VISN 23 IDIQ Site Salient Characteristics.xlsx

XLSX spreadsheet 58 KB Posted

Attached to
VISN 23 RFI: Lights, Booms and Video Integration Equipment and Support Services
Solicitation number
36C26326Q0713
Issued by

About this file

This is a site salient characteristics specification document for Veterans Integrated Service Network 23 (VISN 23) surgical equipment across five VA medical centers (Fargo, St. Cloud, Central Iowa, Black Hills, and Iowa City). The document details technical requirements and current inventory for surgical lights, booms, and video integration systems totaling 123 light units, 116 equipment booms, 32 anesthesia booms, and 55 video integration systems across all sites.

Surgical light specifications require LED technology with adjustable color temperature (3500-5000K), color rendering index greater than 90, minimum 30,000-hour life, 160,000 lux illumination, boom-mounted configuration, automatic focus and dimmer controls (5-100%), and required sterilizable handles with automatic shadow dilution adjustment. Video camera integration is required for most rooms (HD minimum at light center), and all lights must support 360-degree rotation at yoke locations with 80-125 cm vertical adjustment range. Equipment boom specifications include load capacity requirements (150-350 lbs per arm depending on location), multiple electrical receptacles (4-12 duplex outlets), medical gas connections (oxygen, vacuum, compressed air, nitrous oxide, WAG evacuation per NFPA 99 standards), network jacks (4-plex or greater), adjustable shelving, video input connections, and various accessories including monitor brackets, IV supports, and catheter baskets. Video integration system requirements mandate unified routing of disparate video signals (endoscopic, lighthead camera, C-arm, PACS, vital signs), touch-screen control, HD or better output, simultaneous multi-source display capability, 8x8 minimum input/output routing, digital storage (minimum 5TB scalable), automatic device recognition within 120 seconds, and integration control consoles with preset capability for video routing, camera positioning, and surgical light management. Recent upgrades indicate Draeger surgical light and boom installations completed or in-progress (2024), and Karl Storz video integration upgrades completed or planned through 2024-2025.

