KEY LEGEND
x’-x”
REFERENCE
x’-x”
WATER/DRAIN
x’-x”
POWER
NEW FLOOR DRAIN
EXISTING FLOOR DRAIN
# UTILITY LOCATION
DATA
POWER
WATER LOOP
W/SHUT OFF
CCM PSU
AIR
ROCHE DIAGNOSTICS
9115 HAGUE ROAD
INDIANAPOLIS, IN 46256
SCALE: ¼” = 1'-0" SHEET 1 OF 4
GENERAL NOTES:
1.SIZE AND CONFIGURATION OF SYMBOLS
SHOWN ON THE DRAWING ARE USED TO
REPRESENT THE EXISTENCE OF, OR NEED FOR
ADDITIONAL UTILITIES.
2.CUSTOMER IS RESPONSIBLE FOR MEETING ALL
LOCAL CODES.
3.CUSTOMER IS RESPONSIBLE FOR CONTACTING
EACH 3
RD
PARTY VENDOR TO OBTAIN TECHNICAL
EQUIPMENT SPECIFICATIONS RELATED TO
INSTALLATION OF INSTRUMENTS AND REQUIRED
UTILITIES. ROCHE WILL PROVIDE TECHNICAL
SPECIFICATIONS AND GENERAL LOCATIONS FOR
UTILITIES OF ROCHE EQUIPMENT ONLY.
CONNECTION TO 3
RD
PARTY SYSTEMS TO BE
VERIFIED BY CUSTOMER.
4.THIS DRAWING IS SUBJECT TO CHANGE.
DATE: DESCRIPTION:
PROJECT HISTORY:
UTILITIES – CONSTRUCTION PLAN VAMC FORT
HARRISON
HELENA, MT
VAMC_FTHARRISON_CONSTRUCTIONPLAN110222FM-1
LDA: FREDERICK McCARROLL
SWC: LAUREL SPRAGUE
10/25/22 DIM TOOL VERIFIED
U P
S
.0 kV A
S er ie s
B at te ry
C a b in e t
See UPS/Battery Cabinet Specs.
Power
#0
Off-site location TBD. Max distance is 250ft.
Vir tua l Server
Primary
Vir tua l Server
Shadow
Off-site location TBD.
#0
10/28/22 ADDED IT GRID
Existing
Drain
Existing
Drain
Data
Power
#1
Mobi le Table
M o b ile
T a b le
M o b ile
T a b le kg kg kg kg kg kg 7 kg kg kg kg 2 kg kg
R S
F A
O B
O B
S c o b a s o u tp u t s ta tio n kg 1 kg kg kg kg 2 kg kg kg kg kg 4 kg kg kg kg 2 kg kg kg kg 3 kg kg
D S
P A
C U w it h
A C
B
IP
B
H S c o b a s in p u t s ta ti o n
S C
M kg kg kg kg
C R
W cobas 8100 kg kg kg kg kg kg 2 kg kg kg kg 3 kg kg
T L
J
TLJ
kg kg kg kg kg kg kg kg kg kg kg kg kg kg kg kg
TLJ
kg kg kg kg kg kg kg kg
TLJ
N e w
D ra in c o b a s p ro c o b a s L in k 2
R e p o rt P rin te r
Data
Power
Water Loop
C o n tr o l U n it
R e p o rt
P ri n te r
Firewall
N w
D ra in c o b a b a s L in p o rt P rin te r
Data
Power
Water Loop
B R
F c o b a s p ro kg kg kg8kg kg1 kg kg kg c o b a s p ro kg kg kg kg kg kg kg kg kg kg kg 1 kg 1 kg kg kg 2 kg kg kg 4 kg kg kg kg kg kg kg kg kg kg 8 kg kg kg kg c o b a s p ro
IS
E -5
-8
B R
F c o b a s p ro kg kg kg8kg kg1 kg kg kg c o b a s p ro kg kg kg kg kg kg kg kg kg kg kg 1 kg 1 kg kg kg 2 kg kg kg 4 kg kg kg kg kg kg kg kg kg kg 8 kg kg kg kg c o b a s p ro
IS
E -5
-8
15'-6"
8'-7"
8'-7"
Firewall Firewall
14'-10"
#2 #3
11/02/22
LAYOUT/SITE CHANGE; LDA
REMEDIATION
INSTALL
x mm.
x’-x”
REFERENCE
x’-x”
WATER/DRAIN
x’-x”
POWER
NEW FLOOR DRAIN
EXISTING FLOOR DRAIN
# UTILITY LOCATION
DATA
POWER
WATER LOOP
W/SHUT OFF
CCM PSU
AIR
ROCHE DIAGNOSTICS
9115 HAGUE ROAD
INDIANAPOLIS, IN 46256
GENERAL NOTES:
1.SIZE AND CONFIGURATION OF SYMBOLS
SHOWN ON THE DRAWING ARE USED TO
REPRESENT THE EXISTENCE OF, OR NEED FOR
ADDITIONAL UTILITIES.
2.CUSTOMER IS RESPONSIBLE FOR MEETING ALL
LOCAL CODES.
3.CUSTOMER IS RESPONSIBLE FOR CONTACTING
EACH 3
RD
PARTY VENDOR TO OBTAIN TECHNICAL
EQUIPMENT SPECIFICATIONS RELATED TO
INSTALLATION OF INSTRUMENTS AND REQUIRED
UTILITIES. ROCHE WILL PROVIDE TECHNICAL
SPECIFICATIONS AND GENERAL LOCATIONS FOR
UTILITIES OF ROCHE EQUIPMENT ONLY.
CONNECTION TO 3
RD
PARTY SYSTEMS TO BE
EQUIPMENT SPECIFICATIONS
NOT TO SCALE SHEET 2 OF 4
VAMC FORT
HARRISON
HELENA, MT
VAMC_FTHARRISON_CONSTRUCTIONPLAN110222FM-1
LDA: FREDERICK McCARROLL
SWC: LAUREL SPRAGUE
10/25/22 DIM TOOL VERIFIED
10/28/22 ADDED IT GRID
AT UTILITY
LOCATION ROCHE SUPPLIED EQUIPMENT POWERED FROM
NUMBER OF DEVICE
POWER CABLES PRIMARY POWER, VOLTS
POWER PROTECTION,
AMPS PHASE RECEPTACLE TYPE
DEVICE'S UPS
RATING, KVA
HEAT
DISSIPATION,
BTU/HR
0 UPS 40.0 kVA WITH BATTERY CABINET FACILITY 0 208 150 3 PHASE 4 CONDUCTOR WITH GROUND HARDWIRED FROM CUSTOMER'S FACILITIES DEPARTMENT 40.0 12,270
1 COBAS 8100 SYSTEM UPS 40.0 kVA MAIN LUG PANEL (MLP) 5 208 30 SINGLE PHASE NEMA L6-30R 5.2 24,064
1 REPORT PRINTER (from COBAS 8100) FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 2,048
1 COBAS 8100 CONTROL UNIT (PC & MONITOR) UPS 40.0 kVA MAIN LUG PANEL (MLP) 2 120 20 SINGLE PHASE NEMA 5-20R 0.8 1,550
1 FORTIGATE FIREWALL UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 61
2 REPORT PRINTER (from COBAS PRO) FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 1,877
2 COBAS_PRO <ISE/503/801> UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 208 50 SINGLE PHASE HUBBELL 3762C (CORD) OR 3771 (WALL MOUNT) 10.0 19,107
2 COBAS LINK 2 UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 0
2 FORTIGATE FIREWALL UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 61
3 REPORT PRINTER (from COBAS PRO) FACILITY 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 1,877
3 COBAS_PRO <ISE/503/801> UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 208 50 SINGLE PHASE HUBBELL 3762C (CORD) OR 3771 (WALL MOUNT) 10.0 19,107
3 COBAS LINK 2 UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 0
3 FORTIGATE FIREWALL UPS 40.0 kVA MAIN LUG PANEL (MLP) 1 120 20 SINGLE PHASE NEMA 5-20R 0.0 61
82,083
AT UTILITY
LOCATION CUSTOMER SUPPLIED DATA CONNECTION CONNECTION TYPE
1 DATA UTILITY SEE CUSTOMER PROVIDED INTERFACE CONNECTIONS
2 DATA UTILITY SEE CUSTOMER PROVIDED INTERFACE CONNECTIONS
3 DATA UTILITY SEE CUSTOMER PROVIDED INTERFACE CONNECTIONS
AT UTILITY
LOCATION CUSTOMER SUPPLIED WATER CONNECTION CONNECTION SIZE
DEIONIZED WATER
SUPPLY RATE,
LITERS/HR
MINIMUM FLOOR DRAIN
INSIDE DIAMETER,
INCHES
2 WATER UTILITY 5/8" BARBED HOSE FITTING W 1/2" ID 64 2
3 WATER UTILITY 5/8" BARBED HOSE FITTING W 1/2" ID 64 2
CUSTOMER'S POWER RESPONSIBILITY
CUSTOMER'S DATA RESPONSIBILITY
CUSTOMER'S WATER/WASTE RESPONSIBILITY
NOTE: TYPE 1 CLRW, UTILIZING A CONTINUOUS RECIRCULATION /POLISHING LOOP, WITH A MINIMAL VELOCITY OF 5 FT/SEC. AND A 0.2 MICRON OUTPUT FILTER. PLEASE REFER TO THE SPECIFIC INSTALLATION SITE REQUIREMENTS DOCUMENTATION.
TOTAL SYSTEM HEAT LOAD
BTU/HR (VALUES ROUNDED)
82,083
0 1 2 3 4 5 6 7 8 9 10 Line # Communication Type Connection Device l 1 TCP/IP (RJ45) Network to Primary l 2 TCP/IP (RJ45) Network to Shadow l 3 TCP/IP (RJ45) Network to Fortigate/Connect 2/8100 CU l 4 TCP/IP (RJ45) Network to Report Printer l 5 TCP/IP (RJ45) Network to Fortigate/Link 2/c Pro l 6 TCP/IP (RJ45) Network to Report Printer l 7 TCP/IP (RJ45) Network to Fortigate/Link 2/c Pro l 8 TCP/IP (RJ45) Network to Report Printer
2 2 2 2 0 0 0 0 0 0 0 Total ports per Power Pole
** 10/28/2022 RJW_fm v1**
Customer Provided Interface Connections
Power Poles (2 dots on one line
= point to point connection)
11/02/22
INSTALL
x mm.
x’-x”
REFERENCE
x’-x”
WATER/DRAIN
x’-x”
POWER
NEW FLOOR DRAIN
EXISTING FLOOR DRAIN
# UTILITY LOCATION
DATA
POWER
WATER LOOP
W/SHUT OFF
CCM PSU
AIR
ROCHE DIAGNOSTICS
9115 HAGUE ROAD
INDIANAPOLIS, IN 46256
SCALE: ¼” = 1'-0" SHEET 3 OF 4
GENERAL NOTES:
1.SIZE AND CONFIGURATION OF SYMBOLS
SHOWN ON THE DRAWING ARE USED TO
REPRESENT THE EXISTENCE OF, OR NEED FOR
ADDITIONAL UTILITIES.
2.CUSTOMER IS RESPONSIBLE FOR MEETING ALL
LOCAL CODES.
3.CUSTOMER IS RESPONSIBLE FOR CONTACTING
EACH 3
RD
PARTY VENDOR TO OBTAIN TECHNICAL
EQUIPMENT SPECIFICATIONS RELATED TO
INSTALLATION OF INSTRUMENTS AND REQUIRED
UTILITIES. ROCHE WILL PROVIDE TECHNICAL
SPECIFICATIONS AND GENERAL LOCATIONS FOR
UTILITIES OF ROCHE EQUIPMENT ONLY.
CONNECTION TO 3
RD
PARTY SYSTEMS TO BE
INSTALL – CONSTRUCTION PLAN VAMC FORT
HARRISON
HELENA, MT
VAMC_FTHARRISON_CONSTRUCTIONPLAN110222FM-1
LDA: FREDERICK McCARROLL
SWC: LAUREL SPRAGUE
10/25/22 DIM TOOL VERIFIED
10/28/22 ADDED IT GRID
U P
S
.0 kV A
S er ie s
B at te ry
C a b in e t
See UPS/Battery Cabinet Specs.
Power
#0
Off-site location TBD. Max distance is 250ft.
Vir tua l Server
Primary
Vir tua l Server
Shadow
Off-site location TBD.
#0
Existing
Drain
Existing
Drain
Data
Power
#1
Mobi le Table
M o b ile
T a b le
M o b ile
T a b le kg kg kg kg kg kg 7 kg kg kg kg 2 kg kg
R S
F A
O B
O B
S c o b a s o u tp u t s ta tio n kg 1 kg kg kg kg 2 kg kg kg kg kg 4 kg kg kg kg 2 kg kg kg kg 3 kg kg
D S
P A
C U w it h
A C
B
IP
B
H S c o b a s in p u t s ta ti o n
S C
M kg kg kg kg
C R
W cobas 8100 kg kg kg kg kg kg 2 kg kg kg kg 3 kg kg
T L
J
TLJ
kg kg kg kg kg kg kg kg kg kg kg kg kg kg kg kg
TLJ
kg kg kg kg kg kg kg kg
TLJ
N e w
D ra in c o b a b a s L in p o rt P rin te r
Data
Power
Water Loop
C o n tr o l U n it
R e p o rt
P ri n te r
Firewall
N w
D ra in c o b a b a s L in p o rt P rin te r
Data
Power
Water Loop
B R
F c o b a s p ro kg kg kg8kg kg1 kg kg kg c o b a s p ro kg kg kg kg kg kg kg kg kg kg kg 1 kg 1 kg kg kg 2 kg kg kg 4 kg kg kg kg kg kg kg kg kg kg 8 kg kg kg kg c o b a s p ro
IS
E -5
-8
B R
F c o b a s p ro kg kg kg8kg kg1 kg kg kg c o b a s p ro kg kg kg kg kg kg kg kg kg kg kg 1 kg 1 kg kg kg 2 kg kg kg 4 kg kg kg kg kg kg kg kg kg kg 8 kg kg kg kg c o b a s p ro
IS
E -5
-8 m m
678mm.
m m m m
4285mm.
2598mm.
1842mm.
Firewall Firewall
#2 #3
11/02/22
NOT TO SCALE SHEET 4 OF 4
ROCHE DIAGNOSTICS
9115 HAGUE ROAD
INDIANAPOLIS, IN 46256
POWERVAR 3300 SERIES UPS CIRCUIT DIAGRAM CLEARANCE SPECIFICATIONS
RECOMMENDED INSTALLATION OPTIONAL PCDU OPTIONAL PCDU INSTALLATION
| VAMC_FtHarrison_ConstructionPlan110222fm-1.vsdx |
| 1/4" - UTILITIES CONST. PLAN |
| 1/4" - SPECS |
| 1/4" - INSTALL CONST. PLAN |
| 1/4" - UPS/PCDU SPECS |