Attachment J COA (M) RAP Med Equip Inventory.pdf

PDF 748 KB Posted

Attached to
RFR 2025-DMH-3134-01 Medical Equip Inspections, Inventory Serv, and Repairs
Solicitation number
BD-24-1022-DMH02-8420A-99954
Issued by
Bristol County, Massachusetts

About this file

This document is an attachment related to a state and local contract opportunity in Massachusetts for medical equipment inspections, inventory services, and repairs. The contract opportunity is designated as RFR 2025-DMH-3134-01 and covers jurisdictions in Worcester County and Worcester City. The scope of work includes performing bi-annual electrical safety inspections, calibrating and testing medical equipment such as AEDs, blood pressure monitors, pulse oximeters, and other devices. The attachment provides detailed inspection records for the various equipment, including asset tag numbers, manufacturer details, and test results. The contract has a target completion date of June 12, 2024 and may be eligible for renewal options.

The pricing for this contract is not explicitly stated, but the attachment indicates that the awarded vendor must utilize certified testing devices and follow manufacturer specifications for calibration and performance testing. There are no details provided about any set-aside requirements or incumbent information. The funding for this contract likely comes from the state and local government budgets, but the specific amounts are not disclosed.

View the file

Other files

File Type Posted
RFR Evaluation Tool.3.12.2024.docx DOCX document
Attachment I COA (W) RAP Med Equip Inventory~3.pdf PDF
Attachment B Central Mass Area Contract Information.docx DOCX document
Attachment D Central Office Area Contract Information.docx DOCX document
Attachment K Location and Contact Information.docx DOCX document
Business Reference Form~22.docx DOCX document
Attachment H Taunton State Hosp Med Equip Inventory.xlsx XLSX spreadsheet
False Claims Memorandum~14.doc DOC document
OSD Instructions for Vendors Responding to Bids via Commbuys~65.docx DOCX document
SEA Award Letter.pdf PDF
Attachment G CC Islands Comm Mental Health Center Med Equip Inventory.xlsx XLSX spreadsheet
DMH Standard Contract Requirments 2024.docx DOCX document
Electronic Funds Transfer Sign Up Form~32.pdf PDF
CMA Award Letter.pdf PDF
Attachment F Corrigan Mental Health Center Med Equip Inventory.xlsx XLSX spreadsheet
Attachment L WRCH Dexis Contract.pdf PDF
OSD Required Specifications for Commodities and Services~17.docx DOCX document
COA Award Letter.pdf PDF
form_contractor-authorized-signatory-listing-for-contractors~1.pdf PDF
MA Form_W-9~42.pdf PDF
RFR_FINAL_Med Equip Inspections.docx DOCX document
Program Response Form.3.11.2024.docx DOCX document
form_contractor-authorized-signatory-listing-for-sole-proprietors.pdf PDF
Attachment C Southeast Area Contract Information.docx DOCX document
Commonwealth Standard Contract Instructions_Contractor Certifications_Terms~26.pdf PDF
Commonwealth of Massachusetts Standard Contract Form~35.docx DOCX document
Attachment E WRCH Medical Equipment Inventory FY24.xls XLS spreadsheet
Show all 27

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STAT Bio-Medical certifies that the equipment included in the Bi - Annual Electrical Safety Inspection has been tested and/or calibrated using software and instrumentation producing traceable results as outlined by the National Institute Of Standards and Technology (NIST). Evidence of Traceability is on file at our corporate service center.

Acct: 20491 60 Hodges Ave

Taunton MA 02780

Bi - Annual Electrical Safety Inspection Record: 10328

Completion Date: 12/12/2023 Due Date: 6/12/2024

CERTIFICATE OF INSPECTION

STAT Bio-Medical

RECOVERY FOR ADDICTIONS PROGRAM - M-RAP

If you have questions in regards to our testing or need service for your medical equipment, please contact us at (508) 923-8701 bill@statbiomedical.net

Continued on Next Page

If Blank - Electrical Safety Test Not Performed

Electrical Safety Test Results

Inspections Performed Visual

AED

Tag Number: 0752555 Manufacturer: ZOLL Model: AEDPLUS Serial: 010084794600224421X21H405777 Medical Risk Factor: 1

Entry Type Inspd/Tstd

Location South Chamber 02 Code Cart

Result Pass

CERTIFICATE OF INSPECTION

Inspections Performed

EST: NA

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: Pad Date Expires - 10/28/26

Inspection Notes

AED Tag Number: 0752555

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua l

Result: Pass

Electrical Safety Test Results

Inspections Performed Visual

AED

Tag Number: 0752561 Manufacturer: ZOLL Model: AEDPLUS Serial: 010084794600224421X21H406072 Medical Risk Factor: 1

Entry Type Inspd/Tstd

Location North Chamber 02 Code Cart

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: Pad Date Expires - 12/18/26

Inspection Notes

AED Tag Number: 0752561

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Visual

AED

Tag Number: 0752575 Manufacturer: ZOLL Model: AEDPLUS Serial: 010084794600224421X21F381085 Medical Risk Factor: 1

Entry Type Inspd/Tstd

Location North Chamber 03

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: Pad Date Expires - 12/18/26

Inspection Notes

AED Tag Number: 0752575

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Visual

AED

Tag Number: 0752582 Manufacturer: ZOLL Model: AEDPLUS Serial: 010084794600224421X21H405851 Medical Risk Factor: 1

Entry Type Inspd/Tstd

Location South Chamber 03 Code Cart

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: Pad Date Expires - 08/28/26

Inspection Notes

AED Tag Number: 0752582

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0752557 Manufacturer: ADC Model: lifetime Certified Serial: 15167626 Medical Risk Factor:

Entry Type Inspd/Tstd

Location South Chamber 02 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

300 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0752557

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0752563 Manufacturer: ADC Model: certified Serial: 21300772 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 02 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0752563

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0752583 Manufacturer: ADC Model: sphygmomanometer Serial: 20236141 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location South Chamber 03 Code Cart

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0752583

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0748878 Manufacturer: ADC Model: BP Serial: 150108488 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location South Chamber 02 Code Cart

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0748878

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0748876 Manufacturer: Welch Allyn Model: Shock Resistant Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location South Chamber 02 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0748876

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0753780 Manufacturer: ADC Model: Sphygmomanometer Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location South Chamber 03 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0753780

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0753781 Manufacturer: ADC Model: sphygmomanometer Serial: N/A Medical Risk Factor:

Entry Type Inspd/Tstd

Location South Chamber 03 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0753781

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0753795 Manufacturer: ADC Model: sphygmomanometer Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 03 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0753795

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0753794 Manufacturer: Adc Model: sphygmomanometer Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 03 Medicine Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0753794

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0758163 Manufacturer: ADC Model: lifetime certified Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: Adjusted to manufacturer Specifications.

Performance: Blood Pressure Performance Testing.

Visual: Inspected case, gauge, tubing, and cuff.

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0758163

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0754348 Manufacturer: sphygmomanometer Model: Ce1434 Serial: 21337002 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0754348

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0754351 Manufacturer: Adc Model: Ce1434 Serial: 21320920 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0754351

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0760607 Manufacturer: Adc Model: Ce1434 Serial: 21325722 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam 213

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0760607

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0760609 Manufacturer: Adc Model: Ce1434 Serial: 21320630 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0760609

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0760610 Manufacturer: Adc Model: Ce1434 Serial: 21324827 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0760610

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0760615 Manufacturer: Adc Model: Ce1434 Serial: 21336491 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0760615

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0762346 Manufacturer: Adc Model: Na Serial: 20337929 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0762346

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0762347 Manufacturer: Adc Model: Na Serial: 2132438 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0762347

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0754347 Manufacturer: sphygmomanometer Model: Ce1434 Serial: 20281816 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0754347

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Electrical Safety Test Results

Inspections Performed Performance Visual

Blood Pressure - Gauge - Handheld Tag Number: 0762398 Manufacturer: Omron Model: Certified Serial: 820528 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med Room

Result Pass

BP Gauge Calibration

0 mm

ActualExpct'd Tolerance

+/- 3 mm

50 mm +/- 3 mm

100 mm +/- 3 mm

200 mm +/- 3 mm

NIBP: Blood Pressure Simulation was performed. See Test Results.

Sp02: NA EKG Wave: NA Calibration: NA Blood Pressure Performance Testing Visual: NA

Inspection Notes

Blood Pressure - Gauge - Handheld Tag Number: 0762398

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

STAT Bio-Medical Mobilize App Report Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0762400 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD Date/Time: 2023/12/12 11:24 AC Voltage: 126 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Compression System Tag Number: 0762400 Manufacturer: Invacare Model: Na Serial: Na Medical Risk Factor: 2

Entry Type Inspd/Tstd

Location Basement

Result

EST: Electrical Safety Test utilizing Certified Testing Devices. See results enclosed.

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Compression System Tag Number: 0762400

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0762401 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD Date/Time: 2023/12/12 11:25 AC Voltage: 126 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Compression System Tag Number: 0762401 Manufacturer: Invacare Model: Na Serial: Na Medical Risk Factor: 2

Entry Type Inspd/Tstd

Location Basement

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Compression System Tag Number: 0762401

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752565 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD NO SW Date/Time: 2023/12/12 08:55 AC Voltage: 126 V Peak Load Current: 0.2 A Norm Pol Earth Leak:

PASS Norm Cond = 88 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Light Tag Number: 0752565 Manufacturer: Welch Allyn Model: GS00 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 304

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Light Tag Number: 0752565

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0748882 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD NO SW Date/Time: 2023/12/12 10:05 AC Voltage: 125 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 83 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Light Tag Number: 0748882 Manufacturer: Welch Allyn Model: GS300 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 213

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Light Tag Number: 0748882

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0748892 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD NO SW Date/Time: 2023/12/12 08:56 AC Voltage: 126 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 88 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Light Tag Number: 0748892 Manufacturer: Welch Allyn Model: GS300 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 301

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Light Tag Number: 0748892

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Sydney Asset ID: 0753778 Test Eqpt Model: Safe-T ST-1 S/N 3756 Last Cal: 06/02/23 Electrical Safety Test Profile: EST NO GD NO SW Date/Time: 2023/12/12 08:57 AC Voltage: 125 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 90 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 89 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Light Tag Number: 0753778 Manufacturer: Welch Allyn Model: GS300 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 314

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Light Tag Number: 0753778

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752550 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:08 AC Voltage: 124 V Peak Load Current: 0.2 A Gnd Resist:

PASS Test Gnd = 21 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 118 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Table Tag Number: 0752550 Manufacturer: Ritter Model: 224-001 Serial: V2330415 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 212

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Table Tag Number: 0752550

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752568 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 08:54 AC Voltage: 125 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 19 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 9 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Table Tag Number: 0752568 Manufacturer: Ritter Model: 224-001 Serial: V2352181 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 304

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Table Tag Number: 0752568

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752952 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:04 AC Voltage: 125 V Peak Load Current: 0.2 A Gnd Resist:

PASS Test Gnd = 20 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 120 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Table Tag Number: 0752952 Manufacturer: Ritter Model: 224 Serial: V2330694 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 213

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Table Tag Number: 0752952

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Sydney Asset ID: 0748891 Test Eqpt Model: Safe-T ST-1 S/N 3756 Last Cal: 06/02/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 08:55 AC Voltage: 125 V Peak Load Current: 0.2 A Gnd Resist:

PASS Test Gnd = 10 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 117 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 117 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Table Tag Number: 0748891 Manufacturer: Ritter Model: 224 Serial: V2348179 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam Room 301

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Table Tag Number: 0748891

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0762397 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:51 AC Voltage: 125 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 18 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 15 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Exam Table Tag Number: 0762397 Manufacturer: Na Model: Na Serial: Na Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Exam 219

