This document provides the scope of work for a federal contract opportunity to expand sterile processing services at the Hines VA Medical Center in Chicago, Illinois. The solicitation seeks to upgrade existing sterile processing departments by procuring new washer/disinfectors, pass-through windows, conveyor systems, and other equipment. Product specifications and equipment cut sheets are included in an appendix. The Department of Veterans Affairs Technology Acquisition Center in Austin, Texas is listed as the contracting agency.
View the file
Other files
Show all 24
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Edward Hines, Jr. VA Medical Center | 150
Appendix C : Basis of Design Equipment Cut Sheets
Cart Conveyor System
Edward Hines, Jr. VA Medical Center | 152
Edward Hines, Jr. VA Medical Center | 154
AMSCO® 7053HP SINGLE-CHAMBER
WASHER/DISINFECTOR
10283380 Rev E (10/27/20)
Item ________________________
Location(s)___________________
(Typical - details may vary.)
The Selections Checked Below Apply To This Equipment
POWER*
Steam-Heated Unit
Electric-Heated Unit
VOLTAGE*
200 to 208 V, 60 Hz (Steam)
200 to 208 V, 50 Hz (Steam)
380 to 400 V, 60 Hz
380 to 415 V, 50 Hz
460 to 480 V, 60 Hz
LANGUAGE PACKAGE
French (FD355)
Spanish (FD356)
German (FD357)
Italian (FD358)
Japanese (FD383)
Polish (FD384)
* Careful consideration must given to voltage selection prior to ordering. Later changes require substantial field modification.
† Refer to 10726323 for AMSCO® Top Enclosure Panels
Accessories.
Refer to 10370003 for ATLAS WAV System Accessories.
Refer to SD1014 for Material Handling Accessories.
ACCESSORIES†
Impact Printer (FD094)
Thermal Printer (FD012)
Seismic Tie-Down Kit (FD354)
Condensate Return To Drain (FD353)
Two-Level Manifold Rack (FD105)
Two-Level Manifold Rack with Equal Space
(FD106)
Three-Level Manifold Rack (FD107)
Four-Level Manifold Rack (FD108)
Five-Level Manifold Rack (FD110)
Multi-Function Rack for Large Items (FD114)
Multi-Function Rack for Small Items (FD115)
General Purpose Basket (FD116)
RAS 12 Long Rack (FD257)
Self Docking Transfer Cart (FD179)
Transfer Cart Adapter (FD209)
Set of dry Contacts (FD362) Handheld Bar Code reader (FD363) Conductivity Probe (FD350)
Additional Wall Flange (FD351) Installation Kit – Steam (FD088)
Installation Kit – Electric (FD037)
ACCESSORIES (CONT’D)†
Additional Chemical Pump (FD349)
Offset Installation Air Management Kit (FD352)
IQ-OQ-PQ Protocols (FD359)
Machine Identifier (11000866)
Drain Acu-CoolDown (FD361)
ConnectAssure Technology for Data Export
(CAT3030)
ProConnect® Technical Support Services
(Remote Monitoring, Priority Technical Support, Customer Care Center Access, Equipment
Performance Reports). Available in U.S. and
Canada only. (GP09-167)
OPTIONS
Acu-Rinse Reservoir (FD000054)
Acu-Wash Reservoir (FD000055)
40-Amp Min Circuit Protection (FD000056)
RAS Cycle (FD000057)
APPLICATION
The AMSCO 7053HP Single-Chamber Washer/
Disinfector with touch screen control is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans and urinals, trays, bowl, basin, kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
DESCRIPTION
The AMSCO 7053HP Single-Chamber Washer/
Disinfector is a cabinet-type mechanical washer that includes:
• Color touch screen control system and visual progress indicator
• Audible warning system with adjustable volume
• Space saving, power vertical sliding doors
• Eight factory-loaded (Instruments, Orthopedic
Instruments, Utensils, Plastic Goods, Rigid M.I.S., Anesthesia / Respiratory Goods, Gentle and
Decontamination) and up to 20 Custom cycles. Each cycle is effective, fast, energy efficient and may be customized (within allowed range of parameters) to meet specific operating requirements
• LED status indicator lighting
• Compact space saving footprint
• Conforms to seismic building code requirements
• Patented spray arm system providing entire chamber coverage
• 180 degree reverse installable
STANDARDS
The AMSCO 7053HP Single-Chamber Washer/Disinfector complies with the following standards:
As certified by Intertek:
• CAN/CSA-C22.2 No. 61010-1, 3rd Edition (R2017)
• CAN/CSA-C22.2 No. 61010-2-040 2nd Edition Dated 07/01/2016
• UL 61010-1, 3rd Edition (2012-05-11)
Governing Directive for the affixing of the CE mark:
• Medical Devices Directive 93/42/EEC as amended by 2007/
47/EC
• Machinery Directive 2006/42/EC
Standards Applied to Demonstrate Conformity to the
Directive:
• IEC 61010-1 3rd Edition (2012-05-11), EN 61326-1, 2013, IEC 61326-1 Second Edition (July 2012) and IEC 61010-2-
040:2015
• EN ISO 15883-1:2009+A1:2014, ANSI/AAMI ST15583-1:2009/
(R)2014 and CAN/CSA-Z15883-1:09 (R2019) Washer-
Disinfectors – General Requirements, Definitions and Tests
• EN ISO 15883-2:2009, ANSI/AAMI ST15583-2:2013/
(R)2015 and CAN/CSA-Z15883-2:09 (R2019) Requirements and Tests for Washer-Disinfectors Employing Thermal
Disinfection for Surgical Instruments, Anesthesia
Equipment, Holloware, Utensils, Glassware, etc.
SIZE (W X H X L)
Exterior dimensions:
• 42 x 80-3/4 x 38" (1067 x 2051 x 965 mm)
Interior chamber dimensions:
• 29-1/2 x 29-1/4 x 32-1/2" (749 x 743 x 826 mm)
Load height:
• 31" (787 mm) above finished floor
FEATURES
Vertical sliding power door:
• Door is constructed of double pane tempered glass to allow operator to view chamber interior with door closed.
• Color LED light illuminates top portion of each door to indicate machine status.
• Door is pneumatically activated using touch screens located on the control panel (one each side of unit) and is equipped with a built-in safety system.
• Each door is mounted on a compressed seal that reduces heat loss and increases heating capability.
• If a power failure occurs, door can be opened manually. A door interlock feature is provided to prevent cross-contamination.
• Unit can be configured in single- or double-door operation.
Stainless-steel pump is powered by a dual-speed motor. High pump speed provides equivalent capacity of a 14 hp (10.4 kW) motor, 165 U.S. gal/min at 196 ft (625 L/min at 60 m) head pressure.
Low pump speed provides equivalent capacity of a 3.5 HP
(2.6 kW) motor, 90 U.S. gal/min at 46 ft (341 L/min at 14 m) head pressure. Pump impeller is mounted directly on motor shaft and does not require additional bearings. Pump motor is equipped with a drip-proof frame, magnetic starter, overload protection and sealed bearings (not requiring periodic lubrication).
Pump, spray system and all recirculating piping are of stainless-steel construction.
Wash chamber:
• Constructed of 16-gauge, 316L stainless steel (No. 4 finish), argon-welded and polished. Chamber is of sanitary-type design (horizontal fixed surfaces are sloped, without overlapping sheet metal) for complete drainability and reducing hard-to-clean locations.
• Single-walled, insulated construction of chamber exterior reduces heat loss and noise level to the work area.
• Rotary spray assemblies are positioned (one at top and one at bottom of chamber, each measuring 24-3/8" [620 mm] long) to reach all load surfaces. Spray assemblies are easily removable without tools to aid in cleaning and maintenance.
Sump:
• Constructed of 316L stainless steel with 9.0 gal (34 L) capacity with removable stainless-steel filter preventing debris from entering pump and piping system.
• Heating coil (steam or electric) at the bottom of the wash chamber (sump) raises and maintains water temperature up to 185°F (85°C) during the Wash phase and up to 194°F
(90°C) during the Thermal Rinse phase.
Chemical peristaltic injection pumps:
• Pre-Programmed cycles and pumps are optimized to use
Prolystica® Ultra Concentrate (PUC) HP chemicals. Each ultra concentrated product is 10 times the concentration of a traditional product; so,10 times less chemical is injected per cycle.
• Unit comes with one enzyme and one detergent injection peristaltic pump which automatically add a selected quantity of chemical from 1/40 to 2 oz/gal (0.2 to 16 mL/L).
• One pump is dedicated to automatically add a selected quantity of lubricant during Thermal Rinse phase.
• Pumps are positioned near chemical containers. A low-level sensor is included to indicate when detergent level is low, or when insufficient chemical is available for next cycle.
• 100’ (30 m) extension tubes are included to pump chemicals from canisters installed in a remote location to unit.
• Control monitors volume of chemicals injected and indicates if this parameter meets specified criteria during all specific phases.
Top Utility Connections (except drain connections) facilitate installation. All utilities (including vent, steam, electric, water and compressed air) are connected on washer top. Top water injection gives a secure air gap of 31-1/2" (800 mm). Backflow preventers are not required for hot and pure water.
Front Service Panels provide easy access to piping, valves, electrical components and wiring. Unit can be configured as either a Vented System that exhausts chamber vapors to building exhaust system through 3.0" (76 mm) OD vent connection located at top of washer or as a Non-Vented
Drying System where chamber vapors are exhausted through cold water condenser to facility room.
Drying System uses uniquely designed four-sided inflow drying pattern to produce high-flow air curtain. Air curtain provides broad, efficient, drying within wash chamber. Recycled and non-recycled air is manifolded and circulated through piping and accessory providing energy efficient system while ensuring complete chamber air coverage. Fresh air is drawn through a
HEPA filter. Drying system includes a 3 hp (2.2 kW) blower, to remove vapor from chamber prior to opening doors, and three electric heaters totaling 11.4 kW to heat and maintain chamber air temperature. When Acu-Drying is selected, drying time self-adjusts to meet drying time of loaded rack and items.
Pure Water Stainless-Steel Supply Valve is supplied on washer/disinfector so Pure Water Rinse Supply may be used with factory programmed cycles and optional Acu-Rinse
Reservoir.
Drain Discharge Cool Down ensures water drained at end of each phase, from chamber sump to building drain system, does not exceed 140°F (60°C). If water temperature in sump is higher than 140°F (60°C), cold water is automatically added to reduce temperature of water discharged into building drain system.
Sampling Valve is supplied in washer/disinfector recirculation piping for North American Customers that are not able to install sampling valves on facility supply lines near washer/disinfector.
Barrier Wall Kit Flanges for one wall (clean side) installation come standard. An optional kit to enclose second wall (dirty side) is also available (FD351).
Cycle Graph enables monitoring of injection volume, water circulating pressure and water temperature during a cycle.
Graphs are visible during cycle and can also be displayed at cycle completion. Data can be exported using ConnectAssure
Technology or USB port.
CONTROL SYSTEM
The AMSCO 7053HP Single-Chamber Washer/Disinfector is equipped with a touch screen Customer interface. This 8.4"
(213 mm) color touch screen is mounted at eye level beside chamber door (both sides of unit) and tilted for better visibility, allowing easy monitoring of all wash cycles. Control system monitors and controls all phases of each programmed cycle.
The control system features:
• Cycle selection is simplified through use of Icons representing devices and cycle confirmation (on-screen) to ensure correct cycle is selected.
• Service mode for preventive maintenance testing and to facilitate troubleshooting.
• Built-in service diagnostic program to permit system calibration and verification of component operations.
• Security lock-out feature that enables cycles and temperatures to be locked and unchangeable without proper access code.
• Cycle data is stored as a protection against power disruption. Data may be downloaded from controller using supplied USB ports.
• Permits operator to monitor current washer/disinfector status
(including current chamber temperature, circulating pressure and time remaining in phase).
• Indicates any abnormal conditions with audible and visual alarm system.
CYCLE DESCRIPTION
ADVISORY NOTE: STERIS does not intend, recommend or represent in any way that this AMSCO 7053HP Single-Chamber
Washer/Disinfector be used for the terminal disinfection or sterilization of any regulated medical device. AMSCO 7053HP
Washer/Disinfector is intended only to perform an initial step in the processing of soiled, reusable medical devices. If medical devices contact blood or compromised tissues, such devices must be terminally processed in accordance with good hospital practices before each use in human patients.
Once treatment cycle is selected, washer/disinfector automatically processes load through standard phases
(additional phases are included in certain treatment cycles depending on unit configuration) as described on Cycle Charts and in supplied Operator Manual.
AMSCO 7053HP Single-Chamber Washer/Disinfector enables
Customers to use Dual Wash phase combining PUC HP enzyme and PUC HP detergent in the same filling which accelerates cycle time and reduces water consumption.
Custom cycles are available but must be validated by the
Customer (not STERIS responsibility).
SAFETY FEATURES
Vertical chamber doors are equipped with an obstruction sensor to detect any door obstruction. If obstruction is present, door automatically opens.
The washer/disinfector is equipped with a safety lockout feature so a cycle cannot start unless the door is fully closed.
If the door is opened during a cycle, all utility services to chamber are shut off, and the cycle stops.
Door interlock feature is provided to prevent cross-contamination. Door interlock feature allows only one door to be opened at a time whenever power is on. When the cycle is in process, door interlock prevents either door from being opened. Access to the load is then restricted.
Safety stop pushbutton(s), one on the load and one on the unload side, automatically stop all unit operation when pressed.
Building electrical supply disconnect switch must be used to shut off power to the unit before servicing.
ACCESSORIES AND OPTIONS
Printer (FD094 or FD012), if provided, produces an easy-to-read printed record of whether load was properly processed at preset temperature, as well as a complete list of alarm and abort in-cycle messages. A paper out LED is included.
Acu-Rinse Reservoir (if equipped) pre-heats Pure Water for
Rinse and Thermal Rinse phases significantly reducing cycle times. Constructed of 316L stainless steel.
Acu-Wash Reservoir recovers a portion of water used in
Thermal Rinse treatment and stores it for use in Wash treatment of next cycle reducing overall cycle water consumption.
RAS Cycle (Robotic Assisted Surgery Cycle) (FD000057 or
FD258) is used with the RAS 12 Long Rack (FD257) for effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci X/Xi and S/Si EndoWrist instruments.
da Vinci® and EndoWrist® are registered trademarks of Intuitive Surgical, Inc.
For Further Information, contact:
STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 • USA 440-354-2600 • 800-548-4873 www.steris.com
10283380 ©2020, STERIS Corporation. All rights reserved. Rev E (10/27/20)
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
The base language of this document is ENGLISH. Any translations must be made from the base language document.
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.
Steam Condensate Return To Drain (FD353) allows for connection of steam condensate return outlet to drain when steam condensate return line is not available in building. Cold water is also injected in drain piping when condensate return water temperature is too high. Condensate return cool down keeps temperature in drain piping below 140°F (60°C).
Seismic Tie-Down Kit (FD354) includes a seismic report for proper installing and securing of washer to building floor. Unit is designed to comply with seismic building code.
Additional Chemical Pump (FD349; up to two pumps) is available, giving flexibility to wash with neutral process, alkaline process, or to vary chemical used, depending on load type.
Drain Acu-CoolDown (FD361) precisely injects correct amount of potable water into facility drain system, minimizing volume of cold water needed.
Offset Installation Air Management Kit (FD352) controls airflow at top of door assemblies.
ConnectAssure Data Export (CAT3030) exports XML cycle records from equipment to Customer tracking system.
ProConnect® Technical Support Services maximizes operational efficiencies with secure, Internet-based, real-time equipment monitoring. Data from Customer equipment is used by STERIS to provide pro-active Customer alert notifications, technical support and predictive maintenance. (ProConnect
Technical Support Services is available in U.S. and Canada only.) Refer to Tech Data sheet SD983, PROCONNECT
TECHNICAL SUPPORT SERVICES, for details.
INSTALLATION
The washer/disinfector is designed as a fully enclosed cabinet for freestanding or recessed installation. Clearance between top of the unit and the ceiling must be at least 14" (356 mm).
If the system is recessed through one or two barrier walls, stainless-steel barrier flanges are included to provide a finished wall appearance.
Once installed, system is designed to allow for easy access for maintenance purposes.
All configuration modifications are available upon request.
Contact STERIS to receive a quotation.
PREVENTIVE MAINTENANCE
Customers are encouraged to contact STERIS concerning our annual maintenance program. Under the terms of the program, preventive maintenance, adjustments and replacement of worn parts are provided on a scheduled basis to help ensure optimal equipment performance and help minimize untimely or costly schedule interruptions. STERIS maintains a worldwide staff of well-equipped, factory-trained technicians to provide these services, as well as on-site installation, training and expert repair services. Contact STERIS for details.
NOTES
1. Machine segments (unit is bolted onto a wooden skid) are shipped in one cardboard carton (W x H x L): 50 x 91 x 48" (1270 x 2311 x 1219 mm).
2. Customer must ensure the washer/disinfector stands on a noncombustible floor (floor should be level).
3. Customer must provide utility connections with shutoff disconnects within 2’ (0.6 m) of equipment perimeter and below ceiling deck or purchase Installation Kit from STERIS.
4. STERIS recommends vacuum breakers (not provided by STERIS) be installed on service lines, and disconnect switches (with lockout in OFF position; not provided by STERIS) be installed in electric supply lines near equipment.
5. For all ventilation ducting from the washer/disinfector, STERIS recommends installation of a dedicated, corrosion-proof, flexible watertight duct (3" [76 mm] OD) to the exterior of the building, sloped toward the washer/disinfector.
6. Minimum ceiling height for door removal is 94" (2388 mm).
7. STERIS recommends illumination of the service area along with providing a convenience outlet for maintenance.
8. Noise level of 60.6 dB; Operating weight of 1543 lb (700 kg).
9. Unit is shipped with an Operator Manual. Maintenance Manual is available for purchase.
10. ConnectAssure Data Export requires an Ethernet drop and Ethernet connection within 10 ft of the equipment. The Customer must supply the network cable.
UTILITY REQUIREMENTS
IMPORTANT: Refer to proper equipment drawing for installation details and specifications:
• Electric-Heated Unit: 10288798, 10288799, 10288800 or
10288802
• Steam-Heated Unit: 10288787, 10288794, 10288795 or
10288796
Edward Hines, Jr. VA Medical Center | 156
AMSCO® 50 REPROCESSING SINK
SD1039 (05/01/18)
(Typical - details may vary.)
Item ________________________
Location(s)___________________
Ordering Information
SINK CONFIGURATIONS
❑ Double Bay
❑ Triple Bay
❑ Copper Drain:
❑ Double Bay ❑ Triple Bay
ACCESSORIES
❑ Back Splash Mounted Pre-Rinse and Gooseneck
❑ Air Pistol
❑ Water Pistol
❑ Water Pistol Tip Attachments
❑ Pure Water Faucet
❑ Flex Plates (SWA001)
❑ Universal Side Splash (12"[305 mm])
APPLICATION
The AMSCO 50 Reprocessing Sink is designed to assist users in the manual cleaning and rinsing of soiled re-usable utensils and other commonly found items within
Healthcare facilities that can be subjected to water and are required to be cleaned according to the instrument manufacturers Instructions For Use (IFU).
DESCRIPTION
The AMSCO 50 Reprocessing Sink is a free standing, height adjustable unit that is available in double or triple bay configuration. Counter length options range from 77-
120" (2 - 3 m) and right-to-left and left-to-right workflows are available. There are many additional features on this reprocessing sink that help facilities such as:
• Air and water pistols with tip attachments
• Back splash-mounted pre-rinse and faucet combo
• Back splash-mounted treated water faucet
• Magnetic back splash
• Etched gallon markers in all bays
SIZE (W X H X D)
Bay dimensions: 30 x 17 x 10" (762 x 432 x 254 mm)
STANDARDS
The AMSCO 50 Reprocessing Sink meets all applicable requirements of the following standards:
• Underwriters Laboratories (UL) Standard UL: 61010-1, Third Edition (2012)
• Canadian Standards Association (CSA): CAN/
CSAC22.2 No. 61010-1 (2012)
• ANSI/AAMI Guidelines ST79:2013
• ASME A112.19.12
FEATURES
Standard User Features are included on all models: one air pistol with tip attachments, one water pistol with tip attachments, two back splash-mounted pre-rinse and faucet combo, a back splash-mounted treated water faucet and magnetic back splash
Bay Etched Markings are available as bays are lightly etched with gallon markers to ensure correct chemistry dilution.
Hot and Cold Back Splash Mounted Pre-Rinse Faucet provides cleaning flexibility.
Magnetic Back Splash (12"[305 mm]) is supplied to enable workstation customization as the accessories are equipped with magnets. This feature permits varied mounting and organization to employee preference.
Height Adjustability provides enhanced ergonomics and reduces employee back strain.
For Further Information, contact:
STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 • USA 440-354-2600 • 800-548-4873 www.steris.com
SD1039 ©2018, STERIS Corporation. All rights reserved. (05/01/18)
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
ACCESSORIES
Air Pistol helps facilitate quick drying.
Water Pistol helps facilitate quick rinsing.
Pure Water Faucet meets manual cleaning standards for rinsing instruments.
Flex Plates provide an extended work surface for optional workflow. Each stainless-steel plates with ergonomic handles is rated at 35 lb (16 kg) weight capacity and can be placed over or between bays.
PREVENTIVE MAINTENANCE
Customers are encouraged to contact STERIS concerning annual maintenance programs. Under the terms of these programs, preventive maintenance, adjustments and replacement of worn parts are provided on a scheduled basis to help ensure optimal equipment performance and help minimize untimely or costly schedule interruptions. STERIS maintains a worldwide staff of well-equipped, factory-trained technicians to provide these services, as well as on-site installation, training and expert repair services. Contact STERIS for details.
NOTES
1. Customer must ensure sink stands on a noncombustible, level floor.
2. Request Equipment Drawing for Installation Details.
3. The AMSCO 50 Reprocessing Sink ships standard with 1.5"
ABS drain assembled to a P-trap for each basin. Additional parts are included to connect the drains for a diameter of 2".
The sink should be configured and connected according to applicable local plumbing codes.
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.
The base language of this document is ENGLISH. Any translations must be made from the base language document.
UTILITY REQUIREMENTS
IMPORTANT: Refer to Equipment Drawing for details.
Supply Lines
Air Pistol, 3/8" Compression, Silicon
Water Pistol, 3/8" Compression, Stainless-Steel
Pre-Rinse Hot, 3/8" Compression, Stainless-Steel
Pre-Rinse Cold, 3/8" Compression, Stainless-Steel
DI Faucet, 3/8" Compression, Stainless-Steel
Power
120 V, 1 Phase
AMSCO PASS-THROUGH WINDOW
SD1041 (Rev AB)
Item ________________________
Location(s)___________________
The Selections Checked Below Apply To This Equipment
CSSD MODELS*
❑ PTWAMDR1100 (Automated)
Window Width = 28" (711mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on either side of window
❑ PTWAMDR2100 (Automated)
Window Width = 34" (864mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on either side of window
❑ PTWAMDR3100 (Automated)
Window Width = 40" (1016mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on either side of window
ENDO MODELS
❑ PTWAENDO1100 (Automated)
Window Width = 28" (711mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on one side of window and 4" on other side.
❑ PTWAENDO2100 (Automated)
Window Width = 34" (864mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on one side of window and 4" on other side.
❑ PTWAENDO3100 (Automated)
Window Width = 40" (1016mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on one side of the window and 4" on other side.
❑ PTWMENDO1200 (Manual)
Window Width = 28" (711mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on one side of window and 4" on other side.
❑ PTWMENDO2200 (Manual)
Window Width = 34" (864mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on one side of window and 4" on other side.
