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY
UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO
WHICH THE STATIONARY CEILING TRACKS ARE FASTENED.
THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO
ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH
INSTALLATION.
THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE
SPECIFIED ON SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE
OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND
MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS,
ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE
CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED TO MEET SPECIFIC SITE
REQUIREMENTS BY THE ENGINEER OF RECORD.
CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN.
STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS.
THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT
MOVEMENT IN ANY DIRECTION. THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE
BUILDING STRUCTURE TO HOLD THE DEAD WEIGHTS INDICATED. see EQUIPMENT LEGEND "SHEET A1"
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL
PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE
REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING
DRAWINGS IS UNDERTAKEN.
1. CUSTOMER VA BLACK HILLS HEALTH CARE SYSTEM CAN ALSO UTILIZE AN EXISTING SUPPORT
STRUCTURE, ONLY IF THE STRUCTURE IS CERTIFIED BY A
LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.
2. SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS FOR WIDTH,
DEPTH, STRENGTH,
ETC.
3. CUSTOMER VA BLACK HILLS HEALTH CARE SYSTEM IS RESPONSIBLE FOR MEETING SITE
CONDITIONS FOR STATIC LOADS, LEVELNESS, ETC.
GENERAL STRUCTURAL CEILING SUPPORT REQUIREMENTS
GENERAL STRUCTURAL NOTES
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
BE LEVEL TO 1MM IN 10.0'.
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
CUSTOMERS CONTRACTOR
PENETRATIONS IN POST TENSION FLOORS.
CUSTOMERS CONTRACTORS
ALL HARDWARE FOR "THROUGH THE FLOOR" ANCHORING
AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS FLOORS.
THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRILLING THAT
CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUCTION
ENCOUNTERED BY THE INSTALLER SUCH AS REBAR ETC...
IT IS THE RESPONSIBILITY OF THE CUSTOMER
FLOOR OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
ENSURING THAT NO SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR STRUCTURAL SUPPORTS E.G. POST TENSION
CABLES OR REBAR WILL INTERFERE OR COME IN CONTACT
WITH SUBSURFACE PENETRATION OPERATIONS, E.G. DRILLING
AND INSTALLATION OF ANCHORING PERFORMED DURING THE
INSTALLATION PROCESS. TO ENSURE WORKERS SAFETY
INSTALLERS WILL ONLY PERFORM SURFACE PENETRATION
OPERATIONS ONLY AFTER THE CUSTOMERS CONTRACTOR HAS
VALIDATED THE SURFACE TO BE PENETRATED.
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VENTS,
SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT TO EXTEND
MORE THEN 1/4" BELOW THE FINISHED CEILING.
ALL STEEL WORK AND PARTS NECESSARY TO SUPPORT CEILING
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CUSTOMER
VA BLACK HILLS HEALTH CARE SYSTEM
SPANNING MEMBERS:
STEEL CHANNEL STRUT FLUSH WITH FINISHED
CEILING. SPACED AS INDICATED ON SHEET "S1"
RECOMMENDED FABRICATION DETAILS ON SHEET
"S2" LOADS SHOULD BE CALCULATED FROM SHEET
G1-1 (EQUIPMENT LEGEND ITEMS 4 & 5)
AREA OF EXAM ROOM TO BE LEVELED FOR X-RAY
EQUIPMENT, CONTRACTOR TO VERIFY MINIMUM
SPECIFICATION OF FLOOR SLABS ON WHICH
EQUIPMENT IS TO BE INSTALLED MUST BE LEVEL TO
1MM IN 10.0'".
CHEST STAND MOUNTING PLATE / ANCHOR BOLT
PATTERN3
TABLE MOUNTING BOLT PATTERN AND LOCATION4
5"
9"
1'-01 2"
2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 1'-10"
1'-71 4"
'-0
1' -1
03 4"
AA 6'
-0
11'-103 4"
4' -8
71 2"
2" 4'-11
4" 8'-03 4"
9' -4
GENERAL STRUCTURAL NOTES
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
BE LEVEL TO 1MM IN 10.0'.
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
CUSTOMERS CONTRACTOR N/A MUST PROVIDE ALL
PENETRATIONS IN POST TENSION FLOORS.
CUSTOMERS CONTRACTORS N/AMUST PROVIDE AND INSTALL
ALL HARDWARE FOR "THROUGH THE FLOOR" ANCHORING
AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS FLOORS.
THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRILLING THAT
CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUCTION
ENCOUNTERED BY THE INSTALLER SUCH AS REBAR ETC...
IT IS THE RESPONSIBILITY OF THE CUSTOMER VA BLACK HILLS
HEALTH CARE SYSTEM/CONTRACTOR N/A TO PERFORM ANY
FLOOR OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
CUSTOMER VA BLACK HILLS HEALTH CARE
SYSTEM/CONTRACTOR N/A IS ALSO RESPONSIBLE FOR
ENSURING THAT NO SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR STRUCTURAL SUPPORTS E.G. POST TENSION
CABLES OR REBAR WILL INTERFERE OR COME IN CONTACT
WITH SUBSURFACE PENETRATION OPERATIONS, E.G. DRILLING
AND INSTALLATION OF ANCHORING PERFORMED DURING THE
INSTALLATION PROCESS.
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it.