Argo2000.pdf
PDF 14 MB Posted
- Solicitation number
- 36C26025R0039
About this file
This document is an Arjo System 2000 bath instructions manual detailing the specifications, usage, maintenance, and technical details of a medical bathing system designed for assisted bathing of dependent patients in hospitals and care facilities. The comprehensive manual covers various bathtub models (Rhapsody, Primo, Primo Ferro, Harmonie), control panel functions, safety instructions, cleaning and disinfection procedures, technical specifications, and electromagnetic compatibility guidelines.
Key features include multiple optional systems such as Hydromassage, Hydrosound, and Sound & Vision, with detailed instructions for patient transfer, bathing, showering, and equipment maintenance. The system supports patients with reduced mobility, has a maximum patient weight of 182 kg, and includes safety mechanisms like scalding protection and emergency lowering. Technical specifications cover electrical requirements (230/240V in Europe, 120V in North America), water supply parameters, sound levels, and operational conditions, with a recommended expected service life of ten years when properly maintained.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SARRTP AS BUILT DRAWINGS.pdf | ||
| Liko Shop Drawings .pdf | ||
| 36C26025R0039 0002.docx | DOCX document | |
| Sign in sheet B88 job walk.pdf | ||
| 36C26025R0039 0001.docx | DOCX document | |
| Request For Information for Building 88 Responses.pdf | ||
| Nurse Call Manufacturer Specs.pdf | ||
| Wage Determination Douglas County OR20250085 May 9 2025.pdf | ||
| 653-25-100 Remodel Building 88 Drawings FINAL.pdf | ||
| 653-25-100 Remodel Building 88 For CLC Space TIL spec book FINAL.pdf | ||
| 653-25-100 SOW FINAL.docx | DOCX document | |
| 36C26025R0039_2.docx | DOCX document |
Show all 12
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
System 2000
04.AR.12_23EN • 2025-03
INSTRUCTIONS FOR USE
Design Policy and Copyright ® and ™ are trademarks belonging to the Arjo group of companies.
© Arjo 2025
As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.The content of this publication may not be copied either whole or in part without the consent of Arjo.
WARNING
To avoid injury, always read this Instructions for Use and accompanied documents before using the product.
Contents Foreword Intended Use Safety Instructions Preparations System 2000 Bathtubs Parts Designation Control Panels Buttons and Displays Product Description/Functions
Drain Control Knob Autofill Shower Handle Shower/Disinfection System (UK market only) P220 Temperature Display P300 Display Scalding Protection Shampoo/Body Wash (optional) Bath Oil/ArjoSound™ Water Conditioner (optional) Raise and Lower Bathtub Emergency Lowering (optional, not for UK market) Disinfection Unit (optional) Pillow Foot Support Hot Water Flush (Kit)
Hydromassage System Hydrosound System Sound & Vision System Change Default Setting (P300 only) Bathe Patient using a Lift Hygiene Chair Bathe Patient using a Lift Bath Trolley Cleaning and Disinfection Instructions Care and Preventive Maintenance Troubleshooting Error Codes (P300 only) Technical Specifications Labels Electromagnetic Compatibility Parts and Accessories
Foreword
Thank you for purchasing Arjo Equipment The Arjo System 2000® bath is part of a worldwide family of quality products designed especially for efficient patient hygiene and good ergonomics for the caregiver.
Please read this Instructions for Use thoroughly!
Please read this Instructions for Use (IFU) in its entirety before using your System 2000 bath.
Information in this IFU is crucial to the proper operation and maintenance of the equipment, will help protect your product, and make sure that the equipment performs to your satisfaction. Some of the information in this IFU is important for your safety and must be read and understood to help prevent possible injury.
Unauthorized modifications on any Arjo equipment may affect its safety. Arjo will not be held responsible for any accidents, incidents or lack of performance that occur as a result of any unauthorized modification to its products.
Service and Support A service routine has to be done on your System 2000 to make sure the safety and operating procedures of your product. See section Care and Preventive Maintenance on page 53.
If you require further information, please contact Arjo for comprehensive support and service programs to maximize the long-term safety, reliability and value of the product.
Contact Arjo for replacement parts. Contact information appears on the last page of this Instructions for Use (IFU).
Serious incident If a serious incident occurs in relation to this medical device, affecting the user, or the patient then the user or patient should report the serious incident to the medical device manufacturer or the distributor. In the European Union, the user should also report the serious incident to the Competent Authority in the member state where they are located.
Complementary Information A set of Quick Guides/Posters accompanies this IFU.
Please note that the Quick Guides/Posters do not replace the information in this IFU. The IFU contains additional important safety instructions.
Definition in this IFU:
Means:
Safety warning. Failure to understand and obey this warning may result in injury to you or to others.
Means:
Failure to follow these instructions may cause damage to all or parts of the system.
Means:
This is important information for the correct use of this system or equipment.
WARNING
CAUTION
NOTE
Intended Use The System 2000 is intended for assisted bathing and showering of dependant patients and residents in hospitals and care facilities.
The bathing must be assisted by appropriately trained caregivers with an adequate knowledge of the care environment and its common practices and procedures, and in accordance with the guidelines in the Instructions for Use (IFU). Transfer to the bathtub shall take place by means of an appropriate transfer device.
The bathtub must be disinfected according to instructions in the IFU.
The System 2000 bath should only be used for the purpose specified in this IFU.
Any other use is prohibited.
Indication for Use The System 2000 bath can be used for alleviation of, or compensation for an injury or disability associated with dependant persons with reduced mobility.
Patient/Resident Assessment We recommend that facilities establish regular assessment routines.
Caregivers should assess each patient/resident according to the following criteria prior to use:
• The patient/resident’s weight should not exceed the Safe Working Load (SWL) of the lift hygiene chair, lift bath trolley or sling lifters.
• Active and semi-active patient/residents (i.e. able to sit upright self-supported on the side of a bed or toilet) and completely bedridden patient/residents can use the bathtub. Correct transfer equipment shall be used.
