About this opportunity
The U.S. Geological Survey Forest & Rangeland Ecosystems Science Center seeks one Thermo Fisher QuantStudio 5 384-well quantitative PCR system with quotes due August 4, 2026, at 10:00 a.m. Pacific Daylight Time. This is a brand-name-only procurement evaluated on Lowest Price Technically Acceptable basis with full and open competition; questions are due July 30, 2026, at 5:00 p.m. PDT.
Place of performance and delivery is Boise, Idaho (83702), FOB Destination. The NAICS code is 334516, Analytical Laboratory Instrument Manufacturing, with a 1,000-employee size standard; PSC is 6640, Laboratory Equipment and Supplies.
The system must support 384-well capacity with reaction volumes of 5–20 microliters, bright white LED excitation, at least five optical filters, temperature uniformity within 0.4°C, temperature accuracy within 0.25°C, and multiplexing of up to five targets with specific compatible dyes (FAM/SYBR Green, VIC/JOE/HEX/TET, ABY/NED/TAMRA/Cy3, JUN/ROX/Texas Red, Mustang Purple/Cy5/LIZ, and Cy5.5). Amendment One clarifies that an average sample ramp rate of 2.92°C/sec is acceptable; independently programmable temperature zones are not required; the system should not include a computer; and installation, calibration, and training are not required. The successful contractor must be an authorized Thermo Fisher reseller (Catalog Number A28140).
Offerors must submit a signed SF-18, firm fixed-price quote with lead time, product information identifying manufacturer and part number, SAM UEI, and a manufacturer-issued VPAT/Accessibility Conformance Report satisfying Section 508 requirements. Delivery is required 45 days after award. Standard manufacturer warranty is acceptable.
Notice text
2 versions
Update #2 · Latest ·
***AMENDMENT ONE***
The purpose of this Amendment is to answer vendor questions. No further questions will be accepted.
QUOTES
Quotes must be submitted FOB Destination shipping to Boise, ID, 83702.
Quotes are due Tuesday, August 4, 2026, at 10:00am PD.
Quotes evaluated by Lowest Price Technically Acceptable (LPTA)
Please include lead time in your quote.
Quotes accepted by email to the Contracting Officer, Elizabeth Adam, at elizabeth_adam@ios.doi.gov only.
QUESTIONS and ANSWERS
1. The salient characteristics require an average sample ramp rate of 3.66°C/sec or greater. Thermo Fisher's published specification for the QuantStudio 5 384-well system states an average sample ramp rate of 2.92°C/sec. Since this is a brand-name-only procurement, please confirm whether 2.92°C/sec is acceptable or whether the requirement will be corrected? This is acceptable.
2. The brand-name justification refers to independently programmed temperature zones across the heat block. Thermo Fisher's published literature indicates that VeriFlex temperature zones are not applicable to the 384-well QuantStudio 5 configuration. Please confirm whether independently programmable temperature zones are a mandatory requirement? This is not required.
3. Please confirm the required Thermo Fisher catalog number and whether the requested configuration must include a laptop or desktop computer? This should not include a computer. Catalog number: A28140
4. Are installation, calibration, onsite setup or user training required? No
5. Is the manufacturer's standard warranty acceptable, and is any extended service coverage required? Standard warranty is acceptable
6. Must the successful contractor be an authorized Thermo Fisher reseller, and should evidence of authorization be included with the quotation? Yes
7. Please confirm whether a manufacturer-issued VPAT/Accessibility Conformance Report and supporting accessibility narrative will satisfy the Section 508 submission requirements. Yes
***********************************************
Offerors are to include a technical description of the items being offered in sufficient detail to evaluate compliance with the requirements in the solicitation as seen in the Salient Characteristics (see Attachment). This may include product literature or other documents, if necessary, and is to include the country of manufacture.
Offerors shall return the following:
Signed SF-18
Signed Amendment
Firm, fixed price quote
Product information
System for Award Management (SAM) Unique Identification Entity (UEI)
Update #1 ·
Thermo Fisher QuantStudio 5 Quantitative Polymerase Chain Reaction (qPCR) Machine
BRAND NAME ONLY
REQUEST FOR QUOTE
INTRODUCTION
This is a combined synopsis/solicitation for commercial items prepared in accordance with the format in Revolutionary FAR Overhaul (RFO) Part 12 as supplemented with additional information included in this notice.
Full and open competition.
DESCRIPTION
The US Geological Survey (USGS) Forest & Rangeland Ecosystems Science Center (FRESC) of Boise Idaho requires a qPCR machine as described in the attached Salient Characteristics.
RELEVANT INFORMATION
The North American Industry Classification System (NAICS) code 334516 Analytical Laboratory Manufacturing and associated size standard 1,000 applies to this announcement, along with Product Service Code 6640 Laboratory Equipment and Supplies.
QUESTIONS
Questions due Thursday, July 30, 2026, at 5:00pm PD to elizabeth_adam@ios.doi.gov.
QUOTES
Quotes must be submitted FOB Destination shipping to Boise, ID, 83702.
Quotes are due Tuesday, August 4, 2026, at 10:00am PD.
Quotes evaluated by Lowest Price Technically Acceptable (LPTA)
Please include lead time in your quote.
Identify the manufacturer and part number in your quote.
Quotes accepted by email to the Contracting Officer, Elizabeth Adam, at elizabeth_adam@ios.doi.gov only.
Offerors are to include a technical description of the items being offered in sufficient detail to evaluate compliance with the requirements in the solicitation as seen in the Salient Characteristics (see Attachment). This may include product literature or other documents, if necessary, and is to include the country of manufacture.
Offerors shall return the following:
Signed SF-18
Firm, fixed price quote
Product information
System for Award Management (SAM) Unique Identification Entity (UEI)
Attached Pages:
Salient Characteristics
Clauses and Provisions
ICT Accessibility Requirements Statement per Section 508 of the Rehabilitation Act
Brand Name Justification
Files
| File | Type | Posted |
|---|---|---|
| Sol_140G0326Q0060_Amd_0001.pdf | ||
| Brand_Name_Justification_Redacted.pdf | ||
| Sol_140G0326Q0060.pdf |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| 6515--Purchase of One (1) QS3 qPCR System, One (1) iBright FL1500 Imaging System, and Associated Accessories, Brand Name or Equal To for Oklahoma City VA Medical Center (OKCVAMC) | Veterans Integrated Service Network 19 | No set-aside | |
| Brand Name Nikon Microscope and Camera | — | Total Small Business | |
| GEMINI. HP. L. ASY: interferometer, Gemini-Felix, and PDM module | Department of Energy | No set-aside | |
| Rapid, Automated Particle Characterization (Morphologi 4 ID) | Department of Energy | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity