The federal contract opportunity is a solicitation (Solicitation Number: 36C26125R0014) issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 21 to replace an inpatient pharmacy robot at the VA Palo Alto Healthcare System facility in Palo Alto, California. The project will prepare the site for installation of a new inpatient pharmacy robot to replace an outdated model currently in use. This is an SDVOSB set-aside solicitation. The contractor must follow specific requirements during construction to minimize impact on pharmacy operations, including interim life safety plans, infection control measures, and utility shutdown/hot work coordination. The contractor must also comply with all applicable federal and local laws, standards, and codes.

View the file

Other files

Show all 24

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Omnicell Legacy Product Equipment Specifications

Rev BA | 67-2031 Rev-BA

About Omnicell

Omnicell is a leader in medication and supply dispensing automation, central pharmacy automation and IV robotics, analytics software, and medication adherence solutions. Hospitals, post-acute care sites, and pharmacies worldwide rely on Omnicell products to increase patient safety, streamline workflow, and more effectively address drug diversion and regulations.

This guide and accompanying software and/or hardware described in it are protected under copyright laws and may not be reproduced, adapted, or translated, wholly or in part, without the express written consent of Omnicell, Inc. The same proprietary and copyright notices must be attached to any permitted copies as were attached to the original documents.

Omnicell, Inc.

590 E. Middlefield Road Mountain View, CA 94043

(650) 251-6100 www.omnicell.com

The following terms are trademarks of Omnicell, Inc. in the United States, other countries, or both.

Omnicell ivFlex Designer Pharmacy Line

OmniLinkRx AcuDose-Rx Enterprise Medication Manager

MedCarousel Savvy vSuite

PACMED XT Anesthesia Workstation Connect-Rx

PakPlus-Rx Anesthesia-Rx Autobond

ROBOT-Rx Anywhere RN Autogen

PROmanager-Rx SinglePointe Gemini

NarcStation SafetyStock MTS-350

WorkflowRx FastEntry MTS-500

MedShelf-Rx FlexBin Sureseal

Fulfill-Rx OptiFlex PillVue i.v.STATION Pandora SureMed i.v.STATION ONCO PandoraVIA Time My Meds i.v.SOFT Performance Center

All other trademarks and trade names are the property of their respective owners.

© 2008-2019 Omnicell, Inc. All rights reserved.

http://www.omnicell.com/

Technical Guides | iii

© 2008-2019 Omnicell, Inc. | Product Equipment Specifications Rev BA | 67-2031 BA Periodically, check myomnicell.com for updates to this material.

Contents

Introduction

Overview Safety/Quality Certifications Omnicell Hardware Architecture Omnicell Hardware Requirements

G4 Cabinet Specifications Electrical Requirements

G4 Electrical Requirements G4 Battery Backup Customer Supplied UPS Protection G4 Radiated Emission Level Testing

Communication Requirements G4 Network Requirements G4 Touch Remote Access Color Touch Cabinet Web Access

Environment G4 Transport/Storage Conditions G4 Temperature Requirements G4 Heat Dissipation G4 Cabinet Space Requirements G4 Cabinet Module Specifications Omnicell Pharmacy Drawer Module Specifications

G4 Console Specifications G5 Console Specifications G4 Auxiliary Box Specifications G4 Wireless Specifications Refrigeration Options G4 Printers

Medication Label Printer Specifications Codonics Specifications SATO and HP LaserJet Printer Part Numbers SATO CL408NX Printer Specifications

FlexLock 2.0

Central and Satellite Pharmacy Manager Specifications

Savvy Mobile Medication Workstation Specifications Savvy External Battery Charger Specifications

System Logistics Carousel Specifications System Logistics Large Carousel Dimensions https://myomnicell.com/

Technical Guides | iv

System Logistics Special 20-Carrier Carousel Dimensions System Logistics Small Footprint Carousel Dimensions

WorkflowRx Specifications WorkflowRx Hardware Specifications WorkflowRx Electrical Requirements

Carousel Computer Electrical Requirements Packager Electrical Requirements

Setting Up OmniCenter for HTTPS Connectivity WorkflowRx Environment

WorkflowRx Heat Dissipation WorkflowRx Seismic Restraints

WorkflowRx Safety/Quality Certifications WorkflowRx Carousel and Packager Specifications

Yuyama Packager Specifications Carousel Peripherals and Options Packager Options

OmniLinkRx Specifications Hardware Specifications

OmniLinkRx Servers OmniLinkRx Disk Sizing Tool OmniLinkRx Workstation Requirements OmniLinkRx Faxes

OmniLinkRx Software Requirements

OptiFlex Specifications OptiFlex Electrical Requirements OptiFlex Cabinet Specifications OptiFlex Communication Requirements

OptiFlex Network Requirements OptiFlex Remote Access

OptiFlex Environment Specifications OptiFlex Cabinet Transport/Storage Conditions OptiFlex Heat Dissipation

OptiFlex Safety/Quality Certifications OptiFlex Computer Equipment Minimum Specifications

OptiFlex Platform Matrix OptiFlex Inventory Control Module Computers OptiFlex Point of Use Computers OptiFlex Client Hardware/Software

OptiFlex Servers Supply Modules OptiFlex Scanner Mount

SecureVault Specifications SecureVault Server Configuration

Technical Guides | v

Anywhere RN Specifications Anywhere RN System Requirements

Anywhere RN OmniCenter/Server Minimum Requirement Anywhere RN Color Touch Cabinet Minimum Requirements Anywhere RN Remote Workstation Network Requirements Anywhere RN Interfaces Anywhere RN Remote Workstation Requirements Anywhere RN Mobile Cart Requirements

Anywhere RN Security Notes Anywhere RN Network Notes

Controlled Substance Manager (CSM) Specifications CSM Electrical Requirements CSM Communication Requirements CSM Network Requirements CSM Environment Specifications

CSM Transport/Storage Conditions CSM Heat Dissipation

CSM Safety/Quality Certifications CSM Hardware Specifications CSM Cabinet Specifications CSM Supply Modules and Peripherals

Scanner Read Height

Appendix B: G4 Seismic Specifications

Cybernet S19 Workstation Specifications

Appendix D: G4 Revit Drawings

Appendix E: G3/G4 Spare Keys

Revision History

Technical Guides | vi

Technical Guides | 7

Introduction

This document provides specifications for the Omnicell legacy products.

Overview

This document provides specifications for Omnicell products, including automated dispensing cabinets and anesthesia workstation along with specifications on the G5 console upgrade.

Safety/Quality Certifications Omnicell cabinets are certified according to the Canadian Standards Association (CSA) and Conformity European (CE) mark standards.

• Omnicell cabinets meet fire protection standards with their CSA and CE certification.

• Omnicell is ISO 9001:2000 compliant.

• The Anesthesia Workstation (AWS) and Supply Mobile Cart (55CTMC) are CSA and CE certified medical electrical equipment/systems, with a medical grade power supply.

• The G4 products are CSA and CE certified medical electrical equipment/systems, with a medical grade power supply.

SEISMIC SPECIFICATION

Go to the following link for the OSHPD pre-approval list:

http://www.oshpd.ca.gov/FDD/Pre-Approval/OPAIndex.pdf

Omnicell Hardware Architecture Omnicell solutions are designed for use in the materials management department, nursing units, pharmacy, and specialty areas.

• Omnicell cabinets provide the convenience and security of a closed system, while also supporting management of external supplies.

• Omnicell pharmacy drawer modules can be installed along with supply zones, to create mixed-use cabinets. This unique option gives clinicians the convenience of a single point of access to both supplies and medications.

• Omnicell carousels and packagers serve in the pharmacy department. Carousels store incoming medications from vendors. Packagers prepare dose packets for delivery to cabinets.

• Mobile carts provide convenience and flexibility for storage and access of supplies and medications. They can be used for bedside applications.

http://www.oshpd.ca.gov/FDD/Pre-Approval/OPAIndex.pdf

Technical Guides | 8

Omnicell Hardware Requirements Several site requirements must be met in order for Omnicell hardware to be installed. These requirement types are listed in each product section if applicable, including:

• Electrical

• Communication

• Environment

• Safety certification

• Hardware parameters

Technical Guides | 9

G4 Cabinet Specifications

This section covers the cabinet types. Optional cabinet modules and pharmacy drawer modules are included.

The cabinet types include:

• One-, two-, three-cell cabinets

• Half-cell cabinet

• Mobile supply cart

• OmniRx

• OmniTT, Anesthesia TT

• Anesthesia Workstation

• Savvy Mobile Medication Workstation

Auxiliary cabinets are available for one-cell, two-cell, three-cell, half-cell, OmniRx, and OmniTT products. Auxiliary cabinets are connected to the main cabinet and do not have their own user interface.

Electrical Requirements In addition to general requirements, this topic covers UPS and radiated emission level testing requirements.

G4 Electrical Requirements Omnicell requires the facility to provide electrical service according to Omnicell equipment specifications. The facility is responsible for all electrical and cabling costs.

• One 115V AC outlet for each cabinet control unit and auxiliary unit in North America. For all other countries, consult the local electrical authority.

• All units must be within six feet of an electrical outlet.

Product Operating Voltage (VAC) Frequency Operating

Current Power

Consumption Fuse Phase

Anesthesia Workstation, Anesthesia TT

90 - 260 AR

Medical grade 50-60 Hz

1.25 A at 115

V 0.75 A at 230 V

200 W 6.3 A Single

Technical Guides | 10

Product Operating Voltage (VAC) Frequency Operating

Current Power

Consumption Fuse Phase

FlexLock with TempCheck FlexLock is powered by the cabinet, and no separate power is required Single

Half-cell, OmniRx, Omni TT

90 - 260 AR (Main)

90 - 260 AR

(auxiliary) Medical grade

50-60 Hz

1.25 A at 115

V 0.75 A at 230 V

200 W 6.3 A Single

One-, two- three-cell cabinets

85 - 264 AR

(auxiliary) 90 - 250 AR (Main) G4 one-two-cell Medical grade

50-60 Hz 2 A at 115 V - 230 V

150 W (Aux) 200 W (Main)

6.3 A Single

Power and data connections are made to frames at the back right of the computer. The 1-, 2-, and 3-cell computers are 51-55" (129.54 - 139.7 cm) from the floor. Half-cell, OmniRx, and Anesthesia Workstation computers are 44-50" (111.76 -127 cm) from the floor. The OmniRx TT computer in its frame is 15" (38.1 cm) from the surface the TT sits

on. Omnicell power cords are 15 feet (4.57 m) in length.

Cabinets contain a battery backup feature that can supply power for a limited amount of time.

The healthcare facility should have its own emergency power source. Omnicell recommends that emergency power be provided to each unit with connections to red plugs.

G4 Battery Backup To prevent power problems from damaging the computer, cabinets have a built-in battery backup with surge-suppression circuitry. The battery backup can provide the computer with up to 20 minutes of power when voltage levels become erratic or fail.

The facility is responsible for testing and changing the battery..

The battery must be connected during installation because Omnicell is required to disconnect battery cables prior to shipping. When connecting an OmniCenter server to a UPS, do not plug the printer or computer monitor into the UPS.

This table describes the battery backup parameters for the G4 platform.

Technical Guides | 11

Table 1: G4 Platform Internal battery backup Specifications

Parameter Specification

Voltage range 90-250VAC

Frequency 50-60Hz

Average current 1 A

Peak current 6A

Fuse 6.3A/250V Time-lag

Surge ratings EN61000-4-5 Level 3

Battery recharge rate 6 hours

Battery back-up time (typical): half load/full load 40 minutes/20 minutes

Full time EMI/RF filtering Yes

Customer Supplied UPS Protection If customers want to have added runtime in case of a power failure, Omnicell can recommend an external UPS for usage with one- two- three-cell cabinets, half-cell, OmniRx, OmniTT, Anesthesia Workstation, and Anesthesia TT.

The amount of additional time given by the UPS will vary depending on the Omnisupplier for which the UPS is being used. The facility is responsible for testing and changing the battery in UPS units.

This table describes the external UPS parameters.

Table 2: External UPS Specifications

Parameter Specification

Voltage range 91-145 VAC

Frequency 60 Hz

Peak current 10.1A @ 100V 9.3A/110V 8.6A/120V

Power protection Circuit breaker

Battery recharge rate 4-8 hours

Battery back-up time (typical):

Approximately two hours with Anesthesia Workstation (Varies depending on OmniSupplier model and installed options.)

Energy rating 1800 joules

Technical Guides | 12

G4 Radiated Emission Level Testing Omnicell tests its products for EMI, RFI and susceptibility, and conducted emissions. Cabinets meet Class B radiated emission levels, except for Savvy which is a Class A product.

Individual test reports showing test performed and results are in Agile under 62-92xx.

Products are compliant with the following tests:

• VCCI V-3(2010-04)

• FCC §15.307 Conducted Emissions

• FCC §15.305 Radiated Emissions

• Electrostatic Discharge (EN 61000-4-2)

• Electrical Fast Transients (EN 61000-4-4)

• Radiated RF Immunity (EN 61000-4-3)

• Conducted RF Immunity (EN 61000-4-6)

• Power Frequency Magnetic Fields (EN 61000-4-8)

• Surges EN (61000-4-5)

• EN 60601-1-2 36.202, Immunity

• Voltage Dips And Interruptions (EN 61000-4-11)

• CISPR 11 §5.1 Conducted Emissions

• CISPR 11 §5.2 Radiated Emissions

• CISPR 11 Standard, Class B or Class A limits

• ESD (EN 61000-4-2) Air Discharge = +/-8KV, Contact Discharge = +/-6KV

Communication Requirements Communication considerations include network, remote access, OmniCenter Remote Access (OCRA), backup, archives, cabinet web access, and wireless specifications.

G4 Network Requirements The Omnicell equipment resides on the hospital’s network infrastructure, which is responsible for the performance and security of the system.

The equipment should be set up behind the facility’s firewall. If possible, the equipment should be in an isolated local area network segment or LANS. However, Omnicell applications do have their own security features. They also support network connections between the cabinets and OmniCenter.

Technical Guides | 13

To ensure that the Omnicell system can interface properly with the facility network, the facility must provide:

• LAN or WAN with 100 Mbps or higher is preferred. 10 Mbps is the minimum requirement.

• 802.11 a/b/g WLAN infrastructure and configuration for units requiring this option (Anesthesia Workstation, OmniSuppliers, and OmniRx units)

• Category 5 or higher cabling from the switch to the jack for applicable units

• One network port for each cabinet and at least one for every OmniCenter or Interface Server (NIC bridging or teaming are options on current server lines).

• One static IP address, subnet mask, and gateway address for each cabinet and at least one for each server.

• DNS or WINS server addresses for name resolution (optional).

Omnicell does not support Network Access Translation (NAT) IP configurations, DHCP or DHCP reservation. Only static IP addressing is supported.

G4 Touch Remote Access This topic covers communication protocols, options, and requirements for remote access.

COMMUNICATION PROTOCOLS

Omnicell primarily operates over TCP/IP and uses Health Level 7 (HL7) v.2.5 interface standards.

Socket, FTP and MSMQ protocols are also used.

OPTIONS

• VSuite solution is preferred.

• Virtual private network (VPN) solutions are used only if vSuite is not supported

• T1 for service internet connection is recommended

• Telephone line with direct inward dial (DID) is seldom used

REQUIREMENTS

• Web access on Port 443

• Port 80 for access to cabinets and OmniCenter web, prefer hardwired

• 128 encryption SSL (only for vDirector if used)

• One line per cabinet (if phone used)

• One to four lines per OmniCenter for remote diagnostics (if phone used)

Technical Guides | 14

Color Touch Cabinet Web Access Cabinet web access is an optional feature. The cabinet computer can have an embedded version of Windows XP.

OmniExplorer is an optional web extension of OmniCenter, Port 80, and MSMQ. Web access uses a controlled embedded browser.

Environment These general environment requirements must be met. Additional information is provided in separate topics.

• Adequate ventilation: 4" (10.16 cm) clearance around each Omnicell cabinet (from the wall and other cabinets)

• Adequate air flow

• A floor plan showing the location of where each cabinet is placed

G4 Transport/Storage Conditions The maximum cabinet humidity exposure should not exceed Thursday80% (without condensation). It is the customer’s responsibility to maintain room temperature at or below 71°F (22° C).

See the recommended cabinet temperature ranges by item type in the table below. Document areas in the cabinet that are +6°F of the ambient temperature and note not to store sensitive medication in those areas.

Cabinet Degrees Celsius Degrees Fahrenheit

Medications 20-25° 68-77°

General Supplies (with brief deviations) 15-30° 59-86°

G4 Temperature Requirements The operating temperature for most Omnicell cabinets is 5 to 35° C. The storage temperature is from -40 to +85° C.

G4 Heat Dissipation This table contains useful information for determining a site’s heating and ventilation setup so that room temperatures can be maintained with the cabinets installed.

Technical Guides | 15

Table 3: G4 HVAC Table Heat Dissipation Advisories

Cabinet Heat Dissipation

1-, 2-, and 3-cell 200-600 BTU/hr

Half-cell, OmniRx, OmniTT, Anesthesia TT 200-400 BTU/hr

Anesthesia Workstation 200-400 BTU/hr

Supply Mobile Cart 304-443 BTU/hr

G4 Cabinet Space Requirements Cabinet storage capacity is directly proportionate to the square footage of floor space required for each column. Clearance required for pharmacy drawers is equal to clearance required for supply doors. This topic contains standard cabinet specifications. It includes line item capacity and dimensions.

Dimensions may vary slightly in size, due to manufacturing variances. Configurator dimensions may be slightly different to allow for variations.

Width includes screw heads on sides of cabinet. Clearance should be added to facilitate moving the cabinets in and out from the wall. Clearance is also needed for modules mounted on the side of the unit—such as the external return bin (ERB) or SafetyStock scanner. Depth includes the door or drawer handle. Clearances are provided when the door/drawer is open. User access may require additional clearance.

Allow 4" (10.16 cm) clearance around each Omnicell cabinet (from the wall and other cabinets) for adequate ventilation.

Some cabinets are used by Omnicell Supply Specialty and SecureVault. Their part numbers are identified as OPTIxxx and xxxxxSV respectively.

The images in this topic display sample cabinet configurations. Actual configurations and modules may be customized by the hospital. Cabinet weight will vary depending on the configuration.

Estimated maximum weight for a cabinet shelf is 30 lb (13.61 kg).

CABINET WIDTH WITH EXTERNAL RETURN BIN

If an external return bin (ERB) is attached to a cabinet, extend the width by both the depth of the bin (8.25”/20.96 cm) and an additional 6.6” (16.76 cm) to allow the ERB door to open fully

Technical Guides | 16

(14.85”/37.72 cm total). This is the minimum space required to allow the bin to open fully. Add additional space as desired for convenient access.

The acrylic (see-through) panels do not provide UV resistance. This may affect what gets stored in the cabinets.

ONE-CELL COLOR TOUCH G4

Description Specifications Metric

Model #:

SDA-FRM-001, MDA-FRM-001, OPF-FRM-001

(medical electrical equipment)

Item capacity:

96 to 1744

(8 items per shelf, max. 12 shelves or 18 drawers/2 shelves)

W x H x D: 26.75” x 77.56” x 24.47” 69.1 x 196.90 x 62.5 cm

Min. width with ERB 41.60” 121.92 cm

Clearance: 48” depth to include door swing 105.67 cm

Screen height: 56” 142.24 cm

Screen size: 12.1” Color Touch 30.74 cm

Weight: 500 lb empty; varies by configuration 226.80 kg

Shelf (W x L) 22.75” x 21-22” 59.79 x 53.34-55.88 cm

Technical Guides | 17

Description Specifications Metric

(minimum spacing)

3.25” (between light bars) 8.26 cm

4.5” (between shelves) 11.43 cm

ONE-CELL AUXILIARY

Pharmacy modules can be added.

Model #: OX104, OPTIX1, OX104SV

Item capacity:

104 to 1752

(8 items per shelf, max. 13 shelves or 18 drawers/3 shelves)

W x H x D: 26.75” x 77.56” x 24.47” 69.1 x 196.90 x 62.5cm

Min. width with ERB 41.6” 105.67 cm

Clearance: 48” depth to include door swing 121.92 cm

Screen height: N/A

Weight: 500 lb empty; varies by configuration 226.80 kg

Shelf (W x L): 22.75” x 21-22” 59.79 x 53.34-55.88 cm

(minimum spacing)

Technical Guides | 18

TWO-CELL COLOR TOUCH G4

SDA-FRM-002, MDA-FRM-002, OPF-FRM-002

(medical electrical equipment)

Item capacity:

216 to 3128

(8 items per shelf, max. 27 shelves or 32 drawers/7 shelves)

W x H x D: 51.5” x 77.56” x 24.47” 131.9 x 196.90 x 62.5 cm

Min. width with ERB 66.35” 168.53 cm

Clearance: 48” depth to include door swing 121.92 cm

Space req.: 17.33 sq. ft. with door open 1.61 sq. m

Screen height: 56” 142.24 cm

Screen size: 12.1” Color Touch 30.74 cm

Weight: 750 lb empty; varies by configuration 340.19 kg

Shelf (W x L): 22.75” x 21-22” 59.79 x 53.34-55.88 cm

(minimum spacing)

Technical Guides | 19

TWO-CELL AUXILIARY

Model #: OX224, OPTIX2, OX224SV

Item capacity:

224 to 3136

(8 items per shelf, max. 28 shelves or 32 drawers/8 shelves)

W x H x D: 51.5” x 77.56” x 24.47” 131.9 x 196.90 x 62.5 cm

Min. width with ERB 66.35” 168.53 cm

Clearance: 48” depth to include door swing 121.92 cm

Screen height: N/A

Weight: 750 lb empty; varies by configuration 340.19 kg

Shelf (W x L): 22.75 x 21-22” 57.79 x 53.34-55.88 cm

(minimum spacing)

Technical Guides | 20

THREE-CELL COLOR TOUCH G4

SDA-FRM-003, MDA-FRM-003, OPF-FRM-003

(medical electrical equipment)

Item capacity:

336 to 5280

(8 items/shelf max. 42 shelves or 54 drawers/12 shelves)

W x H x D: 76.25” x 77.56” x 24.47” 193.675 x 196.99 x 62.15 cm

Min. width with ERB 91.1” 121.92 cm

Clearance: 48” depth to include door swing 231.4 cm

Screen height: 56” 142.24 cm

Screen size: 12.1” Color Touch 30.74 cm

Weight: 1000 lb empty; varies by configuration 453.59 kg

Shelf (W x L): 22.75” x 21-22” 57.79 x 53.34-55.88 cm

(minimum spacing)

Technical Guides | 21

THREE-CELL AUXILIARY

Model #: OX344, OPTIX3, OX344SV

Item capacity:

344 to 5288

(8 items per shelf, max. 43 shelves or 54 drawers/13 shelves)

W x H x D: 76.25” x 77.56” x 24.47” 193.675 x 196.99 x

62.15 cm

Min. width with ERB 91.1” 231.4 cm

Clearance: 48” depth to include door swing 121.92 cm

Screen height: N/A

Weight: 1000 lb empty; varies by configuration 453.59 kg

Shelf (W x L): 22.75” x 21-22” 57.79 x 53.34 x 55.88 cm

(minimum spacing)

Technical Guides | 22

HALF-CELL COLOR TOUCH G4

MDA-FRM-004

(medical electrical equipment)

Item capacity:

56 items

(8 per shelf, max. 7 shelves) or 80-120 items w/3-drawer carrier.

W x H x D: 27.9” x 57.7” x 26.6” with bumpers 70.9 x 146.6 x 67.6 cm

Clearance: 48” depth to include door swing 121.92 cm

Work surface height: 45.2” 114.81 cm

Screen size: 15” Color Touch 38.1 cm

Weight: 360 lb with med drawers and ERB; varies by configuration 234.96 kg

Shelf (W x L): 22.75” x 21-22” 57.79 x 53.34-55.88 cm

(minimum spacing)

Technical Guides | 23

HALF-CELL AUXILIARY

Model #: 56AXRX, 56AX

Item capacity: 56 (8 items per shelf, max. 7 shelves) or 80-120 w/3 drawer carrier

W x H x D: 27.9” x 45” x 26.6” 70.9 x 114.3 x 67.6 cm

Clearance: 48” depth to include door swing 121.92 cm

Work surface height: 45” 114.3 cm

Screen size: N/A

Weight: 360 lb with med drawers and ERB; varies by configuration 226.8 kg

Shelf (W x L): 22.75” x 21-22” 57.79 x 53.34-

55.88 cm

(minimum spacing)

Technical Guides | 24

CONTROLLED SUBSTANCE MANAGER (CSM)

Model # CSM-FRM-001, CSM-FRM-002, CSM-FRM-003

Item Capacity:

CSM-FRM-001: 104-1752; CSM-FRM-

002: 224-3136; CSM-FRM-003: 344-

W x H x D:

CSM-FRM-001: 26.75" x 77.56" x 24.47"; CSM-FRM-002: 51.5" x 77.56" x 24.47"; CSM-FRM-003: 76.25" x 77.56" x 24.47"

CSM-FRM-001: 69.1 x 196.9 x 62.5 cm;

CSM-FRM-002: 131.9 x 196.9 x 62.5 cm;

CSM-FRM-003: 193.68 x 196.99 x 62.15 cm

Clearance: CSM-FRM-001, 48" depth to include door swing; (all models) 121.92 cm

Minimum width with

ERB:

CSM-FRM-001: 41.6"; CSM-FRM-002:

66.35"; CSM-FRM-003: 91.1"

CSM-FRM-001: 121.92 cm; CSM-FRM-

002: 168.53; CSM-FRM-003:121.92

Weight:

CSM-FRM-001: 500 lb empty; CSM- FRM-002: 750 lb empty, CSM-FRM- 003: 1000 lb empty (varies by configuration)

CSM-FRM-001: 226.80 kg empty; CSM- FRM-002: 340.19 kg empty; CSM-FRM- 003: 453.59 kg empty (varies by configuration)

Minimum spacing:

3.25" (between light bars) 4.5" (between shelves) (all units)

8.26 cm (between light bars) 11.43 cm (between shelves) (all units)

Technical Guides | 25

OMNIRX COLOR TOUCH G4

MDA-FRM-005

(medical electrical equipment)

Item capacity:

13 drawers (typically 350-450 medications)

3 drawers (typically 80-120 medications)

W x H x D: 25.8” x 57.7” x 26.9” with bumpers 65.5 x 146.6 x 68.4 cm

Min. width with ERB 42.75” 111.76 cm

Clearance: 44” depth includes drawer extension 108.59 cm

Work surface height: 45.2” 114.81 cm

Screen size: 15” Color Touch 38.1 cm

Weight: 518 lb empty; varies by configuration 234.96 kg

OMNIRX AUXILIARY

Model #: RXAX (medical electrical equipment)

Technical Guides | 26

Description Specifications Metric

Item capacity:

13 drawers (typically 350-450 medications)

3 drawers (typically 80-120 medications)

W x H x D: 27.9” x 45” x 26.9” with bumpers 70.9 x 115.57 x 68.4 cm

Min. width with ERB 42.75” 108.59 cm

Clearance: 44” depth includes drawer extension 111.76 cm

Work surface height: 45” 114.3 cm

Screen size: N/A

Weight: 500 lb empty; varies by configuration 226.8 kg

Model #: TTCT, ANTT

OMNIRX TT/ANESTHESIA TT G4

MDA-FRM-006, ANT-FRM-002

(medical electrical equipment)

Item capacity: 3 drawers (typically 80-120 medications)

W x H x D: 26” x 27.5” x 23.75” (without bumper) 26.9" (with bumper)

66.04 x 69.85 x 60.33 (without bumper) 68.33 cm (with bumper)

Clearance: 44” depth includes drawer extension 111.76 cm

Work surface height: 15.7” 39.88 cm

Screen size: 15” Color Touch 38.1 cm

Weight: 133 lb without drawers 52.16 kg

Technical Guides | 27

ANESTHESIA WORKSTATION G4

ANT-FRM-001

(medical electrical equipment)

Item capacity: 13 drawers (typically 350-450 items)

W x H x D:

50” x 63.5" x 30" 127 x 161.3 x 76.2 cm

W=64" with user interface arm extended W=162.6 cm with arm extended

With ERB: Add 6.6” to width (W) 16.77 cm

With fold-out shelf: Add 15“ to width (W) 38.1 cm

With trash or Sharps Bin-Single Use: Add 7“ to width (W) 17.78 cm

Clearance: 46” depth includes drawer extension 116.84 cm

Work surface height: 43” 109.22 cm

Keyboard height: Variable: 31” to 48.5” 78.74 cm to 123.19 cm

Screen size: 15” Color Touch 38.1 cm

Weight: 560 lb empty; varies by configuration 254.01 kg

Technical Guides | 28

Description Specifications Metric

Trellis

Tape dispenser

Gooseneck light

Printer

Accessory mounting kit iPod compatible speakers

Sharps bin - Single use (13.75” x 16” x 7”) 34.93 x 40.64 x 17.78 cm

Anesthesia supply drawers

Power strip

Options:

Waste container (13.75” x 13.5” x 7”) 34.93 x 34.29 x 17.78 cm

Tilt bin organizer

IV Pole

Chart holder

External Return Bin (ERB) (13.25” x 13” x 8.25”) (add 6.6” min for door)

33.66 x 33.02 x 20.96 cm (16.77 cm)

Wireless (802.11 g)

Touch and Go/Magnetic Card Reader

Fold-out shelf (15” x 18”) 38.1 cm x 45.72 cm

Pharmacy drawers

50 ft. cable for ERB 15.24 m

Emergency breakaway panel

Technical Guides | 29

MOBILE CART - SAVVY

Model #: MMM-FRM-001

Item capacity: Up to 12 locking drawers

W x H x D:

22” x H (variable) x 22-32” 55.8 x (variable) x 55.8 - 81.28 cm

Width: 22" 55.8 cm

Height: 52” - 64” including monitor 132.08 - 162.56 cm

Depth: 32” with keyboard fully extended 81.28 cm

Work surface area: 17” x 12” 43.18 x 30.48 cm

Work surface height: 33” - 45” (adjustable) 83.82 - 114.30 cm

Base: 16” x 16” 40.64 x 40.64 cm

Weight: 100 - 165 lb 45.36 -74.85 kg

Drawer sizes 9” L x 4” W x 3” D 22.86 x 10.16 x 7.62 cm

9” L x 4” W x 6” D 22.86 x 10.16 x 15.24 cm

G4 Cabinet Module Specifications This section lists Omnicell cabinet module specifications. It includes line item capacity and dimensions.

