S02 0001 Attch 3 2024 Puget Sound VA Medical Surgical Vacuum Pump Inspection Report.pdf
PDF 198 KB Posted
- Solicitation number
- 36C26026Q0674
About this file
This is a Source Evaluation Data document for a surgical vacuum system at the VA Puget Sound Health Care System, dated 20 May 2024. The document provides a comprehensive inspection and compliance assessment of the vacuum pump equipment located in the Energy Plant area serving Facility/Building 100.
The system utilizes a Rogers/KRVP Series unit (Model KRVP-30-VAC, Serial 23H6528852) with a Rogers pump (Model KRVP-30, Serial H652885223) operating at 274 CFM and 460 volts. The vacuum main line pressure is 20.7"HG with a gauge range of 0-30HG. The lead pump activates at 19.8"HG and deactivates at 24"HG, while the lag pump operates at 17"HG on and 22"HG off. The evaluation confirms compliance across 18 major assessment categories with substantially all requirements marked as "Yes," including: dual pump configuration with check valves and isolation valves; adequate receiver with ASME data plate and manual drain; automatic activation of additional units with local audible/visual signals wired to master alarm panels; automatic pump alternation; dedicated disconnect switches and motor starting devices; power supply from the essential electrical system; outdoor exhaust location at least 10 feet from building openings with trapped drip leg and valved drain; hard-drawn copper distribution piping with appropriate labeling; line vacuum low switch connected to master alarm panels on the patient side of the main valve; and demand check fitting with gas-specific configuration. Technical support is available at 503-639-6151. The document contains a Comments section at the conclusion.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
20 May 2024
VA Puget Sound Health Care System
Surgical Vacuum
SOURCE EVALUATION DATA
LOCATION OF VACUUM PUMP: ENERGY PLANT AREA SERVED: FACILITY/BLDG. 100
UNIT MANUFACTURER: ROGERS/KRVP SERIES TECH SUPPORT PHONE: 503-639-6151
UNIT MODEL NUMBER: KRVP-30-VAC SERIAL NUMBER: 23H6528852
PUMP MANUFACTURER: ROGERS CFM: 274 VOLTAGE: 460
VACUUM PUMP DATA MODEL # KRVP-30 SERIAL # H652885223 2023 HOURS: N/A 2024 HOURS: 3.9
THE VACUUM MAIN LINE PRESSURE IS: 20.7”HG GAUGE RANGE IS: 0-30HG
THE LEAD PUMP IS ON AT 19.8” HG AND OFF AT 24” HG LAG IS ON AT 17” HG AND OFF AT 22” HG
Yes No N/A ☐ ☐ ☒ 1. Is the data on this sheet for the following source(s): LEAD & LAG?
☒ ☐ ☐ 2. Is the entire area Facility served by this source?
☒ ☐ ☐ 3. Is the room adequately vented to prevent accumulation of heat per manufacturers’ recommendation?
☒ ☐ ☐ 4. Are there two or more pumps?
☒ ☐ ☐ 5. Does each pump have a check valve to prevent backflow through off-cycle units?
☒ ☐ ☐ 6. Does each pump have isolation valves on intake piping?
☒ ☐ ☐ 7. Does each pump have isolation valves (or equivalent) on exhaust piping? NOTE: If the answer is no, a tube cap can be used when pumps are removed from service.
☒ ☐ ☐ 8. Is there sufficient flexible couplings installed in the piping to and from the equipment?
☒ ☐ ☐ 9. Is there a receiver?
Yes No ☒ ☐ a) Does the receiver have an ASME data plate?
☒ ☐ b) Does the receiver have sufficient manual drain?
☒ ☐ c) Is the receiver valved to allow draining or servicing without interrupting the Vacuum system?
http://www.medgas.com/
VA Puget Sound Health Care System 20 May 2024
Yes No N/A ☒ ☐ ☐ 10. Is system capable of automatically activating additional units to maintain vacuum?
Yes No ☒ ☐ a) Is there a local audible / visual signal when lag pump is in operation?
☒ ☐ b) Does the signal latch, requiring manual reset?
☒ ☐ c) Is it wired to master alarm panels?
☒ ☐ ☐ 11. Is there automatic pump alternation to provide for even wear?
☒ ☐ ☐ 12. Are there dedicated disconnect switches, motor starting devices, and overload protection for each pump?
☒ ☐ ☐ 13. Is power supplied by the equipment branch of the essential electrical system?
14. Vacuum Exhaust:
Yes No ☒ ☐ a) Is the exhaust outdoors and at least 10 ft. from any door, window, air intake, or other opening in the building?
☒ ☐ b) Is the exhaust discharge turned down and screened or otherwise protected from vermin, debris, or precipitation?
☒ ☐ c) Is there a trapped drip leg with valved drain on the exhaust line to discharge condensate or oil?
☒ ☐ d) Is the exhaust piping considered to be of proper size to minimize back pressure?
Exhaust Location: Above mechanical room
☒ ☐ ☐ 15. Is Vacuum distribution piping hard-drawn copper or stainless steel pipe?
☒ ☐ ☐ 16. Is the source piping appropriately labeled for Vacuum?
Yes No ☒ ☐ Is the label background color white?
☒ ☐ ☐ 17. Is the source valve in the immediate vicinity of the source equipment?
Yes No ☒ ☐ Is the source valve labeled?
☒ ☐ ☐ 18. Is there a line vacuum low switch wired to the master alarm panels?
Yes No ☒ ☐ a) Is the switch located on the patient side of the main valve, or source valve if there is no main valve?
☒ ☐ b) Is the switch connected with a demand check fitting?
☒ ☐ c) Is the demand check fitting gas specific?
☒ ☐ d) Is there a gauge installed next to the switch after source/mainline shutoff valve?
Location of switch: Just upstream of Main Valve
COMMENTS:
http://www.medgas.com/
File details come from the government source that posted it. Updated .