This document is a comprehensive product catalog for S::can, a water quality monitoring equipment manufacturer. The catalog focuses on their range of advanced monitoring devices for drinking water, including the pipe::scan, nano::station, micro::station, and various sensor probes.

The pipe::scan is a multi-parameter sensor system that can measure up to 10 water quality parameters simultaneously, including TOC, DOC, UV254, turbidity, color, chlorine, pH, conductivity, temperature, and pressure. The device can be installed directly in pressurized pipes and offers real-time monitoring with features like event detection, hot maintenance, and 6 months of standalone operation. The catalog also details other monitoring stations like the nano::station and micro::station, which offer varying levels of parameter monitoring and are designed for different applications such as disinfection monitoring, intake protection, and general water quality assessment. The devices can be configured with multiple sensors and come with terminals for data logging, remote control, and software for data visualization and analysis.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

david Text Box

TAB A

PIPE SCAN DEVICE

light.

speed ahead.

the pipe::scan

Conduc-tivity UVT

TOC DOC

UV

Chlorine

Color

ORP

Tempe-rature

Pressure

Turbidity

4G

AT

TES

TATION DE

C O

NFORMITÉ SANIT

AI

R E

ACS

con::line - Low Power Terminal

Drinking water quality monitoring in the network

The pipe::scan is a sensor system for monitoring drinking water quality in pipes under pressure. It measures up to 10 parameters in one device: TOC, DOC, UV254, Turbidity, Color, Chlorine, pH/Redox, Conductivity, Tem-perature and Pressure. The water quality data can be sent to any central database via almost any protocol. Multiple pipe::scans are the ideal solution to monitor drinking water at any point in the network.

Only the pipe::scan can:

The pipe::scan Accurate measurement in perfect agreement to standardized lab reference... not just ”trending”

Organics and Turbidity monitoring

Totally flow-independent, even works under stagnating conditions

Hot-maintenance: without interrupting the flow/ pressure, and for each sensor individually

Full-scale event detection with real-time alarms within the drinking water distribution network

6 months service time: Efficient, reliable stand-a-lone operation without maintenance

Enclosure

Additional security for sensors and operator.

i::scan

Multi-parameter spectrophotometer probe.

Parameters:

FTU/NTU, UV254, UVT,

Color, TOC, DOC

Physical sensors

One chlori::lyser and two additional sensors (condu::lyser, pH::lyser or redo::lyser) can be installed.

Parameters:

Conductivity, Free Chlorine, pH, Redox and Temperature

The con::line is a compact computing platform with a built-in datalogger that can store data for up to one year.

It can be operated by battery, making it perfect for areas with limited power supply.

The con::line can be remote controlled from any standard web browser via PC, notebook, tablet, or smartphone. Its user-friendly web interface, Io::Tool, facilitates easy data visualization and configuration. 4G communication to any cloud system through secure SFTP or SCP connections allows fast and reliable data exchange.

Base unit

Flow cell for up to 4 sensors with retractable insertion nozzle, filter, sample valve, automatic bleeder valve, pressure sensor and flow sensor (optional).

Optional autobrush for i::scan

Provides automatic brush cleaning for the i::scan.

Nano-pump

For water flow even during periods of stagnation.

Pipe saddle

2“ pipe saddle for hot tap installation. Available for pipes from DN80 to DN600. Pipe saddle is not NSF certified.

Conduc-tivity

UVT

DOC

TOC

UV

UV

Chlorine

Turbidity

Color pH

ORP

Tempera-ture

Subject to misprint or typographical errors. For the latest version see www.s-can.at © s::can GmbH (2023) Release: March 2023 (V3.0 EN)

Example of an Installation in a Manhole Drinking Water Network Monitoring for Smart Cities

MEXICO

s::can Mexico Sistemas de Medición S. de R.L. de C.V sales@s-can.mx www.s-can.mx Status: Affiliate

PORTUGAL

s::can contact Portugal Vincenzo Rocca T: +351 91 569 4663 vrocca@s-can.at Status: Regional Sales Manager

SPAIN

s::can Iberia Sistemas de Medición S.L.U.

Ciutat de Granada 28 bis, 1a Planta, 08005 Barcelona T: +34 930 218 447 sales@s-can.es, www.s-can.es Status: Affiliate

USA

s::can USA 6 Iron Bridge Drive Collegeville, PA 19426, USA T: +1 (888) 694-3230 F: +1 (888) 469-5402 sales@s-can.us, www.s-can.us Status: Affiliate

HEADQUARTERS

s::can GmbH Brigittagasse 22-24 1200 Vienna, AUSTRIA T: +43 / 1 / 219 73 93 F: +43 / 1 / 219 73 93-12 sales@s-can.at, www.s-can.at

CHINA

Rm D /17F Building B 1118 Changshou Rd.