Technical Guides | 30

9-DRAWER PHARMACY CARRIER

Model #: OSRXU

Item capacity: 9 drawers

W x H x D: 1 zone / 5 shelves

Weight: 200 lb (w/o drawers) 90.72 kg

3-DRAWER PHARMACY CARRIER

(shown in a cabinet and a half cell)

Model #: OS 3DRXU

Item capacity: 3 drawers

W x H x D: 1 zone (w/o drawers)

Weight: 100 lb 45.36 kg

Technical Guides | 31

OMNIDISPENSER

(shown with 3-drawer carrier)

Model #:

ODMPlus3 (includes 3-drawer pharmacy carrier);

ODMTall (includes half-door, up to 6 shelves);

ODMRx (placed in OmniRx)

Item capacity: Up to 45 small and 24 large cassette dispensers [all]

W x H x D: 1 zone / 3 shelves / 6 drawers [by model order]

Weight: 50 lb (w/o rails) [all] 22.68 kg

EXTERNAL RETURN BIN (ERB)

Model #: OERB1

Item capacity: N/A

W x H x D: 13.25” x 13” x 8.25” 33.66 x 33.02 x 20.96 cm

Clearance: 14.85” minimum depth with door open. 37.72 cm

Weight: 17 lb 7 kg

Placement: On wall or side of cabinet

Technical Guides | 32

No minimum height for placement. The ERB can also be mounted on half-cell modules;

the clearance requirement is the same as on 1-, 2, and 3-cell modules. The part # for the Installation Kit is 20-6100

FLEXLOCK (INSTALLED ON REFRIGERATOR)

Model #: 1.0 OCFLTCK

Item capacity: N/A

W x H x D: 5.5" x 5.5" x 12.4" 14.0 x 14.0 x 31.5 cm

Clearance: Refrigerator specs vary; allow 2” side clearance for key access and for 2" rear access for cable.

add 5.08 cm; side and rear clearance

Weight: 8 lb 3.6 kg

OMNITRACK

Model #: OT32

Item capacity: 32

W x H x D: 1 shelf position

Technical Guides | 33

Weight: 4 lb 1.8 kg

PULL-OUT SHELF

Model #: PS1

Item capacity: 8

W x H x D: 1 shelf position (22.75” x 23” x 4.63”) 57.79 x 58.42 x 11.76 cm

Weight: 19 lb 8.6 kg

SUPPLY DRAWER

Model #: OSD24

# of bins: 24 maximum

Bin W x H x D: 2” x 4.25” x 3.75" 5.08 x 10.8 x 9.53 cm

Drawer W x H x D: 22” x 21.5” x 4” (1 shelf position) 55.88 x 54.61 x 10.16 cm

Weight: 30 lb 13.6 kg

Technical Guides | 34

ANESTHESIA WORKSTATION (AWS) TILT BINS - 4 BINS

Model #: ANT-OPT-008

Dimensions: 23.63"W x 8.13" H x 6.63" D 60.02 x 20.65 x 16.84 cm

ANESTHESIA WORKSTATION (AWS) TILT BINS -5 BINS

Model #: ANT-OPT-009

Dimensions: 23.63 "W x 6.5" H x 5.25" D 60.02 x 16.51 x 13.34 cm

Technical Guides | 35

ANESTHESIA WORKSTATION (AWS) TILT BINS - 6 BINS

Model #: ANT-OPT-010

Dimensions: 23.63 "W x 4.5" H x 3.63" D 60.02 x 11.43 x 9.22 cm

CORDLESS BAR CODE SCANNER

Model #: Symbol P370

Range: 100 feet 30.48 m

Programming: Varies by application

Applications: OmniScanner, OptiFlex, WorkflowRx, and Implant Tracking

WIRELESS BAR CODE SCANNER

Technical Guides | 36

Description Specifications Metric

Model #: Symbol MT2070

Range: 300 feet 91.44 m

Programming: Varies by application

Applications: OmniScanner, OptiFlex, WorkflowRx, and Implant Tracking, CSM

Model #: Code Reader 2500

Range: 300 feet 91.44 m

Applications: Carts

Connection: Bluetooth 2.0 + EDR, Class1 Radio, (2.4GHz ~ 2.4835GHz frequency)

Model #: Code Reader 3500

Range: 300 feet 91.44 m

Applications: WorkflowRx

Connection: Bluetooth 2.0 + EDR, Class 1 Radio, (2.4GHz ~ 2.4835GHz frequency)

Technical Guides | 37

2D SCANNER

Description Specifications

Model #: Symbol DS6708 Imager

Applications: Item Scan, SafetyStock

TOUCH AND GO (BIO ID)

PC Box (left) and AWS housing (G3) and G4

Available for Color Touch cabinets; not available for Omnicell Supply Specialty. Refer to the Touch and Go Technical Guide (P/N 67-3014) for system requirements.

There are various housings for Touch and Go. The finger pad is used for Touch and Go.

There is also a slot for reading magnetic or bar code cards.

MAGNETIC AND BAR CODE CARD READERS

Housing for half-cell, OmniTT, OmniRx G3 on left, G4 on right

Omnicell Supply Specialty only uses magnetic card readers. Readers can have two read heights: 0.4 and 0.49 inches (1.02 and 1.24 cm).

Refer to the Omnicell/Omnicell Supply Specialty Card Reader Installation and Configuration Guide (P/N 67-2006) for system requirements.

Technical Guides | 38

The card reader uses the same housing as Touch and Go. These modules can be ordered separately or together. The electronics in the housing determine which modules are functional.

Omnicell Pharmacy Drawer Module Specifications Drawers fit into 1-, 2-, 3-cell cabinets, OmniRx, OmniTT, and in Anesthesia cabinets. Pharmacy carriers can also be used with 1-, 2-, 3-cell, and half-cell cabinets.

The combination of an Omnicell cabinet with Omnicell pharmacy modules allows storage and dispensing of supplies and drugs in unit dose forms of:

• Oral solids

• Oral liquids

• IV syringes

• Ampoules

• Vials

• Bulk stock

In the measurements (W x H x D), H is the height of the drawer or bin from top to bottom, and D is the distance from the front to the back of the drawer or bin.

High Security FlexBin Single-dose Drawer Specifications The FlexBin Single-dose Drawer provides a high level of security for medications. With this drawer, access to medications is restricted to the item and quantity chosen by the user. There is no need to confirm the remaining quantity or conduct end-of-shift counts, which results in fewer discrepancies.

Key features of the FlexBin single-dose drawer are

• Single-deep drawer

• Three bin sizes

• Bins can be used for single-dose

• Three drawer configurations: 96 small bins, 48 medium bins, and 48 small/large combo

• Capacity of up to 96 doses

FLEXBIN 96 SINGLE DOSE DRAWER CONFIGURATION

Specifications (metric)

Best Use Ampoules

Model #: MDA-DRW-003

Technical Guides | 39

# of Bins: 96

Bin W x H x D: 1 wide/ 2"x 1" x 1.4" (5.08 x 2.54 x 3.56 cm)

Drawer W x H x D: 21.5" x 2.5" x 22.5" (54.61 x 6.35 x 57.15 cm)

Weight: 22.5 lb [empty] (10.21 kg)

Comes with or without lights

FLEXBIN 3131 SINGLE DOSE

Best Use Syringes and ampoules

Model #: MDA-DRW-002

# of Bins: 48

Bin W x H x D: 3 wide/ 6.9" x 1" x 1.5" (17.53 x 2.54 x 3.81 cm)

Bin W x H x D: 1 wide/ 2" x 1" x 1.4" (5.08 x 2.54 x 3.56 cm)

Drawer W x H x D: 21.5" x 2.5" x 22.5" (54.61 x 6.35 x 57.15 cm)

Weight: 21.5 lb [empty] (9.75 kg)

FLEXBIN 2222 SINGLE DOSE

Best Use Vials

Model #: MDA-DRW-001

# of Bins: 48

Bin W x H x D: 4.4" x 1" x 1.5" (11.18 x 2.54 x 3.81 cm)

Drawer W x H x D: 21.5" x 2.5" x 22.5" (54.61 x 6.35 x 57.15 cm)

Weight: 21 lb [empty] (9.53 kg)

Technical Guides | 40

High Security Locking Bin Drawer Specifications Locking bins, suitable for controlled substances, provide a high level of security by restricting access to one pre-selected medication at a time. The tamper-proof metal lids of each bin and audio alerts provide additional security.

Omnicell guiding light technology guides users to the drawer and to the appropriate preselected bin within the drawer.

Key features of the locking bin drawers are

• Guiding light technology that provides both drawer location light and lights on individual product bins

• Single and double-deep drawers

• Three bin sizes and drawer configurations: 6 double-deep bins, 12 bins, or 24 bins

• Tamper-proof, locked metal lids

• Supports zone configuration for increased control over pharmacy items

• Capacity of up to 24 line items per drawer

6-BIN LOCKING

Specifications (metric)

Best Use High

Model # OLL6

# of Bins: 6

Bin W x H x D: 5.7" x 3.6"x 8.1" (14.45 x 9.14 x 20.57 cm)

Drawer W x H x D: 21.5" x 5.1" x 22.5" (54.61 x 12.95 x 57.15 cm)

Weight: 36 lb [empty] (16.33 kg)

8-BIN LOCKING

Best Use Introduced with G4, but supported on CT 14.6 and higher

Model #: MDA-DRW-004

# of Bins: 8

Technical Guides | 41

Specifications (metric)

Bin W x H x D: 4 at: 12.79" x 1.8" x 2.92" (32.49 x 4.57 x 7.42cm) and 4 at: 6.1"x 1.8" x 3.4" (15.5 x

4.57 x 8.64 cm)

Drawer W x H x D: 21.5" x 2.5" x 22.5" (54.61 x 6.35 x 57.15 cm)

Weight 31 lb [empty] (14.52 kg)

12-BIN LOCKING

Model #: OLL12

# of Bins: 12

Bin W x H x D: 6.1"x 1.8" x 3.4" (15.49 x 4.57 x 8.64 cm)

Drawer W x H x D: 21.5" x 2.5" x 22.5" (54.61 x 6.35 x 57.15 cm)

Weight: 31.5 lb [empty] (14.29 kg)

24-BIN LOCKING

Best Use Small quantities

Model #: OLL241

# of Bins: 24

Bin W x H x D: 2.7" x 1.8" x 3" (6.86 x 4.57 x 7.62 cm)

Weight: 32.5 lb [empty] (14.74 kg)

Low Security Lighted Matrix and Matrix Drawer Specifications Matrix drawers are available in two models: lighted and unlighted. The lighted matrix drawer provides lower security, with increased flexibility. Lights frame the section and guide the user to the preselected medication, reducing the possibility of errors.

Key features of lighted matrix drawers are:

Technical Guides | 42

• Guiding light technology that provides both drawer location light and lights on individual product bins

• Flexible drawer configurations up to 24 bins

• Adjustable bin dividers

• Single-deep drawer

• Capacity of up to 24 line items

The matrix drawer is a high-capacity, low-security unlighted drawer that is suitable for large quantities of less-controlled medications.

Key features of matrix drawers are:

• Flexible drawer configurations up to 96 bins

• Adjustable bin dividers

• Double-deep drawer 4 dividers

• Single-deep drawer capacity of up to 96 line items

24-BIN LIGHTED MATRIX

Model #: 0LMD24

# of Bins: 4-24

Bin W x H x D: 10" x 1.8" x 9.2" (max.) (25.4 x 4.57 x 23.37 cm)

Bin W x H x D: 4.9" x 1.8" x 2.9" (min.) (12.45 x 4.57 x 7.37 cm)

Weight: 20 lb [empty] (9.07 kg)

96-BIN MATRIX

Model #: OMD48

# of Bins: 4-96

Bin W x H x D: 9.2" (max.) x 1.8" x 10" (23.37 x 4.57 x 25.4 cm)

Bin W x H x D: 2.4" x 1.8" x 1.4" (min.) (6.1 x 4.57 x 3.56 cm)

Technical Guides | 43

Weight: 20 lb [empty] (9.07 kg)

SINGLE DEEP ANESTHESIA

Model #: OASD

# of Bins: Up to 22

Bin W x H x D: 20” x 1.7” x 18.5” [max] (50.8 x 4.32 x 46.99 cm)

Drawer W x H x D: 21.5” x 2.5” x 22.9” (54.61 x 6.35 x 58.17 cm)

Weight: 15.5 lb [empty] (7.03 kg)

DOUBLE DEEP ANESTHESIA

Model #: OASD2

# of Bins: Up to 13

Bin W x H x D: 19.9” x 4.1” x 17.9” [max] (50.55 x 10.41 x 45.47 cm)

Drawer W x H x D: 21.5 x 4.9”” x 22.9” (54.61 x 12.446 x 58.166 cm)

Weight: 25 lb [empty] (11.34 kg)

TRIPLE DEEP ANESTHESIA

Model #: OASD3

# of Bins: Up to 13

Bin W x H x D: 19.75” x 5.25” x 17.75” (maximum)

Drawer W x H x D: 21.5” x 7.75” x 22.9”

Weight: 42 lb [empty] (19.05 kg)

Supports weight of: 50 lb

Technical Guides | 44

G4 Console Specifications The G4 CT Console, Rx, and AWS allow direct interface with a cabinet via a touch screen, keyboard, and monitor. They also communicate with the OmniCenter.

Table 4: G4 Color Touch E-Box Specifications

Item Omnicell Configuration

Processor Intel® I3-540 CPU at 3.06GHz

Memory 2GB, DDR3-1333, PC3-10600, CL9, 240PIN, SODIMM

Hard Drive 500 GB, SATA 7200RPM, WD5000AAKX Enterprise

Operating System Microsoft Windows 7 Embedded (32-bit)

Display 12.1" LCD with LED backlight for automatic dispensing cabinet console; 15.0” LCD with LED backlight for OmniRx and AWS

Video Controller Integrated Graphics GMA HD

Ethernet One (1) RJ45 10/100/1000 Ethernet port (LAN1, Intel 82578DM) and one (1) RJ45 10/100/1000 Ethernet port (LAN2, Intel 82574L)

Keyboard Keyboard, USB, QWERTY

Mouse None

Floppy Disk Drive None

CD-ROM or DVD- ROM Drive None

Sound Single speaker

Printer Thermal receipt roll printer and optional label printer

UPS Built-in (approximately 20 minutes back-up)

PC box Weights

G4 PC Console: 80 lb (36.29 kg)

E-Box: 40 lb (18.14 kg)

OmniRx: 80 lb (36.29 kg)

AWS: 90 lb (40.82 kg)

Notes: Omnicell proprietary interfaces for cabinet-internal communication

Technical Guides | 45

Item Omnicell Configuration

Auxiliary proprietary communication ports for external cabinet communications

Proprietary peripheral ports for external return bin, FlexLock and other devices

Part Number for printer paper (1-, 2-, and 3-cell cabinets) 95-6015 (paper, thermal, 92 mm roll, 88 mm wide)

G5 Console Specifications G4 series system consoles can be upgraded G5 eBox and user interface items.

Table 5: G5 eBox Specifications

Item Specification

Processor Intel

Memory 4 GB

Storage Drive Solid state drive, 150 GB

Operating System Microsoft Windows 10, with TPM1.2 bit encryption

Display 12.1-in. capacitive touchscreen for automatic dispensing cabinet console;

15.0-in capacitive touchscreen for OmniRx and AWS.

Video Controller Intel Integrated Graphics

Ethernet 2 x RJ-45 ports. Intel Gigabit Ethernet Controllers

Keyboard Keyboard, USB, QWERTY

Mouse N/A

Floppy Disk Drive N/A

CD-ROM or DVD- ROM Drive N/A

Sound Single speaker

Printer

Tall cabinet and OmniRx: card reader (optional), fingerprint scanner, label printer and receipt printer.

AWS and OmniRx: card reader (optional), fingerprint scanner, receipt printer.

Technical Guides | 46

Item Specification

Hand scanner Zebra DS4801

UPS Lead-acid backup battery, built-in. Approximately 20 minutes back-up.

Wireless card (option) ASUS AC68, 802.11ac eBox Weight Approximately 40 lb (18.14 kg)

Notes:

Omnicell proprietary interfaces for cabinet-internal communication

Auxiliary proprietary communication ports for external cabinet communications

Proprietary peripheral ports for external return bin, FlexLock and other devices

G4 Auxiliary Box Specifications Cabinets with an auxiliary (aux) box are connected with 25-foot cables to a main cabinet with a CT PC box for communication to the OmniCenter. An aux box does not have direct interface capability.

Table 6: Aux Box Specifications

Item Omnicell Configuration (Dell part #)

Base System Provides power and communication line to OmniSupplier Aux

Output Voltage 12V and 5V

Max Amps 5A for 12V; 10A for 12V

Input Voltage 85-264 VAC

Operating Temperature 5°C to 35 °C

Storage Temperature -40 °C to 85 °C

Dimensions (H x W x D) 6.56” x 22.5” x 21.3” / 16.66cm x 57.15cm x 54.1cm

Feature Setting options to connect to external PC or main cabinet.

G4 Wireless Specifications Wireless communication between the cabinets and the OmniCenter is accomplished using Intel’s Centrino Advanced N-620 for Desktop adapter.

Technical Guides | 47

Omnicell also still supports the Asus PCE-N15 adapter. The following table lists the wireless specifications.

Table 7: G4 Platform Wireless Specifications

Specification Centrino Advanced-N 6205 for Desktop PCE-N15

Connector Type RP-SMA RP-SMA

Certificates UL, C-UL, CB (IEC 60950-1), FIPS, FISMA CE, FCC, C-Tick, IC

Authentication

WPA & WPA2, 802.1X (EAP-

TLS, TTLS, PEAP, LEAP, EAP-

FAST), EAP-SIM, EAP-AKA

WPA & WPA2

Authentication Protocols

PAP, CHAP, TLS, GTC, MS-

CHAP, MS-CHAPv2 None

Encryption 64-bit, 128-bit WEP, AES-

CCMP, TKIP

64-bit WEP, 128-bit WEP, AES-

CCMP, TKIP

Antenna (Form) PIFA (enclosed in plastic) External Dipole (2)

Dimensions H x W x D; Weight

50.85 x 69.53 x 6.37 mm;

40.68g 120.5 x 21.6 x 22 mm; 62.6g

Operating Humidity 50% to 90% non-condensing 5 to 95% non-condensing

Operating Temperature 0 to 80°C 0 to 45°C

Operating System Compatibility

Windows 7 (32-bit), Vista, XP, 8, Linux

Windows 7 (32-bit), Vista, XP, Linux

Manufacturer Intel Asus

Operating Frequency 2.4 GHz, 5 GHz 2.4 GHz

Network Compliant Standards IEEE 802.11 a/b/g/n IEEE 802.11 b/g/n

Data Transfer Rate Up to 300 Mbps Up to 300 Mbps

Interface Type PCIe PCIe

Status Indicators On/Off Link, Tx/Rx

Technical Guides | 48

Refrigeration Options Refrigerators are used with FlexLock or inside cabinets. They must be purchased from a third party. A half-size refrigerator can take up to two zones in a cabinet, using two shelf positions above it (up to 46 inches). Using a dorm-size refrigerator inside a cabinet takes up only one zone.

Dorm-style refrigerators are not CDC-compliant.

Special orders for refrigerator-ready cabinet (with vents for heat dissipation) have a six-week lead time. The following tables list compatible free-standing refrigerators (full size and compact).

Full-Size Refrigerators

Manufacturer Model Exterior Dimensions W x D x H (metric)

Internal Volume; cu ft.

(metric) Notes

GE GTS18GBSW 28.0" x 32.38" x 67.38" (71.12 x

82.25 x 171.15 cm)

18.2 cu. ft. (52

cu. m)

Inglis IT4KXSQ 28.0" x 31.0" x 62.9" (71.12 x 78.74 x 159.77 cm)

14.4 cu. ft. (0.41

cu. m)

Kenmore 27422 28.0" x 31.0" x 59.0" (71.12 x 78.74 x 149.86 cm)

14.1 cu. ft. (0.40

cu. m)

Kenmore 26422 28.0" x 28.5" x 59" (71.12 x 72.39 x

149.86 cm)

14 cu. ft. (0.40

cu. m)

Kenmore 27042 33.3" x 18.6" x 19.5" (84.15 x 47.14 x 49.53 cm)

20.6 cu. ft.

(0.58 cu. m)

Kenmore 26082 32.0" x 28.50" x 70.0" (81.28 x

72.39 x 177.80 cm)

21 cu. ft. (0.59

cu. m)

Panasonic MPR-721-PA 30.30" x 32.70" x 77.0" (77.0 x

83.0 x 196 cm)

24.2 cu. ft.

(0.69 cu. m)

Medical grade vaccine refrigerat or

Whirlpool ER8AHMXR Q

28.0" x 31.0" x 68.0" (172.72 x

71.12 x 78.74 cm)

17.6 cu. ft.

(0.50 cu. m)

Whirlpool ER6AHKXS Q

28.40" x 30.60" x 65.5" (166.37 x

71.6 x 77.62 cm)

15.9 cubic ft.

(0.45 cubic m)

Technical Guides | 49

Manufacturer Model Exterior Dimensions W x D x H (metric)

Internal Volume; cu ft.

(metric) Notes

Whirlpool GR2FHMPQ 29.50" x 31.50" x 66.5" (168.91 x

74.93 x 80.01 cm)

21.7 cubic ft.

(0.06 cubic m)

Table 8: Small Refrigerators

Manufacturer Model Exterior

Dimensions; W x D x H (metric)

Internal Volume; cu

ft. (metric)

Notes

Accucold by Summit Appliance

FFAR24LMED5

17.75" x 17.63" x 19.75" (45.08 x

44.78 x 50.17 cm)

1.7 cu. ft.

(0.05 cu.

m)

Medical grade

Accucold by Summit Appliance

FF511L7MED

19.25" x 21.38" x 33.5" (48.90 x

54.31 x 85.09 cm)

4.1 cu. ft.

(0.12 cu. m) Medical grade

Accucold by Summit Appliance

FF7LBIMEDADA

23.63" x 23.5" x 32.0" (67.64 x

59.69 x 81.28)

5.5 cu. ft.

