943805-Submittal-02_Phase-2_Boom-Replacement.pdf

PDF 9 MB Posted

Attached to
Y1DA--RENOVATE OPERATING ROOMS 541-19-101
Solicitation number
36C25026R0054Amend0001
Issued by

About this file

This submittal package is a technical specification and installation guide for surgical lighting and equipment management systems being provided to Louis Stokes Cleveland VA Medical Center for a Phase 2 Operating Room Boom Replacement Project in Cleveland, Ohio, dated January 19, 2022 (Revision 02).

The package contains detailed specifications for Hillrom surgical equipment including iLED 7 surgical lights, flat panel mounting systems, and an 8-inch interstitial Equipment Management System (EMS). Key products specified include anesthesia booms (model 2300MHL2), perfusion booms (2300MML0), equipment booms (2600HL4), and integrated LED flat panel brackets (I760). The project involves replacement installations across multiple operating rooms with specific mounting requirements, electrical specifications, and medical gas configurations. Design loads are specified at 9,200 N (2067 lbf) maximum force and 12,500 Nm (9,220 ft-lbs) maximum bending moment, with structural deflection criteria limiting horizontal deflection to 0.1 inches and angular deflection to 1/3 degree. Installation requirements include 23" x 23" finished ceiling openings with 8" +/- 1/4" interstitial spacing, transformer enclosures for lighting systems, medical gas risers with DISS connections, and electrical circuits with isolated XHHW wiring. The package specifies contractor responsibilities for structural attachment, medical gas riser installation following NFPA 99C standards, electrical termination, and site preparation, while Hillrom provides mounting plates, transformer enclosures, and final equipment installation. Pre-installation activities must be completed while ceilings remain open, with final equipment certifications and inspections required following installation completion.

View the file

Other files

File Type Posted
S2-1.pdf PDF
1-9.pdf PDF
1-S2.pdf PDF
S06. 36C25026R0054 Amend 0001.pdf PDF
1-S5.pdf PDF
S02. RFI Form.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

1046 LeGrand Boulevard, Charleston, SC 29492 | Phone 888.474.9359 hillrom.com

LIGHTS BOOMS

TABLES PARTNERSHIPS

LOUIS STOKES CLEVELAND VA

PHASE 2 - OR BOOM REPLACEMENT PROJECT

CLEVELAND, OH

01-19-2022

Submittal Package Revision 02: 01/19/2022 : Page 1 of 86

TRUMPF Medical Systems, Inc.

Customer Order Authorization Release Form for Surgical Lights and Equipment Management Systems

F 50.003-01 Page 1 / 1

Ref.:TUSM291sa Date: 2013-01-01

TDM administers all version updates. Paper printouts are not updated

Project:

Trumpf Customer #: Lead Time: 16 weeks

Trumpf Sales Order #: Signed Drawings Due Date:

Buyer PO#: Delivery Date:

Project Schedule:

To meet the Buyer’s delivery expectations as indicated below, TRUMPF Medical Systems, Inc. (TUSM) requires this form to be completed prior to release for production. TUSM makes every reasonable effort to meet customers’ delivery dates as stated in our Terms and Conditions. With standard lead times of 8 weeks for standard products and 16 weeks for non-standard, TUSM needs this form signed by the Buyer no later than as indicated above. Any changes after the Buyer’s signed date below may affect delivery dates, incur additional charges, restocking fees and/or require a new authorization form.

Approval:

Buyer agrees that they have received the attached submittal package from TUSM and approve the drawings contained therein to be released for production. This approval includes but is not limited to (please check):

□ Ceiling Height Confirmation□ Equipment Management System Service Selections□ Med Gas Latch Valve Type, Receptacle Type, Wiring Type, Brake Gas Selection□ Accessories□ Surgical Lights and Flat Panels

□ Head Clearances

□ Handle Types

□ Flat Panel Type

Cables:

If the Buyer requires video and/or data cables to be installed at TUSM, it is the Buyer’s responsibility to deliver cables to TUSM, no later than 6 weeks prior to the Buyer required delivery date. Cable packages should be clearly labeled with the TRUMPF Sales Order # and shipped to: TRUMPF Medical Systems, Inc.; 1046 Legrand Blvd; Charleston, SC 29492.

Delivery:

Buyer agrees to accept delivery of the complete order as indicated in the delivery date above. Buyer should make TUSM aware of any changes as soon as possible. If changes occur within 4 weeks of the delivery date, Buyer agrees to accept delivery of the complete order and store product until installation.

