Fargo Structural Designs.pdf
PDF 9 MB Posted
- Solicitation number
- 36C26326Q0713
About this file
This document comprises construction and architectural design drawings for the VA Fargo Medical Center renovation project, specifically detailing ceiling supply unit installations and building infrastructure modifications.
The project involves installation of medical gas and electrical supply units in operating rooms (OR 4 and OR 5) at the VA Fargo Medical Center in Fargo, North Dakota. The ceiling supply units feature Dräger Ambia and Polaris 600 systems with dual-arm configurations, incorporating electrical outlets (Ambia-Hubbell Hospital Grade receptacles at 125V, 20A), DISS gas terminal units for medical gases (nitrogen, air, oxygen, CO2, VAC), and control panels with touchscreen brake/lift functionality. Specifications include maximum loads of 130-150kg, lifting heights of 600mm, lifting time of approximately 10 seconds, and concrete installation requirements of minimum 200mm thickness with 20/25 grade concrete. The drawings encompass foundation plans, first and second floor framing plans, roof framing plans, and detailed sectional drawings showing electrical, gas, and structural components. Regulatory information notes that installations must maintain 200mm safety distance from oxidizing gas outlets unless electrical installations do not exceed 10VA in fault conditions, and units cannot be installed in MRI rooms with high Tesla levels or areas with combustible/explosive gases. The design documentation was prepared by Rühmann and approved by Brinker/Janack on August 9, 2022, with revision index 00 for drawings numbered 22-1709, 22-1710, 22-1711, and 22-1712.
View the file
Other files
Show all 24
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Internal
22-1710-00
22-1709-00
Fargo
OR 4
Listed numbers are drawing file names
22-1711-00
22-1712-00
Fargo
OR 5
22-1710-00
22-1709-00
Fargo
OR 4
B
A
1 22 3 4 5 6 7 8 119 10 12 13 1514
B
A
M
L
1 2 3
K
J
H
G
F
E
D
C
4 5 6 7 8 119 10 12 13
G
F
E
D
C size A1+210
H
L
K
M
J
16 17
18 19
0~ •• ~ > ~" :a ::: ~ ••o O•~ ---.!!! :l: C 2 _g,~ 0 C • -::: -<C -:;
0 . 0
<C 0 .... ·;:: 0 .... ":.::
0 ~" . - . e'-, • 0
:::: ~ 5.
~ ~ .e • ~.
~ " . -·--- • o,~
• 0
•£ ::::, "i
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Measurements to be checked by Drager on site.
G)
I I
I I
,,--11 , I
I I I I
L - - - - ----c,1~+*~
[;:ible and hose length outside of ceiling bearing 1000mm.
L _________________ J
4, OP75576 Shelf 1,258/480 with US rails(GG16040 !1220)) Powder-coated aluminum shelf for the column of a supply unit with an usable surface of 530x480 mm and side plastic caps with integrilted Fairfield rails, which have a maximum loi!d of 10kg in each case.
G95048 remote control double arm lift
G39960 Handle with sensoric
G39800 Handle with sensoric lift detail "assembly" M. 110
0/ 1SG "" SE
~· G!30C
G.'41l0 G 0!620 I Cl.'700 G
L
-55,'WJ
-~5.i41 0
-55/560
-55,780 minimum concrete thickness 200mm concrete grade (20/25 concrete ceiling brakes:
F = Friction brake E = Electramagnetic brake false ceiling l
I
I I
I I
I
Side view left
Top view
B ready floor separate preinstallation material:
1 ~ G25210 ceiling fixture set l
I
I I
I I
I w
Back view detail "A"
R detail "A"
F ill lm ill n l! LJ
5~1410
55,'560
(l,•40()
-20.'461
D,600
•. ·1 L. . 55!71\D 0.'790
01,=.50 22,5/41',7
I
55/150
55,'3()0
~5,'661
_l
I
I I
I I
I
Variable adjustment in steps
F
• -c ~ -0
E E i<l
~ r- lifting 1179mm
G
Polaris 600
Side view right Front view incl. control panel + prepared sterilizable handle set E j lifting 600mm lifting time approx. 10 sec.
max. load 130KG net carr in ca acit 88KG
Electrical outlets:
Ambia - Hubbell E socKet Hospital Grade (NFPA- 70)
Position Text circuit F -55/260 Duplex Receptacle, 125V, 20A, red, backbox 1 F -55/410 Duplex Receptacle, 125V, 20A, red, backbox 1 F -55/560 Duplex Receptacle, 125V, 20A, red, bacKbox 3 F -55/780 Duplex Receptacle, 125V, 20A, red, backbox 3 B 55/150 Duplex Receptacle, 125V, 20A, white, backbox 4 L 0/300 Duplex Receptacle, 125V, 20A, white, backbox 4
Gas outlets, DISS gas terminal unit Beacon Medaes - US
Position Text B 55/300 N1
B 55/661 AIR
L 0/460 AIR
L 0/620 01
L 0/780 VAC
R 0/600 CO2
R 0/790 VAC
-------Max Weight
~----Tensile Force
Torque at Ceiling
586kg 5744N 8358Nm
11"1
B
0 ~ ~ ' -""3:::.
displ.-adapter illustrative by others displ.-adapter incl. prepared f incl. prepared '° ~ sterilizable handle set~ I - ,,, sterilizable handle set E c,'a' lifting reduced
1067mm as shown
(max. lifting of arm
1076mm) lifting 1076mm
Other outlets:
AK Cutouts
Position Text B 0/450 1362 - Cut out N2-Control Panel US
Ambia - Hubbell [NFPA- 70)
Position Text comment B -55/150 Blind plate 1-gang, backbox without conduit Foot switrh
Maximum short-circuit power over 10 VA: yes F55/410 Blind plate 1-gang, backbox without conduit Maximum short-frcuit power over 10 VA yes F 55/560 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA yes F 551780 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA yes F 55/160 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available, yes L 0/150 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes
Ambia controls
Position Text R 0/400 T ouchpanel brake/lift 2arm R -20/461 Service socket R 22,5/467 Switch lift stop
Ambia rails
Position Text B 0/950 Fairfield rail 280 L 0/950 Fairfield rail 180 R 0/950 Fairfield rail 180
<I
X X X X concrete ceiling false ceilin main supply
I00-240V
I I I I L __ emergency supply
24V
S7 ready floor
I I
I I I I I
I
I :
---~-J
J
Pas.
A
B
D
E
F G
H I J
K
L M
Head
- 3
Column
No/pc Order-no. Description
1 G15796
1 G94401 clamp ring oti110 suspension tube ¢110x460mm r r separate pre installation separate pre installation
1 G97960 axis 3-fold MOX/L1/MOX
1 G28374 spring arm display 17-26 MOX SD/HO 1nclcablng
1 G35467 1 displ.-adap. 24-32" Reus.
I G36699 spring load arm Polaris 100/200/6001nc1.c.ibtlng
I OP75084 Polaris 600/prepared sterilizable handle set E
1 G36993 sterilizable handle set E (2 □ cs)
I G23361 cable terminal box 1,5-6mm2 (2pcs)
1 G36402 PSU 100-240V Polaris 600 r pre installation
1 G33861 front cover wall control panel P6
1 G36707 wall control panel P600
1 G36396 WCU flush-mtd. housing FM 200 r pre installation
Regulatory information
The safety distance from empty installations shall be 200mm if it is intended to use them with electrical installations and medical gas outlets for oxidizing gases, The safety distance of 200mm to oxidizing gases can be undercut if the electrical installations do not reach a power >10VA in the first fault case (e.g. short circuit).
Driiger Medical Supply Units shall not be installed in rooms, where combustible or explosive gases can occur.
Drager Medical Supply Units shall not be installed in MRI rooms with high Tesla levels.
If installations or devices are installed. which have not been approved by Driiger, these installation or devices needs an on-site approval.
date, author: revision h:ixt:
projectname + department, VA Fargo Medical Center
Drager-country code: voltage (V): frequency (Hz): target country:
us 120/240 60 United States
2.1 + 2.2 room-no.+ roomnamE!:
OR 4
quantity of units: 1 pcs.
CD equipment color:
® service plate:
Q) Color frame:
Medical White W1100/S-0500-N anodized
Medical White W1100/S-0500-N scale:
view: 1:20, index:
rest: see details author:
date:
name:
title:
2022-08-04
RUHMANN
Ceiling supply unit approved:
date:
name:
2022-08-09
BRINKER/ JANACK
t Y p e left: A mb j a {artlcl@-number: (i(i16010) type right: Polaris 600 larticle-number:Ci16720I revision index number:
22-1709
□rager
AutoCAD SHX Text 270°
AutoCAD SHX Text Cable and hose length outside of ceiling bearing 1000mm.
AutoCAD SHX Text detail "A"
AutoCAD SHX Text
88KG
AutoCAD SHX Text
130KG
AutoCAD SHX Text max. load net carrying capacity
AutoCAD SHX Text ready floor
AutoCAD SHX Text 4x OP75576 Shelf 1,25B/480 with US rails(GG16040 (1220)) Powder-coated aluminum shelf for the column of a supply unit with an usable surface of 530x480 mm and side plastic caps with integrated Fairfield rails, which have a maximum load of 10kg in each case. G95048 remote control double arm lift G39960 Handle with sensoric G39800 Handle with sensoric lift
AutoCAD SHX Text min.260mm-max.1000mm
AutoCAD SHX Text Front view
AutoCAD SHX Text Back view
AutoCAD SHX Text Side view right
AutoCAD SHX Text Side view left
AutoCAD SHX Text F
AutoCAD SHX Text lifting 600mm lifting time approx. 10 sec.
AutoCAD SHX Text Variable adjustment in steps
AutoCAD SHX Text E
AutoCAD SHX Text
AutoCAD SHX Text 330°
AutoCAD SHX Text E
AutoCAD SHX Text 330°
AutoCAD SHX Text false ceiling
AutoCAD SHX Text 586kg
AutoCAD SHX Text 5744N
AutoCAD SHX Text 8358Nm
AutoCAD SHX Text Max Weight
AutoCAD SHX Text Tensile Force
AutoCAD SHX Text Torque at Ceiling
AutoCAD SHX Text minimum concrete thickness 200mm concrete grade C20/25
AutoCAD SHX Text concrete ceiling
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text brakes: F = Friction brake E = Electromagnetic brake
AutoCAD SHX Text A
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G15796
AutoCAD SHX Text clamp ring 110
AutoCAD SHX Text D
AutoCAD SHX Text E
AutoCAD SHX Text Pos.
AutoCAD SHX Text No/pc
AutoCAD SHX Text Order-no.
AutoCAD SHX Text Description
AutoCAD SHX Text A
AutoCAD SHX Text
AutoCAD SHX Text B
AutoCAD SHX Text
AutoCAD SHX Text C
AutoCAD SHX Text
AutoCAD SHX Text G
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text H
AutoCAD SHX Text
AutoCAD SHX Text G28374
AutoCAD SHX Text G35467
AutoCAD SHX Text G36993
AutoCAD SHX Text spring arm display 17-26 MOX SD/HD
AutoCAD SHX Text incl. cabling
AutoCAD SHX Text 1 displ.-adap. 24-32" Reus.
AutoCAD SHX Text sterilizable handle set E (2pcs)
AutoCAD SHX Text separate pre installation
AutoCAD SHX Text L
AutoCAD SHX Text
AutoCAD SHX Text M
AutoCAD SHX Text I
AutoCAD SHX Text
AutoCAD SHX Text J
AutoCAD SHX Text
AutoCAD SHX Text K
AutoCAD SHX Text
AutoCAD SHX Text I
AutoCAD SHX Text separate pre installation
AutoCAD SHX Text G36402
AutoCAD SHX Text PSU 100-240V Polaris 600
AutoCAD SHX Text pre installation
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G23361
AutoCAD SHX Text cable terminal box 1,5-6mm² (2pcs)
AutoCAD SHX Text separate preinstallation material: 1 * G25210 ceiling fixture set
AutoCAD SHX Text
OP75084
AutoCAD SHX Text
AutoCAD SHX Text Polaris 600/prepared sterilizable handle set E
AutoCAD SHX Text G94401
AutoCAD SHX Text suspension tube 110x460mm
AutoCAD SHX Text B
AutoCAD SHX Text
AutoCAD SHX Text G36707
AutoCAD SHX Text wall control panel P600
AutoCAD SHX Text G33861
AutoCAD SHX Text front cover wall control panel P6
AutoCAD SHX Text G36396
AutoCAD SHX Text WCU flush-mtd. housing FM 200
AutoCAD SHX Text G97960
AutoCAD SHX Text axis 3-fold MOX/L1/MOX
AutoCAD SHX Text 360°
AutoCAD SHX Text 330°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 360°
AutoCAD SHX Text 330°
AutoCAD SHX Text 330°
AutoCAD SHX Text 330°
AutoCAD SHX Text 330°
AutoCAD SHX Text
AutoCAD SHX Text lifting reduced 1067mm as shown (max. lifting of arm 1076mm)
AutoCAD SHX Text 180°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text displ.-adapter incl. prepared sterilizable handle set E
AutoCAD SHX Text emergency supply 24V
AutoCAD SHX Text main supply 100-240V
AutoCAD SHX Text N
AutoCAD SHX Text L
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Output: Light System
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text System number B
AutoCAD SHX Text System number A
AutoCAD SHX Text Initialize system
AutoCAD SHX Text Contact service
AutoCAD SHX Text Initialize in progress
AutoCAD SHX Text Output voltage
AutoCAD SHX Text Emergency power
AutoCAD SHX Text Supply mains
AutoCAD SHX Text All Poles
AutoCAD SHX Text Output
AutoCAD SHX Text Emergency Power
AutoCAD SHX Text Supply mains
AutoCAD SHX Text
AutoCAD SHX Text Master Switch
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Supply mains
AutoCAD SHX Text 24V
AutoCAD SHX Text Power
AutoCAD SHX Text Emergency
AutoCAD SHX Text ready floor
AutoCAD SHX Text false ceiling
AutoCAD SHX Text concrete ceiling
AutoCAD SHX Text C
AutoCAD SHX Text 360°
AutoCAD SHX Text 340°
AutoCAD SHX Text 200°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text F
AutoCAD SHX Text G
AutoCAD SHX Text lifting 1179mm
AutoCAD SHX Text 180°
AutoCAD SHX Text E
AutoCAD SHX Text H
AutoCAD SHX Text 330°
AutoCAD SHX Text D
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text lifting 1076mm
AutoCAD SHX Text
AutoCAD SHX Text D
AutoCAD SHX Text E
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G36699
AutoCAD SHX Text spring load arm Polaris 100/200/600
AutoCAD SHX Text incl. cabling
AutoCAD SHX Text F
AutoCAD SHX Text
AutoCAD SHX Text J
AutoCAD SHX Text K
AutoCAD SHX Text L
AutoCAD SHX Text M
AutoCAD SHX Text pre installation
AutoCAD SHX Text illustrative by others
AutoCAD SHX Text displ.-adapter incl. prepared sterilizable handle set E
AutoCAD SHX Text Polaris 600 incl. control panel + prepared sterilizable handle set E
AutoCAD SHX Text Electrical outlets: Ambia - Hubbell E socket Hospital Grade (NFPA-70) Position Text circuit Text circuit circuit F -55/260 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 F -55/410 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 F -55/560 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 F -55/780 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 B 55/150 Duplex Receptacle, 125V, 20A, white, backbox 4 Duplex Receptacle, 125V, 20A, white, backbox 4 4 L 0/300 Duplex Receptacle, 125V, 20A, white, backbox 4Duplex Receptacle, 125V, 20A, white, backbox 44
AutoCAD SHX Text L
AutoCAD SHX Text R
AutoCAD SHX Text B
AutoCAD SHX Text F
AutoCAD SHX Text Top view
AutoCAD SHX Text detail "A"
AutoCAD SHX Text Gas outlets: DISS gas terminal unit Beacon Medaes - US Position Text Text B 55/300 N2 N2 B 55/661 AIR AIR L 0/460 AIR AIR L 0/620 O2 O2 L 0/780 VAC VAC R 0/600 CO2 CO2 R 0/790 VACVAC
AutoCAD SHX Text Other outlets: AK Cutouts Position Text Text B 0/450 1362 - Cut out „N2-Control Panel US“ 1362 - Cut out „N2-Control Panel US“ N2-Control Panel US“ Ambia - Hubbell (NFPA-70) Position Text comment Text comment comment comment B -55/150 Blind plate 1-gang, backbox without conduit Foot switch Blind plate 1-gang, backbox without conduit Foot switch Foot switch Maximum short-circuit power over 10 VA: yes F 55/410 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Maximum short-circuit power over 10 VA: yes F 55/560 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Maximum short-circuit power over 10 VA: yes F 55/780 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Maximum short-circuit power over 10 VA: yes F 55/260 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes L 0/150 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes Ambia controls Position Text Text R 0/400 Touchpanel brake/lift 2arm Touchpanel brake/lift 2arm R -20/461 Service socket Service socket R 22,5/467 Switch lift stop Switch lift stop Ambia rails Position Text Text B 0/950 Fairfield rail 280 Fairfield rail 280 L 0/950 Fairfield rail 180 Fairfield rail 180 R 0/950 Fairfield rail 180 Fairfield rail 180
AutoCAD SHX Text Regulatory information The safety distance from empty installations shall be 200mm if it is intended to use them with electrical installations and medical gas outlets for oxidizing gases. The safety distance of 200mm to oxidizing gases can be undercut if the electrical installations do not reach a power >10VA in the first fault case (e.g. short circuit). Dräger Medical Supply Units shall not be installed in rooms, where combustible or explosive gases can occur. Dräger Medical Supply Units shall not be installed in MRI rooms with high Tesla levels. If installations or devices are installed, which have not been approved by Dräger, these installation or devices needs an on-site approval.
AutoCAD SHX Text Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
AutoCAD SHX Text Measurements to be checked by Dräger on site.
AutoCAD SHX Text Ceiling supply unit type left: type right:
AutoCAD SHX Text Polaris 600 (article-number: G16720)(article-number: G16720)
AutoCAD SHX Text Ambia (article-number: GG16010)(article-number: GG16010)
AutoCAD SHX Text
2.1 + 2.2
AutoCAD SHX Text
OR 4
AutoCAD SHX Text equipment color: service plate: Color frame:
AutoCAD SHX Text Medical White W1100/S-0500-N
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Head
AutoCAD SHX Text Column
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text quantity of units: 1 pcs.
AutoCAD SHX Text projectname + department:
AutoCAD SHX Text room-no. + roomname:
AutoCAD SHX Text date: author: revision text: index:
AutoCAD SHX Text LV-Pos.:
AutoCAD SHX Text Dräger-country code:
AutoCAD SHX Text target country:
AutoCAD SHX Text voltage (V):
AutoCAD SHX Text frequency (Hz):
AutoCAD SHX Text revision index 00
AutoCAD SHX Text number:
AutoCAD SHX Text title:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text author: approved:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text scale:
AutoCAD SHX Text view: 1:20, rest: see details
AutoCAD SHX Text detail "assembly" M. 1:10
AutoCAD SHX Text VA Fargo Medical Center
AutoCAD SHX Text Medical White W1100/S-0500-N
AutoCAD SHX Text anodized
AutoCAD SHX Text
US
AutoCAD SHX Text United States
AutoCAD SHX Text 120/240
AutoCAD SHX Text
AutoCAD SHX Text 2022-08-04
AutoCAD SHX Text
RÜHMANN
AutoCAD SHX Text 2022-08-09
AutoCAD SHX Text
BRINKER/JANACK
AutoCAD SHX Text 22-1709
AutoCAD SHX Text Use of this document / contents is forbidden without express authority. All rights reserved. Refer to protection notice ISO 16016.
A
1 22 3 4 5 6 7 8 119 10 12 13 1514 16
B
A
M
L
1 2 3
K
J
H
G
F
E
D
C
4 5 6 7 8 119 10 12 13
G
F
E
D
C size A1
J
H
L
K
M
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemenf-ed
Measurements to be checked by Drager on site.
detail "assembly" M. 1:10
Fl ( ·- ••"3
L
Fl C
015'68 ,5/0'1 60
C separate preinstallation material:
1 x G25210 ceiling fixture set r-----------------7 1',,, ::/✓--1c--,,~\ I \
I \ I ~~ \ L___ _ __ _J
Cable and hose length outside of ceiling bearing 1000mm.
I I I I I I I I I I I I I
L-----------------~ without preparation } Installation set Retrofitting for Ambia Rack IRI . .
d b' p 1 !Pl of accessories Is not an Am 1a a e possible!
(D G39800 Handle with sensoric lift
Top view
B
~ R detail "A"
U {Ansicht von untenJ
F nwJE :l 111 G;e I (;) g 1.; • ·
( 1-••·· • ':J___.)
0.'70 0.'170 0.•'270 0/560 01660 0/760
' 2 5 ' O/ffi8,5JN4:l179
00 l!J I
87,5! 0/640 01740 3 ,, 0/70 0/?40 Oi600 0,'460 01320 01,so 3 '
I I I I G 00 I minimum concrete thickness 200mm concrete grade [20/25 concrete ceiling brakes:
E = Electromagnetic brake F = Friction brake false ceiling
Side view left ready floor detail "A"
~: t :11
Back view
Variable
Es: adjustment in steps
0 ;: E c, ~ ~o 'oo __,. • ~--:
.--- C: I( F ~ 'E ~
12' -11"]
Side view right Front view lifting 600mm lifting time approx. 10 sec.
1max. load 1130KG I net carryi □ Q capacity 130KG
,----------------------------------------7
Electrical outlets:
Ambia - Hubbell E socket Hospital Grade INFPA-701
Position Text circuit F 0/560 F 0/660 B 0/70 U 0/640
LI 0/740
F 0/760
Duplex Receptacle, 125V, 10A, red, backbox Duplex Receptacle, 115V, 10A, red, backbox Duplex Receptacle, 125V, 10A, red, backbox Duplex Receptacle, 125V, 10A, red, backbox Duplex Receptacle, 125V, 10A, red, backbox 4 Duplex Receptacle, 125V, 10A, white, backbox 5
Gas outlets:
DISS gas terminal unit Beacon Medaes - US
Position B 0/180 B 0/320 B 0/460 B 0/600 B 0/740 L 0/160 L 0/50
Text
AIR
N20
WAG□
VAC
VAC
Other outlets:
Ambia - Hubbell lNFPA-70)
Position F 0/170 F 0/270 F 0/70
Text 4x RJ45, white, backbox without conduit 4x RJ45, white, backbox without conduit Blind plate 1-gang, backbox without conduit
Ambia controls
Position R 0/140 R 0/179 R 0/65
Ambia rails
Text Switch lift stop Service socket Touchpanel brake/lift 2arm
Position Text L -68,5/111 Fairfield rail 180 R -68,5/111 Fairfield rail 180 comment
Power over Ethernet {PoEI available, yes Power over Ethernet (PoE) available: yes Maximum short-circuit power over 10 VA: yes
L ________________________________________ J
~ ' -N ' m -
~------Max Weight
~---Tensile Force
Torque at Ceiling
378kg 3708N 5038Nm
[2'[ x 600 flat cover
[2'-6"]
~ ::r l 5:_ T ?
t 1 f- - --, I I ~---,J I I r--1
_J __ I_
Head
I'---._ .,,(i") T/
I _, ! Q).._
Column
~ ,.(1) l_ - -®
CD
illustrative by others
No N.
m ac
X X X X
Regulatory information
The safety distance from empty installations shall be 200mm if it is intended to use them with electrical installations and medical gas outlets for oxidizing gases.
The safety distance of 200mm to oxidizing gases can be undercut if the electrical installations do not reach a power >10VA in the first fault case {e.g. short circuit).
Drager Medical Supply Units shall not be installed in rooms, where combustible or explosive gases can occur.
Drager Medical Supply Units shall not be installed in MRI rooms with high Tesla levels.
date, author: revision text:
projectname + department:
VA Fargo Medical Center
Drager-country code: voltage {VI, frequency !Hz), target country:
us 120/240 60 United States
LV-Pos., 2.3 room-no. + roomname:
OR 4
quantity of units, 1 pcs.
G) equipment color Medical White W1100/S-0500-N scale:
CV service plate, anodized
G) Color frame, Medical White W1100/S-0500-N view: 1:20, index:
rest: see details author: approved:
date: 1011-08-04 date: 1011-08-09 name: RUHMANN name: BRINKER/ JANACK title, Ceiling supply unit iJrager ~Am b j a (article-number, GG16010I revision index lnumber: 22-1710 00
I
AutoCAD SHX Text detail "A"
AutoCAD SHX Text
130KG
AutoCAD SHX Text
130KG
AutoCAD SHX Text max. load net carrying capacity
AutoCAD SHX Text Side view right
AutoCAD SHX Text Side view left
AutoCAD SHX Text Back view
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text F
AutoCAD SHX Text lifting 600mm lifting time approx. 10 sec.
AutoCAD SHX Text E
AutoCAD SHX Text
AutoCAD SHX Text E
AutoCAD SHX Text false ceiling
AutoCAD SHX Text 330°
AutoCAD SHX Text Cable and hose length outside of ceiling bearing 1000mm.
AutoCAD SHX Text 378kg
AutoCAD SHX Text 3708N
AutoCAD SHX Text 5038Nm
AutoCAD SHX Text Max Weight
AutoCAD SHX Text Tensile Force
AutoCAD SHX Text Torque at Ceiling
AutoCAD SHX Text ready floor
AutoCAD SHX Text Variable adjustment in steps
AutoCAD SHX Text 330°
AutoCAD SHX Text 330°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text minimum concrete thickness 200mm concrete grade C20/25
AutoCAD SHX Text concrete ceiling
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text brakes: E = Electromagnetic brake F = Friction brake
AutoCAD SHX Text illustrative by others
AutoCAD SHX Text Retrofitting of accessories is not possible!
AutoCAD SHX Text without preparation Installation set for Ambia Rack (R) and Ambia Pole (P)
AutoCAD SHX Text G39800 Handle with sensoric lift
AutoCAD SHX Text
AutoCAD SHX Text Front view
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text separate preinstallation material: 1 * G25210 ceiling fixture set
AutoCAD SHX Text min.208mm- max.1000mm
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Electrical outlets: Ambia - Hubbell E socket Hospital Grade (NFPA-70) Position Text circuit Text circuit circuit F 0/560 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 F 0/660 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 B 0/70 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 U 0/640 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 U 0/740 Duplex Receptacle, 125V, 20A, red, backbox 4 Duplex Receptacle, 125V, 20A, red, backbox 4 4 F 0/760 Duplex Receptacle, 125V, 20A, white, backbox 5Duplex Receptacle, 125V, 20A, white, backbox 55
AutoCAD SHX Text L
AutoCAD SHX Text R
AutoCAD SHX Text B
AutoCAD SHX Text F
AutoCAD SHX Text U (Ansicht von unten)
AutoCAD SHX Text detail "A"
AutoCAD SHX Text Top view
AutoCAD SHX Text Gas outlets: DISS gas terminal unit Beacon Medaes - US Position Text Text B 0/180 AIR AIR B 0/320 N2O N2O B 0/460 WAGD WAGD B 0/600 O2 O2 B 0/740 O2 O2 L 0/160 VAC VAC L 0/50 VACVAC
AutoCAD SHX Text Other outlets: Ambia - Hubbell (NFPA-70) Position Text comment Text comment comment F 0/170 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes F 0/270 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes F 0/70 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Maximum short-circuit power over 10 VA: yes Ambia controls Position Text Text R 0/140 Switch lift stop Switch lift stop R 0/179 Service socket Service socket R 0/65 Touchpanel brake/lift 2arm Touchpanel brake/lift 2arm Ambia rails Position Text Text L -68,5/111 Fairfield rail 180 Fairfield rail 180 R -68,5/111 Fairfield rail 180 Fairfield rail 180
AutoCAD SHX Text Regulatory information The safety distance from empty installations shall be 200mm if it is intended to use them with electrical installations and medical gas outlets for oxidizing gases. The safety distance of 200mm to oxidizing gases can be undercut if the electrical installations do not reach a power >10VA in the first fault case (e.g. short circuit). Dräger Medical Supply Units shall not be installed in rooms, where combustible or explosive gases can occur. Dräger Medical Supply Units shall not be installed in MRI rooms with high Tesla levels.
AutoCAD SHX Text Use of this document / contents is forbidden without express authority. All rights reserved. Refer to protection notice ISO 16016.
AutoCAD SHX Text Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
AutoCAD SHX Text Measurements to be checked by Dräger on site.
AutoCAD SHX Text Ceiling supply unit type:
AutoCAD SHX Text Ambia (article-number: GG16010)(article-number: GG16010)
AutoCAD SHX Text 2.3
AutoCAD SHX Text VA Fargo Medical Center
AutoCAD SHX Text
OR 4
AutoCAD SHX Text Medical White W1100/S-0500-N
AutoCAD SHX Text equipment color: service plate: Color frame:
AutoCAD SHX Text Medical White W1100/S-0500-N
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Head
AutoCAD SHX Text Column
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text anodized
AutoCAD SHX Text quantity of units: 1 pcs.
AutoCAD SHX Text projectname + department:
AutoCAD SHX Text room-no. + roomname:
AutoCAD SHX Text date: author: revision text: index:
AutoCAD SHX Text LV-Pos.:
AutoCAD SHX Text Dräger-country code:
AutoCAD SHX Text target country:
AutoCAD SHX Text voltage (V):
AutoCAD SHX Text frequency (Hz):
AutoCAD SHX Text revision index 00
AutoCAD SHX Text number:
AutoCAD SHX Text title:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text author: approved:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text scale:
AutoCAD SHX Text view: 1:20, rest: see details
AutoCAD SHX Text
RÜHMANN
AutoCAD SHX Text 2022-08-04
AutoCAD SHX Text 22-1710
AutoCAD SHX Text
BRINKER/JANACK
AutoCAD SHX Text 2022-08-09
AutoCAD SHX Text
US
AutoCAD SHX Text United States
AutoCAD SHX Text 120/240
AutoCAD SHX Text detail "assembly" M. 1:10
22-1711-00
22-1712-00
Fargo
OR 5
A
1 22 3 4 5 6 7 8 119 10 12 13 1514
B
A
M
L
1 2 3
K
J
H
G
F
E
D
C
4 5 6 7 8 119 10 12 13
G
F
E
D
C size A1 zu A0
H
L
K
M
J
16 17
18 19 20 21 22
20 21 22 23
I I
I
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper prov1s1ons at the constructions site have been made or will be implemented
Measurements to be checked by Drager on site.
detail "assembly" M. 1:10
0/:30D (;;· I
I
L
-5Gi300
-55.'450
-5",'600
-501300
Top view
B
F
: liiJ =l ']
R
55''300
55.'450 -IJ-'439
55/600 {) ,'!,82
{),750 I_-~ -_I 55''800
I
I I
I I separate preinstallation material:
1 ~ G25210 ceiling fixture set
,--------------------7
I ,,,--r1~~,, \ I - \
I \ I \ L ____ ci"!"51ir!~----J
._., T ....J~.__ T-
[able and hose length outside of ceiling be.iring 1000mm.
L--------------------~
CD l, OP75576 Shelf 1,25B/480 with US rails[GG16040 (122011
Powder-coated aluminum shelf for the column of a supply unit with an usalJle surface of 530x480 mm and side plastic caps with integrated Fairfield rails, which have a maximum load o' 10kg in each case.
G) OP755?6 Shelf 1,258/480 with US rails[GG16040 (122211 Powder-coated aluminum shelf for t-he column af a supply unit- with an usable surface of 530x480 mm and side plast-ic caps wif-h integraf-ed Fairfield rails, which have a maximum load of 10kg in each case.
□ P75587 Handle for brakes(GG16040 11222)) Wireless handle t-o unlock the brakes of the ceiling supply unit via Bluetooth.
©" G39960 Handle wit-h sensoric
G95048 remote control doulJle arm lift
I minimum concrete thickness 200mm concrete grade (20/25 concrete ceiling
F = Friction brake E = Eledrornagnetic brake false ceiling
Side view left detail 'A"
-55.'200
■ 55,'400
0.'550
E~ 551700
I I
I
Elect-rical outlets·
Ambi<1 - Hubbell E socket Hospital Grade INFPA- 70)
Position F -55/300 F -55/450 F -55/600
F -55/800 B 55/200 L 0/300
Gas outlets
Text circuit- Duplex Recepf-acle, 125V, 20A, red, b.ickbox 2 Duplex Recepf-acle, 125V, 20A, red, b<1ckbox 2 Duplex Recepf-acle, 125V, 20A, red, backbox 3
Duplex Recept-acle, 125V, 20A, red, backbox 3 Duplex Recepf-acle, 125V, 20A, white, backbox 4 Duplex Receptacle, 125V, 20A, white, backbox 4
DISS gas terminal unit Beacon Medaes - US
Position B 55/1,.00 B 55/700 L 0/500 L 0/650 L 0/800 R 0/439 R 0/750
Text N1
AIR
AIR
VAC
CO2
VA(
I I
I I detail "A"
I
Variatile adjustment in sf-eps
II
I
II
F
I i I 3
Back view Side view right Front- view
Of-her ouf-let-s
Ak Cutouts
Position Text-
B 0/550 1362 - Cut out N2-Control Panel US
AmlJia - HulJIJell INFPA-70)
Position Text B -55/200 Blind plate 1-gang, backbox without conduit
F 55/450
F 55/600
F 55/800
F 55/300 L 0/150
Blind plate 1-gang, backbox without- conduit
Blind plate 1-gang, backbox without- conduit
Blind plate 1-gang, backbox without- conduit-
4x RJ45, white, backbox wif-hout rnnduit 4x RJ45, white, backbax without conduit
Ambia controls
Position R 0/582 R 20/660
Ambia rails
Position B 0/950 L 0/950 R 0/950
I
Text Touchpanel brake 2arm Service socket
Text Fairfield rail 280 Fairfield rail 180 Fairfield rail 180
I max. load 150KG I net carrvinn canacitvl 108KG I comment Foot switch Maximum short-circuit power over 10 VA
Foot swit-ch yes
Maximum short--circuit power over 10 VA yes Fool S\./if-ch Maximum shorf--circuil power over 10 VA Foot swit-ch yes
Maximum short--circuit power over 10 VA yes Power over Ethernet IPoE] available: yes Power over Ethernet IPoE] available, yes
I I
I
I
II
1250 1250 I 4 · rl [4' 1"]
L
I
I I
~------Max Weight
~---Tensile Force t Torque at Ceiling
~L,~~~~ 735kg 7212N 124-74Nml
[2']x600 flat cover
1 1 E
E
[ 11" I
<;' T jl I_ T ?
I I f--- - ---, I I I- - -I I I
[2'-6"]
I
I illustrative by others
_Jl
[2'-11"]
Front- view lifting 600mm lifting time approx 10 sec.
I I I I detail "B"
I $ I
Side view left- B.ick view Side view right r- i ready
1 _J __ l_ 7 f~r
Top view
B
U Ansicht von untenJ
F
I: I :
I ,'-------------~°'""'"-~"""·"'~'"" -~
0/70
~ ~•0'.'40"-~"c·'~cOS~_cU-'40 C00_~00Coc,O~-"c-'c"c" -~" ~ • l 11 1 ■ a fill
I I
R max. load 130KG lnel carrvinn cariacitvl 130KG I detail "B"
I
,--------------------------------------7
Electrical outlef-s:
Ambia - Hubbell E socket Hospital Grade (NFPA-70)
Posit-ion F 0/560
F 0/660 B 0/70 U 0/620 U 0/720 F 0/760
Gas out-lets:
Text- circuit
Duplex Receptacle, 125V, 20A, red. backbox 2 Duplex Receptacle. 125V. 20A, red, backbox 2 Duplex Receptacle. 125V. 20A, red. backbox 3 Duplex Receptacle, 125V, ZOA, red, IJacklJox 4 Duplex Recept.icle, 125V, 20A, red, backbox 4 Duplex Receptacle, 125V. 20A, whit-e, backbox 5
DISS gas t-erminal unit- Beacon Medaes - US
Posit-ion B 0/180 B 0/320 B 0/460 B 0/600 B 0/740 L 0/160 L 0/50
Text-
AIR
N20
WAG□
VAC
VA(
,--------------------------------------7
Other outlets, Ambia - Hubbell {NFPA-70)
Posit-ion F 0/170 F 0/270 F 0/70
Text 4x RJ45, whit-e. backbox without conduif- 4x RJ45, white, IJacklJox without conduit Blind plaf-e 1-gang, backbox wif-hout- conduit
Ambia controls
Posit-ion R 0/60 R 20/120 R -20/120
Ambia rails
Text- Touchpanel brake/lift 2arm Switch lift stop Service socket-
Posit-ion Text- L -68,5/111 Fairfield rail 180 R -68,5/111 Fairfield rail 180 comment Power over Ef-hernef- [PoEI available: yes Power over Ethernet [PoEI availalJle: yes
Maximum shorf--circuit power over 10 VA: yes
L ______________________________________ ~
I I I I l
I
X X X X
<l
I
Column
I I without- preparat-ion Installation set for Amtiia Rack [RI and Ambia Pole [Pl
I
Ref-rofif-ting of accessories is 101 possiblel
I
(D G39B00 Handle wif-h sensoric
-qi
_(1)
3 1
(11
I
Requlaf-ory information
The safety distance from empty insf-allations shall be 200mm if it is inf-ended ta use them wit-h electrical inst-allat-ions and medical gas out-lef-s for oxidizing gases.
The safely distance of 200mm to oxidizing gases can be undercut if the electrical inst-allations do not reach a power >10VA int-he first- fault- case (e.g. short circuit-).
Drager Medical Supply Units shall not be inst-alled in rooms, where combustible or explosive gases can occur.
Drager Medical Supply Units shall not Ile installed in MRI rooms with high Tesla levels.
If insf-.illations or devices are installed, which have not been approved by Drager, these installation or devices needs an on-site approval date: aut-hor, revision t-ext:
proJeclname + department:
VA Fargo Medical [enter
Drager-country code: voltage IV): frequency [Hz): target counf-ry:
us 120/240 60 United States
LV-Pos: 31 + 3 2 room-no + roomname
OR 5
quantity of units: 1 pcs.
G) equipment color.
Q) service plaJ-e:
CD Color frame:
Mediccil White W1100/S-0500-N anodized Medical White W1100/S-0500-N scale:
view: 1:20, index:
rest· see details author:
dale:
name· title·
2022-08-08 Rl.iHMANN
Ceiling supply unif approved:
dale: 2022-08-09 name· BRINKER/ JANACK type left: Ambia lartide-number:GG16010I type right· Ambia lorricle-number,GG16070) revision index
!number 22-1711 I I iJrager
I
I-
AutoCAD SHX Text 735kg
AutoCAD SHX Text 7212N
AutoCAD SHX Text 12474Nm
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text min.260mm-max.1500mm
AutoCAD SHX Text Front view
AutoCAD SHX Text Back view
AutoCAD SHX Text Side view right
AutoCAD SHX Text Side view left
AutoCAD SHX Text F
AutoCAD SHX Text 3x OP75576 Shelf 1,25B/480 with US rails(GG16040 (1220)) Powder-coated aluminum shelf for the column of a supply unit with an usable surface of 530x480 mm and side plastic caps with integrated Fairfield rails, which have a maximum load of 10kg in each case. OP75576 Shelf 1,25B/480 with US rails(GG16040 (1222)) Powder-coated aluminum shelf for the column of a supply unit with an usable surface of 530x480 mm and side plastic caps with integrated Fairfield rails, which have a maximum load of 10kg in each case. + OP75587 Handle for brakes(GG16040 (1222)) Wireless handle to unlock the brakes of the ceiling supply unit via Bluetooth. G95048 remote control double arm lift 2x G39960 Handle with sensoric
AutoCAD SHX Text
AutoCAD SHX Text false ceiling
AutoCAD SHX Text 270°
AutoCAD SHX Text 330°
AutoCAD SHX Text Cable and hose length outside of ceiling bearing 1000mm.
AutoCAD SHX Text detail "B"
AutoCAD SHX Text
108KG
AutoCAD SHX Text brakes: F = Friction brake E = Electromagnetic brake
AutoCAD SHX Text
150KG
AutoCAD SHX Text max. load net carrying capacity
AutoCAD SHX Text
130KG
AutoCAD SHX Text
130KG
AutoCAD SHX Text max. load net carrying capacity
AutoCAD SHX Text ready floor
AutoCAD SHX Text Retrofitting of accessories is not possible!
AutoCAD SHX Text without preparation Installation set for Ambia Rack (R) and Ambia Pole (P)
AutoCAD SHX Text 330°
AutoCAD SHX Text E
AutoCAD SHX Text 330°
AutoCAD SHX Text Variable adjustment in steps
AutoCAD SHX Text 330°
AutoCAD SHX Text 330°
AutoCAD SHX Text separate preinstallation material: 1 * G25210 ceiling fixture set
AutoCAD SHX Text G39800 Handle with sensoric
AutoCAD SHX Text minimum concrete thickness 200mm concrete grade C20/25
AutoCAD SHX Text concrete ceiling
AutoCAD SHX Text Max Weight
AutoCAD SHX Text Tensile Force
AutoCAD SHX Text Torque at Ceiling
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text illustrative by others
AutoCAD SHX Text E
AutoCAD SHX Text E
AutoCAD SHX Text lifting 600mm lifting time approx. 10 sec.
AutoCAD SHX Text E
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Side view right
AutoCAD SHX Text Side view left
AutoCAD SHX Text Front view
AutoCAD SHX Text Back view
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text min.208mm- max.1000mm
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text F
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text detail "A"
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Electrical outlets: Ambia - Hubbell E socket Hospital Grade (NFPA-70) Position Text circuit Text circuit circuit F -55/300 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 F -55/450 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 F -55/600 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 F -55/800 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 B 55/200 Duplex Receptacle, 125V, 20A, white, backbox 4 Duplex Receptacle, 125V, 20A, white, backbox 4 4 L 0/300 Duplex Receptacle, 125V, 20A, white, backbox 4Duplex Receptacle, 125V, 20A, white, backbox 44
AutoCAD SHX Text Electrical outlets: Ambia - Hubbell E socket Hospital Grade (NFPA-70) Position Text circuit Text circuit circuit F 0/560 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 F 0/660 Duplex Receptacle, 125V, 20A, red, backbox 2 Duplex Receptacle, 125V, 20A, red, backbox 2 2 B 0/70 Duplex Receptacle, 125V, 20A, red, backbox 3 Duplex Receptacle, 125V, 20A, red, backbox 3 3 U 0/620 Duplex Receptacle, 125V, 20A, red, backbox 4 Duplex Receptacle, 125V, 20A, red, backbox 4 4 U 0/720 Duplex Receptacle, 125V, 20A, red, backbox 4 Duplex Receptacle, 125V, 20A, red, backbox 4 4 F 0/760 Duplex Receptacle, 125V, 20A, white, backbox 5Duplex Receptacle, 125V, 20A, white, backbox 55
AutoCAD SHX Text L
AutoCAD SHX Text R
AutoCAD SHX Text B
AutoCAD SHX Text F
AutoCAD SHX Text U (Ansicht von unten)
AutoCAD SHX Text detail "B"
AutoCAD SHX Text Top view
AutoCAD SHX Text Gas outlets: DISS gas terminal unit Beacon Medaes - US Position Text Text B 0/180 AIR AIR B 0/320 N2O N2O B 0/460 WAGD WAGD B 0/600 O2 O2 B 0/740 O2 O2 L 0/160 VAC VAC L 0/50 VACVAC
AutoCAD SHX Text Other outlets: Ambia - Hubbell (NFPA-70) Position Text comment Text comment comment F 0/170 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes F 0/270 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes F 0/70 Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Blind plate 1-gang, backbox without conduit Maximum short-circuit power over 10 VA: yes Maximum short-circuit power over 10 VA: yes Ambia controls Position Text Text R 0/60 Touchpanel brake/lift 2arm Touchpanel brake/lift 2arm R 20/120 Switch lift stop Switch lift stop R -20/120 Service socket Service socket Ambia rails Position Text Text L -68,5/111 Fairfield rail 180 Fairfield rail 180 R -68,5/111 Fairfield rail 180 Fairfield rail 180
AutoCAD SHX Text L
AutoCAD SHX Text R
AutoCAD SHX Text B
AutoCAD SHX Text F
AutoCAD SHX Text Top view
AutoCAD SHX Text detail "A"
AutoCAD SHX Text Gas outlets: DISS gas terminal unit Beacon Medaes - US Position Text Text B 55/400 N2 N2 B 55/700 AIR AIR L 0/500 AIR AIR L 0/650 O2 O2 L 0/800 VAC VAC R 0/439 CO2 CO2 R 0/750 VACVAC
AutoCAD SHX Text Other outlets: AK Cutouts Position Text Text B 0/550 1362 - Cut out „N2-Control Panel US“ 1362 - Cut out „N2-Control Panel US“ N2-Control Panel US“ Ambia - Hubbell (NFPA-70) Position Text comment Text comment comment B -55/200 Blind plate 1-gang, backbox without conduit Foot switch Blind plate 1-gang, backbox without conduit Foot switch Foot switch Maximum short-circuit power over 10 VA: yes F 55/450 Blind plate 1-gang, backbox without conduit Foot switch Blind plate 1-gang, backbox without conduit Foot switch Foot switch Maximum short-circuit power over 10 VA: yes F 55/600 Blind plate 1-gang, backbox without conduit Foot switch Blind plate 1-gang, backbox without conduit Foot switch Foot switch Maximum short-circuit power over 10 VA: yes F 55/800 Blind plate 1-gang, backbox without conduit Foot switch Blind plate 1-gang, backbox without conduit Foot switch Foot switch Maximum short-circuit power over 10 VA: yes F 55/300 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes L 0/150 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes 4x RJ45, white, backbox without conduit Power over Ethernet (PoE) available: yes Power over Ethernet (PoE) available: yes Ambia controls Position Text Text R 0/582 Touchpanel brake 2arm Touchpanel brake 2arm R 20/660 Service socket Service socket Ambia rails Position Text Text B 0/950 Fairfield rail 280 Fairfield rail 280 L 0/950 Fairfield rail 180 Fairfield rail 180 R 0/950 Fairfield rail 180 Fairfield rail 180
AutoCAD SHX Text Regulatory information The safety distance from empty installations shall be 200mm if it is intended to use them with electrical installations and medical gas outlets for oxidizing gases. The safety distance of 200mm to oxidizing gases can be undercut if the electrical installations do not reach a power >10VA in the first fault case (e.g. short circuit). Dräger Medical Supply Units shall not be installed in rooms, where combustible or explosive gases can occur. Dräger Medical Supply Units shall not be installed in MRI rooms with high Tesla levels. If installations or devices are installed, which have not been approved by Dräger, these installation or devices needs an on-site approval.
AutoCAD SHX Text Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
AutoCAD SHX Text Measurements to be checked by Dräger on site.
AutoCAD SHX Text Ceiling supply unit type left: type right:
AutoCAD SHX Text Ambia (article-number: GG16010)(article-number: GG16010)
AutoCAD SHX Text Ambia (article-number: GG16010)(article-number: GG16010)
AutoCAD SHX Text
3.1 + 3.2
AutoCAD SHX Text
OR 5
AutoCAD SHX Text Medical White W1100/S-0500-N
AutoCAD SHX Text equipment color: service plate: Color frame:
AutoCAD SHX Text Medical White W1100/S-0500-N
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Head
AutoCAD SHX Text Column
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text anodized
AutoCAD SHX Text quantity of units: 1 pcs.
AutoCAD SHX Text projectname + department:
AutoCAD SHX Text room-no. + roomname:
AutoCAD SHX Text date: author: revision text: index:
AutoCAD SHX Text LV-Pos.:
AutoCAD SHX Text Dräger-country code:
AutoCAD SHX Text target country:
AutoCAD SHX Text voltage (V):
AutoCAD SHX Text frequency (Hz):
AutoCAD SHX Text revision index 00
AutoCAD SHX Text number:
AutoCAD SHX Text title:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text author: approved:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text scale:
AutoCAD SHX Text view: 1:20, rest: see details
AutoCAD SHX Text
RÜHMANN
AutoCAD SHX Text 2022-08-08
AutoCAD SHX Text 22-1711
AutoCAD SHX Text
US
AutoCAD SHX Text United States
AutoCAD SHX Text 120/240
AutoCAD SHX Text
AutoCAD SHX Text detail "assembly" M. 1:10
AutoCAD SHX Text VA Fargo Medical Center
AutoCAD SHX Text
BRINKER/JANACK
AutoCAD SHX Text 2022-08-09
Use of this document / contents is forbidden without express authority. All rights reserved. Refer to protection notice ISO 16016.
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
AutoCAD SHX Text quantity of units: 1 pcs.
AutoCAD SHX Text projectname + department:
AutoCAD SHX Text revision index 00
AutoCAD SHX Text number:
AutoCAD SHX Text title:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text author: approved:
AutoCAD SHX Text room-no. + roomname:
AutoCAD SHX Text date: author: revision text: index:
AutoCAD SHX Text date:
AutoCAD SHX Text name:
AutoCAD SHX Text scale:
AutoCAD SHX Text LV-Pos.:
AutoCAD SHX Text
RÜHMANN
AutoCAD SHX Text 2022-08-08
AutoCAD SHX Text 22-1712
AutoCAD SHX Text view: 1:20, rest: see details
AutoCAD SHX Text
BRINKER/JANACK
AutoCAD SHX Text 2022-08-09
AutoCAD SHX Text surgical lights system type left: type right:
AutoCAD SHX Text Polaris (article-number: G16720)(article-number: G16720)
AutoCAD SHX Text equipment color: Medical white NCS-S-0500-N
AutoCAD SHX Text Dräger-country code:
AutoCAD SHX Text target country:
AutoCAD SHX Text voltage (V):
AutoCAD SHX Text frequency (Hz):
AutoCAD SHX Text
US
AutoCAD SHX Text United States
AutoCAD SHX Text 120/240
AutoCAD SHX Text
AutoCAD SHX Text Measurements to be checked by Dräger on site.
AutoCAD SHX Text Polaris Multimedia (article-number: G16915)(article-number: G16915)
AutoCAD SHX Text
3.3 + 3.4 + 3.5
AutoCAD SHX Text
OR 5
AutoCAD SHX Text VA Iowa City VISN 23
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G19200
AutoCAD SHX Text G19779
AutoCAD SHX Text flat cover 144+ 144
AutoCAD SHX Text uni flange
AutoCAD SHX Text Pos.
AutoCAD SHX Text No/pc
AutoCAD SHX Text Order-no.
AutoCAD SHX Text Description
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text false ceiling
AutoCAD SHX Text ready floor
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G15796
AutoCAD SHX Text clamp ring 110, outside diamenter 160mm
AutoCAD SHX Text Top View M 1:20
AutoCAD SHX Text G36699
AutoCAD SHX Text spring load arm Polaris 100/200/600
AutoCAD SHX Text incl. cabling
AutoCAD SHX Text lifting reduced 860mm as shown (max. lifting of arm 1179mm)
AutoCAD SHX Text 360°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G36993
AutoCAD SHX Text sterilizable handle set E (2 pcs)
AutoCAD SHX Text suspension tube 110x220mm
AutoCAD SHX Text G26302
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G97964
AutoCAD SHX Text axis 2-fold MOX 1450/L1 1600
AutoCAD SHX Text 360°
AutoCAD SHX Text 330°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G94401
AutoCAD SHX Text suspension tube 110x460mm
AutoCAD SHX Text G28374
AutoCAD SHX Text G35467
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 180°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 330°
AutoCAD SHX Text
AutoCAD SHX Text lifting reduced 956mm as shown (max. lifting of arm 1076mm)
AutoCAD SHX Text spring arm display 17-26 MOX SD/HD
AutoCAD SHX Text incl. cabling
AutoCAD SHX Text 1 displ.-adap. 24-32" Reus.
AutoCAD SHX Text displ.-adapter incl. prepared sterilizable handle set E
AutoCAD SHX Text 200°
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G36708
AutoCAD SHX Text
AutoCAD SHX Text G50741
AutoCAD SHX Text wall box FM 250
AutoCAD SHX Text WCU lights + camera P600
AutoCAD SHX Text G33866
AutoCAD SHX Text front cover WCU P6+P6+camera
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G36402
AutoCAD SHX Text PSU 100-240V Polaris 600
AutoCAD SHX Text N
AutoCAD SHX Text L
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Output: Light System
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text System number B
AutoCAD SHX Text System number A
AutoCAD SHX Text Initialize system
AutoCAD SHX Text Contact service
AutoCAD SHX Text Initialize in progress
AutoCAD SHX Text Output voltage
AutoCAD SHX Text Emergency power
AutoCAD SHX Text Supply mains
AutoCAD SHX Text All Poles
AutoCAD SHX Text Output
AutoCAD SHX Text Emergency Power
AutoCAD SHX Text Supply mains
AutoCAD SHX Text
AutoCAD SHX Text Master Switch
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Supply mains
AutoCAD SHX Text 24V
AutoCAD SHX Text Power
AutoCAD SHX Text Emergency
AutoCAD SHX Text N
AutoCAD SHX Text L
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Output: Light System
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text System number B
AutoCAD SHX Text System number A
AutoCAD SHX Text Initialize system
AutoCAD SHX Text Contact service
AutoCAD SHX Text Initialize in progress
AutoCAD SHX Text Output voltage
AutoCAD SHX Text Emergency power
AutoCAD SHX Text Supply mains
AutoCAD SHX Text All Poles
AutoCAD SHX Text Output
AutoCAD SHX Text Emergency Power
AutoCAD SHX Text Supply mains
AutoCAD SHX Text
AutoCAD SHX Text Master…
This is the start of the file's text. The full file is on GovTribe.