ATTACHMENT B PD NX EQ Laparoscopic Insufflators.pdf
PDF 135 KB Posted
- Solicitation number
- 36C10X26R0009
About this file
This is a Product Description document prepared by the Department of Veterans Affairs (VA) Strategic Acquisition Center for a national single Requirements contract to procure Laparoscopic Insufflators.
The VA seeks vendors to supply Olympus UHI-4 Intra-abdominal Insufflation Units or equal products that meet all specified salient characteristics. Laparoscopic Insufflators are pressure-limited gas-flow regulators used during laparoscopic procedures to produce and maintain a pneumoperitoneum, enlarging the surgical workspace. The contract covers seven Contract Line Items (CLINs 0001-0007): the main Olympus UHI-4 insufflation unit, footswitch (MAJ-1939), medical gas pipeline adapter (MAJ-1084), cylinder hose/switch over valve (MAJ-1985), sterile tubing with filter (WA95005A), insufflation tubing (72-00163-0), and suction tubing set (72-00162-0). The equipment will be distributed to VA medical centers and facilities throughout the Continental United States (CONUS) and Outside the Continental United States (OCONUS). The contract period is a 12-month base with four optional 12-month periods. As a "Brand Name or Equal" requirement, offerors proposing alternative products must provide detailed descriptions of salient characteristics, quality levels, brand names, make/model numbers, and descriptive literature demonstrating compliance with five mandatory salient characteristics: FDA approval for laparoscopic procedures, smoke evacuation capability, CO2 medical gas usage, minimum alarm functions (excessive pressure, tube obstruction, CO2 pressure), and two cavity modes (Normal and Small). Vendors may propose additional ancillary products deemed essential to system functionality, captured in pricing with zero quantity if not required.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| ATTACHMENT H_CERTIFICATION OF RECALL OR SAFETY ISSUES.docx | DOCX document | |
| ATTACHMENT G_Template for Industry Questions_A0001.xlsx | XLSX spreadsheet | |
| ATTACHMENT A_RFP 36C10X26R0009 A0001_NX EQ Laparoscopic Insufflators.pdf | ||
| 36C10X26R0009_2.docx | DOCX document | |
| ATTACHMENT F_PAST PERFORMANCE QUESTIONNAIRE_NX EQ Laparoscopic Insufflators.docx | DOCX document | |
| ATTACHMENT D_PRICE COST SCHEDULE_NX EQ Laparoscopic Insufflators.xlsx | XLSX spreadsheet | |
| ATTACHMENT A_RFP 36C10X26R0009_NX EQ Laparoscopic Insufflators.pdf | ||
| 36C10X26R0009.docx | DOCX document | |
| ATTACHMENT G_Template for Industry Questions.xlsx | XLSX spreadsheet | |
| ATTACHMENT E_PAST PERFORMANCE REFERENCES_NX EQ Laparoscopic Insufflators.docx | DOCX document | |
| ATTACHMENT C_SALES REPORTING TOOL_NX EQ Laparoscopic Insufflators.xlsx | XLSX spreadsheet |
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PRODUCT DESCRIPTION
For
NON-EXPENDABLE EQUIPMENT:
Laparoscopic Insufflators
Prepared By:
Strategic Acquisition Center
Department of Veterans Affairs 10300 Spotsylvania Avenue, Suite 400
Fredericksburg, VA 22408
1. SCOPE OF WORK:
The Veterans Health Administration (VHA) Non-Expendable Equipment National Program has identified the Laparoscopic Insufflators product line as a candidate for a VA-wide (otherwise referred to as “national”) single Requirements contract award.
Laparoscopic Insufflators are specialized; pressure-limited gas-flow regulators used during laparoscopic procedures to produce and maintain a pneumoperitoneum (gas-filled space within the peritoneal cavity). This enlarges the working space, providing room to insert instruments, and allows the surgeon to see the organs more easily.
Pressure regulators and flow-restricting orifices in the device control pressure and gas flow during insufflation.
The objective is to provide Laparoscopic Insufflators to be used by clinicians throughout the VA medical centers and facilities in the Continental United States (CONUS) and Outside the Continental United States (OCONUS). The period of performance is for a base of 12-months with four 12-month option periods.
2. REQUIREMENT
This requirement will be “Brand Name or Equal” which requires the offeror indicate that each product being offered as an “equal” product to the Olympus UHI-4 Intra-abdominal Insufflation Unit. For each equal product, the offeror must include a description reflecting the salient characteristics (SC) and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The offeror must also clearly identify the item by brand name (if any) and make/model number. Finally, the offeror must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the Contracting Officer and clearly describe any modifications it plans to make to a product to conform to the solicitation requirements.
Contract Line Items
Manufacture Part Number Description
0001 Olympus UHI-4 Intra-abdominal Insufflation Unit
0002 Olympus MAJ-1939 footswitch
0003 Olympus MAJ-1084 Med Gas Pipeline Adapter
0004 Olympus MAJ-1985 Cylinder Hose / Switch Over Valve
0005 Olympus WA95005A Tubing with filter, sterile
0006 Olympus 72-00163-0 Insufflation tubing UHI/UHI-2/UHI-3 insufflator
0007 Olympus 72-00162-0 Suction tubing set for UHI/UHI-2/UHI-3 insufflator
The Department of Veterans Affairs (VA) is seeking vendors who can provide Olympus UHI-4 Intra-abdominal Insufflation Unit or equal as listed above which meet all the salient characteristics listed below. Vendors may quote any product solution or configuration so long as they meet the salient characteristics. Products quoted in the technical volume must be captured on the Vendor’s pricing volume (Attachment A, Price Cost Schedule) of the solicitation. Vendors may quote any additional ancillary products which they deem to be essential to the functionality of the proposed solution; these items must be captured in the Vendor’s pricing volume with a quantity of zero.
CLIN’s 0001-0007 apply to the following SC’s:
SC # SALIENT CHARACTERISTICS METHOD OF
EVALUATION
Page # Where SC is Found in Technical Literature
SC 1 Must be FDA approved for laparoscopic procedures to produce and maintain a pneumoperitoneum (gas-filled space within the peritoneal cavity)
Literature Review
SC 2 Must have smoke evacuation mode Literature
SC 3 Must use CO2 for medical gas. Literature
SC 4 Must be equipped with at least the following Alarms:
1. Excessive pressure
2. Tube obstruction
3. CO2 pressure
SC5 Must have 2 cavity modes: Normal and Small