Attachment - VI - VA Final Rinse Water Usage Chart.pdf
PDF 50 KB Posted
- Attached to
- Y1NE--Upgrade Reverse Osmosis Processes / Sterilization 583-22-201 Federal contract opportunity
- Solicitation number
- 36C25022B0006
About this file
This document contains an attachment analyzing the final rinse water usage of sterilization equipment at a VA facility in Indianapolis, along with details of a related federal contract opportunity. The attachment provides specifications for four steam sterilizers, two washer-disinfectors, one cart washer-disinfector, and three reprocessing sinks. It calculates the total gallons per hour and minimum flow rate needed at 306 gallons per hour and 41.55 gallons per minute. The attachment recommends a special water treatment system may be required since the rinse water requirements exceed standard systems. The related federal contract opportunity is a solicitation from the VA Veterans Health Administration VISN 10 to upgrade reverse osmosis processes and sterilization equipment at facility 583, with an opportunity number of Y1NE--Upgrade Reverse Osmosis Processes / Sterilization 583-22-201.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFI Responses.pdf | ||
| 36C25022B0006 0001.docx | DOCX document | |
| Attachment - IX - RFI Form Solicitaton.docx | DOCX document | |
| Attachment - IV - Reverse Osmosis Site Plan Layout Drawing.pdf | ||
| Attachment - I - ABSO Foot Print Drawing.pdf | ||
| Attachment - VII - Wage Determination.pdf | ||
| Attachment - III - Reverse Osmosis Equipment Drawing.pdf | ||
| Attachment - II - Basement Plumbing Plan Drawing.zip | ZIP file | |
| 36C25022B0006.docx | DOCX document | |
| Attachment - V - Reverse Osmosis Workspace Drawing.pdf | ||
| Attachment - VIII - Contractor EMR Certification Form.docx | DOCX document |
Show all 11
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Date: 7/16/2021
VA Indianapolis
H-NUMBER:
Qty Product Description
Number Cycles per Hour
Pure H2O Gallons per Hour
Minimum Flow Rate
(GPM)
TOTAL
Gallons per Hour
(GPH)
TOTAL
Minimum Flow Rate
(GPM)
4 Reliance VISION SC (steam) 3 21.00 5.70 84.00 22.80 2 Innowave PRO 1 38.50 3.00 77.00 6.00
Disinfection Cycle - With Pure Water (Pro) 1 Vision 1327 Cart Washer-Disinfector
(Steam) 4 100.00 12.00 100.00 12.00
Pure Water-All Cycles (No Water Recycle) 3 AMSCO 53 (Adj.Ht.)-3 Bay Repocessing
Sink 106" 1 15.00 0.25 45.00 0.75
GRAND TOTAL: 306.00 41.55
Piping Distance from Product to Water System: 50 Feet
RECOMMENDATION:
REQUIREMENTS exceed Standard STERIS systems.
Therefore, a special Water Treatment System may be required.