943805-Submittal-02_Phase-2_Boom-Replacement.pdf
PDF 9 MB Posted
- Attached to
- Y1DA--RENOVATE OPERATING ROOMS 541-19-101
- Solicitation number
- 36C25026R0054Amend0001
About this file
This submittal package is a technical specification and installation guide for surgical lighting and equipment management systems being provided to Louis Stokes Cleveland VA Medical Center for a Phase 2 Operating Room Boom Replacement Project in Cleveland, Ohio, dated January 19, 2022 (Revision 02).
The package contains detailed specifications for Hillrom surgical equipment including iLED 7 surgical lights, flat panel mounting systems, and an 8-inch interstitial Equipment Management System (EMS). Key products specified include anesthesia booms (model 2300MHL2), perfusion booms (2300MML0), equipment booms (2600HL4), and integrated LED flat panel brackets (I760). The project involves replacement installations across multiple operating rooms with specific mounting requirements, electrical specifications, and medical gas configurations. Design loads are specified at 9,200 N (2067 lbf) maximum force and 12,500 Nm (9,220 ft-lbs) maximum bending moment, with structural deflection criteria limiting horizontal deflection to 0.1 inches and angular deflection to 1/3 degree. Installation requirements include 23" x 23" finished ceiling openings with 8" +/- 1/4" interstitial spacing, transformer enclosures for lighting systems, medical gas risers with DISS connections, and electrical circuits with isolated XHHW wiring. The package specifies contractor responsibilities for structural attachment, medical gas riser installation following NFPA 99C standards, electrical termination, and site preparation, while Hillrom provides mounting plates, transformer enclosures, and final equipment installation. Pre-installation activities must be completed while ceilings remain open, with final equipment certifications and inspections required following installation completion.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| S2-1.pdf | ||
| 1-9.pdf | ||
| 1-S2.pdf | ||
| S06. 36C25026R0054 Amend 0001.pdf | ||
| 1-S5.pdf | ||
| S02. RFI Form.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
1046 LeGrand Boulevard, Charleston, SC 29492 | Phone 888.474.9359 hillrom.com
LIGHTS BOOMS
TABLES PARTNERSHIPS
LOUIS STOKES CLEVELAND VA
PHASE 2 - OR BOOM REPLACEMENT PROJECT
CLEVELAND, OH
01-19-2022
Submittal Package Revision 02: 01/19/2022 : Page 1 of 86
TRUMPF Medical Systems, Inc.
Customer Order Authorization Release Form for Surgical Lights and Equipment Management Systems
F 50.003-01 Page 1 / 1
Ref.:TUSM291sa Date: 2013-01-01
TDM administers all version updates. Paper printouts are not updated
Project:
Trumpf Customer #: Lead Time: 16 weeks
Trumpf Sales Order #: Signed Drawings Due Date:
Buyer PO#: Delivery Date:
Project Schedule:
To meet the Buyer’s delivery expectations as indicated below, TRUMPF Medical Systems, Inc. (TUSM) requires this form to be completed prior to release for production. TUSM makes every reasonable effort to meet customers’ delivery dates as stated in our Terms and Conditions. With standard lead times of 8 weeks for standard products and 16 weeks for non-standard, TUSM needs this form signed by the Buyer no later than as indicated above. Any changes after the Buyer’s signed date below may affect delivery dates, incur additional charges, restocking fees and/or require a new authorization form.
Approval:
Buyer agrees that they have received the attached submittal package from TUSM and approve the drawings contained therein to be released for production. This approval includes but is not limited to (please check):
□ Ceiling Height Confirmation□ Equipment Management System Service Selections□ Med Gas Latch Valve Type, Receptacle Type, Wiring Type, Brake Gas Selection□ Accessories□ Surgical Lights and Flat Panels
□ Head Clearances
□ Handle Types
□ Flat Panel Type
Cables:
If the Buyer requires video and/or data cables to be installed at TUSM, it is the Buyer’s responsibility to deliver cables to TUSM, no later than 6 weeks prior to the Buyer required delivery date. Cable packages should be clearly labeled with the TRUMPF Sales Order # and shipped to: TRUMPF Medical Systems, Inc.; 1046 Legrand Blvd; Charleston, SC 29492.
Delivery:
Buyer agrees to accept delivery of the complete order as indicated in the delivery date above. Buyer should make TUSM aware of any changes as soon as possible. If changes occur within 4 weeks of the delivery date, Buyer agrees to accept delivery of the complete order and store product until installation.
Installation:
Services must be scheduled with TRUMPF Technical Service at least 2 weeks prior to installation.
Installation services can not be scheduled until the Buyer has properly prepared the site per TUSM rough-in requirements and has confirmed the site readiness via the TRUMPF Pre-Installation Checklist.
Authorized Seller’s Signature Authorized Buyer’s Signature
Authorized Seller’s Name/Title Authorized Buyer’s Name/Title
Date Date
Submittal Package Revision 08: 11/08/2021 : Page 2 of 86
William J.
Banner 3555053
Digitally signed by William J. Banner 3555053 Date: 2021.12.10 08:51:56 -05'00'
William Banner, COR
11-29-2021
LOUIS STOKES CLEVELAND VA MEDICAL CENTER - Phase 2 - OR's
943805 73596983
Submittal Package Revision 02: 01/19/2022 : Page 2 of 86
SERVICE SELECTIONS PAGE Department (Optional)__________________
1046 LeGrand Blvd., Charleston, SC 29492 Toll Free: (888) 474 9359 www.trumpfmedical.com
Brake Gas Type: Nitrogen Instrument Air Electric Brakes (Additional Charges Apply)
Gas Latch Valve Type:Manufactured by Tri Tech (Beacon Medaes available upon request. Additional Charges Apply)
Select Latch Style:
Special Requirements: _________________________________________________
Wire Type Selection:
Normal, TEW 105 (black, white, green color wires)
Isolated, XHHW (brown, orange, green color wires)
Duplex Outlet Type/s:Manufactured by Leviton (Hubbell available upon Request. Additional Charges apply)
Red, Standard Hospital Grade, 20A
Ivory, Standard Hospital Grade, 20A
Other (ie. Tamperproof, GFCI, Illuminated available at additional charge)
Special Requirements: _________________________________________________
Duplex Circuit Label Type: Video Monitor Type:
Standard 3M Adhesive (by Trumpf Medical) Make: _________________________
Engraved Metal Face Plate (by electrician) Size: _________________________
Light/Flat Panel Handle Selection Type: Weight: _________________________
Sterilizable Handle *Must be included on quote*
Disposable adapter, Devon Green Slip on
Disposable adapter, Devon Green Screw on
Disposable adapter, DeRoyal Blue Slip on
CONFIRMED BY: ________________________________________________ Date: __________
Submittal Package Revision 08: 11/08/2021 : Page 3 of 86Submittal Package Revision 02: 01/19/2022 : Page 3 of 86
GWB 9' - 6"
ACT-1 8' - 0"
ACT-1 8' - 0"
Cardiac OR
(OR-1)
2C129
Cardiac Pump Room
2C128
EMS Closet 2D280
Alcove 2C100N
ACT-1 8' - 0"
GWB 7' - 0"
4' -1
4"
4' -1
4"
'-9
4"4' 3'-6" 6"8'-8"
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
A: 2300MHL2 - ANESTHESIA BOOM
B: I760 - ILED 7, 32" FLAT PANEL BRACKET
C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"
FLAT PANEL BRACKET
D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"
FLAT PANEL BRACKET
CARDIAC OR 1 (2C129)
A
BC
D
Submittal Package Revision 02: 01/19/2022 : Page 4 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R1000
AutoCAD SHX Text R910
AutoCAD SHX Text R850
AutoCAD SHX Text R930
Cardiac OR
(OR-2)
2C126
Bed Holding 2C100P
Cardiac Equipment
2C127
Alcove 2C100K
Clean Corridor 2C100H
GWB 9' - 6"
2C100M
GWB
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 2"
4' -1
4"
4' -1
4"
'-9
1 2"4' 3'-6" 6"8'-53 4"
3' -6
3'
-6
6" 8'
-2
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
A: 2300MHL2 - ANESTHESIA BOOM
B: I760 - ILED 7, 32" FLAT PANEL BRACKET
C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"
FLAT PANEL BRACKET
D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"
FLAT PANEL BRACKET
CARDIAC OR 2 (2C126)
A
BC
D
Submittal Package Revision 02: 01/19/2022 : Page 5 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R1000
AutoCAD SHX Text R910
AutoCAD SHX Text R850
Robotic OR
(OR-5)
2C114
Electrical Room
2C200B
Bed Holding 2C100C
Scrub 2C100D
GWB 9' - 6"
GWB 9' - 6"
GWB 7' - 0"
5'-01 4" 4'-111
4"10'-61 2"
3' -6
3'
-6
6" 8'
-2
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
E: 2300MHL2 - ANESTHESIA BOOM
F: I760 - ILED 7, 32" FLAT PANEL BRACKET
G: I760 - ILED 7, 32" FLAT PANEL BRACKET
H: 2600HL4 - EQUIPMENT BOOM ROBOTIC OR 5 (2C114)
E F
G
H
Submittal Package Revision 02: 01/19/2022 : Page 6 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
R850
Bed Holding 2C200C
General OR
(OR-6)
2C210
Robotic OR
(OR-5)
Scrub
2C200BGWB 9' - 6"
GWB 7' - 0"
5'-01 4" 4'-111
4"11'-13 4"
3' -6
3'
-6
6" 9'
-2
4"
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
E: 2300MHL2 - ANESTHESIA BOOM
F: I760 - ILED 7, 32" FLAT PANEL BRACKET
G: I760 - ILED 7, 32" FLAT PANEL BRACKET
H: 2600HL4 - EQUIPMENT BOOM GENERAL OR 6 (2C210)
E F
G
H
Submittal Package Revision 02: 01/19/2022 : Page 7 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
General OR
(OR-07)
2C212
Bed Holding 2C200E
Scrub 2C200D
GWB 9' - 6"
GWB 7' - 0"
TYPE B
5'-01 4" 4'-111
4"11'-3"
3' -6
3'
-6
'-4
1 8"
4'
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
E: 2300MHL2 - ANESTHESIA BOOM
F: I760 - ILED 7, 32" FLAT PANEL BRACKET
G: I760 - ILED 7, 32" FLAT PANEL BRACKET
H: 2600HL4 - EQUIPMENT BOOM GENERAL OR 7 (2C212)
E F
G
H
Submittal Package Revision 02: 01/19/2022 : Page 8 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
Ortho OR
(OR-10)
2C217
General OR
(OR-09)
2C216
Ortho OR
(OR-08)
2C215
Bed Holding 2C200K
Scrub 2C200J
GWB 9' - 6"
Bed Holding 2C200H
2-A3-2
Alcove 2C220R
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 0"
TYPE A
TYPE B3'-57
8"3'-61 8"9'-6"
5' -0
1 4" 4'
-1
4"
'-8
8"
4' -1
4"
4' -1
4"
'-9
8"
8" 7'-111 2"
4'-01 8"
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
E: 2300MHL2 - ANESTHESIA BOOM
F: I760 - ILED 7, 32" FLAT PANEL BRACKET
G: I760 - ILED 7, 32" FLAT PANEL BRACKET
H: 2600HL4 - EQUIPMENT BOOM GENERAL OR 9 (2C216)
E
F G
H
Submittal Package Revision 02: 01/19/2022 : Page 9 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
GWB 9' - 0"
Ortho OR
(OR-08)
2C215
Scrub 2C200G
Laboratory, Frozen Section 2C214
Electrical Room 2C213
Ortho Equipment
2C215A
GWB 9' - 6"
Bed Holding 2C200H
2-A3-2
GWB 7' - 0"
TYPE A
TYPE B
GWB 7' - 2"
GWB
5' -0
1 4" 4'
-1
4"
'-8
8"
4' -1
4"
4' -1
4"
'-9
8"
4'3'-6" 9'-43 4"
4'
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
A:2300MHL2 - ANESTHESIA BOOM
B: I760 - ILED 7, 32" FLAT PANEL BRACKET
C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"
FLAT PANEL BRACKET
D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"
FLAT PANEL BRACKET
ORTHO OR 8 (2C215)
A
B C
D
Submittal Package Revision 02: 01/19/2022 : Page 10 of 86
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R1000
AutoCAD SHX Text R910
AutoCAD SHX Text R850
ACT-1 9' - 0"
Anesthesia Workroom
2C218
Ortho OR
(OR-10)
2C217
General OR
(OR-09)
2C216
Bed Holding 2C200M
Scrub 2C200L
Bed Holding 2C200K
Alcove 2C200Q
GWB2
2-A3-2
Alcove 2C220R
GWB 9' - 6"
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 0"
GWB 7' - 2" TYPE A
9'-6"
4' -1
8"
4' -1
8"
'-9
1 4"
3'-117 8"3'-61
8" 7'-111 2"
4'-01 8"
LOUIS STOKES CLEVELAND VA
MEDICAL CENTER
PHASE 2 - OR BOOM REPLACEMENT PROJECT
73596983
01-19-2022 02 SAW
943805-1
ROOM:
DWG. NO.:
DATE:
EQUIPMENT:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
4279-U Crosspoint Drive, Ladson SC 29456
Phone: 1-888-474-9359 Fax:
1-843-534-0206 www.hillrom.com
S/O NO.:
NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.
A:2300MHL2 - ANESTHESIA BOOM
B: I760 - ILED 7, 32" FLAT PANEL BRACKET
C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"
FLAT PANEL BRACKET
D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"
FLAT PANEL BRACKET
ORTHO OR 10 (2C217)
A
B C
D
Submittal Package Revision 02: 01/19/2022 : Page 11 of 86
AutoCAD SHX Text R930
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R930
AutoCAD SHX Text R910
AutoCAD SHX Text R1000
AutoCAD SHX Text R850
AutoCAD SHX Text R1000
AutoCAD SHX Text R910
AutoCAD SHX Text R850
3D-view CSU
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
A: 2300MHL2_MB-52136.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Print: 16.12.21
Submittal Package Revision 02: 01/19/2022 : Page 12 of 86
View CSU
Brake Type Legend:
1) electro pneumatic brake
2) mechanical friction brake, adjustable
3) pneumatic brake; 4) electro magnetic brake
CSU-flange Tandem System flange tandem additional information
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
A: 2300MHL2_MB-52136.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 13 of 86
VAC
C1
E1
C3
C2
C5 E3
W1
A1
VAC
E2
C4
W2
O2
MA
N2O
WAGD
E4
A2
ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS
C HIGH VOLTAGE CIRCUITS
GAS MEDICAL GAS SERVICES
LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)
WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)
SIDE B SIDE CSIDE A
Media Package Length: Standard (1 meter)
WIRE-NUTW
SIDE D
REF QTY P/N DESCRIPTION
A1 2 1636318 530mm SHELF
A2 1 1636396 ADDITIONAL CONTROL HANDLES
REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)
E1 2 L1 SINGLE GANG, TBD
E2 2 L2 SINGLE GANG, TBD
E3 2 L3 SINGLE GANG, TBD
E4 2 L4 SINGLE GANG, TBD
REF QTY P/N DESCRIPTION
REF CIRCUIT CONDUIT/
WIRE-NUT
# OF
RECEPTACLES
RECEPTACLE DESCRIPTION
(TYPE, COLOR, AMP) TYPE
C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C6 CIRCUIT 6 - - BOOM ACCESSORIES CIRCUIT -
REF QTY DESCRIPTION
VAC 2 VACUUM MEDICAL GAS OUTLET
O2 2 OXYGEN MEDICAL GAS OUTLET
MA 1 MEDICAL AIR GAS OUTLET
N2O 1 NITROUS OXIDE MEDICAL GAS OUTLET
WAGD 1 WASTE ANESTHESIA GAS DISPOSAL OUTLET
BRAKE GAS
EB 1 ELECTRIC BRAKES
943805
211216-43438-TUSM291WR
OR 1-4
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
A: 2300MHL2_MB-52136.00
16.12.21
Customer ref. no:
Quotation no:
Room:
Project Customer:
TM order no:
Amount:
Released for production:
Date:
Signature:
Changes after release may cause delay In shipment or price corrections
-Field of application: surgery -Internat.Standard: US
Order drawing no:
Print:
Submittal Package Revision 02: 01/19/2022 : Page 14 of 86
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
B: I760_MB-52137.00
-Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 15 of 86
Room situation with arms in upper position
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
B: I760_MB-52137.00
-Internat.Standard: US
Scale:
1:33
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 16 of 86
Room situation with arms in lower position
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
B: I760_MB-52137.00
-Internat.Standard: US
Scale:
1:33
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 17 of 86
Electrical installation
Control board plate (Power Pack Supply Voltage: 100 - 240 V AC / 22 - 32 V DC) Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines
Pos. 2 iLED7 / Power supply line: 100-240V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V
Load range / add. components VidiaPort
Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)
* Regard adapter + monitor weight!
Wall control panel / Operating unit
Operating unit iLED7: Wall Control set surface + TR Univ. Front interf.
Handles
Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
B: I760_MB-52137.00
-Internat.Standard: US
Scale:
Picture above: 1:50 Picture below: 1:10
Order drawing no:
Print: 16.12.21
Submittal Package Revision 02: 01/19/2022 : Page 18 of 86 wrights5 Text Box
DISPOSABLE ADAPTER - DEVON GREEN SLIP ON
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
C: 2300MML0 & I760_MB-52138.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 19 of 86
1) electro pneumatic brake
2) mechanical friction brake, adjustable
3) pneumatic brake; 4) electro magnetic brake
CSU-flange Tandem OPL System flange tandem OPL additional information
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
C: 2300MML0 & I760_MB-52138.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 20 of 86
VAC
E2
C3
C4
W1
VAC
C2
E1
MA
O2
O2
MA
CO2
VAC
C1
ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS
C HIGH VOLTAGE CIRCUITS
GAS MEDICAL GAS SERVICES
LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)
WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)
SIDE B SIDE CSIDE A
Media Package Length: Standard (1 meter)
WIRE-NUTW
SIDE D
REF QTY P/N DESCRIPTION REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)
E1 2 L1 SINGLE GANG, TBD
E2 2 L2 SINGLE GANG, TBD
REF QTY P/N DESCRIPTION
REF CIRCUIT CONDUIT/
WIRE-NUT
# OF
RECEPTACLES
RECEPTACLE DESCRIPTION
(TYPE, COLOR, AMP) TYPE
C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C3 CIRCUIT 3 H2 / W1 2 DUPLEX, 20A, RED 8300-R
C4 CIRCUIT 4 H2 / W1 2 DUPLEX, 20A, RED 8300-R
C5 CIRCUIT 5 - - BOOM ACCESSORIES CIRCUIT -
REF QTY DESCRIPTION
VAC 4 VACUUM MEDICAL GAS OUTLET
O2 2 OXYGEN MEDICAL GAS OUTLET
MA 2 MEDICAL AIR GAS OUTLET
CO2 4 CARBON DIOXIDE MEDICAL GAS OUTLET
BRAKE GAS
EB 1 ELECTRIC BRAKES
943805
211216-43438-TUSM291WR
OR 1-4
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
C: 2300MML0 & I760_MB-52138.00
16.12.21
Customer ref. no:
Quotation no:
Room:
Project Customer:
TM order no:
Amount:
Released for production:
Date:
Signature:
Changes after release may cause delay In shipment or price corrections
-Field of application: surgery -Internat.Standard: US
Order drawing no:
Print:
Submittal Package Revision 02: 01/19/2022 : Page 21 of 86
Lights / flat panel / VPA
Control board plate (Power Pack Supply Voltage: 100 - 240 V AC / 22 - 32 V DC) Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines
Pos. 2 iLED7 / Power supply line: 230V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V
Load range / add. components VidiaPort
Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)
* Regard adapter + monitor weight!
Handles
Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Wall control panel / Operating unit
Operating unit iLED7: from SAP.Pos.Nr.: MOUNT B
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
C: 2300MML0 & I760_MB-52138.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Print: 16.12.21
Submittal Package Revision 02: 01/19/2022 : Page 22 of 86 wrights5 Text Box
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
D: 2600HL4 & I760_MB-52139.00
-Field of application: Other -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 23 of 86
1) electro pneumatic brake
2) mechanical friction brake, adjustable
3) pneumatic brake; 4) electro magnetic brake
CSU-flange Tandem OPL System flange tandem OPL additional information
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
D: 2600HL4 & I760_MB-52139.00
-Field of application: Other -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 24 of 86
Released for production:
Date:
Signature:
In shipment or price corrections
-Field of application: surgery -Internat.Standard: US
OR 1-4
211216-43438-TUSM291WR
943805
D: 2600HL4 & I760_MB-52139.00
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Order drawing no:
16.12.21
Customer ref. no:
Quotation no:
Room:
Project Customer:
Amount:
TM order no:
Changes after release may cause delay Print:
REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)
E1 2 L1 SINGLE GANG, TBD
E2 2 L2 SINGLE GANG, TBD
E3 2 L3 SINGLE GANG, TBD
REF QTY P/N DESCRIPTION
REF CIRCUIT CONDUIT/
WIRE-NUT
# OF
RECEPTACLES
RECEPTACLE DESCRIPTION
(TYPE, COLOR, AMP) TYPE
C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C6 CIRCUIT 6 H3 / W2 2 DUPLEX, 20A, RED 8300-R
C7 CIRCUIT 7 H3 / W2 2 DUPLEX, 20A, RED 8300-R
C8 CIRCUIT 8 - - BOOM ACCESSORIES CIRCUIT -
REF QTY DESCRIPTION
VAC 2 VACUUM MEDICAL GAS OUTLET
CO2 1 CARBON DIOXIDE MEDICAL GAS OUTLET
IAP 1 INSTRUMENT AIR CONTROL PANEL (1834390)
BRAKE GAS
EB 1 ELECTRIC BRAKES
ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS
C HIGH VOLTAGE CIRCUITS
GAS MEDICAL GAS SERVICES
LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)
WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)
SIDE B SIDE CSIDE A
Media Package Length: Standard (1 meter)
WIRE-NUTW
SIDE D
REF QTY P/N DESCRIPTION
A1 3 1636318 530mm SHELF
A2 1 1636396 ADDITIONAL CONTROL HANDLES
C5
E1
E2
C1
C2
C3
C4
CO2
VAC
A1
E3
W2
IAP
C6
W1
VAC
A2 C7
Submittal Package Revision 02: 01/19/2022 : Page 25 of 86
Lights / flat panel / VPA
Control board plate (Power Pack Supply Voltage: 100 - 240 V AC / 22 - 32 V DC) Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines
Pos. 2 iLED7 / Power supply line: 230V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V
Load range / add. components VidiaPort
Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)
* Regard adapter + monitor weight!
Handles
Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Wall control panel / Operating unit
Operating unit iLED7: from SAP.Pos.Nr.: MOUNT B
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 1-4
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
D: 2600HL4 & I760_MB-52139.00
-Field of application: Other -Internat.Standard: US
Order drawing no:
Print: 16.12.21
Submittal Package Revision 02: 01/19/2022 : Page 26 of 86 wrights5 Text Box
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
E: 2300MHL2_MB-52140.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 27 of 86
1) electro pneumatic brake
2) mechanical friction brake, adjustable
3) pneumatic brake; 4) electro magnetic brake
CSU-flange Tandem System flange tandem additional information
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
E: 2300MHL2_MB-52140.00
-Field of application: Surgery -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 28 of 86
In shipment or price corrections
-Field of application: surgery -Internat.Standard: US
PHASE 2 - OR BOOM REPLACEMENT
OR 5-8
943805
211216-43438-TUSM291WR
. + . / 73596983
E: 2300MHL2_MB-52140.00
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
Order drawing no:
16.12.21
Customer ref. no:
Quotation no:
Room:
Project Customer:
TM order no:
Amount:
Released for production:
Signature:
Date:
Changes after release may cause delay Print:
REF QTY P/N DESCRIPTION
A1 2 1636318 530mm SHELF
A2 1 1636396 ADDITIONAL CONTROL HANDLES
REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)
E1 2 L1 SINGLE GANG, TBD
E2 2 L2 SINGLE GANG, TBD
E3 2 L3 SINGLE GANG, TBD
E4 2 L4 SINGLE GANG, TBD
REF QTY P/N DESCRIPTION
REF CIRCUIT CONDUIT/
WIRE-NUT
# OF
RECEPTACLES
RECEPTACLE DESCRIPTION
(TYPE, COLOR, AMP) TYPE
C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C6 CIRCUIT 6 - - BOOM ACCESSORIES CIRCUIT -
REF QTY DESCRIPTION
VAC 2 VACUUM MEDICAL GAS OUTLET
O2 2 OXYGEN MEDICAL GAS OUTLET
MA 1 MEDICAL AIR GAS OUTLET
N2O 1 NITROUS OXIDE MEDICAL GAS OUTLET
WAGD 1 WASTE ANESTHESIA GAS DISPOSAL OUTLET
BRAKE GAS
EB 1 ELECTRIC BRAKES
ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS
C HIGH VOLTAGE CIRCUITS
GAS MEDICAL GAS SERVICES
LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)
WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)
SIDE B SIDE CSIDE A
Media Package Length: Standard (1 meter)
WIRE-NUTW
SIDE D
MA C4
E2
N2O
WAGD
E4
W2
VAC
O2
A2
C5
W1 VAC
E3
C1
E1
C3
C2
A1
Submittal Package Revision 02: 01/19/2022 : Page 29 of 86
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
F: I760_MB-52141.00
-Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 30 of 86
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
F: I760_MB-52141.00
-Internat.Standard: US
Scale:
1:33
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 31 of 86
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
F: I760_MB-52141.00
-Internat.Standard: US
Scale:
1:33
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 32 of 86
Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines
Pos. 2 iLED7 / Power supply line: 100-240V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V
Load range / add. components VidiaPort
Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)
* Regard adapter + monitor weight!
Wall control panel / Operating unit
Operating unit iLED7: Wall Control set surface + TR Univ. Front interf.
