The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 10 is seeking a contractor to provide annual inspections, preventative maintenance, and unscheduled repairs for ceiling lift systems at the Dayton VA Medical Center. The solicitation (36C25025Q0219) requires comprehensive services for 254 ceiling lift systems, including Guldmann and Liko models, with specific requirements such as 48-hour response time for service calls and electronic submission of typed inspection reports. Contractors must perform weight testing and safety inspections according to ISO 10535 standards, with services to be conducted during normal business hours (Monday-Friday, 8AM-5PM). Evaluation criteria and specific requirements are detailed in the solicitation documents, with questions due by January 15, 2025, at 8:00 AM EST, and proposals due January 21, 2025, at 8:00 AM EST. The contract is classified under NAICS 811210 for Electronic and Precision Equipment Repair and Maintenance.

The solicitation is not designated as a specific set-aside, though it notes a 100% Service-Disabled Veteran-Owned Small Business (SDVOSB) preference. The previous incumbent was Progressive X-Ray, with a contract value of $216,900. The contract structure includes a base year from February 1, 2025, to January 31, 2026, with four additional one-year option periods extending through January 31, 2030. The services will cover 254 ceiling lift systems located across 8 buildings on a single campus, comprising 241 Guldmann GH3+ units, 6 Guldmann GH3 units, 2 Guldmann GH2 units, and 5 Liko units. Contractors must provide unlimited unscheduled services, vendor-supplied parts and batteries, and ensure initial inspections are completed by the end of February 2025. The size standard for the NAICS category is $34 million, providing potential for significant contract opportunities.

Notice text

2 versions

Update #2 · Latest ·

Page 11 of 11

Evaluation Factors

Describe your plan for adhering to manufacturer, ISO 10535, and VA guidelines for inspection and repair of Guldmann and LIKO ceiling lift systems as outlined in the SOW.

Provide evidence that your servicing technicians are qualified to perform repairs and ISO 10535 compliant inspections for Guldmann and LIKO ceiling lift systems as outlined in the SOW.

Describe your plan to provide the 48-hour telephone or email response time for unscheduled service.

Describe your plan to acquire parts within 4 business days of need determination for both scheduled and unscheduled maintenance.

Describe your plan for coordinating servicing visits with VA staff, specifically Healthcare Technology Management.

Describe your plan to provide documentation of testing and repairs as described in the SOW.

Key personnel: provide contact information and qualifications of your field service manager for the Dayton site.

Dayton Ceiling Lift Questions and Answers:

How many Liko and Guldmann lifts are at the site.

Liko (5), Guldman GH2 (2), Guldmann GH3 (6), Guldmann GH3+ (241)

Is there more than one location?

All units are located on one campus across 8 buildings.

How many 'unscheduled services' do you want figured in the quote? Or do you want a 'not to exceed' dollar amount.

Unlimited unscheduled services.

Confirm that 'parts' are in addition to the contract, once approved by the COR

All parts are to be supplied by the vendor under the contract.

Are we to include batteries or will the VA provide.

Batteries are to be supplied by the vendor.

Do you have a list of the systems with manufacturer and location? If list can t be provided, can you confirm manufacturers and quantities of each?

Liko (5), Guldman GH2 (2), Guldmann GH3 (6), Guldmann GH3+ (241). All units are located on one campus across 8 buildings.

Who was the previous incumbent?

Progressive X-Ray

What was the previous contract number and award amount?

36C25020P0469 $216,900.00

Under scope of program 4.2.1.4 section e. it references load testing to 1.5x system capacity but Liko manufacturer specifications is to test to 100% of system capacity using the motor. Please clarify if the ISO standard is to be used for testing for both Liko and Guldmann lift systems?

Use of the ISO standard is required on both Liko and Guldmann lift systems.

When are the lifts due for service?

All units are slated for annual inspections in February. The first round of inspections under the contract are to be completed by the end of February 2025.

Can they confirm that this is for Annual Inspections and a Service Agreement?

Annual inspections are required for all lift systems. Servicing/repair is only required for Guldmann units.

Update #1 ·

RME Tool Sharpening and Repair

Files

File Type Posted
36C25025Q0219 0001.docx DOCX document
36C25025Q0219_1.docx DOCX document

Notice history

Notice Type Posted
This notice Latest solicitation Solicitation
J065--Patient Lift Annual Inspection | Base+4 OY's Original Pre-Solicitation

Similar open opportunities

Opportunity Agency Set-aside Responses due
NTX Elevator Inspection Bonham, Texas 75418, United States · NAICS 541350 Veterans Integrated Service Network 17 Service-Disabled Veteran-Owned Small Business
WTX Elevator Inspections and Load Testing Services Texas 79720, United States · NAICS 238290 Veterans Integrated Service Network 17 Service-Disabled Veteran-Owned Small Business
H212--Annual Test/Inspection of FireFly Device and Door Drop Cleveland VA Medical Center Department of Veteran Affairs Louis Stokes VA Medical Center Cleveland, Ohio 44106, United States · NAICS 541350 Veterans Integrated Service Network 10 Service-Disabled Veteran-Owned Small Business
J061--SOURCES SOUGHT NOTICE - Triennial Electrical Inspection Testing and Maintenance Services Requirement - Louis A Johnson VA Medical Center 36C24527Q0039 Department of Veterans Affairs Louis A Johnson VA Medical Center Clarksburg, West Virginia 26301, United States · NAICS 238210 Veterans Integrated Service Network 5 No set-aside
J042--Fire Alarm, Suppression Systems, and Hood Testing, Inspection, and Certification Services - SAVAHCS Department of Veterans Affairs Southern Arizona VA Health Care System Tucson, Arizona 85723, United States · NAICS 541350 Veterans Integrated Service Network 22 Service-Disabled Veteran-Owned Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors