Calibration_Certificate.pdf

PDF 10 KB Posted

Attached to
Repair of Shaft Failure Indicator
Solicitation number
FA8119-15-R-0001
Issued by

About this file

Calibration Certificate

View the file

Other files

File Type Posted
SFI_J&A_Redacted.pdf PDF
FA811915R0001_amendment_0002.pdf PDF
MEETINGS_AND_CONFERENCES_2.pdf PDF
CONTRACTORS_CORRECTIVE_ACTION_PLAN.pdf PDF
Engineering_Change_Proposal.pdf PDF
FA811915R0001.pdf PDF
CDRL_Report.pdf PDF
Package_Requirement_Report.pdf PDF
CONFERENCE_MINUTES.pdf PDF
MEETINGS_AND_CONFERENCE_1.pdf PDF
END_ITEM_PROD_AND_COMP_ITEM_CONSUM.pdf PDF
Wage_Determination.pdf PDF
Transport_Data_Report.pdf PDF
Counterfiet_Prevention_Plan.pdf PDF
SOW_CAVAF_CFM.pdf PDF
Status_Report.pdf PDF
Depot_Maint_Production_Report_DI-ALSS-80728A.pdf PDF
PWS_29DEC2014.pdf PDF
CAV_AF_DID_81634B 1 .pdf PDF
Show all 19

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DATA ITEM DESCRIPTION

Title: CALIBRATION CERTIFICATE

Number: DI-QCIC-80798B Approval Date: 20050328 AMSC Number: F7554 Limitation:

DTIC Applicable: GIDEP Applicable:

Preparing Activity: AFMETCAL DET 1/MLE Applicable Forms: None Use, Relationships:

This Data Item Description (DID) contains the format and content requirements for calibration reports or certificates accompanying calibrations performed on DoD test, measurement, and diagnostic equipment.

This DID is applicable to contracts requiring calibration of test equipment prior to delivery. This DID is consistent with ISO 17025 requirements for calibration reports or certificate.

This DID supersedes DI-QCIC-80798A.

Requirements:

1. Format. Contractor format is acceptable.

2. Content. The calibration certificate or report shall include:

a. title (e.g., calibration report or calibration certificate);

b. the name and address of the laboratory, and the location where the calibrations were carried out, if different from the address of the laboratory;

c. unique identification of the test report or calibration certificate (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the test report or calibration certificate;

d. the name and address of the client or customer;

e. identification of the method used;

f. a description of, the condition of, and unambiguous identification of the item(s) tested or calibrated;

g. the date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration;

DI-QCIC-80798B

h. reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results;

i. the test or calibration results with, where appropriate, the units of measurement;

j. the name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the test report or calibration certificate;

k. where relevant, a statement to the effect that the results relate only to the items tested or calibrated;

l. the conditions (e.g. environmental) under which the calibrations were made that have an influence on the measurement results;

m. the uncertainty of measurement or a statement of compliance with an identified metrological specification or clauses thereof;

n. evidence that the measurements are traceable.

End of DI-QCIC-80798B