About this opportunity

This pre-solicitation notice seeks capability statements for a Cytek Aurora 5 Laser spectral analyzer system to support research activities at the VA Pittsburgh Healthcare System Research department. The system must include five lasers, the ability to detect up to 40 colors and 64 fluorescence channels, 96-well sampling capability, and full spectrum unmixing without filters or PMTs. Interested parties who are OEMs, authorized dealers, distributors, or resellers of all listed items may submit capability statements by 16:00 EST on June 11, 2021 to Elaine DiBucci at the Department of Veterans Affairs Veterans Health Administration. Responses will be evaluated to determine qualified sources for a potential brand name or equal RFQ to be issued under NAICS code 334516.

There is no set-aside designated for this requirement. The primary NAICS code is 334516 and PSC code is J070. The notice does not name any incumbents. The budget for this procurement is unspecified, but the required equipment includes a Cytek Aurora 5 Laser system and associated components capable of detecting up to 100 rare immune cell subsets through single cell analysis to validate protein expression.

Notice text

The VA Pittsburgh Healthcare System Research department has a need for a spectral analyzer system to support current research activities, Brand Name or equal to the Cytek Auroa 5 and components described below.

Cytek Aurora 5 or equivalent. System consists of:

Cytek Aurora 5 Laser US/V/B/YG/R (64+3 Channel)

5-laser flow cytometer

Workstation w/computer, monitor, keyboard, and mouse

Intel Core i7, 3.6 GHz

64-bit

32 GB RAM

500 GB Hard Drive

530 NVIDIA GeForce Video Card

32-in UHD 4K Monitor

Microsoft Windows 10 Pro

Additional copy of SpectroFloTM

Fluidic reservoirs

QC and consumable parts kit

1-year warranty

Cytek Auto Micro-Sampling System

Salient Characteristics

Five lasers: Violet, Blue, Yellow-Green Red, Ultraviolet.

3 scattering channels

Up to 40 colors detected

64 fluorescence channels

Capable of detecting up to 30 fluorochromes using spectral analysis.

96-well sampling capability

Autofluorescence extraction

High-sensitivity Coarse Wavelength Division Multiplexing (CWDM), 16-channel semiconductor detector

Flat top beam profiles

Full emission spectra

Highly overlapping dyes

Multiparametric data

100 mW 405 nm laser

SSC detector

System must be able to:

Detect up to 40 colors including fluorochromes with overlapping emission spectra.

Immune cell subsets identified by single cell RNA seq need to be validated by VA researchers for protein expression. System must be able to detect up to 100 rare immune cell subsets

The equipment must have high sensitivity, which is optimal to analyze the most challenging cell populations such as cells with high autofluorescence or low levels of expression of biomarkers. The machine must not require changing optical filters for any fluorochromes, and any commercially available fluorochromes must be be excited by the system.

Full spectrum unmixing for detection of fluorescent signals. No band pass filters or photomultiplier tubes (PMTs)

Cytometer must have flat top laser beam profile with narrow vertical beam height for enhanced capacity to detect rare events and particles smaller than cells

Ability to use violet laser line for cell or particle scatter measurements.

Operating voltage can be 100 or 240 V

No special environmental requirements (air, humidity, and temperature)

This is a Sources Sought synopsis only. The purpose of this synopsis is to gain knowledge of potential qualified sources and their size classification (large business, small business, 8(a), SDVOSB, etc.), relative to NAICS code 334516. Responses to this synopsis will be used by the Government to make appropriate acquisition decisions. Based on responses to this synopsis, the Government may issue this RFQ as a Brand Name or Equal requirement in accordance with VAAR Clause 852.211-73 on eBuy or Fed Biz Ops. This notice is not a request for competitive proposals; however, any responsible source who believes it is capable of meeting the requirement and who is the Original Equipment Manufacturer (OEM), authorized dealer, authorized distributor or authorized reseller of all the items listed above, may submit a capability statement to the contracting office no later than 16:00 EST, Friday, June 11, 2021. Interest/capability statements may be sent to Elaine DiBucci at elaine.dibucci@va.gov. No telephone responses will be accepted. A determination not to compete the proposed requirement based upon the responses to this notice is solely within the discretion of the Government.

Files

File Type Posted
36C24E21Q0141.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
Flow Cytometry Reagents Baltimore, Maryland 21201, United States · NAICS 325413 Veterans Integrated Service Network 5 No set-aside
6550--Optilite analyzer reagents See attachments · NAICS 325413 Veterans Integrated Service Network 22 No set-aside
6640--jess simple western automated system NAICS 334516 Veterans Integrated Service Network 6 No set-aside
6540--Carl Zeiss Visulas Combi YAG Green (VA-26-00065932) NAICS 339115 Veterans Integrated Service Network 12 No set-aside
Combined Synopsis Solicitation- Acquire service ag NAICS 811210 Office of Logisitics and Acquisition Operations No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors