The Department of Veterans Affairs (VA) is conducting a Sources Sought Notice to identify potential suppliers for high-throughput superspeed centrifuges and related components for multiple VA Medical Centers. The procurement specifically targets the Sorvall LYNX 4000 centrifuge or equivalent, with detailed technical requirements including 200-240V power, 4L maximum capacity, 24,000 rpm speed, and 68,905 X g rotational force. Potential suppliers must submit comprehensive responses addressing 19 specific questions covering their capabilities, business size status, manufacturing/distribution capabilities, and compliance with various federal regulations. The VA will use these responses to inform its acquisition strategy and determine potential sources for the equipment. Responses must be submitted via email to john.harrison2@va.gov by January 3, 2025, at 4:00 PM EST, with the opportunity falling under NAICS code 339112 and a size standard of 1,000 employees.
No specific set-aside has been designated for this procurement, meaning businesses of all sizes can potentially compete. The notice does not indicate an incumbent contractor or specific award value, though it includes detailed equipment specifications for two primary rotors: a swinging-bucket rotor for 4x1L capacity and a fixed-angle rotor for 6x250mL capacity. The procurement includes not just the centrifuge units, but also various adapters, start-up services, and a total care warranty plan. Suppliers are encouraged to provide information about FDA clearances, leasing options, federal supply schedule contracts, and general pricing details. The VA seeks comprehensive information to ensure potential suppliers can meet the technical and operational requirements of the medical instrument procurement, with a focus on gathering market research to support future solicitation development.
Notice text
Presolicitation Notice
Presolicitation Notice
Page 6 of 6
Presolicitation Notice
*=Required Field
Presolicitation Notice
Page 1 of 6
THIS REQUEST FOR INFORMATION (RFI) / SOURCES SOUGHT NOTICE IS ISSUED SOLELY FOR INFORMATION AND PLANNING PURPOSES.
THIS IS NOT A SOLICITATION.
SUBMISSION OF INFORMATION ABOUT PRICING, DELIVERY, THE MARKET, AND CAPABILITIES IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI FOR PLANNING PURPOSES IN ACCORDANCE WITH (IAW) FAR 15.201(e).
DISCLAIMER
This RFI is issued solely for information and planning purposes and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. IAW FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.
***ALL submissions SHALL Include the following statement: 605 - 36C26225Q0240 Centrifuge
This is NOT a solicitation announcement. This is a Sources Sought Notice / RFI only. The purpose of this Sources Sought Notice / RFI is to gain information about potential qualified sources and their size classification relative to NAICS 339112 (Size standard of 1,000 Employees). Responses to this Sources Sought Notice / RFI will be used by the Government to make appropriate acquisition decisions. After review of the responses to this Sources Sought Notice / RFI, further RFIs and/or a solicitation or other announcements may be published.
The Department of Veterans Affairs (VA) is seeking sources capable of providing Infusion Pump units and components that can address the attached list of DRAFT requirements for multiple VA Medical Centers.
The list of DRAFT characteristics is intended to be descriptive, not restrictive, of the supplies/services that are required.
If your company is interested and capable of providing the required supplies/services, please provide the information indicated below. Response to this notice should include company name, address, point of contact, size of business pursuant to the following questions:
(1) Submit your capabilities statement illustrating how your organization can/cannot meet the list of SALIENT CHARACTERISTICS requirements. For instances where your company cannot meet the SALIENT CHARACTERISTICS requirement(s), please explain. Also, please show how your company meets/exceeds each requirement.
(2) Please review the list of SALIENT CHARACTERISTICS requirements and provide any additional feedback or suggestions. If none, please reply as N/A.
(3) Please indicate the size status and representations of your business, such as but not limited to: Service-Disabled Veteran-Owned Small Business (SDVOSB), Veteran-Owned Small Business (VOSB), HUBZone, Woman Owned Small Business (WOSB), Large Business, etc.
(4) Is your company considered small under the NAICS code identified in this RFI?
(5) Are you the manufacturer, authorized distributor, and/or can your company provide a solution to the required supplies/services described in the list of SALIENT CHARACTERISTICS?
(6) If you are a large business, do you have any designated/authorized distributors? If so, please provide their company name, telephone, point of contact and size status (if available).
(7) If you re a small business and you are an authorized distributor/reseller for the items identified above, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified.
(8) Limitations on Subcontracting: How does your business ensure compliance with the limitations on subcontracting as outlined in 13 CFR § 125.6?
(9) Are the items you are identifying/providing considered Commercial of the Shelf (COTS) items as defined in FAR Part 2.101 under commercial items?
(10) Non-Manufacturer Rule: If applicable, can you confirm your business complies with the Non-Manufacturer rule? Specifically, does your company: Provide a product from a small business manufacturer or processor? Not exceed 500 employees? Primarily engage in the retail or wholesale trade and normally sell the type of item being supplied? Take ownership or possession of the item(s) with its personnel, equipment, or facilities in a manner consistent with industry practice?
