S02 - Exhibit 1. VMU Cage Washer Cutsheet.pdf

PDF 2 MB Posted

Attached to
Y1DA--537-23-102 VMU Equipment Tie-in
Solicitation number
36C25225B0010
Issued by

About this file

This file is a technical specification sheet and engineering drawings for the LYNX Model 410RS (formerly Scientek SW6100/SW6200/SW6300) Cage and Rack Washer, which appears to be related to a Veterans Affairs procurement. The document details a heavy-duty, large-capacity hydro-spray washer designed for cleaning laboratory animal cages, bottles, racks, and debris pans.

The washer specifications include a 48" x 88" x 96" standard chamber size, 10 HP treatment pump, stainless steel construction, PLC control system with color HMI touch screen, and various utility requirements including 480V electrical supply, hot and cold water connections, steam (40-60 PSI), and drain systems. Key features include modular construction, rotating spray delivery system, 10 configurable cycles, automatic self-cleaning screen, and temperature guarantee. The document includes detailed engineering drawings showing installation requirements, utility connections, and dimensional specifications for the Jesse Brown VA facility.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Ensuring efficient cleaning for all your animal care items The LYNX Model 410RS (formerly Scientek SW6100/SW6200/SW6300) Cage and Rack Washers are heavy duty, large capacity reliable hydro-spray washers designed for ease of use, minimal maintenance, and highest cleanability of caging, bottles, racks, debris pans, and other items used in the care of laboratory animals. These washers deliver high performance within a simplistic design. The system is provided with greater flexibility to meet today’s “green” initiatives and high throughput demands. LYNX standardizes on non-proprietary parts on all equipment. Pit mounting is standard.

Optional Features

• Shallow Base Unit (6”)

• Pit or Floor Mounted

• Pass-thru Unit

• Right or Left Hand Services

• Re-usable Throwaway Alkaline Tank

• House Hot Water Heat Exchanger

• Integral Data Collection System

• Drain Discharge Cooling

• Automatic Damper

• SS Exhaust Fan

• Low Steam Pressure Package

• Pass-thru Door Interlock System

• Air Compressor

• Printer

• Seismic Tie Down

• Detergent Dispensing Systems

• pH Neutralization System

• Feeder Bottle Washing System

• Automatic Water Rack Flush System

• Split Base for Special Entry

• Utility Enclosure Panels

• Inclined Floor Grating

• Barrier Wall Flange(s)

Standard Features

• Heavy-duty Hinge Door

• Modular Construction

• Knock-down Shipment

• Rotating Spray Delivery System

• Personnel Safety Features

• PLC Control System with Full Color HMI Touch Screen

• 10 Configurable Cycles

• Automatic Self-cleaning Screen

• Temperature Guarantee

• Insulated Construction

• Incoming Utility Gauges

• Interior Illumination

• SS Recirculation Piping and Components

• Non-proprietary Parts

Construction LYNX 410RS Cage and Rack Washers are provided with a heavy-duty stainless steel door with port view glass window. Recirculated water is heated by a stainless steel steam coil to achieve guaranteed treatment temperatures. A color touch screen control panel provides a user-friendly interface for selecting, programming, and monitoring cycles.

Unit is provided with a stainless steel 10 HP treatment pump with a direct read pump pressure gauge.

Accessories

• Floor Ramps

• Bottle Wash Cart

• Universal Cage Wash Cart

• High Capacity Cage Wash System

LYNX

Cage & Rack Washers Model 410RS

Your Healthcare and Life Science PartnerA Chalmers Group Company

Side Tank

Quicklock Connection

Steam Heated acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight acocco Highlight

Model Standard Chamber Size

W x H x D Standard Unit Size

W x H x D Opt. Door

Widths Opt. Door

Heights Opt. Chamber

Lengths

410RS1 48" x 88" x 96" 88" x 95" x 98"-116" 36", 42", 46" 74", 82" 90", 100", 108"

Representative Drawing

410RS Cage and Rack Washer Sell Sheet_M1040_1.00 2021-0318

Utilities

*CW - Required if drain discharge cool-down package ordered

Washers can be designed to any size to meet your application requirements. Certain options can increase unit sizes.

1) 410RS (formerly Scientek SW6100/SW6200/SW6300)

Spire-is.com | sales@spire-is.com | 877 679 7800 © Spire Integrated Solutions. All rights reserved.

Electrical (E) 3-60-208/480/600V, 12HP

Hot Water (HW) 1" NPT, 35 PSI, 25 GPM @ 140° F ± 5° F (65 gallon sump capacity)

Cold Water (CW)* 1" NPT, 35 PSI, 25 GPM @ 50° F ± 5° F

Steam (S) 11/2" NPT, 40-60 PSI, Dynamic ± 5 PSI, 450-600 #/Hr Flow Rate

Condensate (C) 3/4" NPT

Air (A) 1/4" NPT, 80-100 PSI, 1-2 CFM Flow Rate

Drain (D) 12" x 12" Floor Sink with 4" Floor Drain (minimum), 180 GPM Flow Rate

Vent (V) 8" x 13" Duct, Saturated Vapor @ 190° F , 1 DC (1200 CFM) + 1 IDC (500 CFM)

Top View

Side ViewFront View

BUILD DRAWING

FOR MARKUP OR APPROVAL

Site condition check list:

- Hot & cold water rough ins

- Floor sink & building drain size

- Compressed air rough in

- Electrical rough in

- Exhaust duct rough in ( All welded ss duct, sloped back to washer recommended. )

- Breather vent containment ( vent hood recommended. No direct connection. )

- Pit configuration is correct

- Ceiling height & wall clearances

- Door swings are clear & correct

- Pathways for entry of washer into building ( A non-standard split of washer may be required )

- Washer interior sizes are correct:

- 48" - Clear - Left to right

- 88" - Clear - Top to bottom

- 96" - Clear - Front to back

SIGNED:

DATE:

75 1 4 "

100 1

8 " OVER BODY

100 15 16 "

SOILED SIDE

SERVICE SIDE

CLEAN SIDE

Scientek / SW6000 Cage And Rack Washer

STANDARD FEATURES CONTROL SYSTEM INCLUDES OPTIONS & CONFIGURATIONS ACCESSORIES

STAINLESS STEEL CONSTRUCTION - TYPE 304 -#4 FINISH SERVICE & CONTROL - ON / OFF SWITCHES WIRE FOR 480 VOLT SUPPLY 480V-6 VERTICAL CAGE WASH HEADER SMR-6

INSULATED WASH CHAMBER EMERGENCEY STOP PUSH BUTTON DRAIN COOL DOWN SIDE TANK ( GRAVITY DRAIN) DDCD-6

STAINLESS STEEL - PUMP HOUSINGS ALLEN BRADLEY MICROPROCESSOR 10 HP PUMP MOTOR 10 HP

EMERGENCEY STOP CABLE SYSTEM TOUCH SCREEN - OPERATOR INTERFACE LOW PROFILE LP-6

INTERNAL SIDE PROTECTION RAIL TEN PRE-PROGRAMMED WASH CYCLES KNOCKED DOWN FOR SHIPPING KD-6

REMOVABLE FILTER BASKETS TEMPERATURE - ACTUAL & SET POINT INDICATION DRAIN COOLDOWN DCD-6

MICROPROCESSOR CONTROLLED WASH CYCLES TIME - ACTUAL & PRESET INDICATION PANEL MOUNTED PRINTER PTI-46

DOOR SAFETY HARDWARE ADJUSTABLE CYCLE TIME & TEMPERATURE SET POINTS QUICK LOCK CONNECTION QL-6

THREE STAGE AUTOMATIC DRAIN CHANGEABLE CYCLE NAMES SPLIT BASE

PNEUMATIC CONTROL VALVES CYCLE LOCK OUT CAPABILITY

" AQUAPULSE " ADUJUSTABLE SPRAY PROGRAM MONITORED ALARM EVENTS ( LOGS FOR 50 EVENTS )

INTERIOR LIGHT TIME & DATE DISPLAY INTERIOR DIMENSIONS

CONTROLS & SERVICE ON LEFT HAND SIDE CONTROL SIGNALLING OF UP TO 4 CHEMICAL PUMPS 48" - DOOR OPENING CLEAR - LEFT TO RIGHT

STAINLESS STEEL SUMP HEATER EXTRA EXHAUST TIME AVAILABILITY 88" - DOOR OPENING CLEAR - TOP TO BOTTOM

REMOVABLE SECTIONED FLOOR GRATES AUDIBLE END OF CYCLE SIGNAL 96" - INTERIOR LENGTH CLEAR

COLD WATER PREWASH ( HEALTH CARE ONLY )

D

E

F

C

1 2 3 4

B

A

C

D

A

B

A3

SHEET/DRAWING #:SCALE:

FAB

PART #:

TITLE:

REVISION:DO NOT SCALE DRAWING

DATESIGNATURENAME

DEBURR AND

BREAK SHARP

EDGES

JOB #:UNLESS OTHERWISE SPECIFIED:

DIMENSIONS ARE IN INCHES

TOLERANCES:

LINEAR: 1/32 "

ANGULAR: .5 Deg.

Q.A

MFG

APPV'D

CHK'D

DRAWN

SERIAL #:

E

410RS-PT-RH-STM 48X88X96JAN 18 24

WEIGHT:

Gulshan

NOTE: ALL LAYOUT DRAWINGS ARE - GOOD SIDE OF MATERIAL UP INVENTORY #:

87654321

S1-410RS-083

CUSTOMER:

PROJECT:

VA JESSE BROWN

VA JESSE BROWN

E:\Solidworks Washers\6000 WASHERS\6000- JOB FILES\3654- JESSE BROWN-410RS-PT-RH-LP-STM\3654- 48x88x96-PT-RH-LP-STM

101460

BUILD DRAWING

FOR MARKUP OR APPROVAL

Site condition check list:

- Hot & cold water rough ins

- Floor sink & building drain size

- Compressed air rough in

- Steam rough in

- Electrical rough in

- Exhaust duct rough in ( All welded ss duct, sloped back to washer recommended. )

- Breather vent containment ( vent hood or recommended. No direct connection )

- Pit configuration is correct

- Ceiling height & wall clearances

- Door swings are clear & correct

- Pathways for entry of washer into building ( A non-standard split of washer may be required )

- Washer interior sizes are correct:

- 48" - Clear - Left to right

- 88" - Clear - Top to bottom

- 96" - Clear - Front to back

SIGNED:

DATE:

ISO VIEW

100 31 32 "

FROM FINISHED

FLOOR LEVEL

6 13 16 "

100 5 32 "

8" 8" 64 5 32 "

GENERAL NOTES & RECOMMENDATIONS:

-All services to Washer ( by table ) must comply with applicable local codes and practices regarding final connections, responsibility herein lies with installation company,and not equipment manufacturer.

-Supply lines to be sized larger than utilities shown to ensure adequate service.

-All connection lines to be provided with appropriate shut offs, connections, & regulators.

-Range indicates minimum/maximum pressures, if supply exceeds maximum, provide a regulator and or pressure release valve prior to connection with machine ( by others ) -Electrical supply should be connected with sealtite flex and a fused panel box.

- If condensate lift is required, a condensate pump is recommended.

10 HP PUMP

MOTOR

SERVICE SIDE VIEW

SOILED

SIDE

4 5

CLEAN

SIDE

48" DOOR

OPENING

88" DOOR

OPENING

RH HINGE

CLEAN SIDE

CONTROL PANEL

CLEAN SIDE VIEW

LH HINGE

USER

INTERFACE

SCREEN

EMERGENCEY

STOP

BUTTON

MAIN

DISCONNECT

WINDOW

SOILED SIDE VIEW

223 3

8 " OVER RAMP

100 5

32 " OVER BODY

76"

OVER RAMP

60" RAMP 60" RAMP

8 9

GRAVITY DRAIN

COOLDOWN TANK

RAMP RAMP

SW6000 - Service Requirements & Engineering Data - Steam - No Side Tanks Machine Electrical Draw - 480 Volt - 14 Amp - 3 Ph

CHK. # SERVICE CONNECTIONS DYNAMIC PRESSURE RANGE PEAK FLOW RATES AVERAGE CONSUMPTION X 1 HOT WATER 1" NPT - 120 DEG F / 49 DEG C 40-60 PSIG / 267-414 KPA 25 GPM / 95 LPM 300 G/CYCLE / 1136 L/CYCLE

2 STEAM 1 1/2" NPT 40-60 PSIG / 276-414 KPA 885 LBS / HR / 401 KG / HR 420 LBS / CYCLE / 191 KG /

CYCLE

X 3 CONDENSATE - HEAT EXCHANGER 3/4" NPT 1.9 GPM / 7.0 LPM

X 4 CONDENSATE - SUMP HEATER 3/4" NPT 1.0 GPM / 3.8 LPM

X 5 DRAIN / OVERFLOW 2" NPT 315 GPM / 1193 LPM

X 6 AIR 1/4" NPT 80 PSIG / 552 KPA 1 CU. FT/HR / .03 CU M/HR

X 7 ELECTRICAL 1/2" KO @ CONTROL BOX

X 8 EXHAUST VENT @ SOILED END 8" X 13" - DIRECT CONNECTION .5 SP "WG / .13 KPA 600 -1200 CFM 1200 CFM RECOMMENDED

X 9 BREATHER VENT @ CLEAN END 8" X 22"- (EXHAUST HOOD REQD. ABOVE) .25 SP "WG / .07 KPA 500 CFM / 14.2 M3/MIN

X 10 DRAIN COOLDOWN 1" NPT - 60 DEG F / 16 DEG C MAX 30-60 PSIG / 207-414 KPA 25 GPM / 95 LPM 106 G/CYCLE / 401 L/CYCLE

X 11 FLOOR SINK DRAIN 12"X12"X8" DP 4" WASTE PIPE - DEDICATED (RECOMMENDED) 600 GPH / 2271 LPH

Miscellaneous Engineering Data

CRATED SHIPPING WEIGHT 63500 LBS / 2800 KGS

NO. OF SPLITS 5 ( STANDARD )

HEAT LOSS 34,000 BTU`S / HR

OPERATING SUMP CAPACITY 90 GALS US

D

E

F

C

1 2 3 4

B

A

C

D

A

B

A3

SHEET/DRAWING #:SCALE:

FAB

PART #:

TITLE:

REVISION:DO NOT SCALE DRAWING

DATESIGNATURENAME

DEBURR AND

BREAK SHARP

EDGES

JOB #:UNLESS OTHERWISE SPECIFIED:

DIMENSIONS ARE IN INCHES

TOLERANCES:

LINEAR: 1/32 "

ANGULAR: .5 Deg.

Q.A

MFG

APPV'D

CHK'D

DRAWN

SERIAL #:

E

410RS-PT-RH-STM 48X88X96JAN 18 24

WEIGHT:

Gulshan

NOTE: ALL LAYOUT DRAWINGS ARE - GOOD SIDE OF MATERIAL UP INVENTORY #:

87654321

S1-410RS-083

CUSTOMER:

PROJECT:

VA JESSE BROWN

VA JESSE BROWN

E:\Solidworks Washers\6000 WASHERS\6000- JOB FILES\3654- JESSE BROWN-410RS-PT-RH-LP-STM\3654- 48x88x96-PT-RH-LP-STM

101460

JOB # 3654 - 48x88x96-PT-RH-STM- 480V- SALES.pdf
Sheet1
Drawing View2
JOB # 3654 - 48x88x96-PT-RH-STM- 480V- ENG..pdf
Sheet1
Drawing View6
Drawing View7
Drawing View9
Drawing View10
Drawing View11

File details come from the government source that posted it. Updated .