If a patient/resident does not meet these criteria an alternative equipment or system shall be used.
Installation Requirements The equipment must only be used for the purposes stated above and must be installed according to the Assembly and Installation Instructions, which can only be superseded by local code.
Expected Service Life The expected lifetime of this equipment, unless otherwise stated is ten (10) years, subject to required preventative maintenance as specified in the section Care and Preventive Maintenance, the Assembly and Installation Instructions and the Parts List.
Safety Instructions
WARNING
To avoid injury to both the patient and the caregiver, never modify the equipment or use incompatible parts.
WARNING
To avoid explosion or fire, never use the equipment in oxygen rich environment, in presence of heat source or flammable anaesthetic gases.
WARNING
To avoid electric shock, never use other electrical equipment near this device.
WARNING
To avoid falling, make sure the patient always remains in a seated position.
WARNING
To avoid injury, make sure that the patient is not left unattended at any time.
NOTE
A medical assessment of the patient must be made in order to ensure their suitability to use this product.
Preparations
Site Preparation Prior to installation of the System 2000 bath the floor and walls within the required area must be finished.
Material: The floor should be made of a slip-resistant material, free of bumps.
Floor Slope: When using a floor drain, the slope towards the drain should not exceed 1:50.
Floor attachments: The floor construction must be suitable for anchoring the bolts.
Actions Before the First Use (10 steps) 1 Make sure that the bath has been installed according to the Assembly and Installation Instructions.
2 Check that all parts of the product are supplied.
Compare with section Parts Designation on page 10. If any part is missing or damaged - do NOT use the product.
3 Unpack the plastic from the Sound & Vision system hand control, if any.
4 Read the IFU.
5 Disinfect the product according to section Cleaning and Disinfection Instructions on page 40.
6 Perform a functionality test according to section Care and Preventive Maintenance on page 53.
7 Choose a designated area, where the IFU should be kept and is easily accessible at all times.
8 Make sure the thermostatic mixer has been calibrated according to the Assembly and Installation Instructions by an Arjo service technician.
10 Make sure to have a rescue plan ready in case of a patient emergency.
9 WARNING
To prevent scalding, make sure that the water mixer has been calibrated.
Confirm that responsible organization has established calibration routines.
Actions Before Every Use (5 steps) 1 Check that all parts are in place. See section Parts
Designation on page 10.
2 Check bath and accessories for damage.
3 If any part is missing or damaged - do NOT use the product.
4 Check the liquid level of disinfection bottle (P220 only).
Before every use, make sure the System 2000 bath is cleaned and disinfected according to section Cleaning and Disinfection Instructions on page 40.
5 WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use.
NOTE
If you have any questions, please contact your local Arjo representative for support and service. The contact information is listed on the last page of this IFU.
System 2000 Bathtubs Rhapsody® 1600 mm (63”)
1900 mm (75”)
2150 mm (85”)
White
Primo® 1600 mm (63”)
1900 mm (75”)
White
Primo® Ferro 1870 mm (73 5/8”)
2120 mm (83 1/2”)
Stainless steel
Harmonie™ 1680 mm (66 1/8”)
White
Intentionally left blank
Parts Designation 1 Control panel lid
2 Control panel
3 Patient shower handle
4 Filling spout
5 Dispenser outlet for shampoo and bath oil (ArjoSound water conditioner)
6 Surface overflow outlet and drain control knob
7 Top water level sensor
8 Temperature/Low water level sensors
9 Drain with a drain plug
10 Foot support
11 Bathtub
12 Drainage hose
13 Foot pad for 1680 mm (66 1/8”), 1870 mm (73 5/8”) and 1900 mm (75”) bathtubs.
14 “Duck Foot” for 1600 mm (63”), 2120 mm (83 1/2”) and 2150 mm (85”) bathtubs.
15 Pillow
16 Emergency lowering (optional)
17 Compartment for shampoo/bath oil (ArjoSound water conditioner) and disinfectant.
18 Lock bracket
Disinfection unit (optional) Not NSF approved 19 Disinfection spray handle
20 Disinfectant bottle compartment under shampoo/ bath oil (ArjoSound water conditioner).
21 Flow meter
Hydromassage™ system (optional) 22 Water flow and air mix control
23 Hydromassage system outlet with quick coupling device
24 Suction inlet
25 Hose with nozzle (accessory)
Hydrosound® system (optional) 26 Hydrosound system transducer plate
27 Intensity control knob
Sound & Vision® system (optional) 28 Sound & Vision system light
29 Speakers
30 Remote control
31 USB connector
32 USB memory stick
Type B, Applied part.
Degree of Protection - Electrical shock
27 26
UNDER THE CONTROL PANEL LID
24 25
202118
Control Panels
1 Shampoo/body wash ON/OFF
2 Autofill ON/OFF
3 Shower ON/OFF
4 Mains power indicator
5 Temperature display - bathtub water
P220 Control Panel 1 2 3 4 5 6
7 8 9 10
6 Disinfection ON/OFF
7 Bath oil/ArjoSound water conditioner ON/OFF
8 Hydromassage/Hydrosound system ON/OFF
9 Raise and lower the bathtub
10 Temperature setting knob
P300 Control Panel 2 3 4 5 6 7
8 10
1 Shampoo/body wash ON/OFF
2 Hidden button (change default setting according to Change Default Setting (P300 only) on page 27)
3 Display - temperature, filling, bathtub water and disinfectant, error codes etc.
4 Temperature setting: colder (-), warmer (+)
5 Autofill ON/OFF
6 Shower ON/OFF
7 Disinfection ON/OFF
8 Bath oil/ArjoSound water conditioner ON/OFF
9 Raise and lower the bathtub
10 Hydromassage/Hydrosound system ON/OFF
Buttons and Displays Autofill Button 1 Press Autofill button to fill bath to preset level.
2 LED lights up when filling is active (P300).
No LED is available on P220.
3 Press the Autofill button again to stop the filling.
Shower Button 1 Press the Shower button to start the shower.
2 Press the Shower button again stop the shower.
3 LED lights up when active.
4 The shower function shuts off after 15 minutes. Press Shower button to restart.
Raise Bathtub Button
• Press and hold in the Raise Bathtub button to raise the bathtub.
Lower Bathtub Button
• Press and hold in the Lower Bathtub button to lower the bathtub.
Temperature Display - Bathtub water (P220) The display shows the current water temperature in the bathtub.
• P220 has a mechanical and electronic scalding protection.
Display (P300) The display shows the temperature of the filling water, active shower and disinfectant liquid from the disinfection spray handle.
• Press plus (+) for warmer water and press minus (-) for colder water.
• Press plus (+) or minus (-) to select between preset autofill levels.
• Provides digital adjustment of the Hydrosound system.
• Built-in electronic scalding protection.
P220P300
75.0°F
37.0°C
75.0°F
P300 StandbyMode
• The display will show two dashes when there is no water in the bathtub and no functions are activated.
Shampoo/Body Wash Button
• Press the Shampoo/Body Wash button to dispense shampoo/ body wash from the outlet.
Bath Oil/ArjoSound™ Water Conditioner Button 1 Press the Bath Oil/ArjoSound water conditioner button to dispense Bath Oil/ArjoSound water conditioner from the outlet.
2 Baths equipped with the Hydrosound system have ArjoSound water conditioner instead of Bath Oil.
Hydromassage™/Hydrosound® System Button (optional) The bath is either equipped with Hydromassage system or with the Hydrosound system.
1 Press the Hydromassage/Hydrosound system ON/OFF button to start hydromassage (built-in jets) or hydrosound (low frequency ultra sound).
2 Press the Hydromassage/Hydrosound system ON/OFF button again to stop hydromassage/hydrosound.
3 LED lights up green when active (P300).
LED is placed below the control panel on P220. It lights up green when hydrosound is active (no indication for hydromassage on P220).
4 The Hydromassage/Hydrosound system will automatically shut itself off after 15 minutes.
Disinfection Button (optional) Note that the button is always on the panel but the function might not be included.
• Press the disinfection button to activate disinfection unit.
• LED lights up when active.
• Low disinfectant warning indicators (P300 only).
• Button only works when there is no water in the bathtub and no functions are activated.
P220P300
Product Description/Functions Drain Control Knob Turn the drain control knob (A) to open and close the drainage plug (B) of the bathtub. (See Fig. 1)
• Turn Clockwise to open.
• Turn Counter Clockwise to close.
The drain control knob is equipped with a built-in surface overflow outlet to prevent water to overflow the bathtub.
Autofill The bath is equipped with an autofill function that prefills the bathtub to a preset water level. The preset autofill levels fluctuate depending on water pressure.
1 Press the Autofill button to start prefill. (See Fig. 2)
2 LED lights up when filling is active (P300).
No LED is available on P220.
3 The Autofill stops automatically when prefill level is reached.
4 Press the Autofill button to stop filling at any time during Autofill.
To adjust water level on the P220, call an Arjo service technician.
The P300 has three different Autofill levels, Low, Normal and High. The preset level is Normal. (See Fig. 3)
To change between autofill levels, follow instructions below (P300 only).
Normal level (preset, 30 s over the top sensor) 1 Push the Autofill button
2 The display shows:
After 8 s the filling water temperature is shown and no more level adjustments are possible.
NOTE
If the Autofill button is pressed after the preset water level is reached, bathtub filling continues.
Fig. 1 A
B
Fig. 2
Fig. 3
High Level Normal Level Low Level
Low level (preset, 200 s above the low level sensor) 1 Push Autofill button
2 Push the minus (-) button
3 The display shows:
4 After 8 s the filling water temperature is shown and no more level adjustments are possible.
High level (preset, 60 s over the top sensor) 1 Push autofill button
2 Push the plus (+) button
3 The display shows:
After 8 s the filling water temperature is shown and no more level adjustments are possible.
Each level can be adjusted according to Change Default Setting (P300 only) on page 27.
Shower Handle
• Press the Shower button to start the shower. The LED will turn green when active.
• Press the trigger on the shower handle to let water out.
• Hold the trigger in for constant water flow.
To get continuous water flow without having to hold in the trigger do as follow: (See Fig. 1)
• Press the trigger and push in the lock button on the side. The water will continue to flow without having to press the trigger.
• Release lock button by pressing the handle trigger and the lock button will pop up by itself.
The patient shower can not be activated during bathtub filling.
Shower/Disinfection System (UK market only) For UK version MPA system will flush 24 h after last use.
Fig. 1 Lock Button
Trigger
P220 Temperature Display (See Fig. 1) Temperature display
• Shows the current bathtub water temperature.
P300 Display (See Fig. 2)
• Shows the preset autofill level first.
• Shows the active filling or shower temperature.
• Shows the current bathtub water temperature when filling and shower is not activated.
• During disinfection the display shows the temperature of the disinfectant.
Scalding Protection
If a dangerous temperature level of 45 °C (113 °F) is reached, the scalding protection shuts off the water in less than 10 s.
• P220 has a mechanical (calibration and mixer) scalding protection which disables high temperature settings.
• There is in addition to the mechanical scalding protection an electronic one equipped with two sensors. When scalding protection turns on, all the LEDs flash and a beeping sound will be heard.
• P300 has an electronic scalding protection and is equipped with three sensors. The display shows and a beeping sound will be heard. During this time the P300 recalibrates.
Reset scalding protection (P220 only) (5 steps) 1 Make sure both LEDs at the Disinfection/Autoclean button and the Patient Shower button have stopped flashing.
2 Turn the knob on the thermostatic mixer to COLD.
3 Press the Tub Filling button and fill the bathtub until the temperature display shows 30-35 °C (86-95 °F).
4 Press the Shower button. Lift off the shower handle, direct it away from yourself down into the bathtub. Press the trigger on the shower handle.