View the file

Other files

Show all 24

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Surgical Lights

ItemDescriptionMIN OR1-21NWI OR1NWI OR2NWI OR3NWI OR4NWI OR5NWI OR6NWI CATH LAB ASUX OR1SUX OR2SUX OR3SUX OR4SUX Minor 1SUX Minor 2Fargo-OR 1-4Fargo-OR 5St. Cloud-OR 1-2St. Cloud-OR 3Central Iowa-OR 2Central Iowa-OR 3-5Black Hills-OR 1-3Iowa CityTotal
1LIGHT SOURCE, TYPELEDLEDLEDLEDLEDLEDLEDLEDLEDLEDLEDLEDLEDLEDNo minimum requirementNo minimum requirementLEDLEDLEDLED
1aNumber of lightheads422222221222222836319620123
1bNumber of bulbs or elementsNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsGreater than or equal to 50 LEDs per light
1cVolts110/120110/120110/120110/120110/120110/120110/120110/120DC Voltage, no specified valueDC Voltage, no specified valueDC Voltage, no specified valueDC Voltage, no specified valueDC Voltage, no specified valueDC Voltage, no specified value110/120110/120110/120110/120110/120110/120110/120
1dLife, hrGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LED>40000>40000>40000>40000>40000>40000Greater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LEDGreater than 30,000 LED
2COLOR TEMPERATURE, KAdjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000Adjustable from 3500-5000
2aAdjustableRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-Touch at light and control displayRequired-touch as light and control display
3COLOR RENDERING INDEX>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90>90≥97
3aR9 valueminimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90minimum 90
4FIELD SIZE, cm (in)
4aDiameterNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementMinimum of 2 choices, min, size 5" Max. size 10"Minimum of 2 choices, min, size 5" Max. size 10"No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirements19 - 31 cm
5FOCAL LENGTH, cm (in)≥65≥65≥65≥65≥65≥65≥65≥65>70 cm>70 cm>70 cm>70 cm>60>60≥65≥65≥65≥65≥65≥65≥65≥65
6ILLUMINATION LEVEL, maximum lux at 1 m≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000>140000>140000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000≥160,000
7MOUNTINGBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom OnlyBoom OnlyBoom OnlyBoom OnlyBoom OnlyBoom OnlyBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom onlyBoom only
8CONTROLS
8aDimmerRequired, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%Required, At light and control display. 5 -100%
8bFocusRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayRequired at light head and control displayAt light and control displayAt light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control displayRequired, At light and control display
8biAutomaticRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequired
8cField sizeNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
8dOn sterile handleRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequired
9ROTATION, °The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location.≥420 degrees at the lamp unlimited at the suspension
10VERTICAL ADJUSTMENT RANGE, cm (in)≥80 cm≥80 cm≥80 cm≥80 cm≥80 cm≥80 cm≥80 cm≥80 cm>125 cm>125 cm>125 cm>125 cm>125 cm>125 cm> than 30"> than 30"≥80 cm≥80 cm≥80 cm≥80 cm≥80 cm≥125 cm
11HEAT
11aMaximum irradiance at 1 m, W/m2No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
11bHeat-to-light ratio, mW/m2.luxNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
12LIGHTHEAD MATERIALNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
13LENS COVER MATERIALNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsPolycarbonate? Why? Material strength is better so will prevent being broken.
14FILTER MATERIALNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
15REFLECTOR MATERIALNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
16SHADOW CONTROL
16aAutomatic shadow dilution adjustmentAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light controlAutomatically by in-light control
17SATELLITESNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
18AMBIENT LIGHTINGRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequired
19BATTERY OR EMERGENCY BACKUPRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequired
20STERILIZABLE HANDLERequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequired
21VIDEO CAMERA
21aCompatible (type) or integrated-separate video source incorporated into one of the surgical lights brought to one of the surgical fields streaming live videoYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumRequired, HD MinimumRequired, HD MinimumRequired, HD MinimumRequired, HD MinimumNot RequiredNot RequiredYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimumYes, HD minimum
21bLocation on lightheadLightCenterCenterCenterCenterCenterCenterCenterCenterCenterCenterCenterCenterCenterLightLightLightLightLightLightLightLight
22MINIMUM CEILING HEIGHT, cm (in)9' 11" from floor to intersticial deck. From floor to finished ceiling usually around 9ft. Can vary by area.9' 8"9' 8"9' 8"9' 8"9' 8"9' 8"9' 6.5"120 inch120 inch120 inch120 inch108 inch108 inch9' 11.5"9' 6"9' 4"9' 4"Curent ceiling height 12'-11.5"Curent ceiling height 12'-11.5"10 ftOR 2-6=120 inch

OR 1=122 inch OR 8 =107 inch ICU rooms =108 inch CT room 2 =108 inch CT room 1 =101 inch GI rooms=94 inch

23DIAMETER, cm (in)
23aLightheadNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirement=>30"=>30"No minimum requirementNo minimum requirementCURRENTLY 24", NOT REQURED BUT SOMETHING CLOSENo minimum requirements
24WEIGHT, kg (lb)No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
25SUPPLY VOLTAGE, VACNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
26POWER, WNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
27PURCHASE INFORMATION
27aWarrantyMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 year
28ADDITIONAL SPECIFICATIONS
28aAbility to turn light on/off at lightRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementsNo minimum requirements
28bRecent UpgradesDreager Surgical Lights upgrade completed in 2024Dreager Surgical Lights upgrade completed in 2024Draeger Surgical Lights upgrade in progressDraeger Surgical Lights Phase I upgrade completed in 2024. Phase II and III in progress.

Booms

ItemDescriptionMIN OR1-20NWI OR3NWI OR4NWI OR5NWI CATH LAB ASUX OR1SUX OR2SUX OR3SUX OR4SUX Minor 1 Light Boom OnlySUX Minor 2 Light Boom OnlyFargo-OR 1-4Fargo-OR 5St. Cloud OR 1-3Central Iowa-Cysto RoomCentral Iowa-OR 2-5Black Hiills-OR 1-3Iowa City-OR 1Iowa City-OR 2Iowa City-OR 3Iowa City-OR 4Iowa City-OR 5Iowa City-OR 6Iowa City-GI 1Iowa City-GI 2Iowa City-GI 3Iowa City-CT 1Iowa City-CT 2Iowa City-ICU 1Iowa City-ICU 2Iowa City-ICU 3Iowa City-ICU 4Iowa City-ICU 5Iowa City-ICU 6Iowa City-ICU 7Iowa City-ICU 8Iowa City-ICU 9Iowa City-ICU 10Total
1Number of Anesthesia Booms Per Room8000111111141300311111100000000000000032
2Number of Equipment Booms Per Room671110111100413143221222111111111111111116
3SWIVEL RADIUS, cm (in)No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement>180>180>180>180No Minimum RequirementNo Minimum Requirement330 degrees330 degreesNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
4HEIGHT ADJUSTMENT, cm (in)No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum Requirement24"24"Powered Vertical ArticulationNo minimum requirementNo minimum requirementNo minimum requirement24"24"24"24"24"24"No minimum requirementNo minimum requirementNo minimum requirement24"24"No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
5ELECTRICAL FEATURES
5aCircuit breakerNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirement at boomNo minimum requirement at boomNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
5bElectrical power receptaclesNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement>5 duplex>5 duplex>5 duplex>5 duplexNo Minimum RequirementNo Minimum Requirement5ea. Emer. duplex, 2ea. normal duplex5ea. Emer. duplex, 2ea. normal duplexGreater or equal =>8 and Anesthesia boom shall have 8 duplex power receptacles.No minimum requirementNo minimum requirement101010101010101010104412121212121212121212
5cEquipotential ground busNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
5dGrounding jacksNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
5eTelephone jackNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement000000000000000000000
5fIsolated power systemNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequiredNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
5gCode alarm buttonNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementRequiredNo minimum requirementNo minimum requirementNot required000000111001111111111
5hNurse call systemNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNot required000000000001111111111
5iElapsed timer/clockNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Requirement at BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNot requiredNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
6MEDICAL GAS SERVICES
6aOxygen outletsNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomNo Minimum RequirementNo Minimum Requirement2ea.2ea.2No minimum requirementNo minimum requirement2222222222113333333333
6bOutlet typeNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementDISSDISSDISSDISSNo Minimum RequirementNo Minimum RequirementOhioOhioDISSNo minimum requirementNo minimum requirementDiamondCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRONCHEMETRON
6cVacuum inletsNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomNo Minimum RequirementNo Minimum Requirement2ea. Ohio style2ea. Ohio style2No minimum requirementNo minimum requirement3111111222223333333333
6dMedical compressed airNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomNo Minimum RequirementNo Minimum Requirement2ea. Ohio style2ea. Ohio style2No minimum requirementNo minimum requirement1222222222002222222222
6eNitrous oxideNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirement1No minimum requirementNo minimum requirement1111111000000000000000
6fWAG evacuationNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomRequired on Anesthesia BoomNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirement1No minimum requirementNo minimum requirement1111111000110000000000
6gMedical gas pipingNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementTo Meet NFPA 99 StandardTo Meet NFPA 99 StandardTo Meet NFPA 99 StandardTo Meet NFPA 99 StandardNo Minimum RequirementNo Minimum RequirementConductive- meets NFPA 99 standardConductive- meets NFPA 99 standardNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standardMeets NFPA 99 standard
6hVacuum bottle slidesNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum Requirementslides or comparativeslides or comparitiveNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement222222222223333333333
6iVacuum bottle storageNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNot requiredNot requiredNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement222222222223333333333
7ACCESSORIES
7aMonitor bracketNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementRequiredRequiredRequiredRequiredNo Minimum RequirementNo Minimum Requirementbracket or arm for monitor if requestedbracket or arm for monitor if requestedRequiredNo minimum requirementNo minimum requirementREQUIRED000000111001111111111
7bBed bumperNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNot RequiredNot RequiredNot RequiredNot RequiredNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementRequiredNo minimum requirementNo minimum requirementNo minimum requirementNoNoNoNoNoNoYesYesYesNoNoYesYesYesYesYesYesYesYesYesYes
7cIV supportsNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementRequiredRequiredNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement000000111112222222222
7dCatheter basketNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement000000111112222222222
8H x W x D, cm (in)No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirementHorizontal Anesthesia BoomNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
9WEIGHT, kg (lb)No minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo Minimum RequirementNo minimum requirementNo minimum requirement>=100 lbs per shelfNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirementNo minimum requirement
10LOAD CAPACITY, kg (lb)250 lbs or greater per arm250 lbs or greater per arm250 lbs or greater per arm250 lbs or greater per arm250 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per armNo Minimum RequirementNo Minimum Requirement350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm250 lbs or greater per arm250 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm350 lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm150lbs or greater per arm
11PURCHASE INFORMATION
11aWarrantyMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 2 yearsMinimum of 2 yearsMinimum of 2 yearsMinimum of 2 yearsMinimum of 2 yearsMinimum of 2 yearsMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 yearMinimum of 1 year
12ADDITIONAL SPECIFICATIONS
12aNitrogenMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust Have Regulated Nitrogen outlet w/gaugeMust Have Regulated Nitrogen outlet w/gaugeMust Have Regulated Nitrogen outlet w/gaugeMust Have Regulated Nitrogen outlet w/gaugeNo Minimum RequirementNo Minimum RequirementMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gaugeMust have regulated Nitrogen outlet/w gauge
12bNetwork Jacks4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 Plex or More on each boom4 Plex or More on each boom4 Plex or More on each boom4 Plex or More on each boomNo Minimum RequirementNo Minimum Requirement4 plex or more in each room4 plex or moreAnesthesia boom shall have 5 network jacks4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more4 plex or more
12cKeyboard/mouse11111111111111111111111111111111
12dAdustable Equipment ShelvesRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredNo Minimum RequirementNo Minimum RequirementRequiredRequiredRequired. Equipment boom shall have 4 adjustable shelves with => 100 lb load capacity per shelfRequiredRequiredRequired444444444000000000000
12eCarbon Dioxide on equipment boomRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequiredRequired1 DISS1 DISS1 DISS1 DISS1 DISS2 DISS2 DISS2 DISS2 DISS000000000000
12fMedical Gas ConnectionsEach room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each Room shall have the following Medical Gas connections available. Quantities TBD in each room. -Oxygen, Medical Air, Vacuum, Nitrogen, Nitrous Oxide, WGS, Each Room shall have the following Medical Gas connections available. Quantities TBD in each room. -Oxygen, Medical Air, Vacuum, Nitrogen, Nitrous Oxide, WGS, Each Room shall have the following Medical Gas connections available. Quantities TBD in each room. -Oxygen, Medical Air, Vacuum, Nitrogen, Nitrous Oxide, WGS, Each Room shall have the following Medical Gas connections available. Quantities TBD in each room. -Oxygen, Medical Air, Vacuum, Nitrogen, Nitrous Oxide, WGS, No Minimum Requirement No Minimum Requirement Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.

-Oxygen -Medical Air -Vacuum -Nitrogen

-CO2

-Waste Gas Scavenging

12gUniversal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs
•Anesthesia boom shall have 3 universal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs
•Anesthesia boom shall have 3 universal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs
•Anesthesia boom shall have 3 universal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs
•Anesthesia boom shall have 3 universal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs
•Anesthesia boom shall have 3 universal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs
•Anesthesia boom shall have 3 universal connections for video inputs•Equipment boom shall have 6 universal connections for video inputs