Handles
Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
F: I760_MB-52141.00
-Internat.Standard: US
Scale:
Picture above: 1:50 Picture below: 1:10
Order drawing no:
Print: 16.12.21
Submittal Package Revision 02: 01/19/2022 : Page 33 of 86 wrights5 Text Box
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
G: I760_MB-52142.00
-Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 34 of 86
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
G: I760_MB-52142.00
-Internat.Standard: US
Scale:
1:33
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 35 of 86
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
G: I760_MB-52142.00
-Internat.Standard: US
Scale:
1:33
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 36 of 86
Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines
Pos. 2 iLED7 / Power supply line: 100-240V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V
Load range / add. components VidiaPort
Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)
* Regard adapter + monitor weight!
Wall control panel / Operating unit
Operating unit iLED7: from SAP.Pos.Nr.: MOUNT F
Handles
Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
G: I760_MB-52142.00
-Internat.Standard: US
Scale:
Picture above: 1:50 Picture below: 1:10
Order drawing no:
Print: 16.12.21
Submittal Package Revision 02: 01/19/2022 : Page 37 of 86 wrights5 Text Box
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
H: 2600HL4_MB-52143.00
-Field of application: Other -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 38 of 86
1) electro pneumatic brake
2) mechanical friction brake, adjustable
3) pneumatic brake; 4) electro magnetic brake
CSU-flange Tandem OPL System flange tandem OPL additional information
Customer ref. no:
Quotation no:
Room:
943805
211216-43438-TUSM291WR
OR 5-8
Project:
Customer:
TM order no:
Amount:
PHASE 2 - OR BOOM REPLACEMENT
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
. + . / 73596983
Released for production:
Date:
Signature:
Changes after release may cause delay in shipment or price corrections
H: 2600HL4_MB-52143.00
-Field of application: Other -Internat.Standard: US
Order drawing no:
Submittal Package Revision 02: 01/19/2022 : Page 39 of 86
Signature:
In shipment or price corrections
-Field of application: surgery -Internat.Standard: US
OR 5-8
H: 2600HL4_MB-52143.00
. + . / 73596983
Order drawing no:
211216-43438-TUSM291WR
PHASE 2 - OR BOOM REPLACEMENT943805
LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND
16.12.21
Customer ref. no:
Quotation no:
Room:
Project Customer:
TM order no:
Amount:
Date:
Released for production:
Changes after release may cause delay Print:
REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)
E1 2 L1 SINGLE GANG, TBD
E2 2 L2 SINGLE GANG, TBD
E3 2 L3 SINGLE GANG, TBD
REF QTY P/N DESCRIPTION
REF CIRCUIT CONDUIT/
WIRE-NUT
# OF
RECEPTACLES
RECEPTACLE DESCRIPTION
(TYPE, COLOR, AMP) TYPE
C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R
C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R
C6 CIRCUIT 6 H3 / W2 2 DUPLEX, 20A, RED 8300-R
C7 CIRCUIT 7 H3 / W2 2 DUPLEX, 20A, RED 8300-R
C8 CIRCUIT 8 - - BOOM ACCESSORIES CIRCUIT -
REF QTY DESCRIPTION
VAC 3 VACUUM MEDICAL GAS OUTLET
CO2 1 CARBON DIOXIDE MEDICAL GAS OUTLET
IAP 1 INSTRUMENT AIR CONTROL PANEL (1834390)
BRAKE GAS
EB 1 ELECTRIC BRAKES
ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS
C HIGH VOLTAGE CIRCUITS
GAS MEDICAL GAS SERVICES
LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)
WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)
SIDE B SIDE CSIDE A
Media Package Length: Standard (1 meter)
WIRE-NUTW
SIDE D
REF QTY P/N DESCRIPTION
A1 4 1636319 630mm SHELF
A2 1 1636396 ADDITIONAL CONTROL HANDLES
C6
A2
W1
IAP
W2
VAC
C7
C5
E1
E2
C1
C2
C3
C4
CO2
VAC
A1
E3
Submittal Package Revision 02: 01/19/2022 : Page 40 of 86
24"
20.35"
20.35"
1.02" TYP
24"
23" +/- 1/2"
FINISHED CEILING OPENING
8"
1/
4"
TO
P O
F
TR
U M
PF
M
O U
N
TI
N G
P
LA
T
TO
F
IN
IS
H
ED
C
EI
LN
IG
3" +/- 1/4"
MOUNTING PLATE WELDED, BOLTED, OR
ANCHORED TO STRUCTURE BY OTHER.
ATTACHMENT METHODS MUST BE
PROVIDED AND INSTALLED BY OTHER AND
APPROVED BY ENGINEER OF RECORD
FINISHED
CEILING
DESIGN LOAD:
F max: 9,200 N (2067 lbf) Mb max: 12,500 Nm (9,220 ft-lbs)
NOTES:
1) Components for TRUMPF EMS mounting systems are TRUMPF provided and installed BY OTHERS. If necessary, all hardware required to attach mounting plate to structure is PROVIDED BY OTHERS and INSTALLED BY OTHERS.
TRUMPF recommends installation during the early phase of construction.
2) Engineer of record must approve all fabricated structures and attachment methods.
3) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.
4) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.
5) Required opening in finished ceiling is 23" x 23" 1/2" centered. Ceiling cover (not shown) is installed by TRUMPF during final installation, and is 25.125" x 25.125". A gasket is used to seal the space between the ceiling cover and finished ceiling.
6) M16 all thread must end 3" 1/4" below the room side of the finished ceiling. All thread will be provided with protective plastic covers by TRUMPF. Following installation of the mounting systems by OTHERS, the protective plastic covers shall be reinstalled by OTHERS to protect the all thread during construction.
