6515--Digital Non-Mydriatic Retinal Camera
Open Pre-Solicitation Posted
- Solicitation number
- 36C25027Q0097
- Agency
-
Veterans Integrated Service Network 10
Veterans Health Administration, Department of Veterans Affairs
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 339112 Surgical and Medical Instrument Manufacturing
- Place of performance
- East Indianapolis VA Clinic Indianapolis, Indiana 46229, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of Veterans Affairs, Veterans Health Administration, Veterans Integrated Service Network 10 is requesting market research information on one digital non-mydriatic fundus camera system with retinal image-capture software and vendor installation, with responses due November 13, 2026. No set-aside is used, and no contract will be awarded from this notice. The system is for the new East Indianapolis VA Clinic Tele-Eye Screening room at 9402 East 21st Street, Indianapolis, Indiana 46229. The agency states particular interest in products made in America that meet the domestic end-product test of FAR 25.101, and it asks respondents to submit place of manufacture and socio-economic status.
The required system must be a compact, non-mydriatic imager that captures color retinal photography and fundus autofluorescence images at a minimum of 32 megapixels. It must align and focus using infrared light and image through pupils of approximately 3.3 mm or smaller without routine dilation. Automated alignment and focusing must allow manual override, and the system must flag insufficient image quality. Other requirements include red-free and blue-spectrum viewing modes, adjustable ISO 200-6400 sensitivity, an integrated preview display, DICOM and standard image-format support with electronic medical record integration capability, and a minimum one-year manufacturer warranty. No incumbent is named, and the notice does not reference an industry day.
Responses go by email to Brian.Rosciszewski@VA.gov and must include product information, company details, SAM Unique Entity Identifier, and relevant documentation. Respondents must be registered in SAM to be eligible for future awards, and the government will not reimburse response costs. No anticipated award value or period of performance is stated.
Notice text
Digital Non-Mydriatic Fundus Camera System
This announcement requests information from industry to support market research and preliminary acquisition planning for the procurement of (1) one digital non-mydriatic fundus camera system with retinal image-capture software and vendor installation for the new East Indianapolis VA Clinic Tele-Eye Screening room.
Of particular interest is identifying products that are made in America and meet the domestic end-product test of FAR 25.101.
This is not a solicitation for proposals. No contract will be awarded from this notice.
REQUIREMENT
The agency is seeking information on sources of supply for one digital non-mydriatic fundus camera system with retinal image-capture software.
Description: Compact, non-mydriatic digital retinal imaging system designed to acquire high-resolution color and fundus autofluorescence images of the human retina. The system shall support screening, documentation, and monitoring of ocular conditions without routinely requiring pharmacological pupil dilation.
Salient Characteristics:
Shall provide color retinal photography and fundus autofluorescence imaging.
Shall produce retinal images with a resolution of at least 32 megapixels.
Shall use infrared light for patient alignment and focusing before image capture.
Shall acquire images through pupils approximately 3.3 mm in diameter or smaller without routinely requiring dilation.
Shall provide automated alignment, focusing, exposure adjustment, image capture, and switching between anterior-eye and retinal observation.
Shall permit the operator to override automatic focusing and manually adjust focus when necessary.
Shall automatically evaluate image focus and eye alignment and alert the operator when image quality is insufficient.
Shall provide digital image-processing modes capable of emphasizing retinal vascular and structural features, including red-free and blue-spectrum viewing modes.
Shall include contrast-enhancement processing to improve differentiation between retinal blood vessels and surrounding tissue.
Shall use low-intensity flash illumination to minimize pupil constriction and support sequential image acquisition.
Shall provide adjustable image-sensitivity settings covering a range equivalent to ISO 200 through ISO 6400.
Shall include an integrated display that allows the operator to preview and verify captured images at the imaging device.
Shall include a complete retinal image-capture and management software package supporting image acquisition, patient-data entry or import, image review, processing, storage, and export of retinal images and associated patient information in DICOM and commonly used image formats. The software shall support DICOM worklist and electronic medical record integration when properly configured.
Minimum of one-year manufacturer warranty coverage.
Interested parties who can provide a product that meets the description above are requested to submit the following information:
Product information documenting salient characteristics,
Company name, address and point of contact with phone number,
SAM Unique Entity Identifier (UEI),
Place of manufacture,
Socio-economic status to identify the business class as a large business, SDVOSB, VOSB, small business, Woman Owned, etc.,
Any other pertinent company documentation.
Note: All interested parties must be registered in SAM (System for Award Management) to be eligible for the award of a future Government contract.
Send submissions to Brian.Rosciszewski@VA.gov., reference 36C25027Q0097 in email subject.
This market research is for informational and planning purposes. This is not a solicitation and does not obligate the Government to issue a solicitation. Please be advised that the government will not pay any costs for responses and information submitted.
Respondents may not be notified of the results of this notice; it is at the discretion of the Government to release any results.
Thanks in advance for your time and consideration in supporting America s veterans.
** End of Sources Sought Notice **
Files
| File | Type | Posted |
|---|---|---|
| 36C25027Q0097.docx | DOCX document |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| NX EQ Microscope Spine | VA Strategic Acquisition Center Frederick | No set-aside | |
| DA10--RFI VHA Radiology and Cardiology Enterprise Imaging | VA Technology Acquisition Center | No set-aside | |
| 6515--NX EQ Image Guided Surgical Navigation System (VA-26-00083972) | VA Strategic Acquisition Center Frederick | Service-Disabled Veteran-Owned Small Business | |
| 6540--Carl Zeiss Visulas Combi YAG Green (VA-26-00065932) | Veterans Integrated Service Network 12 | No set-aside | |
| 6525--SEP 2026 Equipment Only Consolidation | National Acquisition Center | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity