This document appears to be a set of technical drawings and details related to the installation of a Liko Overhead Patient Lift System. The drawings provide typical attachment details for wall-mounted Liko Slimline supports with H100, H140, H160, and H180 rail systems, including dimensions, material specifications, and installation requirements. This information is provided for informational purposes only, and project-specific calculations may be required per local building codes.
The related federal contract opportunity is for a solicitation to demolish and replace an MRI unit in Building 84 at the Roseburg VA Medical Center in Roseburg, OR. Key details include the solicitation number (36C26024R0016), a site visit scheduled for 3/27/2024, and advice for offerors to ensure their vendor registration and SAM record are up-to-date to avoid delays. The magnitude of the project is not specified, but interested parties are advised to review the full solicitation and attachments for complete details.
View the file
Other files
Show all 21
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PL-12
This drawing is not to be used as a construction drawing, unless noted as such. Accuracy of this drawing is based upon information provided to hill-rom. Optimal arrangement to be determined at time of installation.
Sheet of
Project #
Drawn by: Date: Scale:
Liko Overhead Patient Lift Systems
Typical Method of Attachment Details - Wall Mtd
Utilizing LIKO-supplied Upright Supports
Liko Slimline Wall Support Details and Material Properites
Makaila Shirley 3/18/2022 As Noted 12 16
8"
Floor
8'-0" Ceiling Height
Floor
Liko Slimline Support
Liko H100 Rail
2'-113 8"
2'-113 8"
Height varies depending upon ceiling height and rail placement height
Finished Ceiling H100 Bayonet Insert Rail
3"
1" Approx.
3-7/8" Ref
16" x 4" Toggle Bolt Typical 3 Places
H100 Bayonet Insert Rail
M8 x 45mm Bolt w/ Locking Nut
4"
2'-117 8"
Enlarged View
Side Elevation ViewFront Elevation View
H100 Bayonet Length 1'-75 8"
Recommended to install with minimum bayonet exposure
Plan View 21 8"2"
8"
Floor
Front Elevation View
8'-0" Ceiling Height
Liko Slimline Support
Liko H140 Rail Shown (Applicable to H160 & H180 Rails)
2'-113 8"
2'-113 8"
Height varies depending upon ceiling height and rail placement height
H140 - H180 Bayonet Insert Rail
1" Approx.
5-1/2" Ref
16" x 4" Toggle Bolt Typical 3 Places
H140 - H180 Bayonet Insert Rail
M8 x 45mm Bolt w/ Locking Nut
4"
2'-117 8"
Enlarged View
Side Elevation View
Floor
Finished Ceiling
Plan View 21 8"2"
H140 - H180 Bayonet Length 3'-33 8"
Recommended to install with minimum bayonet exposure
* FOR INFORMATION PURPOSES ONLY *
PROJECT-SPECIFIC CALCULATIONS (BY
"OTHERS") MAY BE REQUIRED.
REFERENCE LATEST CALIFORNIA BUILDING
CODE OR LATEST BUILDING CODES.
REFER TO OSHPD-APPROVED DRAWINGS
and DETAILS.
Hillrom is now a part of Baxter
AutoCAD SHX Text
AutoCAD SHX Text Liko Slimline Wall Supports with H140, H160, or H180 Rails
AutoCAD SHX Text 1" = 1'-0"
AutoCAD SHX Text
AutoCAD SHX Text Liko Slimline Wall Supports with H100 Rail
AutoCAD SHX Text 1" = 1'-0"