A.02 Statement of Work PD - Bonn Ultrasonic Flow Meter.pdf
PDF 223 KB Posted
- Attached to
- Bonneville Ultrasonic Flow Meter
- Solicitation number
- W9127N26QA166
About this file
Purchase Description Summary
This purchase description outlines the procurement of one Emerson Flexim FLUXUS G631EE Portable Gas Clamp-on Ultrasonic Flow Meter System for the U.S. Army Corps of Engineers' Bonneville Lock and Dam project in Portland District. The contract is a firm fixed price purchase with a 30-day delivery requirement following contract award. The system is designed for non-invasive measurement of both gas and liquid flow rates with integrated thermal energy computation capabilities.
The G631EE transmitter kit (CLIN 0001) is the primary component, featuring dual-channel measurement capability, an integrated thermal energy computer, IP65/IP67 environmental protection, and a high-capacity data logger storing up to 800,000 data points. The unit operates on a rechargeable 7.2V/10Ah Lithium-Ion battery (25+ hours continuous operation) or 100-240V AC mains power and includes a transport case, charger, USB cables, documentation, parameterization software, and three ruggedized NY connection adapters. Additional CLINs (0002-0008) specify matched pairs of specialized transducers for both gas and liquid applications across various pipe diameters and frequencies (ranging from 0.5 MHz to 8 MHz), each with factory calibration, coupling materials, and transducer storage boxes. Support CLINs (0009-0014) provide mounting fixtures, fastening kits, temperature probes, ultrasonic wall thickness probe, and damping material for acoustic noise reduction. CLIN 0015 allows for tariff surcharge application if applicable. All items carry a minimum one-year manufacturer warranty. Delivery must be via Carriage Paid To (CPT Incoterms 2020) to Bonneville Lock and Dam in Cascade Locks, Oregon, during Monday-Thursday business hours (7:00 a.m.-4:00 p.m. PST). Invoices should be submitted to the USACE Finance Center in Willington, Tennessee, with electronic copies to specified email addresses.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Purchase Description Bonneville Lock and Dam
Emerson Flexim FLUXUS G631EE Portable Gas Flow Meter System
1. General Portland District, U.S. Army Corps of Engineers (COE), Bonneville Lock and Dam project requires the purchase of one (1) Emerson Flexim FLUXUS G631EE Portable Gas Clamp-on Ultrasonic Flow Meter System, complete with required accessories, clamp-on transducers, and specialized probes. This is a firm fixed price contract. The delivery date is 30 days after receipt of award.
Item Description 0001 FLUXUS G631EE Dual-Channel Portable Flow & Energy
Transmitter Kit
0002 Shear Wave Liquid Transducers (8 MHz, S Freq) with Coupling Compound
0003 Shear Wave Gas Transducers (4 MHz, Q Freq) with Coupling Foil & Compound
0004 Shear Wave Gas Transducers (2 MHz, P Freq) with Coupling Foil & Compound
0005 Shear Wave Gas Transducers (1 MHz, M Freq) with Coupling Foil & Compound
0006 Lamb Wave Gas Transducers (1 MHz, M Freq) with Coupling Foil & Compound
0007 Lamb Wave Gas Transducers (0.5 MHz, K Freq) with Coupling Foil & Compound
0008 Shear Wave Liquid Transducers (0.5 MHz, K Freq) with Coupling Foil & Compound
0009 Transducer Mounting Fastening Shoe Kit
0010 Portable "Variofix VM" Transducer Mounting Fixture
0011 RTD Strap Fastener Kit for Temperature Probes
0012 Matched Pair Pt1000 Clamp-On Surface Temperature Probes
0013 Ultrasonic Wall Thickness Probe
0014 Pipe Damping Material Kit
0015 Tariff Surcharge (if applicable, merge this pricing into overall contract cost).
2. Salient Characteristics
CLIN 0001: FLUXUS G631EE Dual-Channel Portable Flow & Energy Transmitter Kit
A complete, high-performance portable clamp-on ultrasonic flow and energy transmitter kit designed for the non-invasive measurement of both gas and liquid flow rates and thermal energy computation.