Installation:

Services must be scheduled with TRUMPF Technical Service at least 2 weeks prior to installation.

Installation services can not be scheduled until the Buyer has properly prepared the site per TUSM rough-in requirements and has confirmed the site readiness via the TRUMPF Pre-Installation Checklist.

Authorized Seller’s Signature Authorized Buyer’s Signature

Authorized Seller’s Name/Title Authorized Buyer’s Name/Title

Date Date

Submittal Package Revision 08: 11/08/2021 : Page 2 of 86

William J.

Banner 3555053

Digitally signed by William J. Banner 3555053 Date: 2021.12.10 08:51:56 -05'00'

William Banner, COR

11-29-2021

LOUIS STOKES CLEVELAND VA MEDICAL CENTER - Phase 2 - OR's

943805 73596983

Submittal Package Revision 02: 01/19/2022 : Page 2 of 86

SERVICE SELECTIONS PAGE Department (Optional)__________________

1046 LeGrand Blvd., Charleston, SC 29492 Toll Free: (888) 474 9359 www.trumpfmedical.com

Brake Gas Type: Nitrogen Instrument Air Electric Brakes (Additional Charges Apply)

Gas Latch Valve Type:Manufactured by Tri Tech (Beacon Medaes available upon request. Additional Charges Apply)

Select Latch Style:

Special Requirements: _________________________________________________

Wire Type Selection:

Normal, TEW 105 (black, white, green color wires)

Isolated, XHHW (brown, orange, green color wires)

Duplex Outlet Type/s:Manufactured by Leviton (Hubbell available upon Request. Additional Charges apply)

Red, Standard Hospital Grade, 20A

Ivory, Standard Hospital Grade, 20A

Other (ie. Tamperproof, GFCI, Illuminated available at additional charge)

Special Requirements: _________________________________________________

Duplex Circuit Label Type: Video Monitor Type:

Standard 3M Adhesive (by Trumpf Medical) Make: _________________________

Engraved Metal Face Plate (by electrician) Size: _________________________

Light/Flat Panel Handle Selection Type: Weight: _________________________

Sterilizable Handle *Must be included on quote*

Disposable adapter, Devon Green Slip on

Disposable adapter, Devon Green Screw on

Disposable adapter, DeRoyal Blue Slip on

CONFIRMED BY: ________________________________________________ Date: __________

Submittal Package Revision 08: 11/08/2021 : Page 3 of 86Submittal Package Revision 02: 01/19/2022 : Page 3 of 86

GWB 9' - 6"

ACT-1 8' - 0"

ACT-1 8' - 0"

Cardiac OR

(OR-1)

2C129

Cardiac Pump Room

2C128

EMS Closet 2D280

Alcove 2C100N

ACT-1 8' - 0"

GWB 7' - 0"

4' -1

4"

4' -1

4"

'-9

4"4' 3'-6" 6"8'-8"

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

A: 2300MHL2 - ANESTHESIA BOOM

B: I760 - ILED 7, 32" FLAT PANEL BRACKET

C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"

FLAT PANEL BRACKET

D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"

FLAT PANEL BRACKET

CARDIAC OR 1 (2C129)

A

BC

D

Submittal Package Revision 02: 01/19/2022 : Page 4 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R1000

AutoCAD SHX Text R910

AutoCAD SHX Text R850

AutoCAD SHX Text R930

Cardiac OR

(OR-2)

2C126

Bed Holding 2C100P

Cardiac Equipment

2C127

Alcove 2C100K

Clean Corridor 2C100H

GWB 9' - 6"

2C100M

GWB

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 2"

4' -1

4"

4' -1

4"

'-9

1 2"4' 3'-6" 6"8'-53 4"

3' -6

3'

-6

6" 8'

-2

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

A: 2300MHL2 - ANESTHESIA BOOM

B: I760 - ILED 7, 32" FLAT PANEL BRACKET

C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"

FLAT PANEL BRACKET

D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"

FLAT PANEL BRACKET

CARDIAC OR 2 (2C126)

A

BC

D

Submittal Package Revision 02: 01/19/2022 : Page 5 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R1000

AutoCAD SHX Text R910

AutoCAD SHX Text R850

Robotic OR

(OR-5)

2C114

Electrical Room

2C200B

Bed Holding 2C100C

Scrub 2C100D

GWB 9' - 6"

GWB 9' - 6"

GWB 7' - 0"

5'-01 4" 4'-111

4"10'-61 2"

3' -6

3'

-6

6" 8'

-2

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

E: 2300MHL2 - ANESTHESIA BOOM

F: I760 - ILED 7, 32" FLAT PANEL BRACKET

G: I760 - ILED 7, 32" FLAT PANEL BRACKET

H: 2600HL4 - EQUIPMENT BOOM ROBOTIC OR 5 (2C114)

E F

G

H

Submittal Package Revision 02: 01/19/2022 : Page 6 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

R850

Bed Holding 2C200C

General OR

(OR-6)

2C210

Robotic OR

(OR-5)

Scrub

2C200BGWB 9' - 6"

GWB 7' - 0"

5'-01 4" 4'-111

4"11'-13 4"

3' -6

3'

-6

6" 9'

-2

4"

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

E: 2300MHL2 - ANESTHESIA BOOM

F: I760 - ILED 7, 32" FLAT PANEL BRACKET

G: I760 - ILED 7, 32" FLAT PANEL BRACKET

H: 2600HL4 - EQUIPMENT BOOM GENERAL OR 6 (2C210)

E F

G

H

Submittal Package Revision 02: 01/19/2022 : Page 7 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

General OR

(OR-07)

2C212

Bed Holding 2C200E

Scrub 2C200D

GWB 9' - 6"

GWB 7' - 0"

TYPE B

5'-01 4" 4'-111

4"11'-3"

3' -6

3'

-6

'-4

1 8"

4'

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

E: 2300MHL2 - ANESTHESIA BOOM

F: I760 - ILED 7, 32" FLAT PANEL BRACKET

G: I760 - ILED 7, 32" FLAT PANEL BRACKET

H: 2600HL4 - EQUIPMENT BOOM GENERAL OR 7 (2C212)

E F

G

H

Submittal Package Revision 02: 01/19/2022 : Page 8 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

Ortho OR

(OR-10)

2C217

General OR

(OR-09)

2C216

Ortho OR

(OR-08)

2C215

Bed Holding 2C200K

Scrub 2C200J

GWB 9' - 6"

Bed Holding 2C200H

2-A3-2

Alcove 2C220R

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 0"

TYPE A

TYPE B3'-57

8"3'-61 8"9'-6"

5' -0

1 4" 4'

-1

4"

'-8

8"

4' -1

4"

4' -1

4"

'-9

8"

8" 7'-111 2"

4'-01 8"

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

E: 2300MHL2 - ANESTHESIA BOOM

F: I760 - ILED 7, 32" FLAT PANEL BRACKET

G: I760 - ILED 7, 32" FLAT PANEL BRACKET

H: 2600HL4 - EQUIPMENT BOOM GENERAL OR 9 (2C216)

E

F G

H

Submittal Package Revision 02: 01/19/2022 : Page 9 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

GWB 9' - 0"

Ortho OR

(OR-08)

2C215

Scrub 2C200G

Laboratory, Frozen Section 2C214

Electrical Room 2C213

Ortho Equipment

2C215A

GWB 9' - 6"

Bed Holding 2C200H

2-A3-2

GWB 7' - 0"

TYPE A

TYPE B

GWB 7' - 2"

GWB

5' -0

1 4" 4'

-1

4"

'-8

8"

4' -1

4"

4' -1

4"

'-9

8"

4'3'-6" 9'-43 4"

4'

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

A:2300MHL2 - ANESTHESIA BOOM

B: I760 - ILED 7, 32" FLAT PANEL BRACKET

C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"

FLAT PANEL BRACKET

D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"

FLAT PANEL BRACKET

ORTHO OR 8 (2C215)

A

B C

D

Submittal Package Revision 02: 01/19/2022 : Page 10 of 86

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R1000

AutoCAD SHX Text R910

AutoCAD SHX Text R850

ACT-1 9' - 0"

Anesthesia Workroom

2C218

Ortho OR

(OR-10)

2C217

General OR

(OR-09)

2C216

Bed Holding 2C200M

Scrub 2C200L

Bed Holding 2C200K

Alcove 2C200Q

GWB2

2-A3-2

Alcove 2C220R

GWB 9' - 6"

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 0"

GWB 7' - 2" TYPE A

9'-6"

4' -1

8"

4' -1

8"

'-9

1 4"

3'-117 8"3'-61

8" 7'-111 2"

4'-01 8"

LOUIS STOKES CLEVELAND VA

MEDICAL CENTER

PHASE 2 - OR BOOM REPLACEMENT PROJECT

73596983

01-19-2022 02 SAW

943805-1

ROOM:

DWG. NO.:

DATE:

EQUIPMENT:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

4279-U Crosspoint Drive, Ladson SC 29456

Phone: 1-888-474-9359 Fax:

1-843-534-0206 www.hillrom.com

S/O NO.:

NOT TO SCALEThis drawing contains proprietary information and is property of Hillrom, Inc. It must not be reproduced in whole or in part without written permission from Hillrom, Inc., and is subject to return upon request.

A:2300MHL2 - ANESTHESIA BOOM

B: I760 - ILED 7, 32" FLAT PANEL BRACKET

C: 2300MML0 & I760 - PERFUSION BOOM, ILED 7, 32"

FLAT PANEL BRACKET

D: 2600HL4 & I760 - EQUIPMENT BOOM, ILED 7, 32"

FLAT PANEL BRACKET

ORTHO OR 10 (2C217)

A

B C

D

Submittal Package Revision 02: 01/19/2022 : Page 11 of 86

AutoCAD SHX Text R930

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R930

AutoCAD SHX Text R910

AutoCAD SHX Text R1000

AutoCAD SHX Text R850

AutoCAD SHX Text R1000

AutoCAD SHX Text R910

AutoCAD SHX Text R850

3D-view CSU

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

A: 2300MHL2_MB-52136.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Print: 16.12.21

Submittal Package Revision 02: 01/19/2022 : Page 12 of 86

View CSU

Brake Type Legend:

1) electro pneumatic brake

2) mechanical friction brake, adjustable

3) pneumatic brake; 4) electro magnetic brake

CSU-flange Tandem System flange tandem additional information

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

A: 2300MHL2_MB-52136.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 13 of 86

VAC

C1

E1

C3

C2

C5 E3

W1

A1

VAC

E2

C4

W2

O2

MA

N2O

WAGD

E4

A2

ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS

C HIGH VOLTAGE CIRCUITS

GAS MEDICAL GAS SERVICES

LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)

WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)

SIDE B SIDE CSIDE A

Media Package Length: Standard (1 meter)

WIRE-NUTW

SIDE D

REF QTY P/N DESCRIPTION

A1 2 1636318 530mm SHELF

A2 1 1636396 ADDITIONAL CONTROL HANDLES

REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)

E1 2 L1 SINGLE GANG, TBD

E2 2 L2 SINGLE GANG, TBD

E3 2 L3 SINGLE GANG, TBD

E4 2 L4 SINGLE GANG, TBD

REF QTY P/N DESCRIPTION

REF CIRCUIT CONDUIT/

WIRE-NUT

# OF

RECEPTACLES

RECEPTACLE DESCRIPTION

(TYPE, COLOR, AMP) TYPE

C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C6 CIRCUIT 6 - - BOOM ACCESSORIES CIRCUIT -

REF QTY DESCRIPTION

VAC 2 VACUUM MEDICAL GAS OUTLET

O2 2 OXYGEN MEDICAL GAS OUTLET

MA 1 MEDICAL AIR GAS OUTLET

N2O 1 NITROUS OXIDE MEDICAL GAS OUTLET

WAGD 1 WASTE ANESTHESIA GAS DISPOSAL OUTLET

BRAKE GAS

EB 1 ELECTRIC BRAKES

943805

211216-43438-TUSM291WR

OR 1-4

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

A: 2300MHL2_MB-52136.00

16.12.21

Customer ref. no:

Quotation no:

Room:

Project Customer:

TM order no:

Amount:

Released for production:

Date:

Signature:

Changes after release may cause delay In shipment or price corrections

-Field of application: surgery -Internat.Standard: US

Order drawing no:

Print:

Submittal Package Revision 02: 01/19/2022 : Page 14 of 86

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

B: I760_MB-52137.00

-Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 15 of 86

Room situation with arms in upper position

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

B: I760_MB-52137.00

-Internat.Standard: US

Scale:

1:33

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 16 of 86

Room situation with arms in lower position

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

B: I760_MB-52137.00

-Internat.Standard: US

Scale:

1:33

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 17 of 86

Electrical installation

Control board plate (Power Pack Supply Voltage: 100 - 240 V AC / 22 - 32 V DC) Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines

Pos. 2 iLED7 / Power supply line: 100-240V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V

Load range / add. components VidiaPort

Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)

* Regard adapter + monitor weight!

Wall control panel / Operating unit

Operating unit iLED7: Wall Control set surface + TR Univ. Front interf.

