The document is a comprehensive site preparation support document for a Philips Radiography 7300 C Plus VS X-ray system installation at Lyndon B. Johnson General Hospital in Houston, Texas. The document, dated July 22, 2024, provides detailed technical specifications, equipment layout, electrical requirements, network configuration, and site readiness guidelines for the medical imaging equipment installation. The project involves a complete radiography system setup, including a digital vertical stand, wide tabletop, ceiling suspension system, and associated electrical and network infrastructure. The installation is planned for room EC 1137, with multiple revisions made to the document between December 2024 and August 2025.
The technical specifications outline stringent requirements for site preparation, including electrical configurations (3-phase, 400-480 VAC power), precise room dimensional specifications, ceiling height restrictions (8'-9" to 10'-6"), and specific environmental conditions (temperature 75° ± 11°F, humidity 52.5% ± 22.5%). The document details extensive electrical, network, and physical infrastructure needs, such as conduit routing, cable management, network connectivity for remote diagnostic services, and precise mounting and anchoring specifications. The installation requires coordination between Philips technicians and hospital facilities management, with specific responsibilities outlined for site preparation, electrical work, network configuration, and equipment positioning. The site must meet detailed readiness criteria before the installation can commence, including wall and floor preparation, electrical and network infrastructure, and compliance with safety and building codes.
View the file
Other files
Show all 17
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Table of Contents Revision History
Note for Architects and/or Contractors: If revisions are listed, these drawings must be thoroughly reviewed so that all changes can be incorporated into your project
Rev. Date Revision Descriptions By
00 12/02/2024 Created preliminary site preparation support document per order 6600710479.010000 per PM request. PD
A 03/17/2025 Created final site preparation support document per order 6600710479.010000 per PM request. PD
B 06/04/2025 Updated customer name per PM request. PD
C 06/17/2025 A2, S1, E1: Moved MS table 0'-2" plan east and moved WAP to control room wall per PM request. PD
D 08/14/2025 A2, S1, E1: Moved PS and DGS per PM request. Updated room height. PD
Section A - Equipment Plan
General Notes ----------------------------------------- AN
Section S - Support Plan
Section E - Electrical Plan
Support Notes ------------------------------------------ SN Support Layout - Floor & Wall --------------------- S1 Support Layout - Ceiling ----------------------- S2-S3
Electrical Notes ---------------------------------------- EN
Electrical Details ------------------------------ ED1-ED2
Remote Services Network --------------------- N1 - N2
Checklist ---------------------------------------------- CHK
Site Layout ---------------------------------------- A1
Section N - Network Plan
Ceiling Height Guide -------------------------- A3
Section C - Site Readiness
Equipment Details -------------------------- AD2 - AD3
Support Details ------------------------------ SD3 - SD5
Anchor Reference List ----------------------------- SD1 Ceiling Clearance ----------------------------------- SD2
Transport Information ------------------------------ AD1
Electrical Layout --------------------------------------- E1 Conduit List -------------------------------------------- E2
Equipment Layout --------------------------------- A2
Final Site Preparation Support Document The equipment components shown in this drawing package are based on the current proposed purchase and are subject to change if modifications are made to the configuration.
www.healthcare.philips.com
*Photo shown is not site specific.
C1 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Electrical Legend --------------------------------------- EL
AN
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
HVAC Requirement for General Equipment Locations
Heating, ventilation, air conditioning requirement for general equipment locations must maintain temperature at 75° +/- 11°Fahrenheit (24° +/- 6°Celsius) and non-condensing relative humidity at 52.5%, +/- 22.5%.
(13.0)
Minimum Site Preparation Requirements
A smooth efficient installation is vital to Philips and their customers. Understanding what the minimum site preparation requirements are will help achieve this goal. The following list clearly defines the requirements which must be fulfilled before the installation can begin.
1. Walls shall be painted or covered, baseboards installed, floors shall be tiled and/or covered, ceiling shall have grid tiles and lighting fixtures installed and operational.
2. Doors and windows, especially radiation protection barriers, installed and finished with lock sets operational.
3. All electrical convenience, conduit, raceway, knockouts, cable openings, chase nipples, and junction boxes installed and operational.
4. Incoming mains power operational and connected to room x-ray breaker.
5. 115v convenience outlets operational.
6. All support structure correctly installed. All channels, pipes, beams and/or other supporting devices should be level, parallel, and free of lateral or longitudinal movements.
7. All contractor-supplied cables pulled and terminated.
8. A dust-free environment in and around the procedure room.
9. All HVAC (heating, ventilating and air conditioning) installed and operational as per specifications. BTUs shown on sheet A2 are average heat capacity.
10. Architectural features such as computer floor, wood floor, casework, bulkheads, installed and finished. When technical cabinets are installed in a closet with doors, it is suggested that the customer install a temperature alarm in the event of an air conditioner failure.
11. All plumbing installed and finished.
12. Philips does not install or connect developing tanks, automatic processors or associated equipment, built in illuminators, cassette pass boxes, loading benches and cabinets, lead protective screens, panels or lead glass window and frame. This is to be done by the customer/contractor.
13. Refer to Transport Information page for clear door openings and corridor widths.
Note Once Philips has moved equipment into the suite and started the installation, the contractor shall schedule his work around the Philips installation team on site. It is suggested that a telephone be provided in the room to receive telephone calls. This would alleviate facility staff from answering calls for Philips personnel.