Standard Equipment and Capabilities:
• Specification & Part No: G631EE FLUXUS Transmitter, for the measurement of gases (Part No. DE6-G631EE-NNNNN-2PG3-
NN.CS2.MP2.TT2.IS2/FDG).
• Dual-Channel Measurement: Must feature two (2) independent, synchronized measurement channels to monitor multiple pipes or conduct multi-path averaging.
• Non-Invasive Technology: Uses clamp-on transit-time difference correlation measurement for all acoustically conductive gases and liquids.
• Integrated Thermal Energy Computer: Incorporates calculations for standard volumetric flow rates, thermal energy loads, and heating/cooling rates.
• Environmental Protection: Rated to IP65/IP67 ingress protection, enclosed in a robust glass-fiber reinforced polyamide housing.
• Operating Range: Designed for ambient operation from -20°C to +60°C (-4°F to +140°F).
• Data Logging: High-capacity integrated logger storing up to 800,000 physical and diagnostic data points.
• Inputs and Outputs (Galvanically Isolated):
• Two (2) switchable active/passive current outputs (4-20 mA).
• Two (2) digital outputs (frequency, pulse, or binary).
• Two (2) Pt100/Pt1000 temperature inputs (4-wire connection).
• Two (2) passive current inputs (0/4-20 mA).
• Interfaces: Integrated USB-C service port and USB-A host port.
Standard Accessories Included:
• Transport Case: One (1) ultra-rugged, watertight, impact-resistant carrying case custom-molded for the G631 transmitter, specialized probes, and transducer accessories.
• Charging System: One (1) 100-240 V AC (50/60 Hz) AC battery charger with standard USA mains power cable.
• Data Transmission: One (1) USB-C to USB-A interface connection cable.
• Mounting Hook: One (1) Quickfix pipe mounting hook for temporary pipe mounting of Flexim FLUXUX 63x (belt not included)
• Measuring Tape: One (1) FLEXIM 9-foot measuring tape displaying metric, inch, and foot scales.
• Documentation and Software: One (1) English printed operating manual and one (1) USB stick containing the electronic manual and FluxDiag parameterization software.
• Ruggedized "NY" Connection Adapters:
• One (1) Output Adapter AO62 (Part No. 63101-2) for signal output distribution.
• One (1) Temperature Probe Input Adapter AI61 (Part No. 63103-1) for dual Pt100/Pt1000 RTD connections.
• One (1) Passive Current/Voltage Input Adapter AI62 (Part No. 63103-2).
Power Requirements:
• Must operate on both a rechargeable, high-capacity internal Lithium-Ion battery (7.2 V / 10 Ah, 72 Wh) providing over 25 hours of continuous operation (up to 80 days in low-power mode) and 100-240 V AC (50/60 Hz) mains power.
CLIN 0002: Shear Wave Liquid Transducers (8 MHz, S Freq)
One (1) matched pair of portable clamp-on transducers optimized for liquid flow measurement in small pipe applications.
Standard Equipment and Capabilities:
• Specification & Part No: Shear Wave Liquid Transducers, S Frequency, 8 MHz (Part No. TRD-FSS-NNNN-NNNY-002).
• Pipe Diameter Range: Recommended for inner pipe diameters of 0.39" to 2.8" (extended range covers 0.24" to 2.8").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory flow calibration with at least seven (7) measurement points, supplied with traceable flow calibration certificates.
• Consumables Included: One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +266°F (Part No. 990739-1).
CLIN 0003: Shear Wave Gas Transducers (4 MHz, Q Freq)
One (1) matched pair of portable clamp-on transducers optimized for gas flow measurement in small pipe applications.
Standard Equipment and Capabilities:
• Specification & Part No: Shear Wave Gas Transducers, Q Frequency, 4 MHz (Part No. TRD-GSQ-NNNN-NNNY-003).
• Pipe Diameter Range: Recommended for inner pipe diameters of 0.39" to 0.87" (extended range covers 0.28" to 1.2").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory aperture calibration with at least six (6) measurement points, supplied with traceable calibration certificates.
• Consumables Included:
• One (1) VT-Q Coupling Foil (Part No. 990739-5) for Q-frequency transducers.
• One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +392°F (Part No. 990739-2).
• Transducer Box: One (1) Transducer box, small (M-Q) for portable transducers size M to Q.
CLIN 0004: Shear Wave Gas Transducers (2 MHz, P Freq)
One (1) matched pair of portable clamp-on transducers optimized for gas flow measurement in low-to-medium diameter applications.
Standard Equipment and Capabilities:
• Specification & Part No: Shear Wave Gas Transducers, P Frequency, 2 MHz (Part No. TRD-GSP-NNNN-NNNY-004).
• Pipe Diameter Range: Recommended for inner pipe diameters of 0.79" to 2.0" (extended range covers 0.59" to 2.4").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory aperture calibration with at least six (6) measurement points, supplied with traceable calibration certificates.
• Consumables Included:
• One (1) VT-M Coupling Foil (Part No. 990739-6) for P-frequency transducers.
• One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +392°F (Part No. 990739-2).
• Transducer Box: One (1) Transducer box, small (M-Q) for portable transducers size M to Q (Part No. 63302-1).
CLIN 0005: Shear Wave Gas Transducers (1 MHz, M Freq)
One (1) matched pair of portable clamp-on transducers optimized for gas flow measurement in mid-sized pipe applications.
Standard Equipment and Capabilities:
• Specification & Part No: Shear Wave Gas Transducers, M Frequency, 1 MHz (Part No. TRD-GSM-NNNN-NNNY-004).
• Pipe Diameter Range: Recommended for inner pipe diameters of 1.6" to 5.9" (extended range covers 1.2" to 7.1").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory aperture calibration with at least six (6) measurement points, supplied with traceable calibration certificates.
• Consumables Included:
• One (1) VT-M Coupling Foil (Part No. 990739-6) for M-frequency transducers.
• One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +392°F (Part No. 990739-2).
• Transducer Box: One (1) Transducer box, small (M-Q) for portable transducers size M to Q (Part No. 63302-1).
CLIN 0006: Lamb Wave Gas Transducers (1 MHz, M Freq)
One (1) matched pair of high-efficiency Lamb wave clamp-on transducers optimized for gas flow measurement in mid-sized pipes.
Standard Equipment and Capabilities:
• Specification & Part No: Lamb Wave Gas Transducers, M Frequency, 1 MHz (Part No. TRD-GLM-NNNN-NNNY-004).
• Pipe Diameter Range: Recommended for inner pipe diameters of 1.6" to 5.9" (extended range covers 1.2" to 7.1").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory aperture calibration with at least six (6) measurement points, supplied with traceable calibration certificates.
• Consumables Included:
• One (1) VT-M xL Coupling Foil (Part No. 990739-14) for GLM Lamb wave transducers.
• One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +392°F (Part No. 990739-2).
• Transducer Box: One (1) Transducer box, small (M-Q) for portable transducers size M to Q (Part No. 63302-1).
CLIN 0007: Lamb Wave Gas Transducers (0.5 MHz, K Freq)
One (1) matched pair of high-efficiency Lamb wave clamp-on transducers optimized for gas flow measurement in larger pipe applications.
Standard Equipment and Capabilities:
• Specification & Part No: Lamb Wave Gas Transducers, K Frequency, 0.5 MHz (Part No. TRD-GLK-NNNN-NNNY-005).
• Pipe Diameter Range: Recommended for inner pipe diameters of 3.1" to 11.8" (extended range covers 2.4" to 14.2").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory aperture calibration with at least six (6) measurement points, supplied with traceable calibration certificates.
• Consumables Included:
• One (1) VT-K Coupling Foil (Part No. 990739-0) for GLK Lamb wave transducers.
• One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +392°F (Part No. 990739-2).