7) Gas risers are mounted as shown with DISS connections facing the interior of the mount . Services are then brazed to 1/2" OD copper outside the support structure, above the finished ceiling. Gas risers consist of gas specific DISS nipples brazed to 1/2" OD copper tubing. Gas risers must be installed horizontally.
8) All gas risers have a color coded label according to NFPA requirements.
9) Junction boxes for electrical connections are provided by electrical contractor, and can be mounted to the TRUMPF Interface Plate. Use Raco 265 (4 11/16" x 4 11/16" x 2 1/8") or similar.
VIEW 'A'
MOUNTING PLATE DIMESIONS
VIEW 'B'
MOUNTING SYSTEM ELEVATIONS
PAGE CONTENTS:
DWG. NO.:
DATE:
MOUNT:
CAT. NO:
QUANTITY ON PLAN:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
TRUMPF Medical Systems, Inc.
1046 LeGrand Blvd, Charleston, SC 29492
Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com
NOT TO SCALE
THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!
This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.
LOUIS STOKES CLEVELAND VA MEDICAL
BOOM MOUNTING SYSTEM
8" INTERSTITIAL SPACE
MOUNTING SYSTEM
110106 943805-MA
01 SAW4 12-16-2021A
UNITS: [In.]/mm
S.O. NO.: 73596983
Submittal Package Revision 02: 01/19/2022 : Page 41 of 86
VIEW 'C'
MOUNTING SYSTEM COMPONENTS
(ALL COMPONENTS INSTALLED BY CONTRACTOR)
TIGHTEN M16 NUTS
TO 52 ft-lbs
TIGHTEN M22 x M16
ADAPTER NUTS
TO 89 ft-lbs
LIST OF REQUIRED MATERIALS FOR BOOM MOUNTING SYSTEM
DESIGN LOAD:
F max: 9,200 N (2067 lbf) Mb max: 12,500 Nm (9,220 ft-lbs) NOTES:
A) Components for TRUMPF EMS mounting systems are TRUMPF provided and installed BY OTHERS. If necessary, all hardware required to attach mounting plate to structure is PROVIDED BY OTHERS and INSTALLED BY OTHERS.
TRUMPF recommends installation during the early phase of construction.
B) Engineer of record must approve all fabricated structures and attachment methods.
C) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.
D) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.
E) Required opening in finished ceiling is 23" x 23" 1/2" centered. Ceiling cover (not shown) is installed by TRUMPF during final installation, and is 25.125" x 25.125". A gasket is used to seal the space between the ceiling cover and finished ceiling.
F) M16 all thread must end 3" 1/4" below the room side of the finished ceiling. All thread will be provided with protective plastic covers by TRUMPF. Following installation of the mounting system by OTHERS, the protective covers shall be reinstalled by OTHERS to protect the all thread during construction.
G) Gas risers are mounted as shown with DISS connections facing the interior of the mount. Services are then brazed to 1/2" OD copper outside the support structure, above the finished ceiling. Gas risers consist of gas specific DISS nipples brazed to 1/2" OD copper tubing. Gas riser must be installed horizontally.
H) All gas risers have a color coded label according to NFPA requirements.
I) Junction boxes for electrical connections are provided by electrical contractor, and can be mounted to the TRUMPF Universal Interface Plate. Use Raco 265 (4 11/16" x 4 11/16" x 2 1/8") or similar.
PAGE CONTENTS:
DWG. NO.:
DATE:
MOUNT:
CAT. NO:
QUANTITY ON PLAN:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
TRUMPF Medical Systems, Inc.
1046 LeGrand Blvd, Charleston, SC 29492
Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com
NOT TO SCALE
THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!
This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.
LOUIS STOKES CLEVELAND VA MEDICAL
BOOM MOUNTING SYSTEM
8" INTERSTITIAL SPACE
MOUNTING SYSTEM
110106 943805-MA
01 SAW4 12-16-2021A
UNITS: [In.]/mm
Submittal Package Revision 02: 01/19/2022 : Page 42 of 86
5"
14 1 2 "
ROUGH OPENING
10"
TRUMPF
MOUNTING PLATE
(110036)
(INSTALLED BY
OTHER)
SUB-STRUCTURE SPECIFIED BY
ENGINEER OF RECORD
(PROVIDED AND INSTALLED BY
OTHER)
THREADED ROD AND HARDWARE
APPROVED BY ENGINEER OF RECORD
(PROVIDED AND INSTALLED BY OTHER)
VIEW 'B'
LIGHT MOUNTING PLATE (110036)
14"
14"
10 1 4 "
10 1 4 "
(TYP 4X)
.81" THRU
VIEW 'A'
SIDE VIEW OF MOUNTING SYSTEM
(ALL COMPONENTS INSTALLED BY CONTRACTOR)
FINISHED
CEILING
DESIGN LOAD:
F max: 1,780 N (400 lbf) Mb max: 2,033 Nm (1500 ft-lbs)
VIEW 'C'
ISOMETRIC VIEW OF MOUNTING SYSTEM
(COMPLETED PRE-INSTALLATION)
CONDUIT TO
TRANSFORMER
ENCLOSURE
PAGE CONTENTS:
DWG. NO.:
DATE:
MOUNT:
CAT. NO:
QUANTITY ON PLAN:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
TRUMPF Medical Systems, Inc.
1046 LeGrand Blvd, Charleston, SC 29492
Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com
NOT TO SCALE
THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!