5 Allow a few seconds for the water to reach correct temperature.
WARNING
To prevent scalding if the scalding protection is activated, remove the patient from the bath immediately in a safe manner.
Fig. 1 P220 World
P220
USA
37.0°C
75.0°F
P300 World
P300
USA
37.0°C
37.0°C
75.0°F
37.0°C
75.0°F
Fig. 2
Reset scalding protection (P300 only) (4 steps) P300 resets and recalibrates automatically.
1 Wait until the display indication disappears.
2 Press the Tub Filling button to flush the mixer and pipes with water. Wait until stable filling temperature will be displayed. Press the filling button again to stop the flush.
3 Press the Shower button. Lift off the shower handle, direct it away from yourself down into the bathtub. Press the trigger on the shower handle.
4 Allow a few seconds for the water to reach correct temperature.
Shampoo/Body Wash (optional) The Shampoo/Body Wash bottle is placed under the control panel lid and lock bracket (A). (See Fig. 1)
The bottle should be mounted to its corresponding hose and bottle connection (B). (See Fig. 1)
Arjo recommends to use Arjo Care™ shampoo and body wash, a mild all-in-one soap.
• Press the Shampoo / Body Wash button while holding your hand under the dispenser outlet (C). (See Fig. 2)
The P300 dosage time is preset but can be changed. See section Change Default Setting (P300 only) on page 27.
Bath Oil/ArjoSound™ Water Conditioner (optional) The Bath Oil / ArjoSound water conditioner bottle is placed under the control panel lid and lock bracket (A). (See Fig. 3) The bottle should be mounted to its corresponding hose and bottle connection (B). (See Fig. 3)
Arjo recommends to use Arjo Oil™ as a bath oil or ArjoSound water conditioner if the bath comes with the Hydrosound system.
• Press the Bath Oil / ArjoSound water conditioner button to dispenser directly into the bath water (C). (See Fig. 4)
• A bath system with the Hydrosound system has ArjoSound water conditioner instead of bath oil (Arjo Oil™).
The P300 dosage time is preset but can be changed. See section Change Default Setting (P300 only) on page 27.
BA
C
Fig. 1
Fig. 2
BA
C
Fig. 3
Fig. 4
Raise and Lower Bathtub
The bathtub can be raised and lowered for an ergonomic working height for the caregiver. (See Fig. 1)
• Raise the bathtub to a preferred height by pressing the raise the bathtub button.
• Lower the bathtub to a preferred height by pressing the lower the bathtub button.
Emergency Lowering (optional, not for UK market) In case of a power failure were the bathtub needs to be lowered do as follows:
• Pull the emergency lowering handle on the bottom rear side of the panel. (See Fig. 2)
• The bathtub will move down as long as the handle is being pulled.
Disinfection Unit (optional) The bath has a built-in disinfection unit. The disinfectant bottle is located under the control panel lid and under the shampoo and bath oil bottles. (See Fig. 3) The disinfectant bottle cap is adjustable and regulates the disinfectant/water ratio.
Under the control panel lid, there is a disinfection spray handle and a flowmeter. The flow meter shows the liquid mix ratio between the disinfectant and water.
Continue with the steps on the next page.
WARNING
To avoid entrapment of the patients or caregivers legs or feet, make sure that they are kept clear of all obstacles.
WARNING
To avoid the device in use from tipping, do not raise or lower other equipment close to it and be aware of stationary object when lowering.
NOTE
If a buzzing sound is heard, the battery is low. Please contact your local Arjo service technician.
Fig. 1
To use the disinfectant, see section Cleaning and Disinfection Instructions on page 40.
• The LED will turn green when active.
• The disinfection will only turn on if all other functions are turned off and there is no water in the bathtub.
• If the disinfectant bottle is empty on the P300, all the LED’s will start flashing when pressing the disinfection button. A 5 s long beeping sound will be heard and it is not possible to disinfect the bath. The display will show .
• Replace with a full disinfection bottle and try again (see section Replace Disinfectant Bottle (System 2000 bath with disinfection option only) on page 42).
• P300 has automatic disinfection for the Hydromassage system (P220 optional).
Pillow The pillow is adjustable and attaches to the bathtub with suction cups located on the backside of the pillow. (See Fig. 1)
• Wet suction cups improve the grip.
• Place the pillow in a most convenient and supportive possible way for the patient.
Foot Support The foot support is adjustable and can be placed in three different positions, depending on the patient’s length.
(See Fig. 2)
• Place the foot support in one of the three slots. For Primo® Ferro snap it in place.
Hot Water Flush (Kit) Refer to separate Hot Water Flush Kit P300.
Fig. 1
Fig. 2
Primo® Ferro
Rhapsody® Primo® Harmonie™
Hydromassage System
Use Hydromassage™ System (12 steps) 1 Fill the bathtub with water at desired temperature over the
Hydromassage system outlet and top sensor (A).
(See Fig. 1)
2 Start the hydromassage by pressing the Hydromassage system button on the control panel.
3 The default runtime is 15 minutes after which the hydromassage will automatically end. Wait 15 minutes to restart.
• During the 15 minutes of runtime, the hydromassage can be stopped at any time by pressing the Hydromassage system button.
• Always turn the Hydromassage system off when finished.
4 The water flow is adjustable with the knob located on the upper right side of the bathtub. (See Fig. 2)
• Start at a low intensity by turning the knob counter clockwise.
• Increased intensity of the water flow by turning the knob clockwise.
5 Adjust the direction of the water-jet by turning the Hydromassage system outlet (A). (See Fig. 3)
• Remove the adjustable Hydromassage system nozzle (B) if you want the jet to cover a wider area (just pull it out).
(See Fig. 3)
6 When finished the hydromassage, turn the Hydromassage system off by pressing the Hydromassage system button.
Continue with the steps on the next page.
WARNING
To avoid hair entrapments under water, always install the Hydromassage system suction cover before use.
CAUTION
Turn the hydromassage water intensity knob clockwise at highest intensity in order to prevent water leakage on the floor. The leakage occurs from the built-up water pressure in the Hydromassage system. (See Fig. 2)Fig. 1
A
Fig. 2 Decrease Intensity
Increase Intensity
A B
Continue Hydromassage with Hose (optional) 7 If hydromassage is turned on, stop it by pressing the
Hydromassage system button before the hose is connected.
• Remove the Hydromassage system nozzle (just pull it out).
• Connect the hydromassage hose to the quick coupling device in the bathtub. (See Fig. 4)
• The hydromassage hose and the built-in jet within the bathtub cannot be used simultaneously.
8 Direct the hydromassage hose nozzle into the water. Make sure the air inlet is above the water surface. (See Fig. 5)
9 Hydromassage from the hydromassage hose starts when the Hydromassage system button on the control panel is pressed.
10 To adjust the jet stream intensity, turn the outermost part of the hydromassage hose nozzle. (See Fig. 6)
• When the two dots on the nozzle are aligned, the jet stream is at its highest intensity.
• When turning the nozzle 90°, the jet stream intensity decreases.
• When turning the nozzle additional 90°, the jet stream intensity increases and so on.
11 When finished the hydromassage, turn the Hydromassage system off by pressing the Hydromassage system button.
12 Remove the hydromassage hose from the quick coupling device by pushing in the hose while pulling the hose bracket back. (See Fig. 7)
Fig. 4
Air Inlet
Fig. 5
MAX.
MAX.
MIN.MIN.
Fig. 6
Fig. 7
Hydrosound System The Hydrosound® system is a method of cleaning patient by means of low frequency and low intensity ultra sound.
Working together with water, it is an effective and skin friendly method of cleaning the skin.
The amount of time and intensity setting used for the Hydrosound system cleaning is a matter of choice consistent with the degree of patient cleanliness required and patient comfort. The longer the time of exposure and the greater the intensity setting, the more efficient the cleaning will be.
For patients experiencing Hydrosound system cleaning for the first time, it is recommended that the intensity is set at the lower levels before using the higher intensity settings.
Use Hydrosound® System (10 steps) Before starting the Hydrosound system:
1 Fill the bathtub with water. Make sure that the water level covers the top of the Hydrosound system Transducer Plate (A). (See Fig. 1)
2 Check the water temperature by hand to make sure it is not too cold or hot.
Continue with the steps on the next page.
WARNING
To avoid injury, never use the Hydrosound system function when pregnant.
WARNING
To avoid damage to hearing, make sure to keep the patients head above water when using the Hydrosound system.
CAUTION
Only use ArjoSound water conditioner with the Hydrosound system. Do not use bath oils and other skin conditioners with the Hydrosound system as they may inhibit the ultra sound pressure waves.
Fig. 1 A
3 Add ArjoSound water conditioner by pressing the ArjoSound water conditioner button on the control panel and check that the conditioner mixes into the water.
4 Start the Hydrosound system by pressing the Hydrosound system button on the control panel. The default runtime is 15 minutes after which the Hydrosound system will automatically end. Wait 15 minutes to restart.
5 Stop the Hydrosound system at any time by pressing the Hydrosound system button again.
6 On P220, it is possible to select different intensity levels by turning the knob on the upper right side of the bathtub. (See Fig. 3)
• The intensity is increased by turning the knob clockwise.
On P300 select intensity by pressing the, plus (+) and minus (-) on the display. (See Fig. 4)
• The levels 20, 40, 60, 80 or 100 mW/cm2 and test can be chosen step by step.
7 The LED (A) on P220 or the LED (B) on the Hydrosound button on P300 will show a steady green light and a buzzing sound from the transducer plate will be heard when any active level is selected. (See Fig. 5)
• The buzzing sound will increase with each higher level.
8 The test level is used to check the intensity level of the
Hydrosound and to make sure it’s in its specific range. If it passes the test it turns itself off and the Hydrosound system needs to be restarted.
• P300 will get a TEST text in the display.
9 If it fails the test, the P220 will flash a yellow light when in test mode. The P300 will show -04- in the display as an error. Contact an Arjo service technician if it fails the test.
• The same yellow light will flash in normal mode if problem occurs while using the Hydrosound system.
10 When finished, press the Hydrosound system button to stop.
WARNING
To avoid burns, avoid continuous contact with the Hydrosound system transducer plate while in operation. (See Fig. 2)
Fig. 3
Fig. 2
Fig. 4
Fig. 5
A
B
Sound & Vision System The Sound & Vision® system is a pleasant colourful underwater lighting and soft music stimulus which enhances the benefits of bathing. It calms a stressed or anxious patient.
The Sound & Vision system can be combined with still bath, Hydromassage system or the Hydrosound system.
Use Sound & Vision System (4 steps) 1 To activate the Sound & Vision system:
• Start by opening the control panel lid.
• Then, unscrew the black cap counter clockwise to get to the USB outlet.
• Plug in the Sound & Vision system memory stick.
2 Press the MP3 button on the hand control to start the music. Only MP3 files can be read and only files from the root (not in “folders”).
• the Sound & Vision system always starts of at a very low volume. Use the volume buttons to adjust volume to desired level.
• the Sound & Vision system hand control LED flashes when active.
3 Press the Sound & Vision system light button on the hand control to get colored light in the bathtub.
• Press the one more time to change light colour.
• Choose between red, green, blue, pink and yellow.
4 To start a rotation of different light colours press the automatic light colour change button.
Press the Sound & Vision system light button for 2 s to turn the light off.
WARNING
To avoid electric shock, only use the USB memory stick in the connector on the bath.
NOTE
Operating the hand control or control panel in the vicinity of other electrical appliances may lead to unintended compatibility actions with other equipment.
Sound & Vision System Hand Control 1 LED - activity indicator
2 Sound & Vision system Light - ON/OFF
3 Automatic light colour change - ON/OFF
4 MP3 player ON/PAUSE
5 Reduce (-) volume.
6 Increase (+) volume
7 MP3 track forward
8 MP3 track reverse
The music function will stop automatically 30 minutes after activation. Press the MP3 player button on the hand control for continued listening.