Remote Service Diagnostics Medical imaging equipment to be installed by Philips Healthcare is equipped with a service diagnostic feature which allows for remote and on site service diagnostics. To establish this feature, a RJ45 type ethernet 10/100/1000 Mbit network connector must be installed as shown on plan. Access to customer's network via their remote access server is needed for Remote Service Network (RSN) connectivity. All cost with this feature is the responsibility of the customer.
(24.0)
General Specifications
1. Planning, Design, and Implementation Process The multidisciplinary project team should be assembled as early as possible in the design process. The multidisciplinary team should include administrators, clinicians, infection preventionists, architects and other design professionals, facility managers, safety officers, security managers, users of equipment, and support staff relevant to the areas affected by the project as well as those with knowledge of the organization's functional goal for the project.
Inclusion of patient advocates/consumers, A/E consultants, and construction specialists should be considered. FGI 2018 APPENDIX A1.2-1.2
2. Responsibility The customer shall be solely responsible, at its expense for preparation of site, including any required structural alterations. The site preparation shall be in accordance with plans and specifications provided by Philips. Compliance with all safety electrical and building codes relevant to the equipment and its installation is the sole responsibility of customer. The customer shall advise Philips of conditions at or near the site which could adversely affect the carrying out of the installation work and shall ensure that such conditions are corrected and that the site is fully prepared and available to Philips before the installation work is due to begin. The customer shall provide all necessary plumbing, carpentry work, or conduit wiring required to attach and install products ready for use.
3. Permits Customer shall obtain all permits and licenses required by federal, state/provincial or local authorities in connection with the construction, installation and operation of the products and related rules, regulations, and shall bear any expense in obtaining same or in complying with any ordinances and statutes.
4. Protection The customer or his contractor, at his own expense, shall obtain the service of a licensed radiation physicist to specify radiation protection. (X-Ray Tube output 150 KVp max.)
5. Asbestos and Other Toxic Substances Philips assumes that there is no hazardous material contained in project site. The customer is responsible for the removal of any materials, including but not limited to asbestos, deemed hazardous by local authorities, the EPA, OSHA, or any other authority having jurisdiction over the work. If such materials are discovered at any time that the work is proceeding, the work will immediately cease, the owner will be notified, and the work will again proceed after the owner has removed all of the hazardous material from the job site.
6. Labor In the event local labor conditions make it impossible or undersirable to use Philips' regular employees for such installation and connection, such work shall be performed by laborers supplied by the customer, or by an independent contractor chosen by the customer at the customer's expense, and in such case, Philips agrees to furnish adequate engineering supervision for proper completion of the installation.
7. Schedule The general contractor shall provide Philips with a schedule of work to assist in the coordination of delivery of Philips supplied products which are to be installed by the contractor and delivery of the primary equipment.
8. Extended Installation or Turnkey Work by Philips Any room preparation requirements for Philips equipment indicated on these drawings is the responsibility of the customer. If an extended installation or turnkey contract exists between Philips and the customer for room preparation work required by the equipment represented on these drawings, some of the responsibilities of the customer as depicted in these drawings may be assumed by Philips. In the event of a conflict between the work described in the turnkey contract workscope and these drawings, the turnkey contract workscope shall govern.
9. Electromagnetic Interference and Image Quality The Philips digital X-Ray detectors are highly sensitive receivers. Therefore, when determining where in the facility to install the x-ray system, care must be taken to avoid electromagnetic interference (EMI) to ensure optimal image quality. Possible EMI sources can be high power generators, power transformers, high power mains lines or other electromagnetic sources, which are located in the examination room directly or in the environment around the examination room. These EMI sources may emit excessive magnetic noise that could cause artifacts in the digital detector.
10.Avoid Adjacent EMI Sources Do not locate the digital detector in a room adjacent to such an EMI source (next to, above or below). EMI sources can be located also in the wall, floor or ceiling.
(21.0)
General Conditions (continued)
11. Detector Location and Additional Sheiding If a location next to an EMI source is unavoidable, ensure the table and/or wallstand and proposed locations for portable detector exposures are away from these sources. The magnetic noise interference is reduced dramatically by locating as far away as possible from the EMI sources. It might be necessary to implement additional means to prevent an interference with the x-ray detector (e.g. additional shielding).
(21.0)
(19.0)
Electrical Requirements
M-Cabinet CXA 80
Supply Configuration: 3 phase, 3 wire power, neutral and ground. Wye.
or 3 phase, 3 wire power, ground. Delta.
Nominal Line Voltage: 400, 440, 460 or 480 VAC, 60 Hz
Branch Power Requirement: 112.5 kVA minimum
Circuit Breaker: 3 pole, 50 Amps (@ 480V) (Slow Blow)
Residual Current Device to be installed depending on local regulations.
14'-815 16"
'-7
3 8"
Table Transverse ℄
Ta bl e Lo ng itu di na l ℄
Medical Gases
8'-3"
4' -6
1' 2' 3' 4' 5' 6' 7' 8' 9' 10' 11' 12'
.5 m
1m 1.
5m 2m
2.
5m
3m 3.
5m
'-4
7 8"
2'-0"
6'-45 8" 7'-11 1
16"
5' -1
13 16
1' -1
1"
4'-11 1 16"
1' -2
" 2' -0
4' -0
3' -1
115 16
4'-2 1116 "
115°
2"
'-3
3 8"
4"
5'-101 2"
211 16
A1 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Site Layout 1/4" = 1'-0"
Recommended Minimum Ceiling Height: 8'-9" (2667mm) Recommended Maximum Ceiling Height: 10'-6" (3200mm)
Reported Ceiling Height: 9'-6" (2896mm)
Ceiling Heights other than recommended may impact equipment functionality - see Ceiling Height Guide and consult with Philips.
Ceiling Height measured from finished floor to bottom of Unistrut.
Walls
Soffit
Existing (to be removed)
Beams or other building construction elements
Legend
* Counters and cabinetry shown to be supplied and installed by contractor.
General Notes
* Field to verify all room dimensions.
* Refer to A.D.A. Guidelines for doors and clearances. Verify all other applicable code(s) with the architect of record.
ME
4' -0
" S er vi ce ab ilit y
PS
MS
Tube Location 102" S.I.D.
Table Transverse ℄
Ta bl e Lo ng itu di na l ℄
W al l B uc ky
WAP
ELW
SBC
8'-3"
4' -6
1' 2' 3' 4' 5' 6' 7' 8' 9' 10' 11' 12'
.5 m
1m 1.
5m 2m
2.
5m
3m 3.
5m
4' -0
DGS
DVS
DC
CS7' High Wall
3"
8' -6
Weight (lbs)
Heat Load (btu/hr)
Detail SheetEquipment Designation
Description
Equipment Legend A Furnished and installed by Philips B Furnished by customer/contractor and installed by customer/contractor C Installed by customer/contractor D Furnished by Philips and installed by contractor E Existing F Future G Optional item furnished by Philips
A2 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Site Planning Issues and Considerations
General Notes
Notes to Philips Field Personnel
* Longer cables will need to be ordered for cable runs from:
- "ME" to "MS" [55' (17m)]
- "ME" to "ELW" [88' (27m)]
* Stitching shots at 102" S.I.D. will be available. See layout.
* Any counters, keyboard trays and cabinetry shown to be supplied and installed by contractor. Shelf under control counter for UPS and CPU is required.
* Field to verify all existing Philps and/or third party equipment will not affect the functionality of the system and its components.
Equipment Layout 1/4" = 1'-0"
Recommended Minimum Ceiling Height: 8'-9" (2667mm) Recommended Maximum Ceiling Height: 10'-6" (3200mm)
Reported Ceiling Height: 9'-6" (2896mm)
Ceiling Heights other than recommended may impact equipment functionality - see Ceiling Height Guide and consult with Philips.
Ceiling Height measured from finished floor to bottom of Unistrut.
A ELW Eleva Workspot
- Eleva Examination Control
- Acquisition Workspot (on shelf under counter)
- Uninterruptible Power Supply (on shelf under counter)
- Keyboard and Mouse
- mShield (optional, not shown)
127 536 AD2
ME Generator M-Cabinet CXA 342221A AD2
A WAP Wi-Fi Access Point 2 -- AD2
2 -- AD2A SBC SkyPlate Battery Charger
Drag ChainDC 66 --A AD2
Ceiling Suspension CSM3 Comfort Move LongCS 862 1365A AD2
MS 614593DigitalDiagnost TH/TH2 (Wide Tabletop)A AD3
DVS 614562Digital Vertical Stand (VS2) Tiltable (Right)A AD3
* Location of the Wall Bucky may be in violation of ADA. Verify all applicable code with customer's architect of record.
* Tabletop to be installed before table base is bolted to the floor.
* A portion of the existing floor raceway will need to be filled as required to provide adequate bolt anchorage for table. Verify with Structural Engineer of Record. See sheets S1 and E1 of final drawing package for details.
* Verify exact location and configuration of existing electricals with local Philips service for re-use.
* Field to verify existing conduit lengths, sizes, quantities, and origination and termination points.
* Cross-table shots for angled views will not be available behind the table. Patient will need to be flipped on table in order to obtain angled bilateral cross table views.
* New electrical boxes for ME, MS and CS will be required to be installed.
--154PS Patient SupportA AD3
--16DGS Detector and Grid Storage (wall mounted) AD3D
* Limited stretcher access to Wall Bucky. Recommend to move tabletop for improved access.
Ceiling Height Guide for High Performance VS and Value Room Not Site Specific
8'-9" to 10'-6" * Full Vertical Tube Movement
Limitations 8'-8" to 8'-9" Tube cannot reach Highest Position of Wall Bucky 8'-7" to 8'-8" Limitation noted above, plus horizontal (tilted) Bucky must be lowered for 40" SID 7'-11" to 8'-7" Limitations noted above, plus tabletop must be lowered for 40" SID
Loss of Coverage 7'-10" to 7'-11" 40" SID to horizontal (tilted) Bucky not available 7'-7" to 7'-10" Limitations noted above, plus 40" SID to tabletop not available 7'-7" Absolute Minimum Ceiling Height
* Ceiling Heights higher than 10'-6" - consult with Site Planning for limitations/loss of coverage
40" S.I.D. to table at standard working heightTube can cover full range of movement of the Wall Bucky
Allows comfortable working height for 40" S.I.D. to
Tilted Bucky
1' 2' 3' 4' 5' 6' 7' 8' 9' 10' 11' 12'
.5 m
1m 1.
5m 2m
2.
5m
3m 3.
