Attachment 3 Temperature and Humidity Requirements.pdf

PDF 2 MB Posted

Attached to
Belcourt RTU and Cabinet – CT Area
Solicitation number
75H70124Q00025
Issued by

About this file

This document is a product data sheet for the GE BrightSpeed 16 CT Scanner System. It provides detailed specifications and requirements for the system, including physical dimensions, cooling and power requirements, temperature and humidity ranges, network connectivity, and compatible options.

The key details include:

  • Physical dimensions and weights for the gantry, table, power distribution unit, and operator consoles
  • Cooling and power requirements by subsystem, with minimum, recommended, and growth allowance figures
  • Temperature and humidity ranges for the exam/control rooms (60-75°F, 30-60% RH) and optional equipment room (60-84°F, 30-60% RH)
  • Standard networking capabilities, including DICOM, TCP/IP, and gigabit Ethernet
  • Available compatible options such as UIF, cables, desktop kits, and software packages

This is a technical specification document, not a contract solicitation or performance work statement. It does not contain information about federal procurement requirements, set-asides, pricing, or response dates.

View the file

Other files

File Type Posted
B05 Amendment 2.pdf PDF
B05 Amenment 1.pdf PDF
Attachment 7 Past Performance Questionnaire.docx DOCX document
Attachment 4 WH347 Payroll.pdf PDF
Attachment 5 Release of Claims.pdf PDF
Attachment 6 Relevant Experience Form.docx DOCX document
Attachment 1 Wage Determination ND20240029.pdf PDF
Attachment 8 Self Performed Calculation Sheet.docx DOCX document
B01 24Q00025.pdf PDF
Attachment 2 Design Package.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

BrightSpeed 16 (BrightSpeed Elite Select Edition-US naming)

Product Data Sheet - Ver?. May 2007

• 2 >< Color monitors:

- LCD Color Monitor

- 19 inch diagonal width

- 1280 x 1024 dot resolution

• Scan control keyboard (English language) assembly with intercom speaker. microphone and volume controls.

• Global modem• to allow lnSite connectivity

• 3-Button Mouse

• 3-Button Trackball*

Image Networking

• Standard auto-configuring l00BaseT/l0BaseT Ethernet (UTP connection).

• Supports gigabit ethernet capability.

• Direct network connection; multi-suite ethernet card not required for gateway out of suite

• Protocols supported:

- DICOM 3.0 network send (one IP address at a time) and receive, pull/query, and storage commitment push;

- lnSite point-to-point;

- TCP/IP (for system administration)

Compatible Options

• A DICOM Print Interface is standard on the system

• Chinese UIF & Keyboard

• English UIF &Keyboard

• French UIF & Keyboard

• German UIF& Keyboard

• Scandinavian UIF & Keyboard

• Short Cable

• Long Cable

• Single-piece desktop kit

• Side-by-side desktop kit

• Table Convenience Kit (IV Pole. Table Tray and other accessories)

• ConnectPro HIS/RIS Interface with Performed Procedure Step (PPS)

• Volume Viewer (including Virtual Endoscopy, 3D, Volume Rendering)

• DentaScan

• SmartScore

• CT Perfusion Neuro Package for Operator Console

• CT Perfusion Multi-Organ for Operator Console

• SmartStep (CT lnterventional kit)

• SmartView™ Fluoro

• AutoBone for CT Operator Console

• Advanced Vessel Analysis

• Coronal Head Holder

• Gantry Rear Raceway

• Exam Split FX

• UPS

* Option for Lite16 & Lite16 std.

**Option for Lite16 std. only

GE Healthcare

Siting Requirements.

Recommended Suite Size

Recommended Suite Size

Exam Room w/o Limited access option:

Minimum Size

3. 91m x 5.87 m 112' l0"x 19· 3")

Recommended Size:

3.96 m x 6.30 m (13' x 20 ' 8").

Exam Room w/ Limited access option:

Minimum Size

3.31 m x 5.92 m (10'10" x 19'5")

Control Room:

single piece desktop table:

2.74 m x 3.91 m (9' x 12'10") side-by-side desktop table.

3.86 m x 2.49 m (12'8" x 8'2")

Equipment Room: Not required. Power Distribution Unit heat and noise output under all operating conditions with all available options is low enough to allow inclusion in the patient scanning suite with the table and gantry.