8 7
Change Default Setting (P300 only) (4 steps) 1 To change the default setting, press the hidden button and one of the function buttons (see section Control Panels on page 12). Press the two buttons simultaneously for at least 2 s. When pressing the hidden button a beep will be heard.
Function Button
Alternative Settings (Display Code)
Settings Interval (Preset) Default values
Filling Filling temp. 18-43 °C (Europe, Japan)
Filling temp. 64-109 °F (USA)
Filling time of the preset water level settings can be selected as:
High: 60 s from reaching top level sensor.
Normal: 30 s from reaching the top level sensor.
Low: 200 s from reaching the low level sensor.
Each level can be adjusted individually from -200 s to 500 s
37 °C (Europe, Japan)
99 °F (USA)
60 s High
30 s Normal
-20 s Low
Showering Shower temp. 18-42 °C (Europe, Japan) Shower temp. 18-41 °C (UK) Shower temp. 64-108 °F (USA)
37 °C (Europe, Japan)
37 °C (UK)
99 °F (USA)
Disinfect Disinfect. temp. 20-40 °C (Europe, Japan) Disinfect. temp. 68-104 °F (USA)
25 °C (Europe, Japan)
77 °F (USA)
Bath Oil
ArjoSound water conditioner
Preset Bath Oil/ArjoSound water conditioner dosage time within 1 to 7 s
If set to 0 s, the dosage will continue as long as the Bath Oil/ArjoSound water conditioner button is pressed.
1 s = 4 ml Bath Oil/ArjoSound water conditioner
3 s (12 ml)
Shampoo Preset shampoo dosage time within 1 to 7 s If set to 0 s, the dosage will continue as long as the Shampoo button is pressed.
1 s = 4 ml Shampoo
3 s (12 ml)
Hydrosound System
Preset the Hydrosound system intensity 20, 40, 60, 80 or 100 mW/cm²
60 mW/cm²
2 To toggle between two settings press the hidden button.
3 Adjust values with the plus (+) and minus (-) buttons.
4 Press the selected function button to confirm the setting before continuing.
Bathe Patient using a Lift Hygiene Chair Preparations for Bath (5 steps) Alenti, Calypso and Combilift are the recommended Arjo lift hygiene chairs that can be used with the System 2000 bath.
See Alenti, Calypso and Combilift IFU before use.
1 Lower the bathtub to lowest height. (See Fig. 1)
2 Press the Autofill button to fill the bathtub. Select Autofill level on P300. (See section Product Description/Functions on page 16.)
Place your naked hand in the water stream and make sure the water temperature is not to hot or to cold.
• P220: The display will oscillate in the beginning of the filling. Wait until the desired water temperature is reached on the display. If needed, adjust the temperature with the temperature knob.
• P300: The display will flash the current temperature until the preset temperature is reached. If needed, adjust the temperature with the plus (+) and minus (-) buttons.
4 Close the drainage plug and attach the pillow.
5 Additional filling is achieved by continuously pressing the Autofill button.
3 WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
Transfer and Bathe Patient (13 steps) 1 Undress the patient and cover him or her with towels to keep warm.
2 A sitting patient can be transferred from their bed or wheelchair into the bath with a lift hygiene chair.
(See Fig. 1)
Transfer the patient to the bath.
4 Position lift hygiene chair diagonally to the side of the bath with sufficient sideways clearance. (See Fig. 2)
Raise the lift hygiene chair to where it and any genitals clears the bathtub rim.
Move the lift hygiene chair towards the bath and help the patient to move his or her legs over the bathtub rim while rotating the lift in place. Let the patient feel the water with his or her legs. Adjust temperature, if needed. (See Fig. 3)
Continue with the steps on the next page.
3 WARNING
To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement.
5 WARNING
To avoid entrapment or pinching of genitals, make sure there is enough clearance during movement over the tub edge, toilet, bedpan or other furniture.
6 WARNING
To avoid entrapment of the patients or caregivers legs or feet, make sure that they are kept clear of all obstacles.
Fig. 1
7 Centre chair over the bathtub until it reaches a stop against the bath. Make sure the lift hygiene chair is docked against the bath legs and the bathtub rim. (See Fig. 4)
8 Adjust the lift hygiene chair to an ergonomic working height. (See Fig. 5)
Raise the bath, but make sure that there is enough room for body parts between the bottom of the bath and the seat.
10 If the patient desires and is capable, there is the option of folding away the back rest.
• Lean the patient forward and place their hands on the driving handle.
• Then move the back rest up and out of the way, thus allowing the patient to recline against the back of the bathtub and pillow.
11 Attach the foot support if needed.
12 Bathe the patient.
13 Use the Hydromassage, Hydrosound or Sound & Vision system, if available.
9 WARNING
To avoid entrapment or pinching of genitals, make sure there is enough clearance during movement over the tub edge, toilet, bedpan or other furniture.
Fig. 5
Shower Patient (14 steps) 1 After bathing, open the drain and let the water run out.
2 Let the patient place their hands on the driving handle and lean forward. Move the back rest to its original position.
3 Make sure that the patient keeps holding on to the driving handle.
4 To separate the chair seat from the bottom of the bathtub, lower the bathtub to its lowest height or raise the lift hygiene chair.
5 Lift off the shower handle, hold it over the bathtub.
6 Press shower button on the control panel (the LED lights up green when active).
7 Direct the shower handle away from the patient.
Press the trigger on the shower handle and direct the water stream away from the patient. Place your naked hand in the water stream and make sure the water temperature is not to hot or to cold.
Continue with the steps on the next page.
8 WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the
9 • P220: Check the water temperature by placing your naked hand in the water stream. (See Fig. 1) If needed, adjust the temperature with the temperature knob.
• P300: The display shows the temperature value. Allow a few seconds for the water to reach the preset temperature.
Check the temperature setting on the display. If needed, adjust the temperature on the control panel with the plus (+) and minus (-) buttons.
10 Let the patient to feel the water by hand before directing the water stream on the patient. Adjust water temperature, if needed.
11 To keep the trigger handle in for a longer period of time, activate the lock button by pushing it in. To unlock, push the trigger handle.
12 Shower the patient. (See Fig. 2)
13 To stop the shower, press the shower button on the control panel again.
14 Make sure the shower trigger is unlocked before placing the shower handle back into the holder.
End the Bath (9 steps) 1 Dry the patient, as well as the underside of the seat, to prevent water from dripping on the floor.
2 Place one or two towels on the patient or a bathrobe, if possible, to keep him or her warm.
3 Lower the bathtub to its lowest height.
Raise the hygiene chair to where it and any genitals clears the bathtub rim.
5 Help the patient to move his or her legs over the edge of the bathtub while rotating the lift away from the bathtub. Make sure there is sufficient sideways clearance between the bathtub and lift hygiene chair. (See Fig. 1)
6 Lower the lift hygiene chair.
7 Dry the remaining wet body areas.
8 Adjust the lift hygiene chair to an ergonomic transfer height for the caregiver. (See Fig. 2)
9 When transferring the patient straight from the bathtub to bed, put a towel on the bed under the lift hygiene chair seat to avoid making the bed wet. Dry the patient thoroughly when lying down in bed.
Disinfect After Each Patient (1 step) 1 Make the bath ready for the next patient. The bath and lift hygiene chair must be fully disinfected after each use to prevent cross contamination. See section Cleaning and Disinfection Instructions on page 40. (See Fig. 3)
For Alenti Disinfection, see Alenti IFU.
4 WARNING
To avoid entrapment or pinching of genitals, make sure there is enough clearance during movement over the tub edge, toilet, bedpan or other furniture.
Fig. 1
Bathe Patient using a Lift Bath Trolley Miranti and Bolero are the recommended Arjo lift bath trolleys that can be used with the System 2000 bath.
Compatible with bath sizes: 1870 mm (73 5/8”), 1900 mm (75”), 2120 mm (83 1/2”) and 2150 mm (85”).
See Miranti and Bolero IFU before use.
Preparations for Bath (7 steps) 1 Lower the bath to its lowest height. (See Fig. 1)
2 Remove the foot support and the pillow.
3 Press the Autofill button to fill the bathtub with water.
Select autofill level on P300 (see section Product Description/Functions on page 16).
5 Place your naked hand in the water stream and make sure the water temperature is not to hot or to cold.
• P220: The display will oscillate in the beginning of the filling. Wait until the desired water temperature is reached on the display. If needed, adjust the temperature with the temperature knob.
Continue with the steps on the next page.
CAUTION
To prevent equipment damage when using a lift bath trolley, the foot support and pillow must be removed.
4 WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the
• P300: The display will flash the current temperature until the preset temperature is reached. If needed, adjust the temperature with the plus (+) and minus (-) buttons.
6 Close the drainage plug.
7 Additional after filling is achieved, by continuously pressing the Autofill button.
Transfer and Bathe (13 steps) 1 A bedridden patient can be transferred from bed with the help of a Miranti lift bath trolley.
2 Undress the patient and cover him or her with towels to keep warm.
3 Position the lift bath trolley behind and to the side of the bath.
4 Raise the lift bath trolley stretcher up in order to clear the edge of the bathtub. (See Fig. 1)
5 Move the lift bath trolley with patient diagonally in over the bathtub. (See Fig. 2)
6 Move the trolley centrally over the bathtub until it reaches a stop against the bathtub legs and rim. (See Fig. 3)
Continue with the steps on the next page.
CAUTION
To prevent damage to the equipment, make sure the “Duck Foot” bath legs go underneath the lift bath trolley chassis.
Fig. 1
Manual Drive or Power Drive
7 Make sure the stretcher is centred over the bathtub.
8 Adjust the lift bath trolley to an ergonomic working height.
(See Fig. 4)
9 Raise the bath so that the stretcher is touching the bottom of the bathtub. (See Fig. 5)
10 Lower the back rest to a comfortable position for the patient.
11 Attach the foot support if needed.
12 Bathe the patient.
13 Use the Hydromassage, Hydrosound or Sound & Vision system, if available.
Use the Shower (13 steps) 1 After bathing, open the drain and let the water run out.
2 Raise the back rest and let the patient place their hands on the security grip.
3 To separate the stretcher from the bottom of the bathtub, lower the bathtub to its lowest height or raise the lift bath trolley.
4 Grab shower handle, hold it over the bathtub.
5 Press shower button on the control panel (LED lights up green when active).
6 Direct the shower handle away from the patient.
Press the trigger on the shower handle and direct the water stream away from the patient. Place your naked hand in the water stream and make sure the water temperature is not to hot or to cold.
8 • P220: Check the water temperature by placing your naked hand in the water stream. If needed, adjust the tempera-ture with the temperature knob. (See Fig. 1)
• P300: The display shows the temperature value. Allow a few seconds for the water to reach the preset temperature.
Check the temperature setting on the display. If desired, adjust the temperature on the control panel with the plus (+) and minus (-) buttons.
Continue with the steps on the next page.
7 WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the
9 Let the patient to feel the water by hand before directing the water stream on the patient. If needed, adjust water temperature.
10 To keep the trigger handle in for a longer periods of time, activate the lock button by pushing it in. To unlock, push the trigger handle.
11 Shower the patient. (See Fig. 2)
12 To stop the shower press the shower button again.
13 Make sure the shower trigger is unlocked before placing the shower handle back into the holder.
End the Bath (6 steps) 1 Dry the patient, as well as the underside of the stretcher, to prevent water from dripping on the floor.
2 Place one or two towels on the patient to keep him or her warm.
3 Lower the bathtub to lowest height and raise the stretcher.
(See Fig. 3)
4 Manually pull the lift bath trolley out approximately 30 cm (1 foot) from the bathtub (A). Then activate the power drive on the handle bar at the foot end and move the lift bath trolley along the side of the bath (B).
(See Fig. 4)
5 If there is no power drive, just pull the lift bath trolley away manually along the side of the bath. (See Fig. 5)
6 When moving the patient straight from the bath to bed, put a towel on the bed under the lift bath trolley stretcher at to avoid making the bed wet. Dry the patient thoroughly when lying down in bed.
Fig. 2
Fig. 3
Power Drive
B
A
Manual DriveFig. 5
Disinfect After Each Patient (1 step) 1 Make the bath ready for the next patient. The bath and the lift bath trolley must be fully disinfected after each use to prevent cross contamination. See section Cleaning and Disinfection Instructions on page 40. (See Fig. 1)
For cleaning and disinfection of the Miranti lift bath trolley, see the Miranti IFU.
Cleaning and Disinfection Instructions Thorough disinfection routines are necessary to stop the build-up of bacteria on and in the equipment. This is particularly vital in cases where the System 2000 bath is in a shared facility.
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use.
WARNING
To prevent cross contamination, the caregiver must always be present during the whole disinfection procedure to make sure it is performed correctly.
WARNING
To avoid eye or skin irritation, never disinfect in the presence of a patient.
To avoid eye and skin damage, always use protective glasses and gloves. If contact occurs rinse with plenty of water. If eyes or skin becomes irritated, contact a physician. Always read the material safety data sheet of the disinfectant.
CAUTION
To avoid damage on the equipment only use Arjo branded disinfectants.
Arjo Disinfectants For optimal performance only use Arjo Disinfectants.
If you have any questions regarding disinfecting the equipment or wish to order disinfectant (see section Parts and Accessories on page 69), please contact your Arjo Customer Service located on the last page.
Accessories for Cleaning and Disinfection the System 2000 bath
• Protective glasses
• Protective gloves
• Disinfectant filled spray bottle or the disinfection spray handle.
• Water filled spray bottle or the shower handle
• Cloths – wet and dry
• Disposable towels
• Soft bristled brush
• Long handled soft bristled brush
Microbiological Formation Prevention In order to enhance the on-site curative actions (chlorine or thermal shock), which do not guarantee a reduction of the contamination long term, make sure to follow the instructions below.
Do not forget to contact Arjo for the annual Preventive Maintenance visit.
Installation precautions In order to limit the growth of bacterial flora, it is necessary to take action at four levels:
• Avoid stagnating water and make sure proper water circulation;
• Take precautions against the formation of scale and corrosion based on the water quality;
• Maintain a high water temperature all the way from the hot water dispenser to the different distribution pipes;
• Mix hot and cold water as close as possible to the point of use.
Use
• Make sure that the water circulates in the bath and the shower on a daily basis even if the bathtub is not used; and particular to make sure to remove any water that may be left behind in the hose.
• Let the water flow approximately 5 minutes before the first bath of the day.
• Clean and disinfect the bathtub according to the IFU before the first bath of the day and after the bath of each patient.
Replace Disinfectant Bottle (System 2000 bath with disinfection option only) (12 steps)
1 Put on protective glasses and gloves.
2 Open the compartment for the disinfectant bottle.
(See Fig. 1)
3 Remove bracket and shampoo/bath oil bottles, if any.
4 Disconnect the suction part from the disinfectant bottle.
(See Fig. 2)
5 Discard the old bottle and replace with a new one.
6 Connect the suction part to the new bottle.
7 Grab the disinfection spray handle.
8 Press the disinfection button.
9 Adjust the water temperature until it reaches approximately 25 °C (77 °F) to eliminate disinfectant fumes.
10 Spray disinfectant in the bathtub. Keeping an eye on the flowmeter (See Fig. 3).
11 Check the liquid mix ratio on the flowmeter so that it is in accordance with the setting range stated in this IFU.
(See Fig. 3)
12 If necessary, adjust the bottle cap by hand to get the right mix ratio. (See Fig. 4)
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use.
Fig. 1
Fig. 2
Fig. 3
Arjo Clean™ (60)
Cen-Kleen IV ™(40) Arjo Disinfectant Cleanser IV (35) Arjo General Purpose Disinfectant (35)
Arjo Guard (60)
Clean and Disinfect with Disinfection Spray Handle (Still Bath and Hydrosound System) (37 steps) Remove visual residue (Step 1-6 of 37) 1 Check the level of disinfectant.
2 Raise the bathtub to convenient working position.
3 Make sure all functions are turned off.
4 Open drain to empty any left over water.
5 Detach and place the accessories (if applicable) such as pillow, foot support and toiletry tray at the bottom of the bathtub.
6 Shower and rinse the bathtub and the accessories in the bottom of the bathtub with water to remove visible residue.
Turn the accessories and rinse again. (See Fig. 1)
Clean (Step 7-23 of 37) 7 Put on protective gloves and eye wear. (See Fig. 2)
8 Open the panel lid, lift out the disinfection spray handle and hold it over the bathtub.
9 Press the disinfection button to start.
10 Hold the disinfection spray handle over the bathtub rim, then press the trigger. Keep an eye on the flowmeter.
11 Spray until the flowmeter shows the correct liquid mix ratio.
If the flow meter value is not correct, adjust it according to section Replace Disinfectant Bottle (System 2000 bath with disinfection option only) on page 42.
12 Adjust water temperature so that it reaches approximately 25° C (77 °F) (P220 only) to eliminate disinfectant fumes.
13 Spray disinfectant in the drain. Use a long handled brush to scrub the drain.
14 Close the drain.
Continue with the steps on the next page.
15 Spray disinfectant using the disinfection spray handle on all internal bathtub surfaces and the accessories at the bottom of the bathtub. Turn the accessories and spray on the other side. Pay attention to parts and knobs and sensors attached to the bathtub surface such as drain control knob/surface overflow outlet. (See Fig. 3)
16 Spray into the surface overflow outlet. (See Fig. 4)
17 Press the…