(0.16 cu. m) Medical grade

Accucold by Summit Appliance

FF590SSHHMED

23.63" x 24.25" x 34.13" (67.64 x

61.56 x 86.69 cm)

5.7 cu. ft.

(0.16 cu. m) Medical grade

American BioTech Supply ABT-UCFS-0204G

17.75 " x 19.5" x 29.0" (45.09 x

49.53 x 73.66 cm)

2.5 cu. ft.

(0.07 cu.

m)

Medical grade;

Omnicell/Pyxis compliant

American BioTech Supply ABT-UCFS-0504W

21.5" x 22.5" x 33.5" (54.61 x

64.80 x 85.09 cm)

4.6 cu. ft.

(0.13 cu. m)

Medical grade;

Omnicell/Pyxis compliant

American BioTech Supply ABT-UCFS-0504G

23.75" x 24.5" x 32.2" (60.33 x

62.23 x 81.79 cm)

5.2 cu.ft.

(0.15 cu. m) N/A

American BioTech Supply ABT-UCBI-0404SS

23.75" x 24.5" x 33.4" (60.33 x

62.23 x 84.84 cm)

4.5 cu. ft.

(0.13 cu. m)

Medical grade;

Omnicell/Pyxis compliant

Danby DAR044A4BDD 20.6" x 21.0" x 32.6" (52.6 x 53.0 x 83.1 cm)

4.4 cu. ft.

(0.12 cu. m)

Technical Guides | 50

Manufacturer Model Exterior

Dimensions; W x D x H (metric)

Internal Volume; cu

ft. (metric)

Notes

Follett REF-1 18.6" x 24.0" x 17.75" (47.0 x 61.0 x 45.0 cm)

1 cu. ft.

(0.03 cu.

m)

Medical grade;

priority/most commonly used

Follett REF-4P-xx-xx 23.8" x 27.0" x 31.4" (60.3 x 68.5 x 79.7 cm)

3.9 cu. ft.

(0.11 cu. m)

Medical grade;

priority/most commonly used

Follett REF-5P-00-00 23.75" x 25.62" x 34.0" (60.3 x 65.1 x 86.4 cm)

4.5 cu. ft.

(0.13 cu. m)

Medical grade;

priority/most commonly used

Helmer SLR105 19.50" x 22.0" x 19.75" (49.53 x

55.88 x 50.17 cm)

5 cu. ft.

(0.14. cu m)

Medical grade;

priority/most commonly used

Helmer SLR104-ADA

19.50" x 22.0" x 17.75" (49.53 x

55.88 x 45.09 cm

4 cu. ft. (0.11

cu. m)

Medical grade;

priority/most commonly used

Helmer HLR105 19.50" x 22.0" x 19.75" (49.53 x

55.88 x 50.17 cm)

5 cu. ft.

(0.14. cu m)

Medical grade;

priority/most commonly used

Helmer HLR104-ADA

19.50" x 22.0" x 17.75" (49.53 x

55.88 x 45.09 cm

4 cu. ft. (0.11

cu. m)

Medical grade;

priority/most commonly used

Helmer iLR105 19.50" x 22.0" x 19.75" (49.53 x

55.88 x 50.17 cm)

5 cu. ft.

(0.14 cu. m)

Medical grade;

priority/most commonly used

Helmer iLR104-AD 19.50" x 22.0" x 17.75" (49.53 x

55.88 x 45.09 cm)

5 cu. ft.

(0.14 cu. m)

Medical grade;

priority/most commonly used

Migali GU1R-ADA 23.78" x 25.52" x 32.23" (60.40 x

64.82 x 81.86 cm)

4.3 cu. ft.

(0.12 cu. m)

Medical grade;

Omnicell/Pyxis compliant

Nor-Lake NSLR051WMW0 27.0" x 25.0" x

35.25 (68.6 x 63.5 x 89.5 cm)

5.1 cu. ft.

(0.14 cu. m) Medical grade

Technical Guides | 51

Manufacturer Model Exterior

Dimensions; W x D x H (metric)

Internal Volume; cu

ft. (metric)

Notes

Nor-Lake NSLR041WMW/0M 23.9" x 25.0" x 35.25" (60.6 x

63.5 x 58.1 cm)

4.6 cu. ft.

(0.13 cu. m) Medical grade

Nor-Lake NSLR051WMW0M 27.0" x 25.0" x

35.25 (68.6 x 49.5 x 58.1 cm)

5.4 cu. ft.

(0.15 cu. m) Medical grade

Sanyo SR-L6111W 23.80" x 22.50" x 34.5" (60.0 x

64.80 x 87.6 cm)

6.1 cu. ft.

(0.17 cu. m)

Medical grade;

priority/most commonly used

Thermo Scientific MR05PA-SEEE-TS

24.0" x 24.10" x 34.70" (61 x 61.3 x

86.7 cm)

5.4 cu. ft.

(0.15 cu. m)

Priority/most commonly used

The following two diagrams show the maximum cabinet internal dimensions for a refrigerator.

Allow sufficient space around the unit for ventilation, installation, and refrigerator door-swing clearance. Some refrigerators have hinges that allow the door to swing within the width of the refrigerator. Others may swing wider than the refrigerator width.

Technical Guides | 52

These diagrams allow for either a small “dorm" refrigerator in the bottom zone or a taller refrigerator that extends into the zone above as well.

Technical Guides | 53

Dorm-style refrigerators are not CDC-compliant.

Refrigerators dissipate heat as part of the cooling cycle. Ventilation must be provided to allow the internally generated heat to be dissipated.

Customers must order the tall cabinets in advance if they want to put a refrigerator inside. The frames are manufactured from scratch with special panels that have ventilation holes. These panels are riveted in place as the frames are manufactured and painted.

Refrigerators are allowed in the first cell and the second cell, but not in the third cell of a 3-cell cabinet.

Power consumption and dissipation for small refrigerators ranges from approximately 99-135 BTU/hr and from 253-345 KWhr/year. Note that for proper operation, refrigerators cycle the compressors on and off. Omnicell does not recommend using dorm-style refrigerators for critical refrigeration needs.

Use caution with vaccines as the temperature ranges must be strictly controlled, monitored, and recorded. Low cost, non-medical grade refrigerators may not provide adequate temperature control for these substances.

G4 Printers

Medication Label Printer Specifications The Integrated Medication Label Printer allows nurses to print patient-specific labels during medication issue.

Medication Label Printer Hardware and Software Requirements Following are the hardware and software requirements for the Medication Label Printer.

• OmniCenter and Color Touch cabinets running Omnicell 16.0 software or higher. Medication labels can be printed only after the upgrade to 16.0 software.

• Medication Label Printer (APS CP 290) installed and enabled. For procedures on how to install the printer, see the Medication Label Printer Installation Guide (67-2111).

• Omnicell 16.0 software or higher

Technical Guides | 54

• G4 cabinet consoles—OmniRx or One-, Two-, or Three-cell automated medication dispensing cabinets.

• Windows 7 OS or Windows XP OS

• Medication labels purchased from Omnicell. One medication label roll is provided in the kit.

Additional label rolls must be purchased separately from Omnicell.

• SafetyStock

− Confirm Med option

− Include bar code on label option

− 2D scanner if medication bar codes are 2D

About the Medication Label Printer Labels The use of labels supplied by Omnicell is required. Customer deviations from Omnicell product specifications will void Support Service terms.

The labels are custom-sized to support a printer cutter which cuts the labels between prints to provide a superior user experience. The label adhesive is also FDA Title 21, Section 175.105 compliant.

LABEL CARE AND HANDLING

• For best results, store in a dark, cool, and dry place. Temperature: 60ºF–75ºF, Humidity: 30%– 60% Relative Humidity.

• Avoid direct or indirect exposure to sun light.

• Each…