This document contains questions and answers regarding solicitation number 36C24523Q1220 for 14 slit lamps with imaging software for the Martinsburg VA Medical Center. The solicitation seeks brand name or equal Zeiss SL800 slit lamps, with each unit to include an imaging module. Bidders may provide pricing for the slit lamp and imaging module together in one line item or separately. Confirmation was provided that the imaging module referred to in the solicitation is the Zeiss SL Cam Compact Imaging with software, and bidders may offer a PC option as well. Measuring prisms were clarified to be brand name or equal Zeiss Applanation tonometers. Two technical requirements were removed from consideration as incorrectly included. No additional imaging modules beyond those included with each slit lamp are required.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Amendment_36C24523Q1220_0002.pdf | ||
| Amendment_36C24523Q1220_0001.pdf | ||
| RFQ 36C24523Q1220_Question and Answer_9.27.23_v.2.docx | DOCX document | |
| RFQ 36C24523Q1220_Question and Answer_9.27.23.docx | DOCX document | |
| RFQ_36C24523Q1220.pdf | ||
| 36C24523Q1220.docx | DOCX document | |
| VA Directive 6550_36C24523Q1220.pdf | ||
| ERA Ports_36C24523Q1220.xlsx | XLSX spreadsheet | |
| MDS2_36C24523Q1220.xlsx | XLSX spreadsheet | |
| Topology Directions_36C24523Q1220.pptx | PPTX presentation | |
| ERA Inventory_36C24523Q1220.xlsx | XLSX spreadsheet | |
| Combined Synopsis Solicitation Notice_36C24523Q1220.pdf |
Show all 12
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
RFQ 36C24523Q1220 Question & Answer (All information has been confirmed with the requesting program office) Q: Is imaging needed for all 14 units?
A: Yes, imaging will be needed for all 14 slit lamps.
Q: For multiple quote responses, will two separate quotes of SF 1449 be needed or include both in one SF 1449?
A: Please include one singed SF 1449 with a quote response. Multiple price/cost schedules offering different products can be added in one quote response.
Q: CLIN 0002 - Imaging Module - Zeiss SL Cam. You do not list a Zeiss Part Number so I assume it is the SL Cam Compact Imaging with SL Imaging Software. Could you please confirm.
A: Yes, CLIN 0002 is brand name or equal to the SL Cam Compact Imaging with SL Imaging Software. This will be reflected in an amendment.
Q: CLIN 0002 - You do not state if it is with the PC option so I am assuming it does not have that option. Could you please confirm.
A: CLIN 0002 is to offer a PC option. If not included with CLIN 002, please add an additional line item.
Q: CLIN 0003 - Measuring Prisms. You do not list a Zeiss part number. A) Could you provide the part number so I make sure I am offering the proper Equal to? B) I assume this is the Zeiss tonometer as the SL 800 does not come with a tonometer and this is the only CLIN that Clinical Requirements Criteria “Goldman application tonometer attached for measuring intraocular pressure” would be met with a Zeiss tonometer. Could you please confirm.
A: CLIN 0003 – Measuring Prisms is brand name or equal to: Zeiss Applanation tonometer AT 030. This will be reflected in an amendment.
Q: Technical Requirements Criteria states “Ability to provide both illumination characteristics (LED and Halogen)”. The Zeiss SL 800 Light Source is LED only (SL 800 Brochure attached). Halogen is not a light or illumination source on the SL 800. May I suggest this Technical Requirements Criteria be eliminated as you already cover LED light source under the Clinical Requirements Criteria and you cover filters in another line item in Technical Requirements Criteria.
A: The criteria listed in the SOW of the ability to provide both illumination characteristics (LED and Halogen) will be removed from the SOW and will not be considered in the evaluation of quote responses. This will be reflected in an amendment.
Q: If Slit lamps come with imaging module, are you seeking additional imaging modules in CLIN 0002?
A: No, we are seeking 14 slit lamps with imaging modules. If the Slit lamp and imaging module come as a set, please reflect the price of the set in CLIN 0001 and put “include” or “N/A” for pricing in CLIN 0002. Please ensure all items listed in your quote response are compatible with one another and requirements listed in the statement of work.
9.28.23 Q: A technical requirement in the SOW states “Semi-automatic and fully automatic shock delivery options”. It is unclear what this is referring to. Could you please expand on this requirement and refer to the particular location that you took this requirement from?
A: The technical requirement was incorrectly added to the SOW. It will be removed and will not be considered in the evaluation of quote responses. This will be reflected in an amendment.
Q: The SOW was revised to remove "ability to provide both illumination characteristics (LED Halogen)" since filters and light source were already listed. Will the Government consider reinserting this salient characteristic?
A: No. The solicitation is for a brand name or equal to the Zeiss SL 800 with the features/specifications/salient characteristics listed in the SOW.