(11) Please indicate whether your product conforms to the requirements of the Buy American Act?
(12) What is your lead time to deliver a single unit with all components? Is there scale in lead time with greater quantities? Please elaborate.
(13) What is estimated life span of your solution? What support/services does that entail?
(14) Does your proposed equipment have FDA clearance? Please specify what FDA clearance(s) have been obtained.
(15) Does your organization offer a leasing solution? Please elaborate.
(16) Does your company have a Federal Supply Schedule (FSS) GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract? If so, please provide the contract number(s).
(17) If you are an FSS GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract holder, are all the items/solutions you are providing information about available on your schedule/contract? Please state if all or some items are available on the contract.
(18) General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award.
(19) Please provide your SAM.gov Unique Entity ID/Cage Code number.
Responses to this notice shall be submitted via email to john.harrison2@va.gov Telephone responses will not be accepted. Responses must be received no later than 01/03/2025, by 4:00 PM EST.
All responses to this Sources Sought/RFI will be used for planning purposes only. Responses to this Sources Sought Notice / RFI are not considered a request to be added to a prospective bidders list or to receive a copy of the solicitation. If further RFIs and/or a solicitation or other announcement is issued as a result of the information provided from this RFI, all interested parties must respond to the specific posting separately IAW the specifications of that announcement.
Brand name Information:
Manufacturer & Part Number: Sorvall LYNX 4000
Type: High-throughput Superspeed Centrifuge
Possible Capabilities: Centrifuge
Equal to product Information:
Main component: High-throughput Superspeed Centrifuge
Possible Capabilities: Centrifuge
***ALL submissions SHALL Include the following statement: 605 - 36C26225Q0240 Centrifuge
75006580 1 LYNX 4000 200-240 V 50/60 HZ: Sorvall 2 EA
LYNX 4000: Superspeed Centrifuge,
Capacity: 4 x 1000mL, Max. RCF: 68,905
x g, Max. Speed: 24,000 rpm,
Refrigerated: Yes, Superspeed centrifuge,
Accel/Decel Profiles: 9 Accel, 10 Decel
STARTCCEN 2 START-UP SVC COMPL. CENTR: Unity Lab 2 EA
Services Start-up Service (Standing
Centrifuges), Duration: 1 Day, For Use
With: Floor Model Centrifuges, SpeedVacs,
Warranty and Service Offering: Start-up Service.
TCWCEN3 3 ULS TOTAL CARE WNTY PLN SS LC: Unity 2 EA
Lab Services Total Care Warranty for
Thermo Scientific SuperSpeed and Large
Capacity Centrifuges, For Use With:
SuperSpeed and Large Capacity Centrifuges.
NC3246059 4 SERVICE TRAVEL: Vendor Catalog # 4 EA
SVS-TRAVEL-ZONE 4/NEW
096145075 5 ROTOR LYNX F14-14X50CY: Thermo Scientific 1 EA
Fiberlite F14-14 x 50cy Fixed-Angle Rotor,
For Use With: Sorvall LYNX 6000 and 4000
superspeed centrifuges, Capacity: 14 x 50mL,
Max. Speed: 14,000 rpm, Max. RCF: 33,746 x g,
Carbon fiber composite, Fixed angle
rotor, Angle: 34 deg., Biocontainment: Yes
75100378 6 ADPT 50 TO 15ML CONICAL 2/PK: Thermo 3 EA
Scientific Fiberlite Rotor Adapters,
Description: Adapter for 15mL conical tubes.
75003000 7 ROTOR LYNX BIOFLEX HC 4X1L: Thermo 1 EA
Scientific BIOFlex HC Swinging-Bucket
Rotor, For Use With: Sorvall LYNX 6000
and 4000 superspeed centrifuges,
Capacity: 4 x 1000mL, Max. Speed:
5,500 rpm, Max. RCF: 7,068 x g,
Stainless steel rotor body with
aluminum buckets, Swinging bucket
rotor, Angle: Horizontal,
Biocontainment: Yes
75007306 8 ADAPTER TX-1000 15MLCONICALPK4: 1 EA
Adapters for Thermo Scientific TX-1000
Swinging Bucket Rotor, Capacity: 96 x
15mL, For Use With: 15mL Conical Tube,
Max. Tube Diameter: 17.5mm, Max. Tube
Length: 121mm.
75003672 9 37 X 10 ML ADAPTER 4 PK: Thermo 1 PG
Scientific 10mL Blood Collection Tubes
for BIOLiner Rotor Bucket Adapters,
Description: 10mL Blood Collection
Tubes, No. per Pack: 4.
75003662 10 F14S-6X250 LE FIBERLITE ROTOR: 2 EA
Thermo Scientific Fiberlite F14-6 x
250LE Fixed Angle Rotor, for Sorvall
Legend T Plus, Multifuge 3 Plus, For
Use With: Thermo Scientific Sorvall
Legend T Plus and Multifuge 3 Plus
Centrifuges, Capacity: 6 x 250mL, Max.
Speed: 10,000 rpm, Max. RCF: 15,316 x g,
Angle: 23 deg.
75101073 11 ADPT 250 TO 15ML CON 2/PK: Thermo 3 EA
Scientific Fiberlite F14-6 x 250LE
Rotor Adapters, Description: 15 mL
Conical Tube (Set of 2).
75100136 12 ADPT 250 TO 50ML CON 2/PK: Thermo 6 EA
Scientific Fiberlite F14-6 x 250LE
Rotor Adapters, Description: 50mL
Conical Tube (Set of 2).
FREIGHT 13 FREIGHT 1 EA
TOTAL COST: $
SALIENT CHARACTERISTICS
High-throughput Superspeed Centrifuge, Rotors, and Adaptors
Superspeed centrifuge Must be a high-throughput floor model
Power requirements are 200-240V 50/60hz
Must have maximum capacity of 4L
Must produce centrifugation speed of no less than 24,000 rpm
Must be able yield rotational force of no less than 68,905 X g
Must be able to perform at a temperature range of 10 degree C to 40 degree C
Must have a timer control that can allow 99 hours hold-time
Must have capability of quick and secure changes of rotors
Must use a high torque brushless drive system
Must be able to perform no less than 9 acceleration and 10 deceleration profiles
Must be quiet and produce no more than 60 dBA
Must offer installation and training
Must offer long-term warranty
Rotor 1 for 4 X 1L capacity
Must be a swinging-bucket rotor
Must be able to be used in the above superspeed centrifuge
Must be able to process 35 or more tube styles from microtubes to 1L bottles, conical tubes, microplates, tissue culture flasks, and blood bags
Must be stainless steel rotor body and aluminum bucket
Must be able to handle no less than 7,068 x g and no less than 5,500 rpm
Must support a range of capacities and be able to accommodate adapter for 15 mL conical tube and adapter for 10 mL centrifuge tubes
Adapters for Rotor 1
Must be adapter suitable for use in the above rotor.
Must be adapter for use in swinging-bucket carrying 15 mL conical tubes
Must be adapter for use in swinging-bucket carrying 10 mL centrifuge tubes
Rotor 2 for 6 X 250 mL capacity
Must be a fixed angle rotor 1
Must be able to be used in the above superspeed centrifuge
Must be lightweight, fiberlite carbon fiber rotors and be corrosion- and fatigue-resistant
Must support a range of capacities and be able to accommodate 250 mL centrifuge bottles or 10 mL, 15 mL, 50 mL, and 50 mL conical centrifuge tubes with corresponding adapters
Must provide no less than 15 years warranty
Adapters for Rotor 2
Must be adapter suitable for use in the above rotor 2.
Must be adapter for use in fix-angle rotor with 250 mL capacity carrying 15 mL conical tubes
Must be adapter for use in fix-angle rotor with 250 mL capacity carrying 50 mL conical tubes
Superspeed centrifuge
Must be a high-throughput floor model
Must have maximum capacity of 4L
Must produce centrifugation speed of no less than 24,000 rpm
Must be able yield rotational force of no less than 68,905 X g
Must be able to perform at a temperature range of 10 degree C to 40 degree C
Must have a timer control that can allow 99 hours hold-time
Must have capability of quick and secure changes of rotors
Must use a high torque brushless drive system
Must be able to perform no less than 9 acceleration and 10 deceleration profiles
Must be quiet and produce no more than 60 dBA
Must offer installation and training
Must offer long-term warranty
Rotor for 6 X 250 mL capacity
Must be a fixed angle rotor
Must be able to be used in the above superspeed centrifuge
Must be lightweight, fiberlite carbon fiber rotors and be corrosion- and fatigue-resistant
Must support a range of capacities and be able to accommodate 250 mL centrifuge bottles or 10 mL, 15 mL, 50 mL, and 50 mL conical centrifuge tubes with corresponding adapters
Must provide no less than 15 years warranty
Adapter for the above Rotor
Must be adapter suitable for use in the above rotor
Must be adapter for use in fix-angle rotor with 250 mL capacity carrying 15 mL conical tubes
Must be adapter for use in fix-angle rotor with 250 mL capacity carrying 50 mL conical tubes
Files
| File | Type | Posted |
|---|---|---|
| 36C26225Q0240.docx | DOCX document |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Osmometer Equipment Replacement (OsmoPro Max) | Veterans Integrated Service Network 16 | No set-aside | |
| 6550--Cepheid GeneXpert Testing Supplies | Veterans Integrated Service Network 5 | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity