Attachment 3 - Hach CL17.pdf

PDF 576 KB Posted

Attached to
WATER TREATMENT PLANT (WTP) CLEARWELL REPAIR/REPLACEMENT CONSTRUCTION, GREAT LAKES, IL
Solicitation number
Not on record
Issued by

About this file

This is a product specification document for Hach's Single Parameter Monitoring Panel with SC4500 and CL17sc or ULRCL17sc water quality monitoring systems. The document details a preassembled water quality monitoring panel designed for applications in drinking water, wastewater, food and beverage, pharmaceutical, power generation, semiconductor, and field use settings.

The technical specifications include chlorine measurement ranges of 0-10 mg/L for the CL17sc model and 0.008-5 mg/L Cl2 for the ULRCL17sc model. The system features wall or rack mounting capabilities, ¼" OD tubing sample inlet connection, ½" NPT or hose barb process connection, and operates within sample temperatures of 5-40°C. The unit weighs 9 kg, measures 711 x 330 x 218 mm, requires 100-240 VAC power supply, and handles sample pressures of 4.5-75 psi with flow rates of 60-200 mL/min. Two model options are available for ordering: SPMP-CL (standard range) and SPMP-ULRCL (ultra-low range).

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Single Parameter Monitoring Panel with SC4500 and CL17sc or ULRCL17sc

Expertise + Simplicity

Quality from start to finish Our panel configurations allows you to maximize the inherent quality of the instruments, providing peace of mind while leveraging Hach’s 100 years of water quality innovations.

Up and running in no time Our preassembled, ready-to-use solutions streamline the process between purchasing and startup.

Ease of installation When you buy a water quality monitoring station from Hach, you’ll receive an expertly designed, preassembled panel that makes installation straightforward and user friendly.

Applications

• Drinking Water

• Wastewater

• Food and Beverage

• Pharmaceutical

• Power Generation

• Semiconductor

• Field Use

World Headquarters: Loveland, Colorado USA | hach.com United States 800-227-4224 fax: 970-669-2932 email: orders@hach.com Outside United States 970-669-3050 fax: 970-461-3939 email: intl@hach.com ©Hach Company, 2023. All rights reserved.

In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.

Single Parameter Monitoring Panel with SC4500 and CL17sc or ULRCL17sc

DOC053.53.35346.Dec24

Order Information Panel SPMP-CL Single Parameter Monitoring Panel with SC4500 and CL17sc SPMP-ULRCL Single Parameter Monitoring Panel with SC4500 and Ultra Low Range CL17sc

Technical Data* Range CL17sc: 0 - 10 mg/L free or total residual chlorine

ULRCL17sc: 0.008 - 5 mg/L Cl2 Controller SC4500 Mounting Wall or rack mounting Sample Inlet Connection ¼" OD tubing Process connection Sample drain: ½" NPT or hose barb Weight 9 kg / 20 lbs Dimensions (H x W x D) 711 mm x 330 mm x 218 mm Sample Temperature 5 - 40°C (41 - 104°F) Sample Pressure 4.5 - 75 psi Sample Flow Rate 60 - 200 mL/min.

Power Requirements (Voltage) 100-240 VAC ±10%, 50/60 Hz; 1 A Solution Type Drinking water, raw and treated

*Subject to change without notice.

Dimensions

File details come from the government source that posted it. Updated .