• Transducer Box: One (1) Transducer box, large (G-Q) for portable transducers size G to Q (Part No. 63302-2).
CLIN 0008: Shear Wave Liquid Transducers (0.5 MHz, K Freq)
One (1) matched pair of portable clamp-on transducers optimized for liquid flow measurement in large-diameter applications.
Standard Equipment and Capabilities:
• Specification & Part No: Shear Wave Liquid Transducers, K Frequency, 0.5 MHz (Part No. TRD-FSK-NNNN-NNNY-005).
• Pipe Diameter Range: Recommended for inner pipe diameters of 7.9" to 78.7" (extended range covers 3.9" to 255.9").
• Connection System: Heavy-duty, water-tight NY connector design to prevent water ingress and corrosion in wet turbine galleries.
• Calibration: Must undergo factory flow calibration with at least seven (7) measurement points, supplied with traceable flow calibration certificates.
• Consumables Included:
• One (1) VT-K Coupling Foil (Part No. 990739-0) for G,H,K transducers.
• One (1) 0.03 gal tube of N-Type Acoustic Coupling Compound, temperature range -22°F to +392°F (Part No. 990739-2).
• Transducer Box: One (1) Transducer box, large (G-Q) for portable transducers size G to Q (Part No. 63302-2).
CLIN 0009: Transducer Mounting Fastening Shoe Kit
One (1) set of transducer mounting fastening shoes designed for small transducers.
Standard Equipment and Capabilities:
• Specification & Type: Fastening shoes 601 kit including rail (Part No. MOU-
FSS-DS-C0100).
• Fixation: Equipped with ball chains, optimized for securing S-frequency (8 MHz) transducers to pipes.
CLIN 0010: Portable "Variofix VM" Transducer Mounting Fixture
One (1) transducer mounting fixture designed for small-to-medium transducers.
Standard Equipment and Capabilities:
• Specification & Type: Portable "Variofix VM" mounting rail and fastening shoe kit (Part No. MOU-VMA-DM-G1500).
• Fixation: Equipped with tension belts, optimized for securing all sizes of transducers to pipes.
• Variofix M Foam Insert: Variofix M foam packaging insert for use with Flexim FLUXUS 63x carrying case (Part No. 63303-1)
CLIN 0011: RTD Strap Fastener Kit for Temperature Probes
One (1) strap fastener kit used to secure temperature probes to pipe surfaces.
Standard Equipment and Capabilities:
• Specification & Part No: Strap Fastener Kit (Part No. US-MOU.SK2).
• Application range: Specifically designed strap mounting kit for temperature probes, compatible with pipe sizes from 3" to 6" in outer diameter.
CLIN 0012: Matched Pair Pt1000 Clamp-On Surface Temperature Probes
One (1) matched pair of clamp-on surface temperature probes used to measure thermal energy parameters.
Standard Equipment and Capabilities:
• Specification & Part No: Pair of Pt1000 clamp-on surface probes (Part No.
PTN13NY / 630412-1).
• Accuracy: 4-wire connection, Class A accuracy, matched in accordance with DIN EN 1434-1 to ensure precise thermal energy calculations.
• Connection System: Fitted with heavy-duty, circular NY-compatible connectors for standard connection to the AI61 transmitter input adapter.
CLIN 0013: Ultrasonic Wall Thickness Probe
One (1) ultrasonic wall thickness probe used to accurately determine the pipe wall thickness parameter on-site.
Standard Equipment and Capabilities:
• Specification & Part No: Ultrasonic Wall Thickness Probe (Part No.
DWR1NZ4 / 630522-0).
• Performance Range: Measuring range of 0.04" to 9.8" (1 to 250 mm) with a resolution of 0.01 mm and accuracy of 1% ±0.1 mm.
• Temperature Limits: Operating fluid/surface temperature limits of -20°C to +200°C.
• Functionality: Direct connection to the transmitter for automatic thickness detection, equipped with a 4.9 ft (1.5 m) cable and standard NY connection compatibility.
CLIN 0014: Pipe Damping Material Kit
One (1) roll and installation kit of damping material for gas flow measurement.
