High_Speed_Tensile_Tester_Questions_and_Answers6-6-17.docx
DOCX document 12 KB Posted
- Attached to
- High Speed Tensile Tester Solicitation
- Solicitation number
- FA8224-17-R-0059
About this file
Questions from industry and answers from Government
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 8Mar2017-High_Speed_Tensile_Tester_(HSTT)_PS(Rev_5_8Mar2017).docx | DOCX document | |
| High_Speed_Tensile_Tester_Questions_and_Answers.docx | DOCX document | |
| HSTT_Responses_to_Industry_Questions.docx | DOCX document | |
| HSTT_A008_81446A_Computer_Manual.pdf | ||
| HSTT_A011_81000E_Drawings.pdf | ||
| 8Mar2017-High_Speed_Tensile_Tester_(HSTT)_PS(Rev_5_8Mar2017).docx | DOCX document | |
| HSTT_A009_81443A_Computer_Software.pdf | ||
| HSTT_A004_80033A_Installation.pdf | ||
| HSTT_A005_81646b_Configuration.pdf | ||
| HSTT_A006_80283B_OPTS_and_ATR.pdf | ||
| HSTT_A007_80527C_COTS_Manual.pdf | ||
| Form_-_Blank_1423_for_(CMRS)HSTT.pdf | ||
| Form_-_Blank_1423_(Cal_Cert)HSTT.pdf | ||
| HSTT_A010_81626_Safety.pdf | ||
| Solicitation.pdf | ||
| TM-86-01N_(T)_HSTT.docx | DOCX document | |
| Form_Blank_1423_(Manuals)HSTT.pdf | ||
| 8Mar2017-High_Speed_Tensile_Tester_(HSTT)_PS(Rev_5_8Mar2017).docx | DOCX document | |
| TM-86-01N_(T)_HSTT.docx | DOCX document |
Show all 19
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
High Speed Tensile Tester Questions and Answers
1. In 3.2 (Load Train)(Hydraulic Lifts and Locks), states that the pressure vessel must be raised and locked hydraulically. Does this mean the locking mechanism must be hydraulic? Or only the raising/lowering mechanism?
A: The locking mechanism does not need to be hydraulic.
2. In 3.2, the laser extensometer data rate is specified at 1 KHz. Common commercially available systems that are rated to ASTM E82, Class B1, are typically 100Hz or lower update rate. Is this specification based on a known commercially available instrument, or based on the testing requirements?
A: 100 Hz is not fast enough for the testing, 1kHz is ideal but 200Hz would be the minimum acceptable. Fiedler Optoelektronik GmbH manufactures a laser extensometer capable of 200Hz; others should be able to do the same.