Actual final room size determined by siting

Temperature and Humidity

Exam and Control Rooms: 60°-75°F ll5°-26°C) at 30%-60% relative humidity (non-condensing).

Equipment Room: If a separate equipment room is used to house the PDU. the allowable temperature range is 60°-84°F (15°-29°C) at 30%- 60% relative humidity !non-condensing).

Temperature Rate of Change: 3°C/hour max.

Relative Humidity Rate of Change: 5% RH/hour max.

Power Requirements

The only facility input to the system is 200 to 240 V and 380 to 480 V nominal, 3 phase Delta or Wye. 50/60 Hz. 50kVA / 75kVA** service. 20 kVA average power; main disconnect to be located within 5 feet (1.5 m) of the PDU. The facility must also provide a protective disconnect device with low voltage. low energy local and multi-point remote capability. in the line feeders to the PDU.

Complete. detailed specifications of all power requirements are available upon request. For most installations. the BrightSpeed 16 CT Scanner System does not require any power conditioning equipment to be used in conjunction with the PDU. Regulators are not recommended for use with this system. For those sites with known large power line transients. a suppresser filter for the system computer and peripherals may be useful. In general, suppresser filters are not recommended.

GE Healthcare - America: Milwaukee, USA

GE Healthcare - Asia: Tokyo, Japan; Beijing, Chino

GE Healthcare - Europe: Bue, Fronce;; Paris, Fronce • imagination at work HTR-HVL-CA002 Rev?

tastrand Highlight

BrightSpeed 16 (BrightSpeed Elite Select Edition-US naming) Product Data Sheet -Ver?. May 2007

Cooling Requirements

The cooling requirements do not include cooling for the room lighting.

personnel or non-CT equipment present. Cooling requirements are listed by subsystem to allow planning for each room of the CT suite.

Cooling requirements are given for minimum. recommended and growth allowance scenarios.

• The minimum cooling figures assume patient throughput of 3 patients per hour and 75 scan rotations per patient.

• The recommended cooling requirements assume patient throughput limited by the tube cooling algorithm.

• The suite cooling can be sized for future developments by using the growth allowance figures. This cooling will accommodate more patients per hour and/or potential future system enhancements.

Minimum Allowance

Subsystem Watts BTU/hr

Gantrv 5500 18766*** Table zoo 682

PDU 1500 5118

Ooerator Console 2400 8189

Recommended cooling values should not be used for calculating system input power requirements.

*** Recommended Allowance: 7,200/24.566 Growth Allowance: 9,825/33,520 s iystem C omponents o· 1mens1on Width+/. Depth+/-10% Height+/-10% 'eight+/-10% 10%

Gantrv 2040m 1018 mm 1930 mm 1770 kg Table 650 mn 2347mm*** 1043 mm 320 kg Power 700 mn 550 mm 1062 mm 350 kg Distribution

Unit

Operator 1240 1135-1235 680-885mm 195kg(W/0 Console with mm mm monitor, 1 IG) single-piece desktop

OperatorConsole 1240mrr 705mm 680-810mm 160kg(W/0 w/o single-piece monitor, 1 IG) desktop

Side-by-side deskt 1295mrr 660mm 686mm****- 39kg 1092mm o ***ELEC, up limit o ****Excluding monitor arm

Warranty

The published Company warranty in effect on the date of shipment shall apply. The Company reserves the right to make changes.

All specifications are subject to change.

* Option for Lite16 & Lite16 std.

**Option for Lite16 std. only

GE Healthcare

Regulatory Compliance

This product is designed to comply with applicable standards under the Radiation Control for Health and Safety Act of 1968.

Laser alignment devices contained within this product are appropriately labeled according to the requirements of the Center for Devices and Radiological Health.

CAUTION

LASER RADIATION

DO NOT STARE INTO BEAM

635,670 nm LASER DIODE

1.0 MILLIWATT MAXJMUM OUTPUT

CLASS II LASER PRODUCT

This product is a CE-compliant device which satisfies regulations regarding Electro-Magnetic Compatibility (EMC) and Electro-Magnetic Interference (EMI), pursuant to IEC-60601-1-2.

GE Healthcare - America: Milwaukee, USA

GE Healthcare - Asia: Tokyo, Japan ; Beijing, Chino

GE Healthcare - Europe: Bue, Fronce;; Paris, Fronce• imagination at work HTR-HVL-CA002 Rev?