This is a specifications and salient characteristics document for a portable ultrasound system required by the Cheyenne Veterans Affairs Medical Center. The equipment must operate in multiple imaging modes including B Mode, M-Mode, Color Doppler, Color Power Doppler, Pulsed Wave, and Continuous Wave, with support for linear, phased, and curved array transducers and comprehensive measurement and calculation packages.
The system specifications establish strict dimensional and operational parameters: the stand dimensions must be 23 inches by 25.4 inches with maximum and minimum heights of 65 inches and 42.2 inches respectively, featuring a 9.8-inch height adjustment capability. The system weight cannot exceed 16.6 pounds without the stand or 85.8 pounds with the stand. The unit requires four 4-inch swivel casters with locking capability, a 10-inch by 5.7-inch projected capacitive touch screen interface supplemented by physical buttons, and a 15.6-inch clinical monitor with minimum resolution of 1920 x 1080 pixels and minimum luminance of 215 candlepower per square meter. Additional specifications include a work surface adjustable to 140 degrees, a monitor angle of 15 degrees from horizontal, a work surface angle of 55 degrees from horizontal, and an operating system running Windows 10 IoT Enterprise 2019 LTSC with digital broadband architecture.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| AMENDMENT Questions and Answers.docx | DOCX document | |
| 36C25926Q0751 0001.docx | DOCX document | |
| Attachment 1 - Combined Synopsis - Portable Ultrasounds.docx | DOCX document | |
| 36C25926Q0751.docx | DOCX document | |
| Attachment 3 - Buy American Certificate 52.225-2.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ATTACHMENT 2: Specifications and Salient Characteristics
The equipment must be capable of operating in B Mode, M-Mode, Color Doppler, Color Power Doppler (CPD), Pulsed Wave and Continuous Wave modes. It must support transducers with linear, phased, and curved array formats. Further, it must include a full array of measurement and calculation packages. The specific requirements for this equipment are the following:
System Specifications
1. Stand Dimensions (width x length): 23 inches x 25.4inches (58.4cm x 64.5cm)
2. Height Maximum (system and stand): 65 inches (165 cm) monitor up
3. Height Minimum (system and stand): 42.2 inches (107 cm) monitor down
4. Height Adjustment: 9.8” (24.9 cm) travel
5. Weight (System only) 16.6 lbs. / 7.5 kg
6. Weight (System and Stand) 85.8 lbs. / 38.9 kg
7. Casters: 4 inch Swivel, 4 Locking
8. User Interface: Projected Capacitive (PCAP) Touch Screen 10 inches x 5.7 inches (25.6 cm x
14.5 cm)
9. User Interface: Physical Buttons
10. Work Surface Adjusts: 0 to 140 degrees
11. Monitor Angle: 15 degrees from horizontal
12. Work Surface Angle: 55 degrees from horizontal
13. Clinical Monitor: 15.6 inches (39.6 cm)
14. Clinical Monitor Image Size: 8.1 inches x 6.1 inches (20.7 cm x 15.5 cm)
15. Clinical Monitor Display Size: 13.5 inches x 7.6 inches (34.4 cm x 19.4 cm)
16. Clinical Monitor Resolution (minimum):
a. Display Size: 1920 x 1080 pixels
b. Image Size: 1152 x 864 pixels
17. Clinical Monitor Luminance (minimum): 215cdm2
18. Operating System: Windows 10 IoT Enterprise 2019 LTSC
19. Architecture: Digital Broadband
20. System Boot up:
a. <60 second cold start
b. <20 seconds in Performance Mode
21. Total Hard Drive (HDD/SSD) Capacity: 640 GB
22. Gray Scale: 256 shades
23. Dynamic Range: Up to 183 dB
24. Transducer connection: 3 x system/stand, 1 x system only
25. RFID Filter on linear transducers
26. HIPAA Compliance: Meets Federal Information Processing Standards (FIPS) for Moderate security control selections per FIPS199, FIPS200, and NIST 800-53.
27. The system shall withstand a drop from a height of 1 meter. The system shall be safe and function normally after this drop.
Salient Characteristics Warranty & Service
1. Five-year standard warranty on the system.
2. Five-year standard warranty on the stand head.
3. Five-year standard warranty on transducers.
4. One-year standard warranty on the transesophageal transducer
5. One-year standard warranty on battery, stand and accessories carrying the manufacturer’s label excluding the system and stand head.
6. Provide a service loaner system within 24hrs where available (24-hour shipping not available in all geographical locations)
Imaging Modes
1. 2D/B-mode
a. Broadband Imaging
b. Tissue Harmonic Imaging
c. Pulse Inversion Harmonic Imaging
2. M-mode
3. Color Velocity Doppler (CVD)
4. Color Power Doppler (CPD)
5. Color Variance (Cardiac Exam Types)
6. Pulsed Wave Doppler (PW)
7. Tissue Doppler Imaging (TDI) (Cardiac Exam Types)
8. Continuous Wave Doppler (CW) (Cardiac Exam Types)
9. Triplex (Venous, Arterial and Cardiac Exam Types)
10. High Pulse Repetition Frequency (HPRF)
B-Mode Imaging
1. B- mode Controls:
a. Depth
b. Gain
c. Time Gain Compensation (TGC)
d. Auto Gain
e. SonoMB
f. 2D Mode THI
g. 2D Image Optimization (Res, Gen, Pen)
h. Dynamic Range (Live and Frozen)
i. Image Orientation (U/R, U/L, D/L, D/R)
j. Sector width control (Steerable, Cardiac Only)
k. Thermal Index (TIS/TIB/TIC)
l. LVO
m. Output Power Control
n. Trapezoid (Curvilinear and Linear Transducers)
2. Screen depth: 1 – 35cm
3. Support imaging frame rates up to 100Hz.
4. B-mode Imaging Features
a. speckle reduction algorithm.
b. Imaging Technology: Adjusts imaging parameters based on depth eliminating complicated manipulation of multiple controls
c. SonoMB: Multi-beam technology to effectively reduce speckle artifact, enhance tissue contrast, and improve border detection.
5. 8x live Read Zoom capability
6. Write Zoom on transducers (Abdomen, OB, Early OB, Gynecology, Prostate and Cardiac exam types only)
7. 4x Frozen post processing Zoom
8. Dual Imaging
9. Adjustable Focal Zones for Cardiac Transducers and Lung exam types of transducers
10. Centerline graphic on display for linear and curvilinear transducers
11. On screen needle guide support for transducers
12. Auto Gain adjust and reset
13. Procedure Mode
Auto Steep Needle Profiling Available on transducers and exams:
1. Arterial, MSK, Nerve, Superficial, Venous, PIV, Vascular Access
2. Arterial, Breast, Carotid, MSK, Nerve, Superficial, Venous, PIV
3. Arterial, Breast, Carotid, MSK, Nerve, Superficial, Venous
4. MSK, Nerve, Spine
5. MSK, Superficial, Venous, PIV
M-Mode Imaging
1. M-Mode Controls
1. Depth
2. Gain
3. Time Gain Compensation (TGC)
4. Auto Gain
5. SonoMB
6. THI
7. Image Optimization (Res, Gen, Pen)
8. Dynamic Range (Live and Frozen)
9. M-Mode Display Format- 1/3 and 2/3, 1/2 and 1/2, 2/3 and 1/3, side by side and full screen duplex
10. M-Mode Sweep Speed
11. Anatomical M-Mode on Cardiac Exam Types Color Imaging
1. Color modes:
a. Color Velocity Doppler
b. Color Power Doppler
c. Color Variance (Cardiac Only)
2. Color controls
a. Color Gain
b. Color Scale
c. Color Invert
d. Color Flow Optimization (High, Mid, Low)
e. Color Wall Filter
f. Color Steer Angle (Linear only)
g. Adjust Color box size and position
h. Color Baseline
i. Color Compare
j. Hide/Show Color
3. Max PRF used is 16667Hz.
Spectral Doppler Imaging
1. Doppler modes:
a. Pulsed wave (PW)
b. Tissue Doppler Imaging (TDI) (Cardiac Exam Types)
c. Continuous Wave (CW) (Cardiac Exam Types)
d. High Pulse Repetition Frequency (HPRF) (Cardiac Exam Types)
e. Triplex (Venous, Arterial and Cardiac Exam Types)
2. Doppler controls:
a. Doppler Gain
b. Doppler Gate Size (PW/TDI)
c. Doppler Invert
d. Doppler Auto Trace Type (Peak/Mean/Above/Below/All)
e. Doppler Display Format - 1/3 and 2/3, 1/2 and 1/2, 2/3 and 1/3, side by side and full screen duplex
f. Doppler Scale
g. Doppler Steer Angle (Linear Transducers only)
h. Doppler Sweep Speed
i. Doppler Baseline
j. Doppler Angle Correct
k. Doppler Wall Filter
l. Simultaneous (Linear Transducers only)
m. Doppler Volume
n. High Pulse Repetition Frequency (HPRF)
3. 256 gray levels in the spectral Doppler display
4. Cine review available
User Interface and Programmable Controls
1. Projected Capacitive (PCAP) Touch Screen User Interface
2. Physical Buttons: TGC (Time Gain Compensation), Gain Wheel, Auto Gain, Depth, Select, Update, Touchpad, Annotations, Calculations, Image Save, Clip Save, Freeze, Color, M-Mode, Doppler, and 2D.
3. Clinical Display Information
4. Configurable Start Screen: Start, Scanning, Transducer/Exam Selection, Patient Information
5. User Defined Exam Types: Imaging mode settings and on-screen exam specific optimization locations
6. Virtual QWERTY Keyboard
7. Voice Control of Select Functions Onboard Education
1. How-to videos: Imaging Basics, System Setup, Transducers, and System Use
2. Scan-along Educational Tutorials: Over 100 in English, French, German, Spanish, Japanese, Italian, and Mandarin
3. Tutorial Categories: Acute Care, Anesthesia, Pain Management, Obstetrics and Gynecology, Procedures, and Musculoskeletal Scan-along Educational Tutorials
• Anesthesiology & Pain Management o Abdomen: II/IH Block, Rectus Sheath, TAP, Quadratus Lumborum o Thorax: Paravertebral, Airway, Erector Spinae, PECS I & II o Lower Extremity: Adductor Canal, Fasica Iliaca, Femoral, Popliteal Sciatic, Saphenous, Subgluteal Sciatic o Upper Extremity: Axillary, Infraclavicular, Interscalene, Median & Ulnar, Supraclavicular
• Obstetrics & Gynecology o Gynecology: Ovarian Cyst-TA, Pelvis-TA o Obstetrics: 1st Trimester OB- TA, 1st Trimester OB- TV, AFI/Amniocentesis, Ectopic-TA, Ectopic-TV, Fetal Biometry, Fetal Viability
• Procedures o Vascular Access: Femoral Artery Access, Femoral Vein IJV Long, IJV Transverse, PICC, Radial Artery Access, Subclavian/Axillary o Other Procedures: Abscess Drainage, AFI/Amniocentesis, Foreign Body, Lumbar Puncture, Paracentesis, Pericardiocentesis, Thoracentesis
• Acute Care o EFAST: Lung linear array, Lung Phased Array, Left Upper Quadrant, Pelvis, Right Upper Quadrant, Subxiphoid o EFAST Pathology: Left Upper Quadrant Free Fluid, Pelvis Free Fluid, Right Upper Quadrant Free Fluid, Subxiphoid Free Fluid o Cardiac: Apical, PSAX Heart, PLAX Heart, Subxiphoid o Cardiac Pathology: Infective Endocarditis, Tamponade o Lung/Thorax + Pathology: Airway Lung, 12 Point Lung Exam, Pleural Effusion, Pneumonia, Pneumothorax, Pulmonary Edema o Abdomen: Aorta, Gallbladder Evaluation, Inferior Vena Cava, IVC Volume Status, Left Kidney, Right Kidney o Abdomen Pathology: AAA, Appendicitis, Cholecystitis, Cholelithiasis, Hydronephrosis, Pyloric, Stenosis, Renal Calculus o Pelvis: Bladder Volume, Pelvis-Transvaginal, Ovarian Cyst-TA, Testicular Torsion o Extremity: DVT, Long Bone Fracture, Tibia o Head/Neck + Pathology: Airway, Ocular, ICP, Retinal Detachment, Thyroid Evaluation
• COVID
o EFAST: Lung, Lung Phased, Left Upper Quadrant, Pelvis, Right Upper Quadrant, Subxiphoid o EFAST Pathology: Left Upper Quadrant Free Fluid, Pelvis Free Fluid, Right Upper Quadrant Free Fluid, Subxiphoid Free Fluid o Cardiac: Apical, PSAX Heart, PLAX Heart, Subxiphoid o Cardiac Pathology: Infective Endocarditis, Tamponade o Lung/Thorax + Pathology: Airway, Lung, 12 Point Lung Exam, Pleural Effusion Pneumothorax, Pulmonary Edema o Abdomen: Aorta, Inferior Vena Cava, IVC Volume Status, Left Kidney, Right Kidney o Head/Neck + Pathology: Airway o Extremity: DVT o Procedures: Pericardiocentesis, IJV Transverse, Subclavian Axillary
• Musculoskeletal o Shoulder/Rotator Cuff: Infraspinatus, Rotator Cuff, Rotator Cuff Tear, Tendinosis, Subscapularis Tendon, Supraspinatus Tendon o Shoulder - Other: AC Joint, AC Impingement, Biceps Tendon, Glenohumeral Joint, Posterior Groove o Shoulder Injections: AC Joint, Lateral Shoulder, Median Shoulder o Elbow: Anterior Recess, Cubital Tunnel, Distal Biceps Tendon, Elbow Olecranon, Elbow Ulnar Nerve, Lateral Aspect, Median Aspect o Elbow Pathology: Distal Biceps Tendon Tear, Lateral Epicondylitis, Ulnar Nerve Comp o Wrist/Hand + Pathology: Guyon’s Canal, Wrist Ulnar Nerve, Volar, Carpal Tunnel, DeQuervain’s, Median Nerve Entrapment o Hip, Pelvis Injections: Proximal Hamstring, Pubic Symphysis, Sacroiliac Joint o Knee: Iliotibial Band, Infrapatellar Tendon, LCL, MCL, Meniscus, Quadriceps Tendon, Tibia o Lower Extremity Injections: Achilles Tendon, Baker’s Cyst, Lateral Knee, MCL, Medial Knee, Patellar Tendinosis, Achilles Tendon Tear, Plantar Fasciitis o Ankle/Foot + Pathology: Ankle Ext Tendons, Ankle Flexor Tendons, MTP Joint, Tarsal Joint, Achilles Tendon Tear, Plantar Fasciitis How-to-Videos Transducers System Setup System Use Image Terms
• Transducer Types
• Orientation Marker
• Overview
• Getting Started
• Configuring System
• Adding Patient Information
• Performing an Exam
• Taking Images and Clips
• Annotating Images
• Taking Measurements
• Managing Data
• Terminology Onboard Image and Clip Storage Review
1. 2D Cine Review – minimum 10 seconds (or until number of frames fill up the 384 MB cine memory)
2. PW, CW, M Mode Cine Review – minimum 2 cine frames (slow sweep 23 seconds, fast sweep 6 seconds)
3. Internal Memory – 128 GB
4. 700 maximum number of studies (average of 30 clips/images per study)
5. 500 maximum number of clips/images per study
6. Thumbnail review of saved images and clips
7. Prospective and Retrospective Clip Store
8. Video clip playback at 1, ½ or ¼ of the captured rate
9. Video Clip Save Lengths: 2, 3, 4, 6, 10, 15, 30 and 60 seconds. (Retrospective/Prospective)
10. Image/Clip export format: JPEG, BMP and MP4
11. Report export Format: PDF
12. DICOM export to USB: JPEG, BMP and MP4
13. JPEG compression options: High, Medium and Low Worksheets
1. Acute Care Worksheets: Aorta, Appendix, Bladder, Cardiac, Deep Vein Thrombosis (DVT), EFAST, FAST, Gallbladder, Generic, Gynecology, Musculoskeletal, Obstetrics 2nd /3rd trimester, Obstetrics 2nd /3rd trimester Multiple, Obstetrics, Ocular, Renal, Soft Tissue, and Thoracic.