This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.
LOUIS STOKES CLEVELAND VA MEDICAL
LIGHT MOUNTING SYSTEM
FLAT CEILING COVER
MOUNTING SYSTEM
MS-1 943805-MB
01 SAW4 12-16-2021B
UNITS: [In.]/mm
Submittal Package Revision 02: 01/19/2022 : Page 43 of 86
STRUCTURE BY OTHER
MOUNTING
HARDWARE BY
OTHER
CEILING COVER
(INSTALLED BY
TRUMPF)
LIGHT TUBE
(INSTALLED BY
TRUMPF)
11.875"
.7
5"
8"
1.125" HOLES
9.25"
13 1 16 "
LIST OF PRE-INSTALL MATERIALS FOR LIGHT MOUNTING SYSTEMS
NOTES:
A) TRUMPF recommends installation during the early phase of construction. Light mounting plate is provided by TRUMPF but installed by others.
B) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.
C) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.
D) Engineer of record must approve all fabricated structures and attachment methods.
E) TRUMPF recommends 3/4" ASTM-A307 hardware for attachment of the TRUMPF light mounting plate.
F) Required opening in finished ceiling is 14.5" x 14.5" 1/4" centered. Ceiling cover is installed by TRUMPF during final installation and is 15.75" x 15.75".
G) Transformer enclosures provided by TRUMPF and installed by other in an accessible location. Access panels are required if located above hard ceiling.
H) Refer to Preinstall guide (110720) for required conduit runs for transformer enclosures, wall controls, and camera control.
DESIGN LOAD:
F max: 1,780 N (400 lbf) Mb max: 2,033 Nm (1500 ft-lbs)
FOR 3/4" TRADE
SIZE CONDUIT
TERMINATION
VIEW 'D'
TRANSFORMER ENCLOSURE
(INSTALLED BY OTHER)
VIEW 'E'
ISOMETRIC VIEW OF MOUNTING SYSTEM
PAGE CONTENTS:
DWG. NO.:
DATE:
MOUNT:
CAT. NO:
QUANTITY ON PLAN:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
TRUMPF Medical Systems, Inc.
1046 LeGrand Blvd, Charleston, SC 29492
Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com
NOT TO SCALE
THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!
This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.
LOUIS STOKES CLEVELAND VA MEDICAL
LIGHT MOUNTING SYSTEM
FLAT CEILING COVER
MOUNTING SYSTEM
MS-1 943805-MB
01 SAW4 12-16-2021B
UNITS: [In.]/mm
Submittal Package Revision 02: 01/19/2022 : Page 44 of 86
20.35"
1.02" TYP
24"
23" +/- 1/2"
FINISHED CEILING OPENING
8"
O
F
TR
U M
PF
M
O U
N
TI
N G
P
LA
T
TO
F
IN
IS
H
ED
C
EI
LN
IG
3" +/- 1/4"
MOUNTING PLATE WELDED, BOLTED, OR
ANCHORED TO STRUCTURE BY OTHER.
ATTACHMENT METHODS MUST BE
PROVIDED AND INSTALLED BY OTHER AND
APPROVED BY ENGINEER OF RECORD
FINISHED
CEILING
DESIGN LOAD:
F max: 9,200 N (2067 lbf) Mb max: 12,500 Nm (9,220 ft-lbs)
NOTES:
1) Components for TRUMPF EMS mounting systems are TRUMPF provided and installed BY OTHERS. If necessary, all hardware required to attach mounting plate to structure is PROVIDED BY OTHERS and INSTALLED BY OTHERS.
TRUMPF recommends installation during the early phase of construction.
2) Engineer of record must approve all fabricated structures and attachment methods.
3) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.
4) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.
5) Required opening in finished ceiling is 23" x 23" 1/2" centered. Ceiling cover (not shown) is installed by TRUMPF during final installation, and is 25.125" x 25.125". A gasket is used to seal the space between the ceiling cover and finished ceiling.
6) M16 all thread must end 3" 1/4" below the room side of the finished ceiling. All thread will be provided with protective plastic covers by TRUMPF. Following installation of the mounting systems by OTHERS, the protective plastic covers shall be reinstalled by OTHERS to protect the all thread during construction.
7) Gas risers are mounted as shown with DISS connections facing the interior of the mount . Services are then brazed to 1/2" OD copper outside the support structure, above the finished ceiling. Gas risers consist of gas specific DISS nipples brazed to 1/2" OD copper tubing. Gas risers must be installed horizontally.
8) All gas risers have a color coded label according to NFPA requirements.
9) Junction boxes for electrical connections are provided by electrical contractor, and can be mounted to the TRUMPF Interface Plate. Use Raco 265 (4 11/16" x 4 11/16" x 2 1/8") or similar.
VIEW 'A'
MOUNTING PLATE DIMESIONS
VIEW 'B'
MOUNTING SYSTEM ELEVATIONS
PAGE CONTENTS:
DWG. NO.:
DATE:
MOUNT:
CAT. NO:
QUANTITY ON PLAN:
RELEASED FOR PRODUCTION
(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)
SIGNATURE:
DATE: INSTALLATION DATE: REV.:
TRUMPF Medical Systems, Inc.
1046 LeGrand Blvd, Charleston, SC 29492
Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com
NOT TO SCALE
THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!
This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.
LOUIS STOKES CLEVELAND VA MEDICAL
BOOM…
This is the start of the file's text. The full file is on GovTribe.