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Exam Table Tag Number: 0762397

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752585 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 09:53 AC Voltage: 125 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 33 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 16 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 2 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0752585 Manufacturer: Stryker Model: CHG Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location South Chamber 03 Room 326

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0752585

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753793 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 09:25 AC Voltage: 126 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 23 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 15 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753793 Manufacturer: Stryker Model: CHG Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 03 Patient Room N325

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753793

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753792 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 09:22 AC Voltage: 127 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 19 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 15 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 2 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753792 Manufacturer: Stryker Model: CHG Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 03 Patient Room N308

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753792

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753783 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD Date/Time: 2023/12/12 11:17 AC Voltage: 125 V Peak Load Current: 0.3 A Norm Pol Earth Leak:

PASS Norm Cond = 15 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753783 Manufacturer: Stryker Model: CHG Serial: N/a Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 02 Patient Room N212

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753783

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753801 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:18 AC Voltage: 126 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd =146 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 16 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753801 Manufacturer: Stryker Model: CHG Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Basement Chart Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753801

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753802 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 11:41 AC Voltage: 127 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 18 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 16 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753802 Manufacturer: Stryker Model: CHG Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Basement Chart Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753802

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753803 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 11:39 AC Voltage: 126 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 17 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 16 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753803 Manufacturer: Stryker Model: CHG Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Basement Chart Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753803

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753804 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 11:37 AC Voltage: 126 V Peak Load Current: 0.9 A Gnd Resist:

PASS Test Gnd = 21 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 15 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0753804 Manufacturer: Stryker Model: CHG Serial: NA Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Basement Chart Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0753804

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0760614 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:46 AC Voltage: 124 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 20 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 15 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Hospital Bed Tag Number: 0760614 Manufacturer: Tmotion Model: N/a Serial: N/a Medical Risk Factor:

Entry Type Inspd/Tstd

Location Bedroom 216

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Hospital Bed Tag Number: 0760614

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752556 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 11:01 AC Voltage: 125 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 21 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Medicine Sorting System Tag Number: 0752556 Manufacturer: Omnicell Model: Full Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location South Chamber 02 Med Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Medicine Sorting System Tag Number: 0752556

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0748875 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:29 AC Voltage: 126 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd =250 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 3 uA (< 100 uA) PASS Open Gnd = 3 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Medicine Sorting System Tag Number: 0748875 Manufacturer: Omnicell Model: Full Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location North Chamber 02 Medicine Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Medicine Sorting System Tag Number: 0748875

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752975 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 09:39 AC Voltage: 127 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd =222 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 3 uA (< 100 uA) PASS Open Gnd = 0 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Medicine Sorting System Tag Number: 0752975 Manufacturer: Omnicell Model: Full Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location South Chamber 03 Medicine Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Medicine Sorting System Tag Number: 0752975

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0760613 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD Date/Time: 2023/12/12 10:06 AC Voltage: 126 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Nebulizer Tag Number: 0760613 Manufacturer: Devilbiss Model: 3655D Serial: D2889980 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration:

Performance: NA Visual: NA

Inspection Notes

Nebulizer Tag Number: 0760613

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0762395 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD Date/Time: 2023/12/12 10:36 AC Voltage: 126 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Nebulizer Tag Number: 0762395 Manufacturer: Drive medical Model: 18080 Serial: 212V2005079215 Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration:

Performance: NA Visual: NA

Inspection Notes

Nebulizer Tag Number: 0762395

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0753521 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO GD Date/Time: 2023/12/12 11:22 AC Voltage: 126 V Peak Load Current: 0.1 A Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Oxygen Concentrator Tag Number: 0753521 Manufacturer: Invacare Model: Platinum10 Serial: N/A Medical Risk Factor: 2

Entry Type Inspd/Tstd

Location Basement

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Oxygen Concentrator Tag Number: 0753521

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Vital Signs Testing Profile: URL Preset: Vitals Simulation Date/Time: 2023/12/12 10:39 Sim Time: 0:24 Param Sim Result HR 95 95 Pass SpO2 95 95 Pass ( SpO2 Perfusion:95% )

< PASS >

Info-ST=S/N 1034 Last Cal:11/28/23 Info-SC=S/N 03288 Last Cal:01/11/23 Info-OX=S/N 03455 Last Cal:11/21/23

Electrical Safety Test Results

Inspections Performed Visual Performance Electrical

Pulse Oximeter - Handheld Tag Number: 0752560 Manufacturer: ADC Model: 2200 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 02 Code Cart

Result

NIBP: NA

Sp02: Simulation was performed. See Test Results.

EKG Wave: NA Calibration: NA Pulse Oximeter Performance Testing Visual: NA

Inspection Notes

Pulse Oximeter - Handheld Tag Number: 0752560

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Profile: URL Preset: Vitals Simulation Date/Time: 2023/12/12 09:48 Sim Time: 0:15 Param Sim Result HR 95 95 Pass SpO2 95 95 Pass ( SpO2 Perfusion:95% )

< PASS >

Info-ST=S/N 1034 Last Cal:11/28/23 Info-SC=S/N 03288 Last Cal:01/11/23 Info-OX=S/N 03455 Last Cal:11/21/23

Electrical Safety Test Results

Inspections Performed Visual Performance Electrical

Pulse Oximeter - Handheld Tag Number: 0752584 Manufacturer: ADC Model: 2200 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location South Chamber 03 Code Cart

Result

NIBP: NA

Sp02: Simulation was performed. See Test Results.

EKG Wave: NA Calibration: NA Pulse Oximeter Performance Testing Visual: NA

Inspection Notes

Pulse Oximeter - Handheld Tag Number: 0752584

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Profile: URL Preset: Vitals Simulation Date/Time: 2023/12/12 10:24 Sim Time: 0:12 Param Sim Result HR 95 95 Pass SpO2 95 95 Pass ( SpO2 Perfusion:95% )

< PASS >

Info-ST=S/N 1034 Last Cal:11/28/23 Info-OX=S/N 03455 Last Cal:11/21/23

Electrical Safety Test Results

Inspections Performed Visual Performance Electrical

Pulse Oximeter - Handheld Tag Number: 0753787 Manufacturer: ADC Model: 2200 Serial: N/A Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location North Chamber 02 Medicine Room

Result

NIBP: NA

Sp02: Simulation was performed. See Test Results.

EKG Wave: NA Calibration: NA Pulse Oximeter Performance Testing Visual: NA

Inspection Notes

Pulse Oximeter - Handheld Tag Number: 0753787

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Profile: URL Preset: Vitals Simulation Date/Time: 2023/12/12 09:42 Sim Time: 0:28 Param Sim Result HR 95 95 Pass SpO2 95 95 Pass ( SpO2 Perfusion:95% )

< PASS >

Info-ST=S/N 1034 Last Cal:11/28/23 Info-SC=S/N 03288 Last Cal:01/11/23 Info-OX=S/N 03455 Last Cal:11/21/23

Electrical Safety Test Results

Inspections Performed Visual Performance Electrical

Pulse Oximeter - Handheld Tag Number: 0754350 Manufacturer: Covidien Model: Nellcor Serial: N/a Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result

NIBP: NA

Sp02: Simulation was performed. See Test Results.

EKG Wave: NA Calibration: NA Pulse Oximeter Performance Testing Visual: NA

Inspection Notes

Pulse Oximeter - Handheld Tag Number: 0754350

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Profile: URL Preset: Vitals Simulation Date/Time: 2023/12/12 10:34 Sim Time: 0:10 Param Sim Result HR 95 95 Pass SpO2 95 95 Pass ( SpO2 Perfusion:95% )

< PASS >

Info-ST=S/N 1034 Last Cal:11/28/23 Info-OX=S/N 03455 Last Cal:11/21/23

Electrical Safety Test Results

Inspections Performed Visual Performance Electrical

Pulse Oximeter - Handheld Tag Number: 0760611 Manufacturer: Covidien Model: Nellcor Serial: N/a Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result

NIBP: NA

Sp02: Simulation was performed. See Test Results.

