This is a technical data sheet for a Siemens BiographVision 600 PET/CT imaging system installation at the VA Ann Arbor Healthcare System in Ann Arbor, Michigan. The document provides detailed floor plans and equipment specifications for the facility layout, dated August 1, 2025, and references project manager Nicholas Folk at Siemens HealthCare.

The equipment legend specifies 19 major components for the installation, including a BiographVision 600 gantry (6,701 lbs), patient table (1,587 lbs), common power distribution unit (952 lbs), heat exchanger cabinet with water/air split option (911 lbs), outdoor cooling unit (410 lbs), and various supporting systems such as image reconstruction, PET camera, uniform phantom shield, medrad display and base units, ceiling-mounted medrad injector, and LAP laser monitoring systems. The technical specifications document dimensional requirements (in inches), weight (in lbs), and BTU/hour ratings for each component. The floor plan indicates a pet scanner room, shared control room, and equipment room with specific clearance requirements for safety, service access, and equipment movement. The document includes an authorization section requiring facility stakeholder sign-off confirming understanding of equipment limitations and installation parameters, with responsibility assumed by the customer to ensure proper handling and compliance with specified operational constraints.

View the file

Other files

Show all 18

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Click to acquire Typical PDF technical data

SHEET OF

1 2

DATE:

08/01/25

DRAWN BY:

C. SCHANZ

REF:

30266182

SCALE: 1/8" = 1'-0"

REFERENCE DOCUMENT - NOT FOR CONSTRUCTION

PROJECT MANAGER: NICHOLAS FOLK

PHONE: (248)873-9912

EMAIL: NICK.FOLK@SIEMENS-HEALTHINEERS.COM

2215 FULLER RD, ANN ARBOR, MI 48105-2303

D758 & D754 - BIOGRAPH VISION 600

SIEMENS

D

08/01/25

INJECTOR SHIFTED

SYSTEM SHIFTED

https://siemenshealthineers13.autodesk360.com/g/shares/SH56a43QTfd62c1cd968a69bc65747b00f71 https://siemenshealthineers13.autodesk360.com/g/shares/SH56a43QTfd62c1cd968a69bc65747b00f71

SHEET OF

2 2

DATE:

08/01/25

DRAWN BY:

C. SCHANZ

REF:

30266182

REFERENCE DOCUMENT - NOT FOR CONSTRUCTION

PROJECT MANAGER: NICHOLAS FOLK

PHONE: (248)873-9912

EMAIL: NICK.FOLK@SIEMENS-HEALTHINEERS.COM

2215 FULLER RD, ANN ARBOR, MI 48105-2303

D758 & D754 - BIOGRAPH VISION 600

SIEMENS

I have reviewed the below referenced plan(s) and understand and agree with the proposed room configuration and equipment arrangement. By signing below, I agree that I have the authority to sign on behalf of the customer. I have been fully informed by Siemens about the limitations (if applicable) that exist regarding the proposed equipment installation. Siemens has explained that the listed limitations may interfere with certain functions of the equipment. We agree to assume responsibility to inform employees, agents, contractors, users or any other persons handling this equipment of these limitations and will assure that the equipment is handled in compliance with these parameters.

Signature:_____________________________________________________ Date:______________________

Name Printed:__________________________________________________ Title:____________________________

D

08/01/25

INJECTOR SHIFTED

SYSTEM SHIFTED

Sheets and Views