Handles

Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

B: I760_MB-52137.00

-Internat.Standard: US

Scale:

Picture above: 1:50 Picture below: 1:10

Order drawing no:

Print: 16.12.21

Submittal Package Revision 02: 01/19/2022 : Page 18 of 86 wrights5 Text Box

DISPOSABLE ADAPTER - DEVON GREEN SLIP ON

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

C: 2300MML0 & I760_MB-52138.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 19 of 86

1) electro pneumatic brake

2) mechanical friction brake, adjustable

3) pneumatic brake; 4) electro magnetic brake

CSU-flange Tandem OPL System flange tandem OPL additional information

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

C: 2300MML0 & I760_MB-52138.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 20 of 86

VAC

E2

C3

C4

W1

VAC

C2

E1

MA

O2

O2

MA

CO2

VAC

C1

ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS

C HIGH VOLTAGE CIRCUITS

GAS MEDICAL GAS SERVICES

LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)

WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)

SIDE B SIDE CSIDE A

Media Package Length: Standard (1 meter)

WIRE-NUTW

SIDE D

REF QTY P/N DESCRIPTION REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)

E1 2 L1 SINGLE GANG, TBD

E2 2 L2 SINGLE GANG, TBD

REF QTY P/N DESCRIPTION

REF CIRCUIT CONDUIT/

WIRE-NUT

# OF

RECEPTACLES

RECEPTACLE DESCRIPTION

(TYPE, COLOR, AMP) TYPE

C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C3 CIRCUIT 3 H2 / W1 2 DUPLEX, 20A, RED 8300-R

C4 CIRCUIT 4 H2 / W1 2 DUPLEX, 20A, RED 8300-R

C5 CIRCUIT 5 - - BOOM ACCESSORIES CIRCUIT -

REF QTY DESCRIPTION

VAC 4 VACUUM MEDICAL GAS OUTLET

O2 2 OXYGEN MEDICAL GAS OUTLET

MA 2 MEDICAL AIR GAS OUTLET

CO2 4 CARBON DIOXIDE MEDICAL GAS OUTLET

BRAKE GAS

EB 1 ELECTRIC BRAKES

943805

211216-43438-TUSM291WR

OR 1-4

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

C: 2300MML0 & I760_MB-52138.00

16.12.21

Customer ref. no:

Quotation no:

Room:

Project Customer:

TM order no:

Amount:

Released for production:

Date:

Signature:

Changes after release may cause delay In shipment or price corrections

-Field of application: surgery -Internat.Standard: US

Order drawing no:

Print:

Submittal Package Revision 02: 01/19/2022 : Page 21 of 86

Lights / flat panel / VPA

Control board plate (Power Pack Supply Voltage: 100 - 240 V AC / 22 - 32 V DC) Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines

Pos. 2 iLED7 / Power supply line: 230V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V

Load range / add. components VidiaPort

Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)

* Regard adapter + monitor weight!

Handles

Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Wall control panel / Operating unit

Operating unit iLED7: from SAP.Pos.Nr.: MOUNT B

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

C: 2300MML0 & I760_MB-52138.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Print: 16.12.21

Submittal Package Revision 02: 01/19/2022 : Page 22 of 86 wrights5 Text Box

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

D: 2600HL4 & I760_MB-52139.00

-Field of application: Other -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 23 of 86

1) electro pneumatic brake

2) mechanical friction brake, adjustable

3) pneumatic brake; 4) electro magnetic brake

CSU-flange Tandem OPL System flange tandem OPL additional information

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

D: 2600HL4 & I760_MB-52139.00

-Field of application: Other -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 24 of 86

Released for production:

Date:

Signature:

In shipment or price corrections

-Field of application: surgery -Internat.Standard: US

OR 1-4

211216-43438-TUSM291WR

943805

D: 2600HL4 & I760_MB-52139.00

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Order drawing no:

16.12.21

Customer ref. no:

Quotation no:

Room:

Project Customer:

Amount:

TM order no:

Changes after release may cause delay Print:

REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)

E1 2 L1 SINGLE GANG, TBD

E2 2 L2 SINGLE GANG, TBD

E3 2 L3 SINGLE GANG, TBD

REF QTY P/N DESCRIPTION

REF CIRCUIT CONDUIT/

WIRE-NUT

# OF

RECEPTACLES

RECEPTACLE DESCRIPTION

(TYPE, COLOR, AMP) TYPE

C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C6 CIRCUIT 6 H3 / W2 2 DUPLEX, 20A, RED 8300-R

