Attachment G - 1.docx
DOCX document 23 KB Posted
- Attached to
- Q999 - VISN 12 Home Oxygen Delivery Services
- Solicitation number
- 36C25225R0072
About this file
Attachment G is a detailed document outlining mandatory requirements for thermal fuses in home oxygen delivery systems for VA facilities within VISN 12. The document specifies critical safety protocols, including: (a) thermal fuses must be provided for home oxygen therapy patients unless specific clinical contraindications exist, (b) two thermal fuses must be installed in each oxygen delivery system (at patient end and oxygen source), and (c) thermal fuses must be bidirectional or designed to prevent incorrect installation.
Additional key requirements include patient education about thermal fuse placement and safety, replacement of thermal fuses at the end of their expected life or if involved in a fire, and specific guidelines for different oxygen therapy equipment types. The document is associated with the VA 1624 Clinical Respiratory Assessment for VISN 12, covering medical centers in Wisconsin, Michigan, and Illinois, and was last revised on 6/4/2020. The thermal fuse requirements are part of a broader home oxygen delivery services solicitation (36C25225R0072) managed by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 12.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| P07 - Monroe County WD 2015-4929 Rev 29 07.08.2025.pdf | ||
| P07 - Cook County WD 2015-5017 Rev 30 07.08.2025.pdf | ||
| Attachment B - 1.docx | DOCX document | |
| Attachment H.docx | DOCX document | |
| Attachment B - 2.docx | DOCX document | |
| Attachment E - 2.docx | DOCX document | |
| Attachment E - 1.docx | DOCX document | |
| P07 - Lake County WD 2015-4935 Rev 28 07.08.2025.pdf | ||
| P07 - Dane County WD 2015-4897 Rev 29 07.08.2025.pdf | ||
| Attachment K.docx | DOCX document | |
| Attachment F.docx | DOCX document | |
| Attachment J - 1.docx | DOCX document | |
| Attachment J - 2.docx | DOCX document | |
| Attachment D.docx | DOCX document | |
| S02 - 36C25225R0072.pdf | ||
| P07 - Dickinson County WD 2015-4871 Rev 30 07.08.2025.pdf | ||
| P07 - Vermillion County WD 2015-5021 Rev 26 07.08.2025.pdf | ||
| Attachment A.docx | DOCX document | |
| Attachment I.docx | DOCX document | |
| Attachment C.docx | DOCX document | |
| P07 - Milwaukee County WD 2015-4899 Rev 28 07.08.2025.pdf | ||
| Attachment G - 2.docx | DOCX document |
Show all 22
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment G Thermal Fuses (FSV) (1of 2)
Mandatory requirements for provision of thermal fuses are as follows
a. Thermal fuses must be provided for all patients on home oxygen therapy unless there is a documented clinical contraindication (e.g., patients using heated humidity, patients with flow rates in excess of 25 Liters in accordance with minute, patients with behavioral conditions that make use of thermal fuses impractical). If thermal fuses are not provided due to clinical contraindication, this must be documented in the patient’s medical record.
b. Two thermal fuses must be provided and installed in each home oxygen delivery system (one at the patient end- short cannula- and one at the oxygen source-see diagram). This must be done for the patient’s primary oxygen therapy equipment as well as any portable oxygen therapy equipment.
c. Thermal fuses are not required for portable pulse dose/intermittent oxygen therapy equipment when the equipment does not have a continuous flow setting.
d. Thermal fuses provided must work regardless of their orientation in the oxygen tubing (bidirectional) or have a design that prevents them from being installed in the wrong orientation (e.g., source end of thermal fuse has threads or bayonets that make backwards installation impossible).
e. Existing unidirectional thermal fuses that are capable of being installed in the wrong orientation must be replaced at the next scheduled home oxygen visit or sooner if possible.
f. At the time of installation, education must be provided to the patient describing the proper use and placement of thermal fuses, including that a thermal fuse does not make it safe to smoke, be near people who are smoking, or be near any source of sparks or flame while using oxygen.
g. Thermal fuses must be replaced at the end of their expected life in accordance with manufacturer’s guidelines.
h. Thermal fuses must be replaced if actuated or involved in a fire.