200042 Shanghai T: (+86-21) 34 06 03 11 F: (+86-21) 34 06 03 11 lxiao@s-can.cn, www.s-can.cn Status: Representative Office

FRANCE

s::can France SARL 370 route de Saint Canadet 13100 Aix en Provence T: + 33 4 42 20 35 01 F: + 33 9 82 25 35 01 sales@s-can.fr, www.s-can.fr Status: Affiliate

ITALY

s::can contact Italy Alessandro Morra T: +39 333 983 5634 amorra@s-can.at Status: Regional Sales Manager

Turbidity

Conduc-tivity

UVT TOC

DOCUV

Chlorine Color ORP

Tempe-rature

S pectrom eter P robes ervices & olutions i::scan Ionselective

P robes

Term inals

S oftw are ystem

C onfiguration

M onitoring S tations S pare P arts &

A ccessories P hysical P robes

© s::can GmbH pipe::scan pipe::scan pipe::scan pipe::scan for monitoring of drinking water quality pipe::scan pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan pipe::scan

Conduc-tivity UVT

TOC

DOC

DOC

UV

254 Chlorine pH i::scan

Multi-parameter spectrophotometer probe.

Parameters:

FTU/NTU, UV254, UVT,

Color, TOC, DOC

Physical sensors

One chlori::lyser and two ad-ditional sensors (condu::lyser, pH::lyser or redo::lyser) can be installed.

Parameters:

Conductivity, Free Chlorine, pH, Redox and Temperature

Base unit

Flow cell for up to 4 sensors with retractable insertion nozzle, fi lter, sample valve, automatic bleeder valve, pressure sensor and fl ow sensor (optional).

Optional autobrush for i::scan

Provides automatic brush cleaning for the i::scan.

Nano-pump

For water fl ow even during periods of stagnation.

Pipe saddle

2“ pipe saddle for hot tap installation. Available for pipes from DN80 to DN600.

The pipe::scan is a sensor system for monitoring drinking water quality in pipes under pressure. It measures up to 10 parameters in one device: TOC, DOC, UV254, Turbidity, Color, UVT, Chlorine, pH/

Redox, Conductivity, Temperature and Pressure.

The water quality data can be sent to any central database via almost any protocol. Multiple pipe::scans are the ideal solution to monitor drinking water at any point in the network.

Enclosure

Additional security for sensors and operator.

Certifi cate of Sanitary Conformity

AT

TES

TATION DE

C O

NFORMITÉ SANIT

AI

R E

ACS

TOC

DOC

UV254

Turbidity

Color

Chlorine pH/Redox

Conductivity

Temperature pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan technical specification measurement interval 1 min (minimal) precalibrated ex-works all parameters reference standard distilled water integration via con::cube power supply via con::cube power consumption (typical) 14 W power consumption (max.) 35 W wetted materials POM stainless steel fine brass EN12165 and EN12164 dimensions (width x height x depth) 220 x 450 x 340 mm weight (min.) approx. 8 kg operating temperature 0 ... 40 °C storage humidity 0 ... 95 % operating pressure 1 ... 10 bar installation / mounting on 2“ Hawle pipe saddle (to be ordered separately) other operating limits pipeline must be vented installation must be on top of pipe no direct sunlight pH range 4 ... 12 automatic cleaning autobrush (for i::scan) storage temperature 0 ... 45 °C conformity - EMC EN 61326-1 conformity - safety EN 61010-1

RoHS drinking water safety certificate ACS (Attestation de conformité

Sanitaire) protection class (-000) IP67 i::scan for drinking water parameter turbidity

[NTU/FTU]

color (app) [Hazen] color (tru) [Hazen]

TOC

[mg/l]

DOC

[mg/l]

UV254

[Abs/m]

UV254 f [Abs/m]

UVT10

part number i::scan_NTU/FTU min. 0 Y01-1-d-000 / -075 max. 800 i::scan_NTU/FTU+Color min. 0 0 0 Y02-1-d-000 / -075 max. 800 300 200 i::scan_NTU/FTU+UV254 min. 0 0 0 25 Y03-2-d-000 / -075 max. 800 70 55 100 i::scan_NTU/ FTU+UV254+Color min. 0 0 0 0 0 Y04-1-d-000 / -075 max. 800 300 200 70 55 i::scan_NTU/FTU+TOC_ eq+UV254 min. 0 0 0 0 Y05-3-d-000 / -075 max. 800 25 12 70 i::scan_NTU/FTU+TOC_ eq+UV254+Color min. 0 0 0 0 0 0 Y06-3-d-000 / -075 max. 800 300 200 25 12 70 chlori::lyser (stainless steel version) for drinking water parameter temperature [°C] free chlorine [mg/l] total chlorine [mg/l] part number chlori::lyser (FCl) min. 0 0 E-520-1-s-000 max. 40 2 chlori::lyser (TCl) min. 0 0 E-520-1-s-000 max. 40 2 ph::lyser for drinking water parameter temperature [°C] pH [pH] part number pH::lyser (pH) min. 0 4 E-514-2-000 max. 40 10 condu::lyser for drinking water parameter temperature [°C] conductivity [S/cm] part number condu::lyser (conductivity) min. 0 0 E-511-2-000 max. 40 5000 pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan recommended accessories part number article name D-315-xxx con::cube S-11-xx-moni moni::tool Software

F-160-SP-

SET-DKxxx

Hawle shut off pipe saddle DK75 - DK315, incl.

saddle blade

F-160-SP-

SET-DNxxx

Hawle shut off pipe saddle DN80 - DK600, incl.

saddle blade

S-14-vali vali::tool - s::can data validation software S-15-ana ana::tool - s::can event detection software

743min.