5m
Highest Tube Position
Lowest Tube Position
" 53
9' -6
1' -0
8x45°
125° 12 5°
A3 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co
AD1
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Radiography 7300C (Tube Crane, TH Table, Wall Bucky) Transport Information
Crate Packed Crate Packed
Weight 684 lbs (310kg) 375 lbs (170kg)
Height 2'-9" (0.83m) 1'-10" (0.55m)
Length (L) 4'-6" (1.36m) 7'-4" (2.22m)
Width (W) 3'-2" (0.96m) 2'-6" (0.76m)
Width of Corridor "X" Minimum Width of Corridor/Door "Y"
Minimum Width of Corridor/Door "Y"
3'-4" (1.0m) 5'-1" (1.54m) 6'-1" (1.85m)
3'-8" (1.1m) 4'-6" (1.36m) 5'-8" (1.72m)
4'-0" (1.2m) 4'-1" (1.25m) 5'-3" (1.60m)
4'-4" (1.3m) 3'-10" (1.17m) 4'-11" (1.50m)
4'-8" (1.4m) 3'-9" (1.12m) 4'-7" (1.40m)
5'-0" (1.5m) 3'-8" (1.08m) 4'-4" (1.30m)
5'-3" (1.6m) 3'-6" (1.06m) 4'-0" (1.20m)
5'-7" (1.7m) 3'-6" (1.06m) 3'-9" (1.13m)
5'-11" (1.8m) 3'-6" (1.06m) 3'-6" (1.10m)
6'-3" (1.9m) 3'-6" (1.06m) 3'-4" (1.00m)
6'-7" (2.0m) 3'-6" (1.06m) 3'-2" (0.95m)
6'-11" (2.1m) 3'-6" (1.06m) 3'-0" (0.91m)
Telescopic Carriage Longitudinal Carriage Long
Longitudinal Carriage Short
661 lbs (300kg) 397 lbs (180kg) 207 lbs (94kg)
N/A 1'-4" (0.41m) 1'-4" (0.41m)
4'-0" (1.20m) 13'-9" (4.18m) 8'-0" (2.42m)
2'-7" (0.78m) 2'-7" (0.77m) 3'-1" (0.94m)
Minimum Width of Corridor/Door "Y"
Minimum Width of Corridor/Door "Y"
Minimum Width of Corridor/Door "Y"
3'-7" (1.07m) 12'-3" (3.72m) 7'-11" (2.4m)
3'-3" (.99m) 11'-9" (3.58m) 7'-2" (2.17m)
3'-1" (.93m) 11'-4" (3.45m) 6'-8" (2.02m)
2'-11" (.88m) 10'-11" (3.32m) 6'-3" (1.89m)
2'-11" (.88m) 10'-6" (3.20m) 5'-10" (1.77m)
2'-11" (.88m) 10'-2" (3.10m) 5'-6" (1.66m)
2'-11" (.88m) 9'-10" (2.98m) 5'-2" (1.56m)
2'-11" (.88m) 9'-5" (2.87m) 4'-10" (1.47m)
2'-11" (.88m) 9'-1" (2.76m) 4'-8" (1.40m)
2'-11" (.88m) 8'-9" (2.66m) 4'-4" (1.31m)
2'-11" (.88m) 8'-4" (2.55m) 4'-1" (1.24m)
2'-11" (.88m) 8'-1" (2.45m) 3'-11" (1.18m)
TH Table Wall Bucky CSM3
Corridor X
W
L
C or rid or o r D oo r Y
(13.0)
AD2
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
8 7 16"
[215mm] Top
Side
16"
[275mm]
1' -5
11 16
[4 m m
(24.0)
ELW
Eleva Workspot Weight Heat Dissipation
127 lbs (58 kg) 536 Btu/hr (157 W)
Front CPU Tower
Weight: 32 lbs (15 kg) Heat Diss.: 293 Btu/hr (86W)
Eleva Examination Control Weight: 38 lbs (18 kg) Heat Dissipation: 342 Btu/hr (100W)
1'-73 8"
[492mm]
Top
FrontSide
1'-73 8"
[492mm] 1'
-3 3 8"
[3 m m
4"
[235mm]
Front
Top
Side
1'-7 7 16"
[493mm]
8"
[206mm]
3 7 16"
[88mm]
1' -5
1 4" [4 m m
Uninterruptible Power Supply Weight: 57 lbs (26 kg)
(21.0)
ME
Generator M-Cabinet CXA
Weight Heat Dissipation
221 lbs (100 kg) 342 Btu/hr (100 W)
Noise measured at 1 meter distance at 1 meter high over floor < 52 dB(A).
Front
Top
Side
1 2"
[1 3m m
Service Area if cable outlet is in rear.*
*Cable outlet is not shown. For detailed view, please see ED1 page.
2'-5 7 16"
[748mm]
2'-35 8"
[702mm]
3' -2
9 16
[9 m m
1'-7 3 16"
[488mm]
16"
[50mm]
(24.0)
WAP
Wi-Fi Access Point
Weight Heat Dissipation
2.5 lbs (1.12 kg) --
Side Front
Top
6" [152mm]
3" [76mm]
57 8"
[1 m m
Required clear for installation and repair
57 8"
[1 m m
Top
Side Front
Top
Side
(14.0)
SBC
SkyPlate Detector and Battery Charger Weight Heat Dissipation
2.20 lbs (1 kg) --
Front
Front
Top
Side
SkyPlate Wireless Detector (Large)
Weight: 5.7 lbs (2.6 kg)
Battery Charger Weight: 2.20 lbs (1 kg)
1'-011 16"
[322mm]
17 8"
[4 8m m
16"
[170mm]
1'-07 8"
[328mm]
9 16
[2 m m
1'-31 8"
[384mm]
1' -6
1 8"
[4 m m
8"
[16mm]
8"
[16mm]
SkyPlate Wireless Detector (Small)
Weight: 3.5 lbs (1.6 kg)
Battery Slot
Battery Slot
7'-83 4"
3'-3 7 16"
2'-85 8" 6'-0"
[1829mm]
LongShort
77 8"
311 16
13'-4 12" [4076mm]
LongShort
10 3 16" 103
4"
8x45°
125° 12 5°
2' -1
05 16
4' -1
1 4"
5' -5
1 4"
5' -5
1 4"
5 8"
3 8"
Tube Support Upper Position Tube Support Lower Position
14'-1 5 16"
2'-13 4"615
16"
0°
180°
Top
FrontSide
Drag Chain
Drag Chain
(24.0)
CS/DC
Digital Diagnost CSM3 with Drag Chain Weight Heat Dissipation
Short: 785 lbs (356 kg)* 1365 Btu/hr (444 W)
Long: 862 lbs (390 kg)*
* Full length extension rails add 101 lbs (46 kg).
* Drag Chain is 66 lbs(30kg/110lbs (50kg) with extension)
AD3
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Weight of Wall Connection Box: 36 lbs (16 kg)
FrontSide
51 2"
[1 m m
1' -1
05 8"
[5 m m
2' -8
1 2"
[8 m m
3' -6
5 16
[1
5m m
1'-4" [406mm]
6' -9
7 8"
[2
0m m
2 7 16"
[61mm]
Optional Spacer
13 16
[3 m m
16"
[227mm]
(24.0)
DVS
Digital Vertical Stand VS2 Tiltable Weight Heat Dissipation
563 lbs (255kg) 614 Btu/hr (180 W)
Top
3' -2
13 16
[9 m m
6' -1
07 8"
[2
5m m
4' -1
1 16
[1
0m m
13 16
[3 m m
Top (with Optional Spacer)
4' -2
5 8"
[1
5m m
1'-11 7 16"
[596mm]
1' -7
9 16
[4 m m
9'-101 8"
[3000mm]
(24.0)
MS
DigitalDiagnost TH/TH2 (Wide Tabletop) Weight Heat Dissipation
593 lbs (269 kg) 614 Btu/hr (180 W)
FrontSide
Top
1'-115 8"
[600mm] 1'-115
8" [600mm]
2"
[115mm]
1' -6
11 16
[4 m m
2'-9 7 16"
[850mm] 9 3 16"
[233mm]
9 3 16"
[233mm]
3'-23 4"
[985mm]
3' -0
[9 m m
3'-111 4"
[1200mm]
9'-101 8"
[3000mm]
7'-101 2"
[2400mm]
Minimum clearance required for tabletop insertion - standard method
(either side, not both)
1'-13 8"
[340mm]
Table Transverse ℄
3'-115 8"
[1210mm]
Service Space
Detector Clearance
3' -7
11 16
[1
0m m
3' -3
3 8"
[1
0m m
(23.0)
DGS
Detector and Grid Storage Weight Heat Dissipation
15.75 lbs (7.15 kg) --
4"
[82mm]
97 8"
[2 m m
1'-911 16"
[550mm]
Front
Top
Side
415 16
[1 m m
16"
[125mm]
16"
[24mm]
Front
Top
Side
Optional Spool Attachment
(13.0)
PS
Patient Support Weight Heat Dissipation
154 lbs (70 kg) 0 Btu/hr (0 W)
Side Front
Maximum Patient Weight of 496 lbs (225 kg) X-Ray filtration of < 0.9mm AI equivalent
3'-711 16"
[1110mm]
Top
2'-1011 16"
[881mm]
6' -6
3 8"
[1
0m
SN
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Equipment Support Information
1. General The customer shall be solely responsible, at its expense, for preparation of the site, including any required structural alterations. The site preparation shall be in accordance with this plan and specifications, the architectural/construction drawings and in compliance with all safety and building codes. The customer shall be solely responsible for obtaining all construction permits from jurisdictional authority.
2. Equipment Anchorage Philips provides, with this plan and specifications, information relative to equipment size, weight, shape, anchoring hole locations and forces which may be exerted on anchoring fasteners. The customer shall be solely responsible, through the engineer of record for the building, to provide on the architectural/construction drawings, information regarding the approved method of equipment anchoring to floors, wall and/or ceiling of the building. Any anchorage test required by local authority shall be the customer's responsibility. Stud type anchor bolts should not be specified as they hinder equipment removal for service. Consult with Philips service prior to specifying anchor methods.
3. Floor Loading and Surface Philips provides, with this plan and specifications, information relative to size, weight and shape of floor mounted equipment. The customer shall be solely responsible, through the engineer of record for the building, to provide on the architectural/construction drawings confirmation of the structural adequacy of the floor upon which the equipment will be placed.