C7 CIRCUIT 7 H3 / W2 2 DUPLEX, 20A, RED 8300-R

C8 CIRCUIT 8 - - BOOM ACCESSORIES CIRCUIT -

REF QTY DESCRIPTION

VAC 2 VACUUM MEDICAL GAS OUTLET

CO2 1 CARBON DIOXIDE MEDICAL GAS OUTLET

IAP 1 INSTRUMENT AIR CONTROL PANEL (1834390)

BRAKE GAS

EB 1 ELECTRIC BRAKES

ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS

C HIGH VOLTAGE CIRCUITS

GAS MEDICAL GAS SERVICES

LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)

WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)

SIDE B SIDE CSIDE A

Media Package Length: Standard (1 meter)

WIRE-NUTW

SIDE D

REF QTY P/N DESCRIPTION

A1 3 1636318 530mm SHELF

A2 1 1636396 ADDITIONAL CONTROL HANDLES

C5

E1

E2

C1

C2

C3

C4

CO2

VAC

A1

E3

W2

IAP

C6

W1

VAC

A2 C7

Submittal Package Revision 02: 01/19/2022 : Page 25 of 86

Lights / flat panel / VPA

Control board plate (Power Pack Supply Voltage: 100 - 240 V AC / 22 - 32 V DC) Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines

Pos. 2 iLED7 / Power supply line: 230V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V

Load range / add. components VidiaPort

Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)

* Regard adapter + monitor weight!

Handles

Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Wall control panel / Operating unit

Operating unit iLED7: from SAP.Pos.Nr.: MOUNT B

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 1-4

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

D: 2600HL4 & I760_MB-52139.00

-Field of application: Other -Internat.Standard: US

Order drawing no:

Print: 16.12.21

Submittal Package Revision 02: 01/19/2022 : Page 26 of 86 wrights5 Text Box

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

E: 2300MHL2_MB-52140.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 27 of 86

1) electro pneumatic brake

2) mechanical friction brake, adjustable

3) pneumatic brake; 4) electro magnetic brake

CSU-flange Tandem System flange tandem additional information

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

E: 2300MHL2_MB-52140.00

-Field of application: Surgery -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 28 of 86

In shipment or price corrections

-Field of application: surgery -Internat.Standard: US

PHASE 2 - OR BOOM REPLACEMENT

OR 5-8

943805

211216-43438-TUSM291WR

. + . / 73596983

E: 2300MHL2_MB-52140.00

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

Order drawing no:

16.12.21

Customer ref. no:

Quotation no:

Room:

Project Customer:

TM order no:

Amount:

Released for production:

Signature:

Date:

Changes after release may cause delay Print:

REF QTY P/N DESCRIPTION

A1 2 1636318 530mm SHELF

A2 1 1636396 ADDITIONAL CONTROL HANDLES

REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)

E1 2 L1 SINGLE GANG, TBD

E2 2 L2 SINGLE GANG, TBD

E3 2 L3 SINGLE GANG, TBD

E4 2 L4 SINGLE GANG, TBD

REF QTY P/N DESCRIPTION

REF CIRCUIT CONDUIT/

WIRE-NUT

# OF

RECEPTACLES

RECEPTACLE DESCRIPTION

(TYPE, COLOR, AMP) TYPE

C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C6 CIRCUIT 6 - - BOOM ACCESSORIES CIRCUIT -

REF QTY DESCRIPTION

VAC 2 VACUUM MEDICAL GAS OUTLET

O2 2 OXYGEN MEDICAL GAS OUTLET

MA 1 MEDICAL AIR GAS OUTLET

N2O 1 NITROUS OXIDE MEDICAL GAS OUTLET

WAGD 1 WASTE ANESTHESIA GAS DISPOSAL OUTLET

BRAKE GAS

EB 1 ELECTRIC BRAKES

ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS

C HIGH VOLTAGE CIRCUITS

GAS MEDICAL GAS SERVICES

LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)

WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)

SIDE B SIDE CSIDE A

Media Package Length: Standard (1 meter)

WIRE-NUTW

SIDE D

MA C4

E2

N2O

WAGD

E4

W2

VAC

O2

A2

C5

W1 VAC

E3

C1

E1

C3

C2

A1

Submittal Package Revision 02: 01/19/2022 : Page 29 of 86

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

F: I760_MB-52141.00

-Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 30 of 86

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

F: I760_MB-52141.00

-Internat.Standard: US

Scale:

1:33

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 31 of 86

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

F: I760_MB-52141.00

-Internat.Standard: US

Scale:

1:33

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 32 of 86

Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines

Pos. 2 iLED7 / Power supply line: 100-240V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V

Load range / add. components VidiaPort

Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)

* Regard adapter + monitor weight!