29,252"

15,103" interface pipe saddle

13,78"

9,843"

16,434"

2, 2" pipe::scan versions part number article name P-450-

PS-SET-024

pipe::scan base unit: flow cell incl. adapter plate, nano pump, insertion nozzle, vent valve, enclosure, pipe::scan cable hub incl. 10 m connection cable to con::cube, con::cube V3 24 Vincl. D-303LX, S-08-moni, S-11-basic-plc

P-450-

PS-SET-230

pipe::scan base unit: flow cell incl. adapter plate, nano pump, insertion nozzle, vent valve, enclosure, pipe::scan cable hub incl. 10 m connection cable to con::cube, con::cube V3 230 V incl. D-303LX, S-08-moni, S-11-basic-plc

P-FTU FTU with i::scan incl. Autobrush for pipe::scan P-FTU-COL FTU + COLOR with i::scan incl. Autobrush for pipe::scan

P-FTU-

COL-UV254

FTU + COLOR + UV245 with i::scan incl. Autobrush for pipe::scan

P-FTU-

COL-UV254-TOC

FTU + COLOR + UV254 + TOC with i::scan incl.

Autobrush for pipe::scan

P-FTU-UV254 FTU + UV254 with i::scan incl. Autobrush for pipe::scan

P-FTU-

UV254-TOC

FTU + UV245 + TOC with i::scan incl. Autobrush for pipe::scan

P-FCL Free Chlorine sensor, 0-2 mg/l, pressure resistant P-TCL Total Chlorine sensor, 0-2 mg/l, pressure resistant P-pH pH sensor, pressure resistant P-ORP ORP sensor, pressure resistant P-EC Conductivity sensor, pressure resistant P-P Pressure sensor for pipe::scan F-450-PS-P incl.

D-315-in-mA pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH david Text Box

TAB A

STATION SCAN DEVICE

Monitoring Stations Drinking Water www.s-can.at

BOD

COD

BTX

TOC

DOC

UV254

NO3

NO2

NH4

K+

Chlorine

ClO2

H2O2

PAA

F-

TSS

Turbidity

Color pH

ORP

EC

Temperature

O2

O3

H2S

AOC

Fingerprints

Contaminant Alarm

© s::can GmbH (2022)

Online Monitoring Stations

Pre-assembled The modular monitoring stations combine s::can instruments to a compact and versatile system. For every station the required components - s::can terminal, flow cells, mounting fittings and pipework - are factory assembled on a compact panel. Simply insert the probes and sensors in the flow cells, plug in cables and start measuring.

Choose your Parameters The s::can stations designed for online monitoring of drinking water are available in different versions:

From standard turbidity and color measurements to monitoring a complex set of parameters with advanced event detection.

Choose the station and parameter combination that fits best your needs!

Save up to 35 % By buying a complete station instead of all parts separately, you save up to 35 % of the list price. Furthermore you receive a complete package without any need for time consuming configuration.

micro::station

The modular micro::station combines s::can instruments into a compact and versatile system. It presents a complete solution, as the user only has to connect water supply and -discharge (“plug & measure”) in order to receive a previously unheard variety of imme-diately available information and parameters at no extra cost.

The s::can micro::station is designed for online monitoring of water quality parameters in clean media, such as drinking water.

The required components - spectro::lyser, s::can sensors and controller - are factory assembled with all required flow cells, mounting fittings and pipework on a compact panel.

micro::station - the s::can solution for water analysis, compact and easy like never before.

6 Inlet strainer

The inlet strainer ascertains that no coarse material enters the micro::station.

With screw cap for sieve removal/cleaning

1 Terminal con::cube V3 terminal with moni::tool software for data acquisition, data display and station control

3 Flow cell for spectrometer probe

Including auto brush cleaning device to provide cleaning of the optical measuring windows

2 Spectrometer probe

All s::can spectrometer probes are multi-parameter instruments that can measure a variety of water quality parameters

Possible parameters:

AOC, BOD, BTX, COD, Color, DOC, FTU/

NTU, H2S, NO2-N, NO3-N, O3, TOC, TSS,

UV254, Temperature, Fingerprints and Spectral Alarms

5 Flow detector

The flow detector is set to give an alarm if the flow rate decreases below a critical value

7 Pressure transmitter (optional)

Mounting position for pressure transmitter

4 System tubing

Included in panel assembly; Material PU, inside diameter 6 mm, outside diameter 8 mm

10 Physical probes

Up to four s::can physical probes can be installed in one flow cell

Possible parameters:

Conductivity, FCl/TCl, ClO2, H2O2, PAA, pH, Salinity, Redox and Temperature

13 Flow cell for ISE probe

Flow cell for one s::can ISE probe

11 Physical probe or ISE probe

Place for oxi::lyser, soli::lyser or s::can ISE probe (e.g. ammo::lyser)

Possible parameters:

F-, K+, NH4-N, NO3-N, O2, pH and Temperature

12 Flow cell for physical probes

Combined flow cell for up to four s::can physical probes. Provides quick connect/ disconnect design by safety pins to reduce offline time during sensor maintenance

9 Flow restrictor unit

For automatic flow restriction and back-flow prevention in by-pass

8 Main panel

Material: PP Weight of the station (fully equipped):

20 kg (+/- 1 kg) nano::station

The fully modular nano::station combines s::can instruments to a super-compact and versatile system. It presents a complete solution, as the user only has to connect water supply and discharge (“plug & measure”) in order to receive at no extra cost a previously unheard variety of immediately available information and parameters.

The s::can nano::station offers a monitoring solution for many different applications:

From very cost sensitive applications down to highly resolved “Smart Water Grids”, in small unmanned plants, or even in single building protection.

The required components - i::scan, s::can sensors and s::can controller - are factory assembled with required flow cells, mounting fittings and pipework on a super-compact panel.

The s::can nano::station - compact, precise and affordable!

1 Terminal

With con::cube V3 or con::lyte termi-nal. con::cube V3 is equipped with moni::tool software for data acquisition, data display and station control

3 System tubing

Included in panel assembly;

Material: PU, inside diameter 6 mm, outside diameter 8 mm

4 Autobrush for i::scan

Provides automatic cleaning for i::scan

2 i::scan

One i::scan can be installed on every nano::station

Possible parameters:

Color, FTU/NTU, UV254, TOC, DOC, Transmission

6 Physical probes

Up to three s::can physical probes can be installed additionally to the i::scan in one flow cell (e.g. condu::lyser, pH::lyser or chlori::lyser)

Possible parameters:

Conductivity, FCl, TCl, ClO2, H2O2, PAA, pH, Redox and Temperature

5 Main panel

Material: PE Weight of the station (fully equipped):

11 kg (+/- 1 kg)

10 Flow cell for i::scan and physi-cal probes

Combined flow cell for one i::scan and up to three s::can physical probes. Pro-vides quick connect/disconnect design by safety pins to reduce offline time during maintenance.

A flow restrictor (optional) can be installed in the flow cell.

9 Pressure sensor (optional)

Mounting position for pressure transmitter

8 Inlet strainer

The inlet strainer ascertains that no coarse material enters the nano::station.

With screw cap for sieve removal/ cleaning

7 Flow detector (optional) m ic ro -s m ic ro -m m ic ro -l m ic ro -x l m ic ro -x xl micro::station

Classical intake protection, event detection, and general drinking water monitoring in non-chlorinated systems micro-s

Parameters:

NTU/FTU, NO3, TOC, DOC, Color, UV254, pH, EC, Temp. & Alarms

Terminal:

con::cube V3 (optional: modem, digital I/Os and analog I/Os) moni::tool (8 parameters) and event detection software (vali::tool & ana::tool)

Software:

Panel, 2 flow cells, autobrush, pressure sensor, strainer, tubes and fittings, 2 free sensor spaces

Included Components:

Spectrometer Probe

NTU/FTU

Color

UV254

TOC

DOC

NO3

NO2

Temperature

ISE Probe

NH4

pH

Physical Probe pH

Conductivity

Free or Total Chlorine

ClO2

H2O2

PAA

Event Detection

Online Validation

Online Alarm

Parameter Overview The micro::station is available in two versions for non-chlorinated systems and three versions for chlorinating / chloraminating systems.

Advanced intake protection and event detection in non-chlorinated systems micro-m Complete drinking water monitoring in chlorinating systems micro-l

Parameters:

NTU/FTU, NO3, TOC, DOC, Color, UV254, NH4, pH, Temp. & Alarms

Parameters:

NTU/FTU, NO3, TOC, DOC, Color, UV254, pH, EC, FCl, TCl, ClO2, H2O2, PAA, Temp. & Alarms

Terminal:

con::cube V3 (optional: modem, digital I/Os and analog I/Os) moni::tool (8 parameters) and event detection software (vali::tool & ana::tool)

Software:

Panel, 2 flow cells, autobrush, pressure sensor, strainer, tubes and fittings

Included Components:

Terminal:

con::cube V3 (optional: modem, digital I/Os and analog I/Os) moni::tool (24 parameters) and event detection software (vali::tool & ana::tool)

Software:

Panel, 2 flow cells, autobrush, pressure sensor, strainer, tubes and fittings, 1 free sensor space

Included Components:

micro::station

Advanced drinking water monitoring in chloraminating systems micro-xl Complete drinking water monitoring in chloraminating systems micro-xxl

Parameters:

NTU/FTU, NO3, NO2, TOC, DOC,

UV254, NH4, pH, EC, FCl, TCl, ClO2, H2O2, PAA, Temp. & Alarms

Parameters:

NTU/FTU, NO3, NO2, TOC, DOC,

UV254, pH, EC, FCl, TCl, ClO2, H2O2, PAA, Temp. & Alarms

Terminal:

con::cube V3 (optional: modem, digital I/Os and analog I/Os) moni::tool (24 parameters) and event detection software (vali::tool & ana::tool)

Software:

Panel, 2 flow cells, autobrush, pressure sensor, strainer, tubes and fittings

Included Components:

Terminal:

con::cube V3 (optional: modem, digital I/Os and analog I/Os) moni::tool (24 parameters) and event detection software (vali::tool & ana::tool)

Software:

Panel, 3 flow cells, autobrush, pressure sensor, strainer, tubes and fittings, 1 free sensor space sn o-sn o-in o-in o-sno-1 & sno-3 nano::station

Drinking water monitoring for disinfection with 1 or 3 parameters

Parameters:

sno-1 (choose 1 parameter) sno-3 (choose 3 parameters):

pH, Redox, EC, TCl, FCl, ClO2, H2O2

& PAA

Terminal:

con::lyte eco or con::lyte pro1 (optional output modules: analog 0/4-20 mA, Modbus or Profibus)

Panel, 1 flow cell, strainer, tubes and fittings

Included Components:

Parameter Overview The nano::station is available in three versions.