Any load test required by local authority, shall be the customer's responsibility.
a. The floor surface upon which Philips equipment (except floor rails) is to be placed/anchored shall be flat and level to within plus or minus 1/16 inch (2mm) over a length of 39" (1m).
b. The floor surface upon which Philips floor rails are to be placed/anchored shall be flat and level to within plus or minus 1/4 inch (6mm) over the length of the floor rails.
4. Ceiling Support Apparatus
a. Philips provides, with this plan and specifications, information relative to size, weight and shape of ceiling supported equipment. The customer shall be solely responsible, through the engineer of record for the building, to provide on the architectural/construction drawings, information regarding the approved method of structural support apparatus, fasteners and anchorage to which Philips will attach equipment. Any anchorage and/or load test required by local authority shall be the customer's responsibility.
b. Contractor to clearly mark Philips equipment longitudinal centerline on bottom of each structural support.
c. The structural support apparatus surface to which Philips equipment is to be attached, shall have horizontal equipment attachment surfaces parallel, square and level to within plus or minus 1/16 inch (2mm).
d. Any drilling and/or tapping of holes required to attach Philips equipment to the structural support apparatus shall be the responsibility of the customer.
e. Fasteners, anchors (e.g. bolts, spring nuts, lock and flat washers), and strip closures shall be provided by the customer.
5. Lighting Lighting fixtures shall be placed in such a position that they are not obscured by equipment or its movement, nor shall they interfere with Philips ceiling rails and equipment movement or otherwise adversely affect the equipment. Such lighting fixture locations shall be the sole responsibility of the customer.
6. Ceiling Obstructions There shall be no obstructions that project below the finished ceiling in the area covered by ceiling suspended equipment travel.
7. Seismic Anchorage (For Seismic Regions Only) All seismic anchorage hardware, including brackets, backing plates, bolts, etc., shall be supplied and installed by the customer/contractor unless otherwise specified within the support legend on sheet S1. Installation of electronic cabinets to meet seismic anchorage requirements must be accomplished using flush mounted expansion type anchor/bolt systems to facilitate the removal of a cabinet for maintenance. Do not use threaded rod/adhesive anchor systems. Consult with Philips regarding any anchor system issues.
8. Floor Obstructions/Floor Coverings There shall be no obstructions on the floor (sliding door tracks, etc.) in front of the Philips electronic cabinets. Floor must be clear to allow cabinets to be pulled away from the wall for service. Contractor to verify the preferred floor covering installation method with Philips.
PS
MS
Floor Surface to be flat and level to within plus or minus 116" (2mm) over a length of 39" (1m).
Cable Outlet
Table Transverse ℄
Ta bl e Lo ng itu di na l ℄
W al l B uc ky
WAP
8'-3"
4' -6
1' 2' 3' 4' 5' 6' 7' 8' 9' 10' 11' 12'
.5 m
1m 1.
5m 2m
2.
5m
3m 3.
5m
2'-1 5 16" 2'-1 5
16"
2' -2
2'
-2
1' -0
9 16
1' -0
9 16
5'-5"
DGS
DVS
4' -0
2'-6"
Portion of existing flush mounted raceway to be filled
1' -0
1'-01 2"
3'-7"
6" 21 2"
4"
8' -6
S1 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Floor & Wall Support Legend
Item Number
Description
Detail Sheet
Notes:
1. Anchors for items that are installed/anchored by customer/contractor shall be provided by customer/contractor.
2. Anchors for items that are installed/anchored by Philips shall be provided by Philips. If customer's engineering documents specify anchors other than those listed in this document, the anchors shall be provided by customer/contractor and installed by Philips.
3. In all instances, the wall and/or floor support are the sole responsibility of the customer/contractor. The customer's architect/engineer of record shall specify wall and/or floor support sufficient for the bolt forces shown on the details.
A Furnished and installed/anchored by Philips (see exceptions - Note 2, below) B Furnished by customer/contractor and installed/anchored by customer/contractor C Installed/anchored by customer/contractor D Furnished by Philips and installed/anchored by contractor E Existing F Future G Optional item furnished by Philips
All dimensions must be off of the finished wall.
If a wall is furred out to hide electrical duct or boxes, the dimensions included in this plan must come off of the finished furred wall.
Floor & Walls Support Layout 1/4" = 1'-0"
Reported Ceiling Height: 9'-6" (2896mm)
B MS Anchors in floor for Diagnost TH SD5
B DVS Support in wall for BuckyDiagnost VS SD4
B DVS Anchors in floor for BuckyDiagnost VS SD4
A VSC Anchors in wall for BuckyDiagnost VS Connection Box (not shown) SD4
A WAP Wi-Fi Access Point - optional wall mounting SD5
A DVS Anchors in wall for BuckyDiagnost VS SD4
SD3PSA Optional Anchors in floor for Patient Support Floor Fixing Plates
D DGS Detector and Grid Storage - optional wall mounting SD5
Table Transverse ℄
Ta bl e Lo ng itu di na l ℄
4'-3" 4'-3" 4'-3"
'-0
UNI
Typ.
8'-3"
4' -6
7'-4"
3' -1
0"
1' 2' 3' 4' 5' 6' 7' 8' 9' 10' 11' 12'
.5 m
1m 1.
5m 2m
2.
5m
3m 3.
5m
S2 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Ceiling Support Legend
Item Number
Description
Detail Sheet
A Furnished and installed by Philips B Furnished by customer/contractor and installed by customer/contractor C Installed by customer/contractor D Furnished by Philips and installed by contractor E Existing F Future G Optional item furnished by Philips
All dimensions must be off of the finished wall.