2. Procedures: Abscess Drainage, Arterial Line, Arthrocentesis, Central Venous Line, Paracentesis, Pericardiocentesis, Peripheral Venous Line, Peritonsillar Abscess Drainage
3. Customizable worksheets: Available with Synapse Synchronicity workflow manager integration Labelling
1. Predefined labels and pictograms
2. Customizable labels
3. On-screen keyboard label entry
4. Arrows (Up to 5)
5. Labeling of saved images Labelling Packages
• MSK, Foot, Ankle, Knee, Hip, Shoulder, Elbow, Wrist, Hand, Spine, Nerve
• Breast, Superficial, Ophthalmic
• Transcranial, Carotid, Arterial, Venous, PIV
• Abdomen, Lung, Focused Cardiac, Cardiac
• Gynecology, Early OB, Focused OB, OB, Prostate Basic Measurements
1. 2D Measurements: Distance, Curved Distance, Area, Circumference, Ellipse, Manual Trace, Volume, Target Depth, Angle, Volume Flow Diameter
2. Doppler Measurements: Velocity, 2-point velocity, Slope, Time, Manual Trace, Auto Trace, Heart Rate, Resistive Index, Pulsatility index, Systolic/Diastolic Ratio, Volume Flow, Velocity Time Integral, TAM, TAP
3. Doppler measurements can be traced manually or automatically.
4. Automatic trace results (determined by exam type): Velocity Time Integral, Peak Velocity, Mean Pressure Gradient, Mean Velocity on Peak Trace, Pressure Gradient, Cardiac Output, Peak Systolic Velocity, Time Average Mean, Systolic/Diastolic Ratio, Pulsatility Index, End Diastolic Velocity, Acceleration Time, Resistive Index, Time Average Peak, Gate Depth, Heart Rate.
5. M-mode Measurements: Distance, Time, Slope, Heart Rate
6. Measurement on saved images Calculation Packages Abdomen
1. 2D Calculations and Measurements Bladder Volume, Post-Void Bladder Volume, IVC Collapse, Aorta, CBD, CHD, Gallbladder, Liver, Spleen and Renal
2. Doppler Calculations: Renal Aortic Ratio
3. M-Mode Calculations: IVC Collapse Cardiac
1. 2D Cardiac Calculations and Measurements: Teichholz Ejection Fraction, Fractional Shortening, Simpson's method of discs for single and bi-plane ventricular volumes and ejection fraction, Left and Right Fractional Area Change, Planimetry for Valve Area (AVA, MVA), Vena Contracta (AV, MV), LV Mass, LV Mass Index, LA/Ao, IVSd/LVPWd, IVSs/LVPWs, IVS FT, LVPW FT, LA Single and Biplane Volume, LA Biplane Volume Index, LA Area, RA Simpson's Volume, RA Volume Index, LV Dimensions, RV Dimensions, Aorta Dimensions, IVC Collapse, IVC Distensibility Index
2. M-Mode Cardiac Calculations and Measurements: Teichholz Ejection Fraction (EF), Fractional Shortening, LV Dimensions, TAPSE, MAPSE, EPSS, IVC Collapse, IVC Distensibility Index, DE Slope, E-F Slope, Aortic Cuspal Separation, LVET, IVS FT, LVPW FT, LV Mass, LV Mass Index, IVSd/LVPWd, IVSs/LVPWs, LA/Ao, Heart Rate
3. Doppler Cardiac Calculations and Measurements: Cardiac Output, Cardiac Index, Stroke Volume, Stroke Index, Isovolumic Relaxation Time, Aortic Valve Area (By VTI and Vmax), Aortic Valve Area Index, AV Velocity Ratio, Mitral Valve Area (by PHT and VTI), Pressure Half Time (MV, TV, AI), Right Atrial Pressure, Right Ventricular Systolic Pressure (RVSP), Pulmonary SV, Deceleration Time, Delta Pressure: Delta Time (dP: dT), Pulmonary AT(Acceleration Time), PV VTI, PV VMax, TV VTI, TV E and A, Qp/Qs, RIMP, MV E/A, Sep E/e', Lat E/e', Pulmonary vein velocities (S, D, A, Adur), VTI % Change, SV % Change, CO % Change, VMax % Variation, VTI % Variation
4. Tissue Doppler Calculations and Measurements: Sep e’, Sep a’, Lat e’, Lat a’, Ant e’, Inf e’, Sep E/e’, Lat E/e’, RV s’ OB/Gyn
1. 2D Gynecology Calculations and Measurements: Ovary Volume (right and left), Endometrium, Uterus Volume, 10 Follicular Measurements, (right and left)
2. 2D Early Obstetric Calculations and Measurements: YS, Mean Gestational Sac, NT, CRL, BPD, Myometrial Mantle, Cervical Length, EDD by LMP, EDD by AUA, GA by EDD, AUA, Ovary Volume (right and left)
3. 2D Obstetric Calculations and Measurements: BPD, OFD, HC, AC, FL, Cerebellar Diameter, Cisterna Magna, OOD, IOD, Lateral Ventricle, Cervical Length, Amniotic Fluid Index, EDD by LMP, EDD by AUA, GA by LMP, GA by EDD, AUA, EFW, EFW Percentile, HC/AC, FL/BPD, FL/HC, Cephalic Index (HC)
4. M-Mode Calculation: FHR
5. Doppler Obstetric Calculations: MCA Trace (RI, PI, S/D, EDV, PSV), Umbilical Artery Trace (RI, PI, S/D, EDV, PSV), Umbilical Artery S/D Ratio (S/D, PSV, EDV, RI), FHR
6. Biophysical Profile
7. User-Selected Gestational Age References: GS, CRL, BPD, OFD, HC, AC, FL
8. Multiple Gestation Package: Quads Small Parts/MSK
1. 2D Calculations: d:D Hip Ratio, Hip Angles Transcranial Doppler (TCD)
1. Doppler Basic Measurements: TAP, PSV, EDV, MDV, RI, PI, S/D Ratio, Gate Depth Vascular
1. Doppler Basic Measurements: Volume Flow including VTI, Peak Systolic Velocity, End Diastolic Velocity, Resistive Index, Pulsatility Index, S/D Ratio, Time Average Peak Velocity
2. 2D Calculations: Catheter to Vessel Ratio (CVR) Carotid
1. Doppler Basic Measurements: Volume Flow including VTI, Peak Systolic Velocity, End Diastolic Velocity, Resistive Index, Pulsatility Index, S/D Ratio, Time Average Peak Velocity
2. 2D Carotid Measurements and Calculations: CCA Diameter
3. PW Carotid Measurements and Calculations: CCA VTI, PSV, EDV, RI, Time, Heart Rate, Carotid Blood Flow VExUS Scoring
1. 2D Calculations: IVC diameter, IVC assessment, Hepatic vein assessment, Portal vein assessment, Renal vein assessment.
