This is a specification sheet for the procurement, delivery, installation, and certification of an audiometric testing system to support military occupational health and readiness testing.

The contractor shall deliver a 6-Station Single Wall Audiometric Booth (Model CL-M6 PR) with an add-on door featuring a window, or alternatively, a Wireless Automated Hearing System. The wireless option must provide 30-40 decibels of attenuation across octave bands from 250-8000Hz and comply with ANSI S3.6, OSHA 1910.95, IEC 60645-1, ANSI S12.71, and DODI 6055.12 standards. The system includes a Primary Audiometer Package (CCA-200 Mini Audiometer with TDH-39 Headset, response switch, 12-foot serial extension cable, mounting hardware, calibration certificate, USB components, and software key); an Expansion Audiometer Package (Benson Medical CCA-200mini with similar components but 6-foot extension cable); and a BAS-200 Bio-Acoustic Simulator. For wireless systems, headsets, tablets, and software must be included. The contractor is responsible for on-site assembly and installation of the booth, integration of all audiometers and simulation equipment, establishment and verification of daily calibration references, and SR-Ambient Noise Certification to ensure compliance with regulatory and operational standards. The customer shall provide an on-site dumpster for disposal of installation waste, shipping crates, and packaging materials. Note that a personal computer is not included and must be supplied by the customer.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPEC SHEET FOR: 6 MAN AUDIOBOOTH or 6 WIRELESS AUTOMATED HEARING SYSTEM

Project Title: Audiometric Testing Booth and Equipment Installation or Wireless Automated Hearing Test System

1. Objective

The purpose of this SPEC Sheet is to define the requirements for the procurement, delivery, installation, and certification of a multi-station audiometric testing system and associated audiology equipment to support military occupational health and readiness testing.

2. Scope of Work

The Contractor shall provide all necessary labor, equipment, and materials to install and integrate a 6-Station Single Wall Audiometric Booth, primary and expansion audiometers, and a bio-acoustic simulator or Wireless Automated Hearing System. The Contractor is also responsible for ambient noise certification and establishing daily calibration references.

3. Equipment Specifications and Deliverables

The Contractor shall deliver the following specified equipment:

3.1 Audiometric Booth or Wireless Automated Hearing Test System

• Model: CL-M6 PR (6 Station Single Wall Audiometric Booth)

• Door Add-On: EAO-D-004 Add-On Door featuring a Window in Door (20"W x 22"H)

• Wireless: headset provides 30-40dB of attenuation across octave bands from 250- 8000Hz. With compliance system standards meeting at minimum ANSI S3.6, OSHA 1910.95, and IEC 60645-1 for hearing testing. Fitment testing must meet at minimum ANSI S12.71 and DODI 6055.12.

3.2 Primary Audiometer Package

• Model: CCA-200 Mini Audiometer

• Included Components:

o CCA-200mini unit o TDH-39 Headset o Response Switch o 12 ft. Serial Extension Cable o Booth Mounting Plate and Screws o Calibration Certificate o USB CIC, Power Supply, and Power Cord o Software and Plus Package Key

• Note: PC is NOT INCLUDED and must be provided by the customer.

• Wireless High attenuation: Provides 32-40 dB of attenuation, allowing for effective testing at an ambient noise level of 60 dBA or 73 dBC as per ANSI S12.71-2018 that is equal to or greater than a single wall audio booth.

3.3 Expansion Audiometer Package

• Model: Benson Medical CCA-200mini Expansion Audiometer

• Included Components:

o CCA-200mini unit o TDH-39 Headset o Response Switch o 6 ft. Serial Extension Cable o Booth Mounting Plate and Screws o Calibration Certificate

3.4 Simulation Equipment

• Model: BAS-200 Bio-Acoustic Simulator

• Wireless systems: Must include headsets, tablets and software

4. Contractor Responsibilities (Installation & Certification)

The Contractor shall perform the following services (inclusive of "Sound Room Installation ESTIMATE" and "LBR – Installation"):

• Assemble and install the CL-M6 PR audiometric booth on site.

• Install all primary and expansion audiometers, as well as the BAS-200 simulator, directly into the completed booth.

• Set and verify the daily calibration reference for the integrated systems.

• Perform and provide SR-Ambient Noise Certification to ensure testing environments meet regulatory and operational standards.

5. Government / Customer Responsibilities

• Waste Management: The customer shall provide an on-site dumpster for the Contractor to dispose of all garbage, shipping crates, and packaging materials generated during installation.