If a wall is furred out to hide electrical duct or boxes, the dimensions included in this plan must come off of the finished furred wall.
Ceiling Support Layout - Unistrut 1/4" = 1'-0"
Recommended Minimum Ceiling Height: 8'-9" (2667mm) Recommended Maximum Ceiling Height: 10'-6" (3200mm)
Reported Ceiling Height: 9'-6" (2896mm)
Ceiling Heights other than recommended may impact equipment functionality - see Ceiling Height Guide and consult with Philips.
Ceiling Height measured from finished floor to bottom of Unistrut.
B UNI Unistrut (P1001 or equal) SD3
See Ceiling Clearance Detail (sheet SD2) for information on restricted areas for objects that project below the finished ceiling.
Table Transverse ℄
Ta bl e Lo ng itu di na l ℄
8'-3"
8'-0"
4' -6
6"
6' -3
1 8"
1' 2' 3' 4' 5' 6' 7' 8' 9' 10' 11' 12'
.5 m
1m 1.
5m 2m
2.
5m
3m 3.
5m
2'-1015 16"
2' -0
1 4"
CSR
DC
S3 THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Ceiling Support Legend
Item Number
Description
Detail Sheet
A Furnished and installed by Philips B Furnished by customer/contractor and installed by customer/contractor C Installed by customer/contractor D Furnished by Philips and installed by contractor E Existing F Future G Optional item furnished by Philips
All dimensions must be off of the finished wall.
If a wall is furred out to hide electrical duct or boxes, the dimensions included in this plan must come off of the finished furred wall.
Ceiling Support Layout - Equipment 1/4" = 1'-0"
Recommended Minimum Ceiling Height: 8'-9" (2667mm) Recommended Maximum Ceiling Height: 10'-6" (3200mm)
Reported Ceiling Height: 9'-6" (2896mm)
Ceiling Heights other than recommended may impact equipment functionality - see Ceiling Height Guide and consult with Philips.
Ceiling Height measured from finished floor to bottom of Unistrut.
A CSR Philips Equipment Rails SD3
See Ceiling Clearance Detail (sheet SD2) for information on restricted areas for objects that project below the finished ceiling.
A DC Drag Chain SD2
Equipment Option Anchor Style (provided by Philips) Anchor Size (provided by Philips) Qty. Support Size & Material (provided & installed by customer/contractor)
Eleva Examination Control (Wall Mounted)
A SPAX Multipurpose Flat Head Screw #10 x 11 2"L 4 Drywall with min. 3 4" plywood backing
B Toggler Snaptoggle with Round Head Screw and flat fender washers and nuts #BA with #10-24 x 2 12"L 4 Drywall (5 8"D min)
Digital VM Guide Rail with Wall Holders
A A307 bolts (min. ASME Grade 5) 3/8", 3"L 6 Drywall with min. 16 Gauge Steel Backing Thru bolted with head of bolt behind backing and threads sticking out of wall for attachment to wall holders
B Lag Screw 3/8", 3"L 6 Drywall with min 4"x4" Douglas Fir # 2 Grade backing
C Hex Head Tek Screw #14 x 3"L 6 Drywall with min. 10 Gauge Steel Backing
D Hilti KB-TZ 3 8" x 3"L 6 Concrete Wall (min. embedment = 2")
Digital VM Floor Rail A Toggler Alligator with SPAX Multipurpose Flat Head Screw #A6 with SPAX #10 x 11
2"L 20 Concrete Floor
B SPAX Multipurpose Flat Head Screw #10 x 11 2" L 20 Wood Frame Floor
Patient Support Floorplates
A Simpson or Hilti Drop-in with Stainless Steel Phillips Oval Head Screw with dress washer
4" x 1"L 4 Concrete Slab
B Hex Head Lag Bolt with flat washer #14 x 1"L 4 Wood Frame Floor
Wireless Detector Docking Station
A SPAX Multipurpose Flat Head Screw #10 x 11 2 "L 2 Drywall with minimum 3 4" plywood backing
B Round Phillips Head Self Drilling Screw #10-16 x 11 2"" 2 Drywall with min. 10 Gauge Steel Backing
C Toggler SnapToggle with Round Head Screw #BA with #10-24 x 21 2"L 2 Drywall (5 8"D min)
D Toggler Alligator with SPAX Multipurpose Flat Head Screw Toggler Alligator #A6 with SPAX #10 x 11 2"L 2 Concrete Wall
BuckyDiagnost VS (Advanced and Digital)
Wall Anchorage
A Through Bolt 3 8" diameter 4 Minimum 10 gauge steel plate¹, or Minimum 4"x 4" Douglas Fir # 2 grade
B Self Drilling Tek Screw # 14 4 Minimum 10 gauge steel plate¹
C Lag Screw 7 16" 4 Minimum 4" x 4" Douglas Fir # 2 Grade
D Toggler Snaptoggle Anchor 3 8" diameter expansion screw 4 Minimum 10 Gauge Steel Plate¹
BuckyDiagnost VS Connection Box
A SPAX Multipurpose Flat Head Screw #10 x 11 2"L 4 Drywall with Wood Backing
B Round Phillips Head Self Drilling Screw #10-16 x 11 2L" 4 Drywall with min. 10 Gauge Steel Backing
