Attachment 3_Requirements_and_Technical_Data_LargeTabletop.docx

DOCX document 22 KB Posted

Attached to
OCONUS Steam Sterilizers
Solicitation number
DHN-EURSterilizersFY26
Issued by

About this file

This is a technical specifications document for a large-capacity tabletop steam sterilizer (autoclave) that the Defense Health Agency seeks to procure for use at OCONUS facilities in the United Kingdom, Germany, and Turkey. The equipment must be FDA cleared/approved and CE marked as a medical device, with an exterior footprint not exceeding 26.0" W x 20.7" H x 34.5" D. The sterilization chamber must be at least 15 inches in diameter and 30 inches deep with a minimum volume of 84 liters, accommodating at least two trays of minimum dimensions 11" x 26" and 14" x 26", and supporting solid instrument loads of at least 14.0 lbs and textile/porous loads of at least 4.7 lbs.

The autoclave must feature pre-programmed sterilization cycles at 273°F (134°C) and 250°F (121°C) with dedicated programs for unwrapped instruments, wrapped instruments, porous loads, and glassware. Equipment must include closed-door forced-air drying with an air compressor and HEPA air filter (0.2 μm), with a drying cycle supporting a minimum of 30 minutes. Additional requirements include a minimum 7.5-liter mineral-free water reservoir compliant with EN 13060 water quality standards (maximum conductivity 15 μS/cm), cycle temperature documentation capability, safety interlocks preventing operation with improper door closure or insufficient water, system monitoring for cycle status and fault conditions, operation on 230 VAC single-phase 50/60 Hz drawing no more than 15 A, and a minimum 10-year warranty on the sterilization chamber with standard manufacturer warranty of at least two years for parts and labor.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sterilization Autoclave (Large-Capacity Tabletop)

SPECIFICATIONS

Sterilization Autoclave (Large-Capacity Tabletop)

MARKET RESEARCH NOTICE — NOT A SOLICITATION

1. General Description / Scope The requirement is for a large-capacity table-top steam sterilizer (autoclave) for sterilizing medical and surgical instruments in healthcare facilities, supporting unwrapped solid instruments, wrapped instruments, porous loads, and glassware, with automated closed-door forced-air drying.

2. Pertinent requirements and technical characteristics The equipment offered must meet or exceed the following:

1. Regulatory Approval: The equipment offered must be FDA cleared/approved AND CE marked as a medical device for steam sterilization use.

1. Maximum Outside Dimensions: The offered equipment's exterior footprint must NOT EXCEED 26.0" W x 20.7" H x 34.5" D (660 mm x 525 mm x 875 mm). Smaller or equal dimensions are acceptable; larger dimensions in any axis are not.

1. EU Water Quality Standard: The offered equipment's steam generator water supply must comply with EN 13060 water quality requirements for steam sterilization.

1. Cycle Temperature Documentation: The offered equipment must have the ability to produce a printout (or equivalent permanent, retrievable digital record) detailing the temperature achieved during each sterilization cycle, for record-keeping and validation purposes.

1. Any product must meet or exceed the following specifications.

Category
Parameter
Minimum Requirement / Specification
Physical Dimensions
Chamber Dimensions
Chamber must be at least 15 inches in diameter and 30 inches deep (approx. 38 cm x 76 cm), minimum volume of 84 liters.
Physical Dimensions
Exterior Dimensions
Exterior footprint must not exceed 26.0" W x 20.7" H x 34.5" D (660 x 525 x 875 mm).
Physical Dimensions
Tray Capacity
Must accommodate at least two trays with minimum dimensions of 11" x 26" and 14" x 26".
Physical Dimensions
Water Reservoir
Minimum mineral-free water reservoir capacity of 7.5 liters (2.0 US gal).
Performance
Load Capacity (Solid)
Minimum load capacity for solid instruments of 14.0 lbs (6.4 kg).
Performance
Load Capacity (Textile)
Minimum load capacity for textile/porous loads of 4.7 lbs (2.1 kg).
Sterilization Cycles
Temperature Profiles
Must provide pre-programmed cycles at 273°F (134°C) and 250°F (121°C).
Sterilization Cycles
Program Diversity
Must include dedicated programs for unwrapped instruments, wrapped instruments/porous loads, and glassware, plus an auxiliary drying-only program.
Air Removal & Drying
Drying Method
Must be equipped with an air compressor and HEPA air filter (0.2 µm) to perform a closed-door, forced-air drying cycle.
Air Removal & Drying
Drying Duration
Drying cycle time must support a minimum of 30 minutes.
Efficiency
Standby/Ready Mode
Must support a low-power standby heating mode between cycles to reduce subsequent cycle start-up time.
Water Quality
Feed Water Standard
Steam generator water supply must comply with EN 13060 water quality requirements (or equivalent), maximum conductivity 15 µS/cm.
Operation & Safety
System Monitoring
Must provide clear notifications for cycle status, including low water, low heat, low pressure, and high temperature conditions.
Operation & Safety
Safety Interlocks
Must prevent cycle start if the door is improperly closed or if water reservoir volume is insufficient.
Regulatory
Medical Device Certification
Must be FDA cleared/approved and CE marked as a medical device for steam sterilization; must comply with EN 13060 (Small Steam Sterilizers) and ISO 17665 (Moist Heat Sterilization).
Electrical
Power Supply
Must operate on 230 VAC, single-phase, 50/60 Hz, drawing no more than 15 A.
Warranty
Chamber Warranty
Sterilization chamber must carry a minimum 10-year warranty against defects in materials and workmanship.

Warranty and Support

· Standard manufacturer warranty (minimum of two years for parts and labor) is required.

Delivery / Schedule

· Standard delivery and setup required.