657A4-24-105-2024-12-10_SD Report_RF_Room 2030_JJP VAMC_FH22194_JSM_DLD.pdf

PDF 792 KB Posted

Attached to
Z2DA--657A4-24-105, Site Prep for Fluoro and Xray, PB
Solicitation number
36C25525R0013
Issued by

About this file

This is a radiation shielding design evaluation report prepared by West Physics for a radiographic and fluoroscopic suite at John J. Pershing VA Medical Center in Poplar Bluff, MO. The evaluation specifies shielding requirements for a Philips ProxiDiagnost N90 unit in Room 2030, with detailed barrier specifications for walls, doors, windows, floor, and ceiling.

The report provides comprehensive shielding specifications including 1/32" lead or equivalent for most barriers, with specific requirements for doors (16-gauge steel for corridor door, 7-gauge steel for X-ray room door) and concrete requirements for floor (2-1/8") and ceiling (2"). The facility's expected workload is 176 radiographic patients per week at 2.5 mA-min per patient and 9 fluoroscopic patients per week at 13 mA-min per patient. The evaluation includes detailed facility diagrams, regulatory requirements, and installation recommendations certified by John Muryn, M.S., DABR (Missouri Qualified Expert – License No. QE295) on December 10, 2024.

View the file

Other files

Show all 17

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

The Medical Physics Experts

W E S TW E S T P H Y S I C S

3 0 3 3 9S u i t e 2 5 0 A t l a n t a , G A3 8 2 5 P a c e s W a l k S E 866-275-WEST (9378) www.westphysics.com

December 10, 2024

Lee Lovette John J. Pershing VA Medical Center 1500 North Westwood Blvd.

Poplar Bluff, MO 63901

RE: Radiographic & Fluoroscopic Suite Shielding Design Evaluation – John J. Pershing VA Medical Center 1500 North Westwood Blvd., Poplar Bluff, MO 63901

Dear Lee Lovette, Please find enclosed the radiation shielding design evaluation for your Radiographic & Fluoroscopic suite.

The first page of the shielding plan summarizes the total minimum required shielding for each applicable room barrier. Additional information supporting the shielding recommendations constitutes the remaining sheets. You should retain a copy of this shielding design evaluation at your facility in the event that a regulatory agency wishes to verify the room’s shielding. We expect that a regulatory audit of the installation will take place, and an inspector will likely request this document at that time.

Thank you for the opportunity to take care of your radiation protection needs. Should you have any questions or comments about this shielding design, we would be happy to speak with you at your convenience. Please feel free to contact us at (866) 275-9378 or at shielding@westphysics.com.

Thank you, John Muryn, M.S., DABR Senior Medical Physicist

Dylan DuCharme, B.S.

Medical/Health Physicist

Enclosure: Radiation Shielding Design Evaluation

Radiation Shielding Design Evaluation Performed by

John Muryn, M.S., DABR

Site: John J. Pershing VA Medical Center 1500 North Westwood Blvd.

Poplar Bluff, MO 63901

Room: 2030

Unit: Philips ProxiDiagnost N90

Shielding Design Results:

Barrier Function of

Space Behind Barrier

Controlled or Uncontrolled

Occupancy Factor (T)

Distance (ft, in)

Total Minimum Shielding Required

A-B Outdoor Area Uncontrolled 0.025 8' 8-3/4" 1/32" lead or equivalent

B-C Staff Room Uncontrolled 0.200 11' 9-1/2" 1/32" lead or equivalent

C-D Control Room Controlled 1.000 7' 1-7/8" 1/32" lead or equivalent

C-D (Window)

Control Room Window Controlled 1.000 7' 1-7/8" 1/32" lead equivalent glass

E-F Corridor Uncontrolled 0.200 7' 6-1/4" 1/32" lead or equivalent

E-F (Door) Corridor Door Uncontrolled 0.125 7' 9-3/8" 1/32" lead, or 16-gauge steel, or equivalent

F-A X-Ray Room Controlled 1.000 9' 9-1/2" 1/32" lead or equivalent

F-A (Door)

X-Ray Room Door Controlled 1.000 4' 8-3/8" 1/32" lead, or 7-gauge steel, or equivalent

Floor Primary Care Uncontrolled 0.500 12' 1" 1/32" lead, or 2-1/8" concrete, or equivalent*

Ceiling Isolation Rooms Uncontrolled 0.500 10' 8" 1/32" lead, or 2" concrete, or equivalent*

*Required concrete floor and ceiling may already be in place. Please double check concrete floor and ceiling thickness.

Facility Diagram:

Note: Cardinal directions are approximate only. Shielding should be installed carefully and with respect to the function of the surrounding areas to ensure proper barrier thicknesses.

Other Recommendations and Requirements

• If the X-Ray room ceiling and the ceiling of adjacent rooms does not meet the minimum requirement specified in the “Shielding Design Results” table, then all vertical barriers shall extend to overlap the ceiling shielding. Otherwise, all vertical barrier shielding shall extend to a height of at least 7 feet.

• Doors, windows, frames, conduits, and wall openings must be of the specified shielding equivalencies or greater and should have no voids.

• Joints between lead sheets should be constructed so that their surfaces are in contact and with an overlap of at least 1 cm.

• Any penetrations in a given barrier should be designed to afford the same shielding equivalency as that specified for the barrier. For example, pipes, electrical outlets, or other barrier penetrations must be wrapped, covered, or backed up with appropriate shielding-equivalent material.

• A qualified contractor with experience in shielded construction should be utilized and a follow-up inspection performed to verify the adequacy of the installation after completion.

• This shielding design is for the equipment layout and occupancy of adjacent areas as indicated on the attached drawing. If any of these parameters are changed, the shielding design will require re-evaluation by a qualified physicist.

• The shielding requirements specified in this plan review were based on the stated conservative workload(s). If the patient volume increases significantly, the shielding will require re-evaluation by a qualified physicist.

Regulatory Recommendations and Requirements

• The registrant must ensure that each radiation machine is labeled in a conspicuous manner, which cautions individuals that radiation is produced when it is energized.

• Viewing system(s) shall be provided to permit continuous observation of the patient during irradiation and shall be located so that the operator can observe the patient from the control panel.

• The operator shall be allotted no less than 7.5 ft2 of unobstructed floor space in the booth, exclusive of any encumbrance by the X-Ray control panel such as overhangs, cables, etc.

• The X-Ray control shall be permanently mounted in a protected area so that the operator must remain in that protected area during the entire exposure. This control shall be at least 40 inches from any point subject to direct scatter, leakage or primary beam radiation. If a mobile barrier is used to create the protected area, the barrier must be permanently mounted to the wall and/or floor.

Additional Shielding Design Calculation Information John J. Pershing VA Medical Clinic

All calculations were performed in accordance with industry-standard guidance as detailed in NCRP Report No. 147, “Structural Shielding Design and Evaluation for Medical X-Ray Imaging Facilities.” Wherever possible, vendor and model-specific information for the x-ray units and/or radioactive sources in question were used. Where appropriate, calculations account for attenuation by primary beam pre-shielding using the equivalent thickness values listed in Table 4.6 of NCRP Report No. 147. Patient throughput estimates as well as workload kVp distributions were used in the following calculations, based on average values contained in NCRP Report No. 147. The recommended shielding amounts are specified in a highly conservative fashion, which will allow for a broad range of patient workloads, technique factors, etc.

without creating a radiation exposure issue. However, if any major changes occur (i.e., replacement of the x-ray unit, facility/x-ray room structural layout and/or changes in the usage of the x-ray room surrounding areas), the facility administration is advised to seek the services of a qualified medical physicist to determine whether shielding additions would be needed in those cases.

All shielding design goals, occupancy factors and workloads are as specified in NCRP Report No. 147 unless stated otherwise.

Equipment Specifications and Workload (W):

Equipment Type Radiographic Patients per Week

Radiographic mA-min per patient Workload (W)

Philips ProxiDiagnost N90 176 2.5 440 mA-min/week

Equipment Type Fluoroscopic Patients per Week

Fluoroscopic mA-min per patient Workload (W)

Philips ProxiDiagnost N90 9 13 117 mA-min/week

Shielding Design Goals (P):

Area Type Shielding Design Goal (P) (mGy/week)

Controlled 0.10 Uncontrolled 0.02

Methodology:

For this shielding design evaluation, the fitting parameter method for shielding calculation was employed, based on manufacturer data and information provided by the user. The room will be used primarily for general purpose exams directed towards a table-top and chest bucky image receptor. Therefore, barriers in this room are considered to be both primary and secondary barriers; scatter and leakage radiation values were incorporated. This shielding evaluation assumed a conservative number of exposures per week, based on facility estimates of its typical weekly workload. Equation A.3 of NCRP Report No. 147 was used to compute required barrier thicknesses. Fitting parameters for this equation for lead, concrete, glass, and gypsum wallboard thicknesses were found in Tables B.1 and C.1. Patient exam technique was based upon the typical workload distributions defined in Table 4.3 of NCRP Report No. 147 along with facility estimates.

All barrier thickness calculations were performed in Excel. These are available from West Physics and may be furnished upon request of the detailed calculations.

Summary:

The above referenced calculations and recommendations are made by West Physics based on data provided by the facility planners and unit manufacturer. As such, the recommendations made herein are subject to that information’s accuracy. This shielding design is hereby certified and executed:

John Muryn, M.S., DABR Senior Medical Physicist State of Missouri Qualified Expert – License No. QE295

12/10/2024 Date