For disinfection monitoring, intake protection and for general drinking water monitoring.

i::scan

NTU/FTU

Color

UV254

TOC

DOC

Temperature

Physical Probe pH or Redox

Conductivity

Free or Total Chlorine

ClO2

H2O2

PAA

PID -Control (optional)

Event Detection

Online Validation

Online Alarm na no

-3 p ar am et er s) na no

-4 p ar am et er s) na no

-6 p ar am et er s) na no

-8 p ar am et er s) ino-2 & ino-3 nano-3 & nano-4 Classical intake protection and drinking water protection with i::scan

Advanced general drinking water monitoring including organics

Parameters:

ino-2: NTU/FTU, Color ino-3: NTU/FTU, Color, UV254

Terminal:

con::lyte pro1 (optional output modules: analog 0/4-20 mA, Modbus or Profibus)

Panel, 1 flow cell, autobrush, strainer, tubes and fittings

Included Components:

Parameters:

nano-3 (choose 3 parameters) nano-4 (choose 4 parameters):

FTU/NTU, Color, UV254, Transmis-sion, EC, pH, Redox, TCl, FCl, ClO2, H2O2 & PAA

Terminal:

con::lyte pro1 (optional output modules: analog 0/4-20 mA, Modbus or Profibus)

Panel, 1 flow cell, autobrush, strainer, tubes and fittings nano::station nano-6 Very advanced drinking water monitoring including organics nano-8 Complete drinking water monitoring including organics and event detection

Parameters:

Choose 6 of: FTU/NTU, Color, UV254, Transmision, EC, pH, Redox, FCl, TCl, ClO2, H2O2 & PAA

Terminal:

con::lyte pro1 (no output modules included)

Panel, 1 flow cell, autobrush, strainer, tubes and fittings

Included Components:

Parameters:

Choose 8 of: FTU/NTU, Color, UV254, TOC, DOC, Transmission, EC, pH, Redox, FCl, TCl, ClO2, H2O2, PAA & Alarm

Terminal:

con::cube V3; moni::tool & event detection software included

Panel, flow indicator, 1 flow cell, autobrush, strainer, tubes and fittings

Included Components:

moni::tool (8 parameters) and event detection software (vali::tool & ana::tool)

Software:

Subject to misprint or typographical errors.

We worked with greatest accuracy though data can be outdated.

For the latest version see www.s-can.at We do not take any liability for content and data.

© s::can GmbH (2022)

HEADQUARTERS

s::can GmbH Brigittagasse 22-24 1200 Vienna, AUSTRIA T: +43 / 1 / 219 73 93 F: +43 / 1 / 219 73 93-12 sales@s-can.at www.s-can.at

CHINA

Rm D /17F Building B 1118 Changshou Rd.

200042 Shanghai T: (+86-21) 34 06 03 11 F: (+86-21) 34 06 03 11 lxiao@s-can.cn www.s-can.cn Status: Representative Office

FRANCE

s::can France SARL 370 route de Saint Canadet 13100 Aix en Provence P: + 33 4 42 20 35 01 F: + 33 9 82 25 35 01 sales@s-can.fr www.s-can.fr Status: Affiliate

ITALY

s::can Contact Italy Alessandro Morra T: +39 333 983 5634 amorra@s-can.at Status: Regional Sales Manager

MEXICO

s::can Mexico Sistemas de Medición S. de R.L. de C.V sales@s-can.mx www.s-can.mx Status: Affiliate

PORTUGAL

s::can Contact Portugal Vincenzo Rocca T: +351 91 569 4663 vrocca@s-can.at Status: Regional Sales Manager

SPAIN

s::can Iberia Sistemas de Medi-ción S.L.U.

Ciutat de Granada 28 bis, 1a Planta, 08005 Barcelona P: +34 930 218 447 sales@s-can.es www.s-can.es Status: Affiliate

USA

s::can Measuring Systems LLC 38C George Leven Drive North Attleboro, MA 02760 T: +1 (888) 694-3230 F: +1 (888) 469-5402 sales@s-can.us www.s-can.us Status: Affiliate pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan i::scan

∙ s::can plug & measure

∙ turbidity: measurement according to EPA 180.1 and ISO 7027, 90° scattering (35 mm path length)

∙ new light emitting technology

∙ no consumables, no moving parts

∙ special, non-fouling optical window material

∙ low power consumption (less than 1 W typical)

∙ dual-beam compensated optics

∙ optional automatic cleaning compressed air (InSitu, only for version -075 with fixed cable) or autobrush

∙ long term stable, 100 % corrosion free

∙ plug connection or fixed cable

∙ 5000 hours maintenance free operation

∙ mounting and measurement directly in the media (InSitu) or in a flow cell (monitoring station)

∙ can be mounted directly in a mains pipe / pressure pipe

∙ operation via s::can terminals & s::can software

38,5

5, ~3

Messgeräte Sonstige Daten recommended accessories part number article name B-32-xxx s::can compressor D-315-xxx con::cube F-110-iscan carrier i::scan, for easy horizontal attachment F-446-m-iscan Brush unit AutoBrush, Pom-C (for i::scan) S-11-xx-moni moni::tool Software i::scan monitors depending on the application an individual selection of: turbidity, TOC, DOC, UV254, UV254 f, color, UVT10, UVT10 f, UVT100 f and temperature i::scan pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are onfi guration

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH chlori::lyser

∙ s::can plug & measure

∙ measuring principle: amperometric (membrane covered)

∙ ideal for drinking water

∙ long term stable and lowest maintenance in operation

∙ replacement of membrane only once a year

∙ readings stable even at high fl uctuations of pH, temperature and fl ow

∙ compensates fl uctuations of pH in an unmatched way pH range from 4 to 10 FCl; pH range from 4 to 12 TCl (<5% per pH between 6 and 10/12 pH, <1% at <6 pH)

∙ low cross sensitivity to many surfactants

∙ cross sensitivities to chlorine dioxide, ozone can be compensated by using readings from spectro::lyser™

∙ factory precalibrated

∙ mounting and measurement in a fl ow cell

∙ operation via s::can terminals & s::can software

∙ plug connection or fi xed cable

27,6

23,7

,1

Messgeräte Sonstige Daten recommended accessories part number article name C-1-05-sensor 0.5 m connection cable for s::can physical probes D-315-xxx con::cube D-319-xxx con::lyte F-12-sensor carrier s::can physical probes F-45-four fl ow cell for four s::can physical probes F-46-four-iscan i::scan fl ow cell for up to 3 additional s::can probes F-45-sensor fl ow cell for s::can sensor S-11-xx-moni moni::tool Software chlori::lyser™ monitors free chlorine (Cl2 + HOCl + OCl-) or total chlorine (free chlorine + combined chlorine)

P hysical P robes P hysical P robes pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH condu::lyser

∙ s::can plug & measure

∙ measuring principle condu::lyser: 4-electrode, direct-contact measurement

∙ multiparameter sensor

∙ ideal for surface water, ground water and drinking water, also waste water

∙ long term stable and maintenance free in operation

∙ factory precalibrated

∙ mounting and measurement directly in the media (InSitu) or in a fl ow cell

∙ operation via s::can terminals & s::can software

∙ plug connection or fi xed cable

27,6

27,6

Messgeräte Sonstige Daten recommended accessories part number article name D-315-xxx con::cube D-319-xxx con::lyte F-12-sensor carrier s::can physical probes F-48-sensor s::can Sensor fl ow-cell (by-pass setup), PVC S-11-xx-moni moni::tool Software condu::lyser monitors conductivity, temperature & salinity

P hysical P robes P pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH pH::lyser

∙∙ s::can plug & measure

∙∙ measuring principle: unique, non-porous / non-leaking combined reference electrode for technically unrivalled and consistent pH performance

∙∙ multiparameter sensor

∙∙ ideal for surface water, ground water, drinking water and waste water

∙∙ long term stable and maintenance free in operation

∙∙ factory precalibrated

∙∙ mounting and measurement directly in the media (InSitu) or in a flow cell

∙∙ operation via s::can terminals & s::can software

∙∙ optional: automatic cleaning with compressed air

∙∙ plug connection or fixed cable

27,6

27,6

Messgeräte Sonstige Daten recommended accessories part number article name D-315-xxx con::cube D-320-xxx con::lyte F-12-sensor carrier s::can physical probes F-48-sensor s::can Sensor flow-cell (by-pass setup), PVC S-11-xx-moni moni::tool Software pH::lyser eco monitors pH & temperature pH::lyser pro: high temperature range

P pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH technical specification measuring principle potentiometric measuring principle detail combined, non-porous reference electrode resolution 0.01 pH accuracy (standard solution) 0.1 pH automatic compensation instrument temperature response time (T90) 30 ... 0 sec.

integrated temperature sensor 0 ... 90 °C integration via con::cube con::lyte con::nect power supply 9 … 18 VDC power consumption (typical) 0.8 W power consumption (max.) 1 W interface to s::can terminals sys plug (IP67), RS485 cable length 7.5 m fixed cable (-075) or plug connection (-000) cable type PU jacket housing material stainless steel 1.4404/1.4401, POM-C weight (min.) 400 g dimensions (Ø x l) 33 x 257 mm operating pressure 0 ... 10 bar installation / mounting submersed or in a flow cell process connection quick connect flow velocity 3 m/s (max.)

0.01 m/s (min.)

automatic cleaning media: compressed air permissible pressure: 3 ... 6 bar conformity - EMC EN 61326-1 conformity - safety EN 61010-1 operating temperature (eco) 0 ... 70 °C operating temperature (pro) 0 ... 90 °C storage temperature (electrode) -5 ... 30 °C storage temperature (sensor) -10 ... 60 °C protection class (-000) IP67 protection class (-075) IP68 measuring range parameter pH [pH] temperature [°C] part number pH::lyser eco (pH, temp) min. 2 0 E-514-2-000 / -075 max. 12 70 pH::lyser pro (pH, temp) min. 0 0 E-514-3-000 / -075 max. 13 90

P pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH technical specifi cation measuring principle 4-electrode, direct-contact measuring range application 0 ... 500,000 µS/cm

2 ... 42 PSU

accuracy (standard solution) 1% of reading automatic compensation instrument temperature integrated temperature sensor -20 ... 130 °C integration via con::cube con::lyte con::nect power supply 7 ... 30 VDC power consumption (typical) 0.06 W power consumption (max.) 0.15 W interface to s::can terminals sys plug (IP67), RS485 cable length 7.5 m fi xed cable (-075) or plug connection (-000) housing material Stainless steel 1.4435, FDA-approved PEEK, FDA-approved

EPDM, FKE, PMO-C

weight (min.) 240 g dimensions (Ø x l) 33 x 237 mm operating temperature (immersed) -20 ... 60 °C operating temperature (in fl ow cell) -20 ... 110 °C storage temperature 0 ... 60 °C operating pressure 0 ... 5.8 bar installation / mounting submersed or in a fl ow cell process connection quick connect fl ow velocity 3 m/s (max.)

automatic cleaning media: compressed air permissible pressure: 2 ... 6 bar cleaning interval: depending on application conformity - EMC EN 61326-1

EN 300328 V 1.7.1:2006

EN 301489-17 V2.1.1:2009

EN 301489-1 V 1.8.1:2008

conformity - safety EN 61010-1 protection class (-000) IP67 protection class (-075) IP68 municipal WWTP infl uent concentration ranges and sensor/probe type for this application conductivity [μS/cm] temperature [°C] salinity

[PSU]

part number condu::lyser min. 0 0 2 E-511-2-000 / -075 max. 500000 70 42 municipal WWTP aeration concentration ranges and sensor/probe type for this application conductivity [μS/cm] temperature [°C] salinity

[PSU]

part number condu::lyser min. 0 0 2 E-511-2-000 / -075 max. 500000 70 42 municipal WWTP effl uent concentration ranges and sensor/probe type for this application conductivity [μS/cm] temperature [°C] salinity

[PSU]

part number condu::lyser min. 0 0 2 E-511-2-000 / -075 max. 500000 70 42

P hysical P robes P pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH technical specifi cation measuring principle amperometric measuring principle detail potentiostatic 3-electrode system measuring range application E-507-1: 0 … 2 mg/L FCl

E-507-2: 0 … 10 mg/L FCl E-507-3: 0 … 2 mg/L TCl E-507-4: 0 … 10 mg/L TCl resolution 0.001 mg/l (for 10mg/l ... 0.01 mg/l) automatic compensation instrument temperature automatic compensation cross sensitivities pH, fl ow response time 2 min.

integration via con::cube con::lyte 1 con::lyte 2 con::lyte 4 con::nect power supply 9 … 18 VDC interface connection to s::can terminals sys plug, IP68, RS485, 12 VDC cable length 7.5 m fi xed cable (-075) or plug connection (-000) housing material PVC, stainless steel weight (min.) 300 g dimensions (diameter x length) 33 x 267 mm operating temperature 0 ... 45 °C storage temperature 0 ... 45 °C operating pressure 0 ... 1 bar installation / mounting fl ow cell process connection quick connect fl ow velocity 0.015 m/s (min.)

0.06 m/s (max.)

pH range 4 ... 10 conformity - EMC EN 61326-1 conformity - safety EN 61010-1 protection class (-000) IP67 protection class (-075) IP68 drinking water typical concentration ranges for this application free chlorine [mg/l] total chlorine [mg/l] part number chlori::lyser (FCl) min. 0 E-507-1-000 / -075 max. 2 chlori::lyser (FCl) min. 0 E-507-2-000 / -075 max. 10 chlori::lyser (TCl) min. 0 E-507-3-000 / -075 max. 2 chlori::lyser (TCl) min. 0 E-507-4-000 / -075 max. 10

P hysical P robes P pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan technical specification measuring principle 35 mm optical path length:

spectrometry combined 180° absorption and 90° scattering turbidity: according to EPA 180.1 and ISO 7027 5 mm optical path length:

absorption resolution turbidity (35 mm): 0.001 NTU/FTU turbidity (5 mm): 0.01 NTU/FTU (0.1 above 1000 NTU/FTU) color: 0.01 Hazen UV254: 0.015 Abs/m TOC: 0.01 mg/l accuracy (standard solution) turbidity submersed (5 mm): 2 NTU/ FTU or +/- 5 %* turbidity in flow cell (35 mm): 0.02 NTU/FTU or +/- 2.5 %* color: 1 Hazen or +/- 2.5 %* TOC: 0.1 mg/l or +/- 2.5 %* UV254: 0.1 Abs/m or +/- 2.5 %* (*whichever is greater) automatic compensation instrument dual-beam and 180° path precalibrated ex-works all parameters reference standard distilled water onboard memory 512 MB integrated temperature sensor -20 ... 70 °C resolution temperature sensor 0.06 °C integration via con::cube con::lyte con::nect power supply 10 ... 18 VDC power consumption (typical) 20 mA @ 12V power consumption (max.) 200 mA @ 12V interface to s::can terminals RS485, MODBUS cable length 7.5 m fixed cable (-075) or plug connection (-000) housing material PEEK, POM-C weight (min.) approx. 330 g dimensions (Ø x l) 35 mm path length:

38.5 x 325 mm

5 mm path length:

38.5 x 295 mm operating temperature 0 ... 45 °C operating pressure 0 ... 8 bar installation / mounting submersed or in a flow cell flow velocity 3 m/s (max.)

automatic cleaning with autobrush or compressed air (only possible for version (-075) with fixed cable) permissible pressure: 3 ... 6 bar storage temperature -20 ... 60 °C conformity - EMC EN 61326-1

EN 61326-2-3

conformity - safety EN 61010-1 protection class (-000) IP67 protection class (-075) IP68 pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan i::scan directly submersed in the water

The i::scan can be mounted and removed in the pipe without shutting of the water flow.

nano::station with i::scan i::scan installed in-pipe pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can GmbH pipe::scan surface water parameter turbidity

[NTU/FTU]

color (app) [Hazen] color (tru) [Hazen]

TOC

[mg/l]

DOC

[mg/l]

UV254

[Abs/m]

UV254 f [Abs/m]

UVT10

part number i::scan_NTU/FTU min. 0 Y01-1-r-000 / -075 max. 800 i::scan_NTU/FTU+Color min. 0 0 0 Y02-1-r-000 / -075 max. 800 500 500 i::scan_NTU/FTU+UV254 min. 0 0 0 25 Y03-2-r-000 / -075 max. 800 70 70 100 i::scan_NTU/ FTU+UV254+Color min. 0 0 0 0 0 25 Y04-2-r-000 / -075 max. 800 500 500 70 70 100 i::scan_NTU/FTU+TOC_ eq+UV254 min. 0 0 0 0 Y05-3-r-000 / -075 max. 800 25 25 70 i::scan_NTU/FTU+TOC_ eq+UV254+Color min. 0 0 0 0 0 0 Y06-3-r-000 / -075 max. 800 500 500 25 25 70 i::scan_NTU/FTU (5 mm path length) min. 0 Y13-1-c-075 max. 3000 drinking water parameter turbidity

[NTU/FTU]

color (app) [Hazen] color (tru) [Hazen]

TOC

[mg/l]

DOC

[mg/l]

UV254

[Abs/m]

UV254 f [Abs/m]

UVT10

part number i::scan_NTU/FTU min. 0 Y01-1-d-000 / -075 max. 800 i::scan_NTU/FTU+Color min. 0 0 0 Y02-1-d-000 / -075 max. 800 500 500 i::scan_NTU/FTU+UV254 min. 0 0 0 25 Y03-2-d-000 / -075 max. 800 70 70 100 i::scan_NTU/ FTU+UV254+Color min. 0 0 0 0 0 Y04-2-d-000 / -075 max. 800 500 500 70 70 i::scan_NTU/FTU+TOC_ eq+UV254 min. 0 0 0 0 Y05-3-d-000 / -075 max. 800 25 25 70 i::scan_NTU/FTU+TOC_ eq+UV254+Color min. 0 0 0 0 0 0 Y06-3-d-000 / -075 max. 800 500 500 25 25 70 pectrom eter P robes i::scan Ionselective

P robes

Term inals

S oftw are

M onitoring S tations S

P hysical P robes

© s::can Messtechnik GmbH technical specification measuring principle potentiometric measuring principle detail combined, non-porous reference electrode resolution 0.01 pH accuracy (standard solution) 0.1 pH automatic compensation instrument temperature response time (T90) 30 ... 0 sec.

integrated temperature sensor 0 ... 90 °C integration via con::cube con::lyte con::nect power supply 9 … 18 VDC power consumption (typical) 0.8 W power consumption (max.) 1 W interface to s::can terminals sys plug (IP67), RS485 cable length 7.5 m fixed cable (-075) or plug connection (-000) cable type PU jacket housing material stainless steel 1.4404/1.4401, POM-C weight (min.) 400 g dimensions (Ø x l) 33 x 257 mm operating pressure 0 ... 10 bar installation / mounting submersed or in a flow cell process connection quick connect flow velocity 3 m/s (max.)

0.01 m/s (min.)

automatic cleaning media: compressed air permissible pressure: 3 ... 6 bar conformity - EMC EN 61326-1 conformity - safety EN 61010-1 operating temperature (eco) 0 ... 70 °C operating temperature (pro) 0 ... 90 °C storage temperature (electrode) -5 ... 30 °C storage temperature (sensor) -10 ... 60 °C protection class (-000) IP67 protection class (-075) IP68 measuring range parameter pH [pH] temperature [°C] part number pH::lyser eco (pH, temp) min. 2 0 E-514-2-000 / -075 max. 12 70 pH::lyser pro (pH, temp) min. 0 0 E-514-3-000 / -075 max. 13 90

P

File details come from the government source that posted it. Updated .