Attachment 3 - Gas Sample Verification.xlsx
XLSX spreadsheet 48 KB Posted
- Attached to
- Compressed Gases Federal contract opportunity
- Solicitation number
- 1305M326Q0071
About this file
This is a gas sample verification attachment that establishes testing requirements and procedures for compressed gases submitted under solicitation 1305M326Q0071 for supply to the National Oceanic and Atmospheric Administration's Global Monitoring Laboratory in Boulder, Colorado.
The document specifies verification criteria for two gas types: Research-Grade Helium must demonstrate nitrogen concentrations below 3% and oxygen concentrations below 2% relative to mass 69, as measured by GCMS analysis at NOAA's lab; UHP Nitrogen must meet 99.999% purity specifications, be clean for use as a Nafion purge gas, be supplied in minimum 200 cubic foot high-pressure cylinders with a minimum fill pressure of 2000 psi, and include standard CGA fittings. Both gas types require submission of four samples for testing at NOAA's laboratory, with a minimum sample size of N200. Gases must be delivered no later than one week after the Government issues the request. Notably, contractors are prohibited from revising their quotations based on the results of this verification process, indicating that sample testing is a compliance and quality assurance measure rather than a negotiation point.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sol_1305M326Q0071_Amd_0001.pdf | ||
| Attachment 1 - Statement of Need (SON) Revision 1.pdf | ||
| RFQ 1305M326Q0071 Industry Questions-Answers.pdf | ||
| Attachment 2 - Pricing Schedule.xlsx | XLSX spreadsheet | |
| RFQ (CSS) 1305M326Q0071.pdf | ||
| Attachment 1 - Statement of Need (SON).docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Gas Sample Requirements
| Gas Sample Verification Instructions: |
| 1. All submitted gases must meet the criteria listed below, including specification requirements, minimum sample size, and number of samples. |
| 2. Gases must be delivered no later than one week after the Government has provided the request. |
| 3. Your company will not be allowed to revise your quotation as a result of this verification. |
| Gas Description | Specification Requirements | Minimum Sample Size | Number of Samples |
| Research-Grade Helium | Have N2 concentrations below 3% and O2 concentrations below 2% relative to mass 69 (respectively) as measured by GCMS at NOAA lab | N200 | 4 samples to test at NOAA lab |
| UHP Nitrogen | Meet 99.999% purity specs, Clean for multiple compound use as Nafion purge gas as tested by NOAA GCMS, Supplied in minimum 200 cu ft capacity high pressure cylinder, Have a minimum fill pressure of 2000 psi, Have standard CGA fittings | N200 | 4 samples to test at NOAA lab |