❑ PTWMENDO3200 (Manual)
Window Width = 40" (1016mm)/Height = 45" (1143mm), 120V
Shelf extends 25" on one side of the window and 4" on other side.
APPLICATION
The AMSCO line of pass-through windows come in manual and automated configurations. The Automated windows have two configurations: CSSD/MDR and ENDO.
DESCRIPTION
The AMSCO Pass-Through Window is an in-line vertical lift window with a single operable (movable) panel and a single non-operable (fixed) panel, an integral continuous stainless steel shelf. The automated pass-through windows are equipped with a microwave no-touch sensor plate for opening and closing. All windows are fully assembled and ready to install.
STANDARD FEATURES
The AMSCO Pass Through Window is constructed from anodized aluminum extrusions, close-tolerance moving parts and tempered glass. All frames have a clear anodized aluminum finish, fastened with screws, joints are sealed and ¼" (6.4mm) clear tempered glass is dry glazed. All moving panels are guided using industry proven dual rope and counterbalance system.
The Shelf has been developed specifically for the needs of this window.
The shelf extends 25" (635mm) on either side of the window along the full width of the window (CSSD Models) or 4"
(102mm) on the decontamination side and 25" (635mm) on the clean side of the window (ENDO Models). The continuous shelf provides a smooth surface through the window. The shelf is secured by two stainless steel brackets under the shelf on each side of the window (total of four brackets). The manual window is available in the ENDO configuration.
The Automated Window features automated opening and closing activated by a microwave, no-touch, sensor plate located on each side of the window at the edge of the shelf.
Operation of the motorized window ensures that opening and closing of the window is hands-free.
STANDARDS
The AMSCO Automated Pass-Through
Window meets the applicable requirements of the following standards:
• Underwriters Laboratories (UL)
Standard 61010.
• Canadian Standards Association
(CSA) CAN/CSA-22.2 No. 68.
MDR/CSSD Models
FINISHED OPENINGS, SERVICE WIDTHS & WEIGHTS FOR MDR / CSSD WINDOWS
UNIT
X
FINISHED OPENING
WIDTH
Y
FINISHED OPENING
WIDTH
Z
SERVICE
WIDTH
SHIPPING
WEIGHT
(LBs,Kg)
OPERATING
WEIGHT
(LBs, Kg)
PTWAMDR1100 28-1/4" (718mm) 45-1/4" (1149mm) 24" (610mm) 315, 142 270, 122
PTWAMDR2100 34-1/4" (870mm) 45-1/4" (1149mm) 30" (762mm) 335, 151 280, 127
PTWAMDR3100 40-1/4" (1022mm) 45-1/4" (1149mm) 36" (914mm) 375, 170 300, 136
Edward Hines, Jr. VA Medical Center | 158
ENDO Models
FINISHED OPENINGS, SERVICE WIDTHS & WEIGHTS FOR ENDO WINDOWS
UNIT
X
FINISHED OPENING
WIDTH
Y
FINISHED OPENING
WIDTH
Z
SERVICE
WIDTH
SHIPPING
WEIGHT
(LBs,Kg)
OPERATING
WEIGHT
(LBs, Kg)
PTWAENDO1100
PTWMENDO1200
28-1/4" (718mm) 45-1/4" (1149mm) 24" (610mm) 280, 127 250, 113
PTWAENDO2100
PTWMENDO2200
34-1/4" (870mm) 45-1/4" (1149mm) 30" (762mm) 300, 136 260, 118
PTWAENDO3100
PTWENDO3200
40-1/4" (1022mm) 45-1/4" (1149mm) 36" (914mm) 350, 159 275, 125
Item ________________________
Location(s)___________________
AMSCO® EVOLUTION® STEAM STERILIZERS
SD889 Rev D
Typical - details may vary.
Selections Checked Below Apply To This Equipment
TYPE
❑ Prevacuum
❑ Steam Flush Pressure Pulse (SFPP)
MODEL NUMBER/SIZE
❑ HC900 / 26 x 37.5 x 42" (660 x 950 x 1067 mm)
❑ HC1200 / 26 x 37.5 x 54" (660 x 950 x 1372 mm)
❑ HC1500 / 26 x 37.5 x 66" (660 x 950 x 1676 mm)
STERILIZER ELECTRIC SERVICE
❑ 208/240 V ac, 50/60 Hz, 3-Phase, 12 A/Phase
❑ 480 V ac, 50/60 Hz, 3-Phase, 6 A/Phase
STEAM SOURCE
❑ Facility Steam1
❑ Integral Steam Generator (72 kW)
❑ Carbon Steel Stainless Steel
❑ 208V, 50/60 Hz, 3-Phase, 203 A
❑ 240V, 50/60 Hz, 3-Phase,176 A
❑ 480V, 50/60 Hz, 3-Phase, 89 A
STEAM PIPING
❑ Copper/Brass Stainless Steel
❑ Optional Right-hand Piping2
DOOR CONFIGURATION
❑ Single-Door (A - B, following graphic)
❑ Hinged Door
❑ Left-Hand (A)
❑ Right Hand (B)
❑ Horizontal-Sliding3
❑ Double-Door (C, D, E, F)
Note: Operating End hinge position listed first, Non-operating End hinge listed second.
❑ Hinged Door
❑ Left-Hand (OE) /Right-Hand (NOE) (C)
❑ Left-Hand (OE) /Left-Hand (NOE) (D)
❑ Right-Hand (OE) /Left-Hand (NOE) (E)
❑ Right-Hand (OE) /Right-Hand (NOE) (F)
❑ Horizontal-Sliding Door3
SINGLE-DOOR MOUNTING
❑ Cabinet Enclosed/Freestanding
❑ Recessed
DOUBLE-DOOR MOUNTING (RECESSED)
❑ Through One Wall Through Two Walls
ACCESSORIES
❑ Loading Car
❑ Transfer Carriage
❑ Chamber Track Assembly
❑ Single Door ❏ Double Door
❑ Loading Car, Transfer Carriage and Track
Assembly
❑ Single Door Double Door
❑ Chamber Rack and Two Shelves
❑ Single Door Double Door
Extra Shelf
❑ 42" Chamber Length
❑ Seismic Tie-Down Kit
❑ ALUS (Automatic Load/Unload System)
REMOTE MONITORING
❑ ProConnect® Response Center (Remote
Monitoring, Priority Technical Support, Customer Care Center Access, Equipment
Performance Reports). Available in U.S. and
Canada only. (GP09160)
WATER CONSERVATION OPTIONS
❑ STERI-GREEN Water Re-Circulating System
❑ (UE0000000000004)
❑ STERI-GREEN PLUS Chilled Water Loop
Re-Circulating System
❑ (UE0000000000002)
Notes:
1. External Supplied Steam (Facility Steam/Stand- Alone Steam Generator)
2. Left-hand piping is standard
3. Direction of door movement is right to left as viewed from sterilizer operating end and left to right as viewed from non-operating end.
APPLICATION
AMSCO® Evolution® Steam Sterilizers are configured for prevacuum sterilization of heat- and moisture-stable materials used in healthcare facilities.
These units can be equipped with prevacuum, Steam Flush Pressure Pulse
(SFPP), gravity, liquid, and daily air removal test cycles.
DESCRIPTION
AMSCO Evolution Steam Sterilizers are equipped with the latest features in both state-of-the-art technology and ease of use.
Primary Product Features
• Hinged door: door may be hinged on either left side or right side.
• Powered Door Options: door travels horizontally right to left to open.
• All plumbing components are mounted to a free-standing, modular rack (stand). The stand connects to the core sterilizer assembly during installation.
• An advanced Allen-Bradley1 PLC control system employs user-friendly interface screens, with enhanced graphics.
• 7" (178 mm) color touch screen display.
• Thermal printer
• Standard communication interface with most PC-compatible peripheral
26 x 37.5" (660 x 950 mm) Horizontal Sliding Door
26 x 37.5" (660 x 950 mm) Manual Hinged Door
1. The Allen-Bradley brand-name is owned by Rockwell Automation, Inc.
devices (e.g., data collection systems, printers)
• Automatic check of control program and cycle data maintains process integrity
• Control designed to accommodate integrated remote monitoring and instrument tracking interfaces
STANDARDS
Each sterilizer meets applicable requirements of the following listings and standards, and carries the appropriate symbols:
• ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form
U-1 is furnished. Shell and door are constructed to withstand working pressure of 45 psig (3.1 bar).
• Underwriters Laboratory (UL) Standard 61010-1 as certified by Intertek Testing services.
• Canadian Standards Association (CSA) Standard C22.2, No. 1010 as certified by Intertek Testing services.
• ANSI/AAMI ST8:2001– American National Standards
Institute/Association for the Advancement of Medical
Instrumentation, Standard 8 for hospital steam sterilizers. All standard sterilization cycles are programmed into the sterilizer control and validated to this standard.
• EN285:2006 Sterilization-Steam Sterilization-Large – the European harmonized standard.
• CRN (Canadian Registration Number) – design submitted for registration under The Boiler and Pressure Vessel Act to state compliance with the Act, Regulations, and relevant codes and standards.
FEATURES
26 x 37.5" (660 x 950 mm) Chamber Cross-section sized to allow for efficient, high-volume processing of sterilization containers, trays and packs.
Fast-operating, low-effort manual door lock mechanism
(hinged door models) allows door to be locked or unlocked using a single 30° handle motion.
Power Door operates quietly, and consists of a motor-driven cable and pulley mechanism. Vertical- and horizontal-sliding door is controlled from control panel push buttons. Double door configurations are supplied with controls at both ends of the sterilizer to help prevent the possibility of cross-contamination.
Software calibration is performed in the Service Mode, accessible through the touch screen displays. Control system provides printed record of all calibration data for verification to current readings.
Pneumatic valves are fitted in piping for steam, water and exhaust control, and require little or no maintenance.
Principle piping components and the primary control assembly are mounted to a separate, modular support rack
(plumbing stand). The plumbing stand connects during installation to core chamber and frame assembly, allowing for increased access for service and maintenance procedures when necessary.
ProConnect® Technical Support Services - Maximize operational efficiencies with secure, internet-based, real-time equipment monitoring. Data from your equipment is used by
STERIS to provide pro-active Customer alert notifications, technical support, and predictive maintenance. Online parts ordering, equipment performance dashboards, and online service scheduling at steris.com is also available. (ProConnect
Technical Support Services is available in U.S. and Canada only.) Refer to Tech Data Sheet SD983, PROCONNECT TECHNICAL
SUPPORT SERVICES, for details.
UTILITIES CONSERVATION FEATURES
Electronic water saving control monitors and minimizes the amount of water used in condensing the exhausted chamber steam and condensate. The control includes a resistance temperature detector (RTD).
Automatic utilities start-up/shutdown permits utilities conservation. Shutdown may be programmed to activate at the end of any designated cycle or time of day. When activated, the control system automatically shuts off all utility valves, conserving steam and water usage. Sterilizer utilities can be restarted either by programmed time or manual operation. A different shutdown and restart time can be programmed for each day.
Insulation sleeve is fitted around exterior of the sterilizer vessel to conserve heat and limit heat loss to the surrounding environment. The sleeve is sealed and held in place by hook-and-loop closures. Insulation is asbestos-free and chloride-free, silicone impregnated, oil- and water-resistant fiberglass.
Two-stage vacuum pump is supplied on all units to effectively pull chamber to specified vacuum levels, reduce cycle time by shortening conditioning and exhaust times; as well as reduce water consumption.
STERI-GREEN Water Re-Circulating System option uses up to 40% less water per cycle than units without a re-circulating system. Water consumptions are based on the water temperature, pressure and flows listed in the consumption table. Sterilizer water consumptions and vacuum efficiency may vary based on facility feed water tolerances. This system recovers water to a reservoir to be recycled and reused for next cycle.
STERI-GREEN PLUS Chilled Water Loop System option uses 99% less water per cycle than standard units. This system maximizes water savings by connecting into facility’s chilled water system in a closed loop.
PROCESSING CYCLES
All standard sterilization cycles are programmed into the sterilizer control and validated to AAMI/ANSI ST8.
