This document contains a specification for a water quality monitoring station panel and related federal contract opportunity. The specification requires a pre-assembled single panel with mounted, wired and plumbed instruments including an online turbidimeter, chlorine residual analyzer, conductivity cell, pH sensor and pressure sensor. The panel must fit within standard enclosure dimensions and operate within specified pressure and voltage parameters. The contract opportunity is a solicitation to upgrade plumbing and install monitoring stations and mixing valves at two buildings of the Clement J. Zablocki VA Medical Center to prevent Legionella. The project scope includes removing cross-connected lines and dead legs, installing recirculation lines and balance valves, adding thermostatic mixing valves to fixtures, and providing four monitoring stations. Asbestos abatement and architectural patching will be performed.
View the file
Other files
Show all 22
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specification for the Water Quality Monitoring Station
I. General Description
A. The Water quality Monitoring station shall be a single panel with instruments pre-mounted, wired and plumbed.
1. The back plane of the panel shall be not greater than 22” wide by 48” long and made of corrosion resistant material.
2. The back plane shall be affixed to rigid corrosion resistant struts approximately 52” long centered to extend approximately 2” past either end of the back plane
3. The struts shall be predrilled to permit wall mounting of the entire panel using 1/4” steel bolts; one at each end of the two struts.
4. The panel shall be fully assembled with all instruments, except controller, mounted, plumbed and wired prior to shipment. The controller for the panel shall be mounted separately.
5. The panel when fully assembled shall fit inside a standard NEMA 4X 36”X60” enclosure without modifications to the enclosure or panel except to provide for hydraulic and electrical connections.
6. The sample and drain system for the panel shall accommodate flows of 400 to 600 ml/min.
7. A common drain system shall be provided for all instruments. The drain manifold shall be terminated with 1/2” hose barb fitting for connection of a flexible drain tube.
8. Pressure and flow control shall be via a fixed orifice and constant head devices.
i. The constant head device shall provide pressure and flow regulation for all sensors down steam from the orifice device.
ii. No rotameters or other flow control devices shall be required.
iii. Within a pressure range of 20-100 psi, no additional pressure regulation shall be required.
iv. The constant head device shall provide a mounting mechanism for the pH and conductivity sensor as well as one additional probe- type sensor.
B. The panel shall include the following on-line analytical instruments (See Section II for complete instrument descriptions): A turbidimeter, a chlorine residual analyzer, a conductivity cell and a pH sensor. The conductivity and pH sensors shall have built-in temperature measurement and compensation.
1. All analytical instruments shall be pre-wired to a common junction box via quick-connect connectors.
2. All analytical sensor inputs and sensor power shall be supplied from the junction box.
3. The junction box shall provide all sensor outputs to the controller and receive all sensor inputs from the controller via a single shielded cable.
4. All devices on the panel shall operate at 100-115/230Vac ~50/60 Hz.
5. Total power requirement shall not exceed 500 VA.
6. All instruments shall be powered using a single connection using a modular cord or wiring with conduit.
7. All devices on the panel shall be capable of operating at a pressure of 15 - 60 psig.
C. The controller for the panel shall be:
1. Input: 95-240 VAC, 50/60 Hz
2. Control: Equal to input voltage (95-240VAC) fused at 2.5A per relay, prewired units are supplied with an 8’ (248.84 cm) power cord and 8” (20.32 cm) output receptacles.
3. Display: 240 X 128 Graphic LCD.
4. Conductivity Control: uS/cm, mS/Cm, PPM scales with ranges up to 50,000 depending on sensor selection. Boiler sampling: continuous, timed or sample and hold.
5. pH: 0-14 scale
6. ORP: +/- 1000 millivolts scale, The ORP setpoint can be tied to an optional 28-day feed timer as an override.
7. Accuracy: +/- 1%
8. Feed Timers are all selectable from:
• Pulse - water meter activated w/accumulator Recycle - OFF and ON cycle;
HH:MM:SS
• Limit - Limits feed with bleed
• Post - 1-100% of post bleed or other time with an overall limit.
• 28 Day - biocide timer.
9. Enclosure: Heavy duty NEMA 4X high impact ABS with pad lockable gasketed Lexan viewing door.
10. Environment: Ambient temperature: 0o to 125oF (-17o to 52oC); Relative humidity: 0 to 100%.
11. Electrodes: Tower Standard electrodes are supplied in 3/4” (1.91 cm) SCH 80 PVC female slip tees with quick release nut.
• Conductivity: TE-4A 150 psi (10.3 bar) / 140oF(58oC) Max
• pH: TPE-2 100 psi (6.8 bar) / 140oF(58oC) Max
• ORP: TOE-2 100 psi (6.8 bar) / 140oF(58oC) Max
• Temperature: TC-1 150 psi (10.3 bar) / 140oF (58oC) Max
• Conductivity: BE-32, 250 psi @ 400°F / 350 psig @ 265°F PEEK core
D. The panel shall include a pressure sensor for monitoring water pressure at the sample manifold.
1. The pressure sensor shall be a CVD Pressure Transducer with a 2 wire, 24 AWG, shielded PVC cable to carry the variable 4 – 20 mA output signal.
2. The pressure sensor shall have a measuring range of 0 – 150 psi.
3. The pressure sensor shall have an accuracy of +/-0.25% with +/- 1.5% thermal error band.
II. Analytical Instruments. All analytical instruments installed on the panel shall be from a single manufacturer to minimize maintenance and support issues.
A. On-line Turbidimeter:
1. The On-line Turbidimeter sensor shall meet requirements of US EPA method 180.1.
2. Range of the turbidity sensor shall be 0-100 NTU.
3. The analyzer shall be designed to eliminate the need for a sample cell to eliminate errors due to dirt and scratches.
4. The detector shall be immersed in the sample to prevent fogging.
5. The sensor shall be a digital design and receive sensor inputs from the panel’s controller and provide sensor outputs to the controller.
6. All sensor calibration information and sensor diagnostics shall be stored in the sensor not in a remote controller or other device.
7. All optical components – light source, lenses and detector – shall be mounted on a head assembly removable from the turbidimeter body without the use of tools or removal of any fastening device.
8. The turbidimeter body shall have a built-in bubble removal system.
9. Parts comprising the bubble removal system must be removable for cleaning without the use of tools.
B. Chlorine Residual Analyzer:
1. The analyzer shall be capable of measuring free or total residual chlorine utilizing the DPD colorimetric method.
2. The analyzer shall be configured for either free or total residual chlorine measurement by simply changing the reagent package.
3. Range of measurement shall be 0-5 mg/l for either free or total residual.
4. Reagents for the analyzer shall be prepared by the instrument manufacturer, and supplied in three components: Indicator powder, Indicator solvent and buffer.
i. The indicator powder and indicator solvent shall be designed to be mixed by the user immediately prior to use for optimum reagent shelf life and stability.
ii. The analyzer shall use approximately 1 pint each of indicator solution and buffer per month.
iii. The instrument manufacturer shall offer an automatic reagent replacement program for shipment of reagents once per month to eliminate warehousing costs.
5. The light source for the colorimeter shall be a long-life solid-state device.
6. The colorimeter cell shall have constant mixing provided by a magnet immersed in the sample and caused to rotate by an electromagnetic mixing device with no moving parts.
7. The analyzer shall have digital network capability based on LonWorks as well as standard 4-20 mA outputs.
8. The analyzer shall have an LCD display.
9. The enclosure of the analyzer shall be non-metallic and molded construction.
10. All components of the analyzer shall be accessible for servicing from the front of the instrument. No tools shall be required for access.
11. The door of the analyzer enclosure shall have shatter-resistant, plastic viewing windows for the LCD display and reagent levels.
12. The Cl-17 shall communicate with the panel controller via an analog/digital gateway device.
C. On-line pH Analyzer
1. The On-line pH Analyzer shall consist of a control unit and a pH sensor
2. Range of measurement shall be –2 to14 pH units
3. The pH sensor shall utilize the differential electrode technique.
i. The pH sensor shall have a long-life filling solution, which needs to be replenished not more frequently than once every five years.
ii. The pH sensor shall be constructed of PEEK
iii. The pH sensor shall be mounted in a pipe tee for a continuous flowing sample.
iv. The pH sensor mounting shall be designed for easy removal without the use of tools to expedite cleaning and calibration.
v. The pH sensor shall have built-in temperature sensor and temperature compensation.
4. The sensor shall be a digital design and receive sensor inputs from the panel’s controller and provide sensor outputs to the controller.
5. All sensor calibration information and sensor diagnostics shall be stored in the sensor not in a remote controller or other device.
D. On-line Conductivity Analyzer:
1. The On-line Conductivity Analyzer shall consist of a control unit designed to operate at 115Vac and a conductivity cell.
2. The conductivity cell shall have a cell constant of 1.0 and range of 0-2000 µS/cm.
i. The cell shall be calibrated by the DRYCAL calibration method.
ii. The cell shall be mounted in a pipe tee for a continuous flowing sample.
iii. The cell mounting shall be designed for easy removal without the use of tools to expedite cleaning and calibration.
iv. The conductivity cell shall have a built-in temperature sensor.
3. The sensor shall be a digital design and receive sensor inputs from the panel’s controller and provide sensor outputs to the controller.
4. All sensor calibration information and sensor diagnostics shall be stored in the sensor not in a remote controller or other device.
IV. Performance Testing –
A. All instruments offered as part of the panel must have been manufactured and offered for sale for a minimum of 24 months prior to the date of award of the bid.
V. Manufacturer’s Start Up and Training.
A. The supplier of the panel shall provide the services of a technical representative or service technician and employee from the instrument manufacturer for a minimum of 1 hour but not to exceed 2 hours for start -up and configuration of each panel provided.
B. The technical representative or service technician and employee of the manufacturer also shall offer a minimum of one (1) two-hour training sessions on care, calibration and maintenance of all instruments on the panel.
VI. Warranty
A. All analytical instruments on the Water Distribution System Monitoring Panel shall carry a warranty for a minimum period of 1 year from the date of invoice for all analytical instruments.
B. The warranty shall be the responsibility of the supplier awarded the bid.