Wall control panel / Operating unit

Operating unit iLED7: Wall Control set surface + TR Univ. Front interf.

Handles

Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

F: I760_MB-52141.00

-Internat.Standard: US

Scale:

Picture above: 1:50 Picture below: 1:10

Order drawing no:

Print: 16.12.21

Submittal Package Revision 02: 01/19/2022 : Page 33 of 86 wrights5 Text Box

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

G: I760_MB-52142.00

-Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 34 of 86

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

G: I760_MB-52142.00

-Internat.Standard: US

Scale:

1:33

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 35 of 86

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

G: I760_MB-52142.00

-Internat.Standard: US

Scale:

1:33

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 36 of 86

Required space (LxWxH): 300x215x110mm with 1 power supply line, 2 times 300x215x110mm with 2 power supply lines

Pos. 2 iLED7 / Power supply line: 100-240V (1 power supply line) Max. power consumption: 160 VA at 24V / 100-240V

Load range / add. components VidiaPort

Pos. 1 VidiaPort 3000 - Load Range Springarm *: 17-26kg VidiaPort Holder Monitor Single (own weight: 5,5kg)

* Regard adapter + monitor weight!

Wall control panel / Operating unit

Operating unit iLED7: from SAP.Pos.Nr.: MOUNT F

Handles

Pos. 1 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Pos. 2 Amount of sterilizable central handle sets #0337642 (3 handles per set):0

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

G: I760_MB-52142.00

-Internat.Standard: US

Scale:

Picture above: 1:50 Picture below: 1:10

Order drawing no:

Print: 16.12.21

Submittal Package Revision 02: 01/19/2022 : Page 37 of 86 wrights5 Text Box

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

H: 2600HL4_MB-52143.00

-Field of application: Other -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 38 of 86

1) electro pneumatic brake

2) mechanical friction brake, adjustable

3) pneumatic brake; 4) electro magnetic brake

CSU-flange Tandem OPL System flange tandem OPL additional information

Customer ref. no:

Quotation no:

Room:

943805

211216-43438-TUSM291WR

OR 5-8

Project:

Customer:

TM order no:

Amount:

PHASE 2 - OR BOOM REPLACEMENT

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

. + . / 73596983

Released for production:

Date:

Signature:

Changes after release may cause delay in shipment or price corrections

H: 2600HL4_MB-52143.00

-Field of application: Other -Internat.Standard: US

Order drawing no:

Submittal Package Revision 02: 01/19/2022 : Page 39 of 86

Signature:

In shipment or price corrections

-Field of application: surgery -Internat.Standard: US

OR 5-8

H: 2600HL4_MB-52143.00

. + . / 73596983

Order drawing no:

211216-43438-TUSM291WR

PHASE 2 - OR BOOM REPLACEMENT943805

LOUIS STOKES CLEVELAND VA MEDICAL, CLEVELAND

16.12.21

Customer ref. no:

Quotation no:

Room:

Project Customer:

TM order no:

Amount:

Date:

Released for production:

Changes after release may cause delay Print:

REF QTY CONDUIT DESCRIPTION (TYPE, CONDUIT DIAMETER)

E1 2 L1 SINGLE GANG, TBD

E2 2 L2 SINGLE GANG, TBD

E3 2 L3 SINGLE GANG, TBD

REF QTY P/N DESCRIPTION

REF CIRCUIT CONDUIT/

WIRE-NUT

# OF

RECEPTACLES

RECEPTACLE DESCRIPTION

(TYPE, COLOR, AMP) TYPE

C1 CIRCUIT 1 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C2 CIRCUIT 2 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C3 CIRCUIT 3 H1 / W1 2 DUPLEX, 20A, RED 8300-R

C4 CIRCUIT 4 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C5 CIRCUIT 5 H2 / W2 2 DUPLEX, 20A, RED 8300-R

C6 CIRCUIT 6 H3 / W2 2 DUPLEX, 20A, RED 8300-R

C7 CIRCUIT 7 H3 / W2 2 DUPLEX, 20A, RED 8300-R

C8 CIRCUIT 8 - - BOOM ACCESSORIES CIRCUIT -

REF QTY DESCRIPTION

VAC 3 VACUUM MEDICAL GAS OUTLET

CO2 1 CARBON DIOXIDE MEDICAL GAS OUTLET

IAP 1 INSTRUMENT AIR CONTROL PANEL (1834390)