EKG Wave: NA Calibration: NA Pulse Oximeter Performance Testing Visual: NA

Inspection Notes

Pulse Oximeter - Handheld Tag Number: 0760611

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Profile: URL Preset: Vitals Simulation Date/Time: 2023/12/12 10:30 Sim Time: 0:03 Param Sim Result HR 95 95 Pass SpO2 95 95 Pass ( SpO2 Perfusion:95% )

< PASS >

Info-ST=S/N 1034 Last Cal:11/28/23 Info-OX=S/N 03455 Last Cal:11/21/23

Electrical Safety Test Results

Inspections Performed Visual Performance Electrical

Pulse Oximeter - Handheld Tag Number: 0760612 Manufacturer: Adc Model: 2200 Serial: N/a Medical Risk Factor: 3

Entry Type Inspd/Tstd

Location Med room

Result

NIBP: NA

Sp02: Simulation was performed. See Test Results.

EKG Wave: NA Calibration: NA Pulse Oximeter Performance Testing Visual: NA

Inspection Notes

Pulse Oximeter - Handheld Tag Number: 0760612

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752562 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:28 AC Voltage: 125 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd =250 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 0 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Refrigerator Tag Number: 0752562 Manufacturer: Accucold Model: half size under counter Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location North Chamber 02 Medicine Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Refrigerator Tag Number: 0752562

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0748877 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 11:00 AC Voltage: 126 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 17 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Refrigerator Tag Number: 0748877 Manufacturer: Accucold Model: 1/2 Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location South Chamber 02 Medicine Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Refrigerator Tag Number: 0748877

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Sydney Asset ID: 0748894 Test Eqpt Model: Safe-T ST-1 S/N 3756 Last Cal: 06/02/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 08:58 AC Voltage: 124 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 5 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 3 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 0 uA (< 100 uA) PASS Open Gnd = 3 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Refrigerator Tag Number: 0748894 Manufacturer: Danby Model: 1/2 Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location Exam Room 301

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Refrigerator Tag Number: 0748894

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0748898 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 09:11 AC Voltage: 125 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd = 17 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 1 uA (< 100 uA) PASS Open Gnd = 1 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Refrigerator Tag Number: 0748898 Manufacturer: Accucold Model: 1/2 Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location North Chamber 03 Medicine Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Refrigerator Tag Number: 0748898

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0752974 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 09:38 AC Voltage: 127 V Peak Load Current: 0.1 A Gnd Resist:

PASS Test Gnd =223 mOhm (<500 mOhm) Norm Pol Earth Leak:

PASS Norm Cond = 1 uA (< 500 uA) Norm Pol Chassis Leak:

PASS Norm Cond = 3 uA (< 100 uA) PASS Open Gnd = 0 uA (< 500 uA) < Electrical Safety Overall Result: PASS >

Electrical Safety Test Results

Inspections Performed Electrical

Refrigerator Tag Number: 0752974 Manufacturer: Accucold Model: 1/2 Size Serial: N/A Medical Risk Factor: 4

Entry Type Inspd/Tstd

Location South Chamber 03 Medicine Room

Result

NIBP: NA

Sp02: NA EKG Wave: NA Calibration: NA Performance: NA Visual: NA

Inspection Notes

Refrigerator Tag Number: 0752974

If Blank - No Asset Image Acguired

Asset Test Image

Electrica l

Calibratio n

Performance Visua

Facility: Recovery for Addictions Program - M-RAP Technician: Brittany Asset ID: 0758162 Test Eqpt Model: Safe-T ST-1 S/N 1034 Last Cal: 11/28/23 Electrical Safety Test Profile: EST NO SW Date/Time: 2023/12/12 10:10 AC Voltage: 125 V Peak Load Current: 0.1 A…