C Toggler SnapToggle with Round Head Screw #BA with #10-24 x 21 2"L 4 Drywall (5 8"D min)
D Round Phillips Head Self Drilling Screw # 10-16 x 11 2"L 4 Metal junction box or raceway
E Toggler Alligator with SPAX Multipurpose Flat Head Screws Toggler Alligator #A6 with SPAX #10 x 11 2"L 4 Concrete Wall or Hollow Concrete Block Wall
Parking Frame for Accessories (RAD)
A SPAX Multipurpose Flat Head Screw #10 x 11 2" L 4 Drywall with min. 3 4" plywood backing
B Phillips Round Head Self Drilling Screw #10-16 x 11 2"L 4 Minimum 16 Gauge Steel Backing
C Toggler SnapToggle with Round Head Screw #BA with #10-24 x 21 2"L 4 Drywall (5 8"D min)
D Toggler Alligator with SPAX Multipurpose Flat Head Screw #A6 with SPAX #10 x 11 2"L 4 Concrete Wall
Pediatric Support Parking Stand
A Hilti KB-TZ with OD flat washer and hex nut 1 2 x 33
4"L TBD Concrete Slab
B Long Lag Bolt with ID and OD flat washer 1 2" x 21
2"L TBD Wood Frame Floor
Stretch Grip Parking Bracket
A SPAX Multipurpose Flat Head Screw #10 x 11 2" L 3 Drywall with min. 3 4" plywood backing
B Toggler SnapToggle with Round Head Screw #BA with #10-24 x 21 2"L 3 Drywall
Tube Crane Box Cover A Round Phillips Head Self Drilling Screw #10-16 x 11 2"L 4 Metal junction box or raceway
¹ Unistrut, angle, or c -channel material is preferred.
Pre-Evaluated and -Approved Anchor Reference List for Philips Installers
Anchors for items that are installed/anchored by customer/contractor shall be provided by customer/contractor.
Anchors for items that are installed/anchored by Philips shall be provided by Philips. If customer's engineering documents specify anchors other than those listed below, the anchors shall be provided by customer/contractor and installed by Philips.
In all instances, the wall and/or floor support are the sole responsibility of the customer/contractor. The customer's architect/engineer of record shall specify wall and/or floor support sufficient for the bolt forces shown on the details.
(15.0) SD1
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED.
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co
SD2
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
G E
IS
P
R O
VI
D
ED
A
S A
C U
ST
O
M
ER
C O
N
VE
N
IE
N C
E, A
N D
IS
N
O T
TO
B
E C
O N
ST
R
U
ED
A S
AR
C
H
IT
EC
TU
R
AL
D R
AW
IN
G S
O R
C O
N
ST
R U
C
TI
O N
D O
C U
M
EN
TS
Ph ilip s as su m es n o lia bi lit y no r o ffe rs a ny w ar ra nt y fo r t he fi tn es s or a de qu ac y of th e pr em is es o r t he u til iti es a va ila bl e at th e pr em is es in w hi ch th e eq ui pm en t i s to b e in st al le d, u se d, o r s to re d.
Ly nd on B
. J oh ns on
G en er al
H os pi ta l H ou st on , T
X R oo m
: E C
K on in kl ijk e Ph ilip s
El ec tro ni cs N
.V . 2
9.
A ll rig ht s re se rv ed . R ep ro du ct io n in w ho le o r i n pa rt is p ro hi bi te d w ith ou t p rio r w rit te n co ns en t o f t he c op yr ig ht h ol de r.
D ra w n
By : P at ric ia D el la
S er a
Pr oj ec t M an ag er
C on ta ct
N um be r:
Em ai l:
M ic ha el
G ar rie
(2
) 9
-7 m ic ha el
.g ar rie ph ilip s.
co m
Drag Chain measurements scale(1:10)
3 1 16"
37 16
1'-2 16"
8"
Bolt M10
Cable Outlet
Option 2 (to wall)
Cable Outlet Option 1 (to ceiling )
Clamping block (min 3 nos to support mounting rail)
Note: Additional 1m cable at cable outlet for future service
Drag Chain Tray
(24.0)
Detail - Drag Chain
DC
A
CSM3 Long - Comfort Track/Move
16"
[125mm]
13'-37 8"
[4060mm]
2"
[114mm]
3' -8
3 16
[1
3m m
6'
-0
[1
9m m
3 16
[1 m m
3' -8
1 4"
[1
4m m
'-4
7 16
[4
6m m
513 16
[1 m m
A
97 8"
[2 m m
B
B
B
A
Ceiling Clearance Detail - Restricted Ceiling Area for Objects that Project Below Finished Ceiling (Not site specific)
A
B
Objects that project below finished ceiling are NOT allowed in this area (e.g. lights, smoke detectors, sprinkler heads, soffits, ceiling rails, A/C vents, etc).
Objects that project more than 3.5" below finished ceiling are NOT allowed in this area (e.g. lights, smoke detectors, sprinkler heads, soffits, A/C vents, etc).
SD3
Pr oj ec t D et ai ls Ph ili ps
C on ta ct s Pr oj ec t D ra w in g
N um be r
N -S
O U
_R
EV
D
D at e D ra w n:
/1 4/
Q uo te
O rd er
R ad io gr ap hy
C
Pl us
V S
7.22.2024
Q -0
.0
TH
E
IN
FO
R M
AT
IO
N
IN
T H
IS
P
AC
KA
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it. Updated .