This is a Minimum Specification document for an Ion Chromatograph (IC) system to be delivered to the U.S. Naval Ship Repair Facility & JRMC Yokosuka, Japan.
The required Ion Chromatograph system must meet specific physical and operational parameters, including maximum external dimensions of 550mm width × 1,000mm depth × 700mm height with a weight not exceeding 50 kg. The system shall operate in ambient temperatures of 15-35°C with 40-80% relative humidity and require AC power supply of 100-240 volts at 50/60 Hertz. The instrument must feature a gradient elution method using hydroxide or carbonate eluents, a dual plunger pump system with flow rates of 0.1-2.0 ml/minute and minimum pressure resistance of 35 MPa, vacuum degassing, an electric conductivity detector with measuring range of 0-5,000µS/cm, and either a gel-type or electronically regenerated suppressor. The column oven must maintain temperatures between 25-40°C with ±0.5°C accuracy. The automatic sampler shall accommodate 50 vials minimum of 1.5 ml volume with automatic dilution and variable loop injection capabilities. The system must analyze fluoride, chloride, and bromide ions with seven consecutive analyses at 0.1 mg/L achieving relative standard deviation less than 5%.
The contractor shall install the system on-site, conduct operational and performance tests witnessed by SRF-JRMC representatives, and provide on-site training in Japanese covering mechanism, operating procedures, maintenance, and troubleshooting. A complete workstation with specified hardware (Intel Core i5 CPU, 8GB memory, 1TB HDD/SSD, Full HD monitor, Windows 11 Pro) and OEM-recommended software in English or Japanese shall be included. Two sets of technical manuals (English and Japanese versions) with comprehensive documentation must be provided. The warranty period is one year from installation with a minimum 7 business days response time for defective parts, which will be replaced free of charge. The system must utilize standard, factory-supported configurations without post-manufacture, third-party, or undocumented custom hardware modifications. Delivery is required to U.S. Naval Base Yokosuka Japan, Building A-61, Receiving Warehouse.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Minimum Specification Revison-1
SDN:
ERP PR:
JML Document Number:
1. Minimum Specification of Ion Chromatograph system.
a. External dimensions and weight:
(1) Width = 550 mm max.
(2) Depth = 1,000 mm max
(3) Height = 700 mm max.
(4) Weight = 50 kg max.
Note: The stated dimensions apply only to the main unit. If auto sampler is physically separated from the main unit, the stated dimensions apply to autosampler.
b. Environmental Resistance:
(1) Ambient temperature: 15-35ºC
(2) Relative humidity: 40-80% (no condensation)
c. Power Supply: AC100 - 240 volts, 50/60 Hertz,
d. Gradient Elution Method :
(1) Eluent type: Hydroxide or Carbonate
e. Pump Unit for Feeding Elution:
(1) Dual plunger system
(2) Flow rate range, minimum: 0.1 – 2.0 ml/minute
(3) Pressure resistance, minimum: 35 MPa
(4) Flow accuracy: ±2%
(5) Flow precision: ±0.2%
f. Vacuum degassing system
g. Detector:
(1) Electric conductivity detector
(2) Measuring range, minimum = 0 – 5,000µS/cm
(3) Volume of cell, minimum = 1 µL
(4) Noise < 0.5 nS/cm
h. Suppressor
Either (1) or (2) below:
(1) Gel type suppressor
(2) Electronically regenerated suppressor
i. Column oven:
(1) Range of temperature control, minimum: 25 - 40ºC
(2) Accuracy of temperature setup: ±0.5ºC
(3) Precision or stability of temperature control: ±0.2°C
j. Automatic Sampler:
(1) Number of samples being set: 50 vials minimum of 1.5 ml volume
(2) Automatic dilution function
(3) Loop injection at variable volume
k. IC columns for anion analysis recommended by the OEM:
(1) One (1) analytical column
(2) One (1) guard column
l. Workstation: Hardware and software (English or Japanese version) recommended by the OEM for operating the IC system, processing the test data and information from the IC analyzer and outputting the test data and information by displaying on a liquid crystal monitor and also printing out by a color printer in
English or Japanese language.
PC minimum spec
CPU: intel Core i5
Memory: 8GB
HDD (SSD is acceptable): 1TB
Optical Drive: DVD-ROM
Monitor: Full HD (1920*1080)
OS Windows 11 Pro
m. Technical Manuals:
(1) Two (2) sets of technical manuals, English version and Japanese one. The English version may be a
CD/DVD manual.
(2) Includes the followings as minimum information:
(a) Safety guidance, information, warnings and/or cautions
(b) Equipment specifications, capabilities and/or functions
(c) Structure and component names
(d) Operations
(e) Maintenance and troubleshooting
n. Target anion: Floride, Chloride, Bromide ion
(1) Instrument shall analyze fluoride, chloride, bromide ion, (7 consecutive analysis at 0.1 mg/L RSD
< 5%)
2.
Operational and Performance Tests as Receipt Inspection. The contractor shall install the instrument on-site and demonstrate the IC system performances and capabilities conformance to the requirements in this minimum specification and the OEM’s specifications. The contractor shall test samples for the performance verification.
SRF-JRMC Code 134 may provide test samples. The operational and performance tests shall be witnessed by the
SRF-JRMC representatives.
3. On-site and Off-site Trainings. The contractor shall provide on-site trainings on mechanism, operating procedure, maintenance, trouble shooting and other necessary information for SRF-JRMC Code 134 personnel in Japanese along with the installation of IC system.
4. To ensure long-term maintainability and service continuity, the requested analytical instrument must meet all specified analytical and performance requirements utilizing a standard, factory-supported configuration.
Proposals requiring post-manufacture, third-party, or undocumented custom hardware modifications to achieve these capabilities will not be accepted.
5. Warranty
Warranty period: 1 year from the installation data.
Minimum response time: 7 business days
Preventive maintenance visits: None
Spare parts: None
Post warranty support: None
In the event of defective devices, any parts in the IC system will be replaced free of charge during warranty