About this file

36E77618R0094 ATTACHMENT 3 - SOW Appendix 1 - Hot Springs - Samsung Equip Dwgs - 6-14-17.pdf

View the file

Other files

File Type Posted
36E77619C0014-000.docx DOCX document
36E77618R0094-0002002.docx DOCX document
36E77618R0094-0002000.docx DOCX document
36E77618R0094-0002001.docx DOCX document
36E77618R0094-0001001.docx DOCX document
36E77618R0094-0001000.docx DOCX document
36E77618R0094-008.zip ZIP file
36E77618R0094-005.pdf PDF
36E77618R0094-010.zip ZIP file
36E77618R0094-009.zip ZIP file
36E77618R0094-011.docx DOCX document
36E77618R0094-007.pdf PDF
36E77618R0094-001.docx DOCX document
36E77618R0094-002.docx DOCX document
36E77618R0094-003.docx DOCX document
36E77618R0094-012.docx DOCX document
36E77618R0094-013.docx DOCX document
36E77618R0094-006.pdf PDF
36E77618R0094-000.docx DOCX document
Show all 19

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

S1

SHEET INDEX

SHEET NUMBER SHEET DESCRIPTION

N1 GENERAL NOTES

N2 SITE READINESS NOTES

A1 EQUIPMENT LAYOUT

A2 TABLE AND STAND MOUNT

CS1 & CS2 CEILING STRUCTURE

E1 ELECTRICAL REQUIRED

SHEET INDEX

DATE:

REVISION:

SHEET NUMBER:

OF SHEET

HOT SPRINGS, SD

6/14/17

E2 ELECTRICAL REQUIRED

E3 ELECTRICAL REQUIRED

1. THESE DRAWINGS ARE FOR INFORMATION PURPOSES ONLY. DRAWINGS SUGGESTS LOCATION OF

X-RAY EQUIPMENT AND ASSOCIATED DEVICES, ELECTRICAL DETAILS AND ROOM LAYOUT. THE

PREPAIRER OF THESE PLANS CANNOT ACCEPT RESPONSIBILITY FOR ANY CHANGES TO THE EQUIPMENT OR THE ROOM OR ANY DAMAGES RESULTING FROM. ACTUAL CONSTRUCTION DRAWINGS SHOULD BE PROVIDED BY CUSTOMERS ARCHITECT OR GENERAL CONTRACTOR. EVERY EFFORT IS MADE TO GIVE

ACCURATE DETAIL OF ACTUAL EQUIPMENT TO BE INSTALLED.

2. IMAGING SOLUTIONS IS ONLY RESPONSIBLE FOR INSTALLATION OF IMAGING EQUIPMENT PURCHASES

FROM OR CONTRACTED FOR IMAGING SOLUTIONS. ROOM PREPARATION INCLUDING NECESSARY

CONSTRUCTION, NETWORKING AND SHIELDING TO BE SUPPLIED BY OTHERS.

3. CUSTOMER AND GENERAL CONTRACTOR ARE RESPONSIBLE FOR PREPARING THE ROOM AND COSTS

INVOLVED.

4. CUSTOMER AND GENERAL CONTRACTOR ARE RESPONSIBLE FOR ALL, BUT NOT LIMITED TO, BUILDING, ELECTRICAL, CONTRACTING, PROFESSIONAL FEES AND PERMITS.

5. CUSTOMER IS RESPONSIBLE FOR OBTAINING LEAD SHIELDING REQUIREMENTS FROM A LICENSED

PHYSICIST PER STATE AND/OR LOCAL REGULATION

6. CUSTOMER IS RESPONSIBLE FOR LIGHTING (NOT ON THIS PLAN). DIMMING LIGHTS MUST BE PROVIDED

IN X-RAY AND CONTROL ROOMS BY ARCHITECT PLAN OR ELECTRICAL CONTRACTOR.

7. CEILING HEIGHT MUST BE MAINTAINED AS INDICATED ON THIS PLAN. IF UNOBTAINABLE CUSTOMER OR

CONTRACTOR MUST NOTIFY IMAGING SOLUTIONS

8. ALL DIMENSIONS ARE FROM FINISHED SURFACES UNLESS NOTED.

9. WALL BAKING FOR EQUIPMENT SUPPORT TO BE MINIMUM OF 1/2" WOOD SECURED TO MINIMUM OF 3

WALL STUDS WITH BACKING FLUSH WITH FACE OF STUDS.

10. CONSTRUCTION MUST COMPLY WITH BUILDING CODES.

11. CUSTOMER IS RESPONSIBLE FOR FURNISHING AND INSTALLATION AT CUSTOMERS EXPENSE OF ELECTRICAL JUNCTION BOXES, TROUGH, CONDUIT, BUSHINGS, WIRE DUCT, COVER PLATES, MAIN SWITCHES , PANEL BOARDS, WIRES, UNISTRUT (WALL FLOOR OR CEILING), OR ANY SATISFACTORY

INSTALLATION OF X-RAYEQUIPMENT.

12. IMAGING SOLUTIONS MUST PROVIDE A SIGNED PURCHASE ORDER PRIOR TO ANY WORK OR MATERIALS PROVIDED AT IMAGING SOLUTIONS EXPENSE. IMAGING SOLUTIONS ASSUMES NO RESPONSIBILITY FOR

ANY CONSTRUCTION COSTS RELATED OR NOT RELATED TO THE INSTALLATION OF ANY X-RAY

EQUIPMENT.

GENERAL NOTES

GENERAL NOTES

N1

DATE:

REVISION:

6/14/17

1. TO ENSURE THE BEST ENSURE THE BEST ENVIRONMENT FOR INSTALLATION AND

PROTECTION OF MECHANICAL EQUIPMENT, COMPUTERS, MONITORS, ELECTRICAL

DEVICES, X-RAY TUBES, IMAGE INTENSIFIERS, AND CAMERAS THE FOLLOWING

GENERAL CONDITIONS ARE REQUIRED PRIOR TO DELIVERY. RESCHEDULING BY

IMAGING SOLUTIONS MANAGEMENT COULD BE REQUIRED IF CONDITIONS ARE NOT

MET.

2. THE CUSTOMER IS REQUIRED TO OBTAIN ALL PERMITTING, SHIELDING AND APPROVALS

REQUIRED BY LOCAL OR STATE GOVERNING AGENCIES.

3. CONSTRUCTION COMPLETE WITH LEAD BARRIORS INSTALLED, DOORS HUNG, FLOOR

COVERING INSTALLED , CEILING COMPLETED, WALLS COMPLETE AND PAINTED.

4. A DUST FREE ENVIRONMENT MUST BE KEPT IN THE X-RAY, CONTROL AND

SURROUNDING ROOMS.

5. HVAC COMPLETED, TESTED AND FUNCTIONING PRIOR TO DELIVERY.

6. POWER AT DESIGNATED PANEL AND EQUIPMENT PRIOR TO DELIVERY.

7. ALL CONDUIT, TROUGH, JUNCTION BOXES, WIRE RACEWAYS INSTALLED

ACCORDINGLY

8. LIGHTING TO BE INSTALLED AND FUNCTIONING.

9. BASE PLATES, REINFORCEMENT PLATES, OVERHEAD GRID INSTALLED AS DESIGNATED.

IT IS THE RESPONSIBILITY OF THE CUSTOMERS DESIGN ENGINEER TO RECOMMEND

METHODS OF ANCHORING AND ATTACHING IMAGING SOLUTIONS EQUIPMENT.

10. NETWORK, NETWORK DROPS, PACS, ROUTERS, COMPUTERS, PROCESSORS, PRINTERS,

LAST IMAGERS, AND/OR CAMERAS ARE INSTALLED AND TESTED FUNCTIONAL PRIOR

TO INSTALL. CUSTOMER IS RESPONSIBLE FOR OBTAINING AND MAINTAINING ANY AND

ALL BACK-UP POWER, ANTI-VIRUS SOFTWARE, FIREWALLS, OR OTHER SOFTWARE AND

HARDWARE. IMAGING SOLUTIONS WILL CONNECT TO CUSTOMERS NETWORK BUT

ASSUMES NO RESPONSIBILITY FOR THE NETWORK RELATED PROBLEMS. IMAGING

SOLUTIONS IS NOT RESPONSIBLE FOR COMPUTER ISSUES DUE TO CUSTOMER OR

PATIENTS UNAUTHORIZED US OR INSTALLATION OF COMPUTER PROGRAMS FOR THE

PERSONAL USE.

11. IMAGING SOLUTIONS SUPPLIED SERVERS ARE NOT TO HAVE UPDATES INSTALLED

WITHOUT IMAGING SOLUTIONS PRIOR CONSENT. VPN AND PHONE SERVICE MUST BE

INSTALLED PRIOR TO INSTALL.

12. A CLEAR, UNOBSTRUCTED, PATH INCLUDING ROOM ENTRY MUST BE PROVIDED BY THE

CUSTOMER FOR DELIVERY.

13. REMOVAL OF OLD EQUIPMENT IS A SEPARATE SERVICE. PLEASE CONTACT IMAGING

SOLUTIONS FOR ALL YOUR X-RAY NEEDS.

SITE READINESS NOTES

SITE READINESS NOTES

N2

DATE:

REVISION:

EXAM ROOM

SAMSUNG GC-85

18'-4"

'-6

/2

'-3

DATE:

REVISION:

6/14/17

1 SAMSUNG GENERATOR CABINET

SAMSUNG XGEO WORKSTATION2

SAMSUNG TILTING WALL STAND

3 SAMSUNG TABLE

SAMSUNG 4M CEILING RAILS5

29.5"L 19.3"W 43"H

21"L 15"W

25"L 32"W 87"H

96"L 32"W VARIABLE H

157"L

4 SAMSUNG 3M METER BRIDGE

118"L 24"W VARIABLE H

EQUIPMENT LAYOUT

A1

220 LBS

85 LBS

330 LBS

484 LBS

X

11,798 BTU

1200 BTU

120 BTU

N/A BTU

120 BTU

800LBS 873

RECOMMENDED

CEILING HEIGHT 9'-6"

EQ

UI

PM

EN

T

LE

G

EN

D

8"

8' -1

0"

10'-0"

4' -4

FOCAL SPOT

TRAVEL

USE APPROPRIATE ANCHORS AND

WASHERS TO MOUNT TO THE FLOOR. USE

OF ANY SMALLER SIZE MAY CAUSE THE

BASE TO BREAK FREE FROM THE FLOOR

UNDER EXTREME LOAD AND LEVERAGE

CONDITIONS.

IT IS UP TO THE INSTALLER AND THE CUSTOMER TO

DETERMINE THE BEST METHOD FOR MOUNTING THE

BASE TO THE FLOOR. BEFORE MOUNTING THE BASE,

CONSULT WITH THE BUILDING'S MAINTENANCE

SUPERVISOR ABOUT DRILLING HOLES IN THE FLOOR.

MAKE SURE THAT THERE ARE NO HAZARDS UNDER

THE FLOOR SUCH AS PIPES, CONDUITS, OR

STRUCTURAL CABLES WHICH CAN BE DAMAGED BY

DRILLING HOLES IN FLOOR.

TABLE & STAND MOUNT

A2

DATE:

REVISION:

6/14/17

RECOMMENDED

CEILING HEIGHT 9'-6"

CEILING STRUCTURE

CS1

STRUCTURAL SUPPORT TO BE DESIGNED, SUPPLIED

AND INSTALLED BY GENERAL CONTRACTOR WITH

ALL HARDWARE FOR CEILING MOUNTED X-RAY

EQUIPMENT. STRUCTURAL SUPPORT MUST BE

RIGID AND UNABLE TO SWAY/BUCKLE/BEND AT

ANY POINT. SUPPORT ALSO MUST BE LEVEL,

SQUARE, PARALLEL AND COPLANER TO ITSELF.

DATE:

REVISION:

6/14/17

RECOMMENDED

CEILING HEIGHT 9'-6"

EXISTING CEILING STRUCTURE TO

BE UTILIZED, VERIFY POSITIONING

AND STRUCTURAL CAPACITY

EQUIPMENT VENDOR OR VENDORS DESIGNER IS IN NO WAY

RESPONSIBLE FOR THE DESIGN OR INSTALLATION OF

STRUCTURAL SUPPORTS FOR ANY CEILING MOUNTED X-RAY

EQUIPMENT. THESE DRAWINGS ARE FOR ILLUSTRATION

PURPOSES ONLY. PURCHASE R OR GENERAL CONTRACTOR

SHOULD CONSULT A STRUCTURAL ENGINEER AND/OR

UNISTRUT REP TO COORDINATE SUPPORT STRUCTURE, ANY

STRUCTURAL DETAILS SHOWN ARE FOR EXAMPLE ONLY AND

NOT TO BE USED FOR ACTUAL CONSTRUCTION. ENTIRE

CEILING STRUCTURE MUST BE ABLE TO WITHSTAND 1.2 TONS.

CEILING STRUCTURE

CS2

DATE:

REVISION:

6/14/17

ENTIRE CEILING STRUCTURE MUST BE

ABLE TO WITHSTAND 1.2 TONS

2' -5

/2

D D

3' -5

7'-0"

CONTROL ROOM

ELECTRICAL REQUIRED

E1

NOTE: #6 GROUND WIRE REQUIRED

VOLTAGE FLUCTUATION ± 10 FULL

LOAD. ALL WIRES TO BE THHN OR TFFN

STRANDED COPPER THERMOPLASTIC

600 VOLT OR EQUIVALENT, UNLESS

OTHERWISE NOTED. MAXIMUM

LENGTH FROM DISTRIBUTION

TRANSFORMER TO ROOM MAIN

BREAKER NOT TO EXCEED 200 FEET.

THREE

PHASE

MAIN

FACILITY

DIST XFMR

INCOMING

WIRE SIZE

X-RAY

ROOM

BREAKER

480 VAC 105 KVA #2 #00 250MCM 100 AMP

50' 100' 200'

DATE:

REVISION:

* SEE SHEET E2 FOR ELECTRICAL CALLOUT INFORMATION

INCOMING LINE POWER AND SHUNT TRIP BREAKER.

SEE MSP-E2 SPECIFICATION BLOCK

E.P.O. SWITCH IS CONTRACTOR SUPPLIED AND INSTALLED

PROVIDES SHUNT TRIP FOR INCOMING

8"X8"X8" JUNCTION BOX 1" ABOVE FINISHED CEILING WITH

CONDUIT OR FLEX TO TROUGH WITH (1) 3" GROMMETED

HOLE IN COVER FACING DOWN *CONTRACTOR SUPPLIED*

12" WALL TROUGH WITH (1) 3" GROMMETED HOLE

OR ACCESS SPACE IN COVER 18" FROM FLOOR

4" x 4" JBOX LOCATED FLUSH W/ FINISHED FLOOR WITH

(1) 2" GROMMETED HOLE IN COVER

4" X 4" J-BOX CENTERED 18" A.F.F. WITH

(1) 2" GROMMETED HOLE IN COVER

WARNING LIGHT ABOVE DOOR. WIRES TO BE ROUTED

TO WALL TROUGH 2 WITH 12 FOOT WHIP AT FLOOR

TO MAKE CONNECTIONS

DOOR SWITCH(ES) ARE CONTRACTOR SUPPLIED AND

INSTALLED WITH (2) #14 WIRES TO WALL TROUGH 2

WITH 12 ' WHIP FOR MAKING FINAL CONNECTIONS

(1) DEDICATED QUAD OUTLET 115VAC 20 AMP (1)

NETWORKING DATA DROPS LOCATED 12” A.F.F10

(1) DEDICATED

QUAD OUTLET 115VAC 20 AMP LOCATED 12” A.F.F11

LINE POWER @ ITEM

NOTE: #6 GROUND WIRE REQUIRED

VOLTAGE FLUCTUATION ± 10 FULL

LOAD. ALL WIRES TO BE THHN OR TFFN

STRANDED COPPER THERMOPLASTIC

600 VOLT OR EQUIVALENT, UNLESS

OTHERWISE NOTED. MAXIMUM

LENGTH FROM DISTRIBUTION

TRANSFORMER TO ROOM MAIN

BREAKER NOT TO EXCEED 200 FEET.

THREE

PHASE

MAIN

FACILITY

DIST XFMR

INCOMING

WIRE SIZE

X-RAY

ROOM

BREAKER

480 VAC 105 KVA #2 #00 250MCM 100 AMP

50' 100' 200'

8"x8"x8" J-Box

FINISHED CEILING

FLEX

HOLE

DATE:

REVISION:

6/14/17

NOTE: #6 GROUND WIRE REQUIRED

MINIMUM RECOMMENDED

DISTRIBUTION

TRANSFORMER RATING

OPERATIONAL CLIMATETHREE PHASE

INCOMING VOLTAGE

INCOMING

WIRE SIZE

X-RAY ROOM

MAIN

BREAKER

480 VAC 105 KVA #2 #00 250MCM 100 AMP

50' 100' 200'

CONTRACTOR TO SUPPLY POWER WIRES

NECESSARY FROM ROOM BREAKER TO

GENERATOR

VOLTAGE FLUCTUATION ±10% FULL LOAD.

ALL WIRES TO BE THHN OR TFFN STRANDED

COPPER THERMOPLASTIC 600 VOLT OR

EQUIVALENT, UNLESS OTHERWISE NOTED.

MAXIMUM LENGTH FROM DISTRIBUTION

TRANSFORMER TO ROOM MAIN BREAKER

NOT TO EXCEED 200'.

TEMPERATURE

HUMIDITY

PRESSURE

10 - 40 DEGREES C OR

50 - 104 DEGREES F

30% - 75%

70 kPa - 106 kPa

ELECTRICAL CALLOUT INFO

ELECTRICAL REQUIRED

E2

2” HOLE CORE DRILLED THROUGH FLOOR AS

CLOSE TO WALL AS POSSIBLE

D D

2" CONDUIT FROM

WALL STAND TO WALL

TROUGH. MUST BE AS

DIRECT AS POSSIBLE.

2" CONDUIT FROM MAINS

BOX 1 TO WALL TROUGH

AS NEEDED

CONDUIT FOR EPO

TO BREAKER

AS NEEDED CONDUIT

FOR WARNING AND

DOOR SWITCH WIRE TO

GENERATOR

2" CONDUIT FROM

CONTROL J-BOX 6

TO WALL TROUGH.

EXISTING CONDUIT

TO BE UTILIZED

3" CONDUIT OR

FLEX CONDUIT

FROM J-BOX 5

TO EXISTING WALL

TROUGH TOP

TROUGH

EXISTING CONDUIT

TO BE UTILIZED

CONTROL ROOM

DATE:

REVISION:

6/14/17

* SEE SHEET E2 FOR WIRE INFORMATION

ELECTRICAL REQUIRED

E3

1
2
3
4
5
6
7
8
9
10