Standard Equipment and Capabilities:
• Specification & Part No: Pipe Damping Material Kit (Part No. US-
DMPKIT.1).
• Functionality: Specifically utilized in gas measurements to reduce acoustic cross-talk, pipe-wall noise, and extraneous vibrational signal drift.
• Components Included: Includes one (1) roll of self-adhesive damping material (layer thickness 0.03" / 0.8 mm) and one (1) ergonomic rolling installation tool.
CLIN 0015: Surcharge One (1) surcharge for tariff fees.
• Specification & Part No: Service charge for tariff fees (Part No. IMP-
TARIFF-F).
3. Warranty
All items shall be covered by the manufacturer’s standard warranty on all parts, labor, and workmanship from factory defects. Warranty period shall be a minimum of one year.
4. Shipment and Delivery
Deliver Carriage Paid To (CPT Incoterms 2020) to the project address listed below no later than 30 days after contract award. The contractor shall be responsible for packing and shipping the components in a manner that will ensure undamaged and complete arrival to the powerhouse. No shipments are allowed other than normal business hours of Monday-Thursday 7:00 a.m. to 4:00 p.m. PST, excluding federal holidays. Please note, if a federal holiday occurs on a Sunday, it is observed the Monday after.
All shipments shall be delivered to:
US Army Corps of Engineers Bonneville Lock and Dam Interstate 84 Exit 40 NE Bonneville Way Cascade Locks, OR 97014
5. Invoice Information
Invoices shall be submitted to the following:
USACE FINANCE CENTER
5722 INTEGRITY DRIVE
WILLINGTON, TN 38054-5005
Electronic copy also sent to USACE, Finance Center at:
CEFC-G2INVOICES@USACE.ARMY.MIL and roger.a.montavon@usace.army.mil
6. Points of Contact (POC)
Technical POC: Tyler Comstock, tyler.comstock@usace.army.mil Delivery POC: Roger Montavon, roger.a.montavon@usace.army.mil
Phone: (541) 374-3616 or (541) 419-7052 Contracting Officer: To be determined Contract Specialist: John Scukanec, john.r.scukanec@usace.army.mil mailto:CEFC-G2INVOICES@USACE.ARMY.MIL mailto:roger.a.montavon@usace.army.mil mailto:tyler.comstock@usace.army.mil mailto:roger.a.montavon@usace.army.mil
| Purchase Description |
| Bonneville Lock and Dam |
| Emerson Flexim FLUXUS G631EE Portable Gas Flow Meter System |
| 1. General |
| 2. Salient Characteristics |
| CLIN 0001: FLUXUS G631EE Dual-Channel Portable Flow & Energy Transmitter Kit |
| CLIN 0002: Shear Wave Liquid Transducers (8 MHz, S Freq) |
| CLIN 0003: Shear Wave Gas Transducers (4 MHz, Q Freq) |
| CLIN 0004: Shear Wave Gas Transducers (2 MHz, P Freq) |
| CLIN 0005: Shear Wave Gas Transducers (1 MHz, M Freq) |
| CLIN 0006: Lamb Wave Gas Transducers (1 MHz, M Freq) |
| CLIN 0007: Lamb Wave Gas Transducers (0.5 MHz, K Freq) |
| CLIN 0008: Shear Wave Liquid Transducers (0.5 MHz, K Freq) |
| CLIN 0009: Transducer Mounting Fastening Shoe Kit |
| CLIN 0010: Portable "Variofix VM" Transducer Mounting Fixture |
| CLIN 0011: RTD Strap Fastener Kit for Temperature Probes |
| CLIN 0012: Matched Pair Pt1000 Clamp-On Surface Temperature Probes |
| CLIN 0013: Ultrasonic Wall Thickness Probe |
| CLIN 0014: Pipe Damping Material Kit |
| 3. Warranty |
| 4. Shipment and Delivery |
| 5. Invoice Information |
| 6. Points of Contact (POC) |
File details come from the government source that posted it. Updated .