Attachment - VI - VA Final Rinse Water Usage Chart.pdf

PDF 50 KB Posted

Attached to
Y1NE--Upgrade Reverse Osmosis Processes / Sterilization 583-22-201
Solicitation number
36C25022B0006
Issued by

About this file

This document contains an attachment analyzing the final rinse water usage of sterilization equipment at a VA facility in Indianapolis, along with details of a related federal contract opportunity. The attachment provides specifications for four steam sterilizers, two washer-disinfectors, one cart washer-disinfector, and three reprocessing sinks. It calculates the total gallons per hour and minimum flow rate needed at 306 gallons per hour and 41.55 gallons per minute. The attachment recommends a special water treatment system may be required since the rinse water requirements exceed standard systems. The related federal contract opportunity is a solicitation from the VA Veterans Health Administration VISN 10 to upgrade reverse osmosis processes and sterilization equipment at facility 583, with an opportunity number of Y1NE--Upgrade Reverse Osmosis Processes / Sterilization 583-22-201.

View the file

Other files

Show all 11

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Date: 7/16/2021

VA Indianapolis

H-NUMBER:

Qty Product Description

Number Cycles per Hour

Pure H2O Gallons per Hour

Minimum Flow Rate

(GPM)

TOTAL

Gallons per Hour

(GPH)

TOTAL

Minimum Flow Rate

(GPM)

4 Reliance VISION SC (steam) 3 21.00 5.70 84.00 22.80 2 Innowave PRO 1 38.50 3.00 77.00 6.00

Disinfection Cycle - With Pure Water (Pro) 1 Vision 1327 Cart Washer-Disinfector

(Steam) 4 100.00 12.00 100.00 12.00

Pure Water-All Cycles (No Water Recycle) 3 AMSCO 53 (Adj.Ht.)-3 Bay Repocessing

Sink 106" 1 15.00 0.25 45.00 0.75

GRAND TOTAL: 306.00 41.55

Piping Distance from Product to Water System: 50 Feet

RECOMMENDATION:

REQUIREMENTS exceed Standard STERIS systems.

Therefore, a special Water Treatment System may be required.

File details come from the government source that posted it. Updated .