2. Doppler Calculation: Portal vein pulsatility fraction DICOM Image Management
1. Print, Store, Modality Worklist, Perform Procedure Step (PPS), Storage Commitment
2. Comprehensive Structured Reports (SR)
3. DICOM USB export
4. Allow saving DICOM configuration via USB for easy replication or restoration.
5. DICOM 3.0 compliant
6. Allow selection of Archiver on completion of exam Connectivity and External Data Management
1. 4 USB 2.0 Ports (stand only)
2. 2 USB 3.0 Ports (system only)
3. 1 USB Port for optional printer
4. Ethernet Port
5. 1080p digital video output (compatible with 1080p displays) – stand only
6. Storage capacity alert if internal storage is less than 10%
7. Auto Delete function
8. Automatic time server reconciliation
9. IPv4 and IPv6 Networking Optional Medical Informatics
1. Barcode Expressions: barcode scanner customization software (English only)
2. Integration of optional Synapse Synchronicity workflow manager for credentialing, POCUS billing, and customizable worksheets
3. Onboard Qview application for Qpath E Wireless Solutions
1. 2.4GHz and/or 5GHz Embedded Wireless Option:
a) Wireless Protocols supported: 802.11a/b/g/n/ac
b) Wireless Roaming supported: 802.11k and 802.11r (with FT-802.1x authentication only / without support for FT-PSK)
c) Shared Key authentication: 802.11i (with the exception of WPA3)
2. Key management protocols: WEP64, WEP128, and WPA/WPA2 (Personal and Enterprise)
3. FIPS 140-2 wireless security protocol Data Security
1. System meets Federal Information Processing standards (FIPS) for moderate security control selections per FIPS199, FIPS200, and NIST 800-53
2. Secure Boot
3. User and role-based authentication
4. LDAP support
5. Security Policies: WEP, Radius, WPA, WPA2)
6. Complex Password Rules: Complex passwords, minimum password length, automatic password aging, and password history re-use limit rules
7. Federal System Use Notification
8. Password Protection: username and password capability
9. Emergency Mode Support
10. Secure DICOM communication support
11. Data encryption for data at rest and in transit: FIPS 140-2
12. System wide white-listing
13. Hardened OS and services
14. System firewall
15. MDS2
Admin User Settings
1. Manage, import, create new, edit and delete user profiles
2. Connectivity settings including network status
3. Configure connection to directory server, manage password requirements
4. Data security mode, data import and export control, create notifications
5. Audio and display settings
6. Location, date and time
7. Power and battery settings
8. System and Log Information Stand & Storage
1. Adjustable stand height and stand head
2. 4 transducer holders
3. Triple Transducer Connect
4. Stand Battery
5. Dedicated holders for disinfectant wipes and gel
6. Large storage basket
7. Lockable drawer
8. AC cord retainer
9. I/O connections Optional Accessories
1. Stand with stand battery
2. ECG Module with patient leads and accessory cable with adaptor
3. Medical grade B&W printer
4. Barcode Scanner
5. Voice Assist Microphone
6. PowerPark Battery Management
1. System operates via battery or AC power
a) Battery charge indicator light on lid
b) Battery charge indicator on clinical monitor
c) Battery Charge Warning: At 5 minutes of charge left or lower
2. System Battery and Stand Battery
a) Battery Life: 3.0 hours
b) Battery Charging Time: charges from 15% to 80% in 2 hours 35 minutes
3. System Battery
a) Battery: rechargeable lithium-ion
b) Battery Capacity: 129.6 Watt hours
c) Battery Life (system only): 1.0 hour
d) Battery Charging Time (system only): charges from 15% to 80% in 90 minutes
e) Battery Lifecycle: 80% capacity after 500 charge cycles
4. Stand Battery
a) Battery: rechargeable lithium-ion
b) Battery Capacity: 259.2 Watt hours
c) Battery Life: 2.0 hours
d) Battery Lifecycle: 80% capacity after 500 charge cycles Portable Power Supply
1. Power Input: 100-240 VAC, 50-60 Hz, 3.4 – 1.3 A
2. Power Output: 26.7 VDC, 8.24 A, 220 Watts max; Class 1, continuous use Stand Rating
1. Power Input: 100-240 VAC, 50-60 Hz, 6.0-2.5 A
2. Power Output: (for optional printer): 100-240 VAC, 50-60 Hz, 2.5-1.0 A Heat Output
1. While operating 120W - 409.2 BTUs
2. While operating and charging (max power) 200W - 682 BTUs Infection Control Standards
1. Cleaners and Disinfectants Tool
2. Cleaning Mode available for ease of disinfection
3. System is IP22 rated for liquid ingress
4. System work surface is sealed edge-to-edge
END SPECIFICATIONS AND SALINET CHARACTERISTICS