IMPORTANT: Applicable cycles have been validated to satisfy the requirements outlined below. If cycle parameters (sterilize time, dry time, temperature) other than those listed are required, it is the responsibility of the healthcare facility to consult and follow the device manufacturer’s written instructions.
Dry Times for Prevacuum and SFPP are based upon maximum load.
• four full shelves of 25 lb (11 kg) instrument trays
Prevacuum configuration sterilizers are factory programmed with the following cycles: Prevacuum, Gravity, Liquid, and Test
Cycles. Prevacuum cycles are intended for efficient, high-volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of vacuum pulses followed by pressure pulses to condition the porous load prior to sterilization. See cycle descriptions below for more details.
Default Prevac Cycles
• 270°F (132°C) Prevacuum Cycle: For sterilizing double-wrapped instrument trays and fabric packs.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 4 minutes
» Dry time: 30 minutes
• 270°F (132°C) Prevacuum Cycle: For sterilizing single fabric packs.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 4 minutes
» Dry time: 5 minutes
• 275°F (135°C) Prevacuum Cycle: For sterilizing double-wrapped instrument trays.
» Sterilize temperature: 275°F (135°C)
» Sterilize time: 3 minutes
» Dry time: 30 minutes
• 250°F (121°C) Gravity Cycle: For sterilizing fabric packs.
» Sterilize temperature: 250°F (121°C)
» Sterilize time: 30 minutes
» Dry time: 15 minutes
• 270°F (132°C) Gravity Cycle: For sterilizing double-wrapped instrument trays.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 15 minutes
» Dry time: 30 minutes
• 250°F (121°C) Gravity Cycle: For sterilizing double-wrapped instrument trays.
» Sterilize temperature: 250°F (121°C)
» Sterilize time: 30 minutes
» Dry time: 30 minutes
• 270°F (132°C) Gravity Cycle: For sterilizing fabric packs.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 25 minutes
» Dry time: 15 minutes
• Liquid Cycle: This cycle is used for sterilizing liquids in borosilicate containers with vented closures.
» Sterilize temperature: 250°F (121°C)
» Factory programmed sterilize time: 45 minutes
» Dry time: not applicable
IMPORTANT: It is inappropriate for a healthcare facility to sterilize liquids for direct contact with patients.
Steam Flush Pressure Pulse (SFPP) configuration sterilizers are factory programmed with the following cycles: SFPP, Prevacuum, Gravity, Liquid and Test Cycles. SFPP cycles are intended for efficient, high volume processing of double wrapped instrument trays and fabric packs. This dynamic air removal cycle incorporates a series of steam flushes and pressure pulses at pressures above atmospheric levels to condition load prior to sterilization. No vacuum is drawn during conditioning, therefore, no Bowie-Dick test is required. .
Prevacuum cycles are also intended for efficient, high volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of vacuum pulses followed by pressure pulses to condition the porous load prior to sterilization.
NOTE: Prevac and SFPP cycles may be used interchangeably.
See cycle descriptions below for more details:
Default SFPP Cycles
• 270°F (132°C) SFPP Cycle: For sterilizing double-wrapped instrument trays and fabric packs.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 4 minutes
» Dry time: 30 minutes
• 275°F (135°C) SFPP Cycle: For sterilizing double-wrapped instrument trays.
» Sterilize temperature: 275°F (135°C)
» Sterilize time: 3 minutes
» Dry time: 30 minutes
• 270°F (132°C) SFPP Cycle: For sterilizing single fabric packs.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 4 minutes
» Dry time: 5 minutes
• 270°F (132°C) Prevacuum Cycle: For sterilizing double-wrapped instrument trays and fabric packs.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 4 minutes
» Dry time: 30 minutes
• 275°F (135°C) Prevacuum Cycle: For sterilizing double-wrapped instrument trays.
» Sterilize temperature: 275°F (135°C)
» Sterilize time: 3 minutes
» Dry time: 30 minutes
• 250°F (121°C) Gravity Cycle: For sterilizing fabric packs.
» Sterilize temperature: 250°F (121°C)
» Sterilize time: 30 minutes
» Dry time: 15 minutes
• 270°F (132°C) Gravity Cycle: For sterilizing double-wrapped instrument trays.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 15 minutes
» Dry time: 30 minutes
• Liquid Cycle: This cycle is used for sterilizing liquids in borosilicate containers with vented closures.
» Sterilize temperature: 250°F (121°C)
» Factory programmed sterilize time: 45 minutes
» Dry time: not applicable
IMPORTANT: It is inappropriate for a healthcare facility to sterilize liquids for direct contact with patients.
Edward Hines, Jr. VA Medical Center | 160
TESTING CYCLES
• Warm-up Cycle: This cycle is used to warm chamber to operating temperature prior to performing a Bowie-Dick Test cycle (or other operating cycles).
• Bowie-Dick Test Cycle: This cycle is used for conducting
Bowie-Dick tests. Recommended load is a Dart® testing apparatus from STERIS or a properly prepared Bowie-Dick test pack. Preprogrammed cycle parameters cannot be adjusted by the user.
» Sterilize temperature: 270° (132°C)
» Sterilize time: 3 minutes and 30 seconds
» Dry time: 1 minute
• Vacuum Leak Test: This cycle is used for testing the vacuum integrity of the sterilizer piping. Sterilizer chamber must be empty while running this test cycle. All timing is preprogrammed and cannot be adjusted.
CONTROL SYSTEM
Design Features
The Allen-Bradley PLC control system monitors and controls all sterilizer operations and functions. The control system is factory-programmed with standard sterilizing cycles. Each cycle is adjustable to meet specific processing requirements.
All operator-accessible control functions can be changed using the touch screen control.
IMPORTANT: If cycle parameters (sterilize time, dry time, temperature) other than those listed are required, it is the responsibility of the healthcare facility to consult and follow the device manufacturer’s written instructions.
Cycle values and operating features may be adjusted and verified prior to cycle operation. Cycle parameters are retained in control memory for repeated use.
Once cycle is started, cycles and cycle values cannot be changed until cycle is complete. If chamber temperature drops below set point during the exposure phase, the timer is set to stop and automatically reset once normal operating temperature is reached.
Critical control system components are housed within a sealed compartment to protect the components from moisture and heat generated during the sterilization process.
Operator interface control panel, consisting of a touch screen, is located on the operating (i.e., load or non-sterile) end of the sterilizer.
The operator interface consists of a color touch-screen. The display allows for control communications, graphics and excellent visibility in most environments. The display panel, in conjunction with the control, is used as the monitor for the operator. All sterilizer functions, including cycle initiation and cycle configuration, are operated by pressing the touch-sensitive areas on the display. Display indicates appropriate control buttons, operator prompts, and status messages necessary to assist in sterilizer operation. All displayed messages are complete phrases with no codes to be cross-referenced. Display also indicates any abnormal conditions that may exist either in or out of a cycle.
Thermal printer, located below touch screen, is a high-resolution (8 dots per mm) printer. It is fast and quiet, printing at 25 lines per minute on industrial grade thermal paper. The device provides an easy-to-read printed record of all pertinent cycle data on 2.25" (57 mm) wide paper. Data is automatically printed at the beginning and end of each cycle and at transition points during the cycle.
Three paper tape rolls are furnished with each unit. Printouts have a guaranteed lifetime of 25 years and are resistant from exposure to steam, alcohol, UV and visible light, oil, heat and water.
Non-operating end (NOE) control panel, on double-door sterilizers only, includes a touch sensitive screen identical to the operating end screen. Preprogrammed cycles can be started from the NOE control panel. Display concurrently shows the same information as the operating end screen display.
Other controls located at the non-operating end include door control pushbuttons (if power door), jacket and chamber gauges and an emergency-stop button (if power door).
Cycle configuration is performed by accessing the Change
Values menu through the operating end touch screen.
In addition to adjustment of cycle values, the following operating parameters can also be changed through the
Machine Setup menu:
• Time display and printout units 24-hour or AM/PM.
• End-of-cycle signals and alarm signals have three adjustable volume levels available through the control and display panel.
• Temperature display and printout units Fahrenheit (°F) or
Celsius (°C). Temperature is set, displayed, controlled and printed to the nearest 0.1°. Recalibration is not required when changing temperature units.
• Pressure/vacuum display and printout units psig/InHg or bar. Recalibration is not required when changing pressure units.
SAFETY FEATURES
Emergency stop button located on the front panel, below the sterilizer control touch pad (Power Door Units Only). When pressed, immediately shuts off all outputs on the sterilizer. A key is used to reset the switch.
Control lockout switch equipped on chamber door(s), senses when door seal is energized and tight against the door. Control prevents cycle from starting until the limit switch signal is received. If control loses appropriate signal during cycle, alarm activates, cycle aborts and chamber safely vents with a controlled exhaust.
Chamber float switch activates alarm, aborts cycle and safely vents chamber with a controlled exhaust if excessive condensate is detected in the vessel chamber.
Pressure relief valve limits the amount of pressure buildup so that the rated pressure in the vessel is not exceeded.
Power door safety feature causes door drive to slip if the sliding door encounters an obstruction during its movement.
CONSTRUCTION
Shell Assembly Two fully fabricated Type 316L stainless-steel shells, welded one within the other, form the sterilizer vessel. Type 316L stainless-steel end frame(s) is welded to door end. On single door units, back of chamber is fitted with welded, 316L stainless-steel dished head.
Sterilizer vessel is ASME and PED rated at 45 psig (3.1 bar) and insulated. Vessel includes one 1"-NPT chamber port for Customer use.
Steam-supply opening inside the chamber is shielded by a stainless-steel baffle.
Chamber Door(s)
Door is constructed of Type 316L stainless steel.
During cycle operation, door is sealed by a steam-activated door seal. Door seal is constructed of a special long-life rubber compound. When sterilize cycle is complete, the seal retracts under vacuum into a machined groove in the sterilizer end frame.
A proximity switch is used by the control to determine if door is closed. An additional seal pressure switch prevents inadvertent cycle initiation if door is not sealed.
The door assembly is equipped with a mechanical locking mechanism that ensures the door cannot be opened as long as the seal is intact and energized and more than 2 psig
(0.14 bar) pressure is in the chamber.
The sterilizer door is fitted with a stainless-steel panel that insulates the operator from the chamber end frame, reducing the chance of accidental contact with a hot metal surface.
Chamber Drain System Drain system is designed to prevent pollutants from entering into the water-supply system and sterilizer.
The automatic condensing system, consisting of a heat exchanger, converts chamber steam to condensate and disposes condensate to waste. Cooling water flow is regulated by the waste line RTD to minimize water usage. Water supply shutoff valve is located in the recessed area of the unit.
Vacuum System Two-stage vacuum pump reduces chamber pressure during prevacuum and post-drying phases. Air is drawn from the chamber through the vacuum system. Following the dry phase, chamber vacuum is relieved to atmospheric pressure by admitting air through a bacteria-retentive filter.
Piping Rack (Plumbing Stand)
All piping is located on a modular plumbing rack (stand).
Plumbing stand is located on left side (standard) of the sterilizer, but may be changed to right side (optional). Piping comes in copper/brass (standard), but is available in stainless-steel
(optional).
Steam Source Sterilizers are piped, valved and trapped to receive facility-supplied steam delivered at 50 to 80 psig (3.5 to 5.6 bar) dynamic.
Standard steam piping is constructed of copper and brass and includes a shutoff valve and pressure regulator. (An optional, integral electric steam generator is also available.)
Options:
• Facility Steam
• Stand-alone steam generator
• Integral steam generator
» carbon steel for standard copper-and-brass piping
» stainless-steel for the optional stainless-steel piping (see following note)
NOTE: Reverse Osmosis (RO) or deionized (DI) water system required for stainless-steel piping system.
MOUNTING ARRANGEMENT
Sterilizers are arranged for either freestanding or recessed installation, as specified. Each sterilizer is height-adjustable.
Sterilizer subframe is equipped with a synthetic rubber gasket to ensure tight fit…