BRAKE GAS

EB 1 ELECTRIC BRAKES

ACCESSORY ITEMSA A ACCESSORY ITEMS E LOW VOLTAGE PREPARATIONS

C HIGH VOLTAGE CIRCUITS

GAS MEDICAL GAS SERVICES

LATCH VALVE TYPE: TRI-TECH: DISS (CO2, N2, AND IA ARE DISS ONLY)

WIRING FOR ALL CIRCUITS: ISOLATED, XHHW (BROWN, ORANGE, GREEN)

SIDE B SIDE CSIDE A

Media Package Length: Standard (1 meter)

WIRE-NUTW

SIDE D

REF QTY P/N DESCRIPTION

A1 4 1636319 630mm SHELF

A2 1 1636396 ADDITIONAL CONTROL HANDLES

C6

A2

W1

IAP

W2

VAC

C7

C5

E1

E2

C1

C2

C3

C4

CO2

VAC

A1

E3

Submittal Package Revision 02: 01/19/2022 : Page 40 of 86

24"

20.35"

20.35"

1.02" TYP

24"

23" +/- 1/2"

FINISHED CEILING OPENING

8"

1/

4"

TO

P O

F

TR

U M

PF

M

O U

N

TI

N G

P

LA

T

TO

F

IN

IS

H

ED

C

EI

LN

IG

3" +/- 1/4"

MOUNTING PLATE WELDED, BOLTED, OR

ANCHORED TO STRUCTURE BY OTHER.

ATTACHMENT METHODS MUST BE

PROVIDED AND INSTALLED BY OTHER AND

APPROVED BY ENGINEER OF RECORD

FINISHED

CEILING

DESIGN LOAD:

F max: 9,200 N (2067 lbf) Mb max: 12,500 Nm (9,220 ft-lbs)

NOTES:

1) Components for TRUMPF EMS mounting systems are TRUMPF provided and installed BY OTHERS. If necessary, all hardware required to attach mounting plate to structure is PROVIDED BY OTHERS and INSTALLED BY OTHERS.

TRUMPF recommends installation during the early phase of construction.

2) Engineer of record must approve all fabricated structures and attachment methods.

3) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.

4) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.

5) Required opening in finished ceiling is 23" x 23" 1/2" centered. Ceiling cover (not shown) is installed by TRUMPF during final installation, and is 25.125" x 25.125". A gasket is used to seal the space between the ceiling cover and finished ceiling.

6) M16 all thread must end 3" 1/4" below the room side of the finished ceiling. All thread will be provided with protective plastic covers by TRUMPF. Following installation of the mounting systems by OTHERS, the protective plastic covers shall be reinstalled by OTHERS to protect the all thread during construction.

7) Gas risers are mounted as shown with DISS connections facing the interior of the mount . Services are then brazed to 1/2" OD copper outside the support structure, above the finished ceiling. Gas risers consist of gas specific DISS nipples brazed to 1/2" OD copper tubing. Gas risers must be installed horizontally.

8) All gas risers have a color coded label according to NFPA requirements.

9) Junction boxes for electrical connections are provided by electrical contractor, and can be mounted to the TRUMPF Interface Plate. Use Raco 265 (4 11/16" x 4 11/16" x 2 1/8") or similar.

VIEW 'A'

MOUNTING PLATE DIMESIONS

VIEW 'B'

MOUNTING SYSTEM ELEVATIONS

PAGE CONTENTS:

DWG. NO.:

DATE:

MOUNT:

CAT. NO:

QUANTITY ON PLAN:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

TRUMPF Medical Systems, Inc.

1046 LeGrand Blvd, Charleston, SC 29492

Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com

NOT TO SCALE

THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!

This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.

LOUIS STOKES CLEVELAND VA MEDICAL

BOOM MOUNTING SYSTEM

8" INTERSTITIAL SPACE

MOUNTING SYSTEM

110106 943805-MA

01 SAW4 12-16-2021A

UNITS: [In.]/mm

S.O. NO.: 73596983

Submittal Package Revision 02: 01/19/2022 : Page 41 of 86

VIEW 'C'

MOUNTING SYSTEM COMPONENTS

(ALL COMPONENTS INSTALLED BY CONTRACTOR)

TIGHTEN M16 NUTS

TO 52 ft-lbs

TIGHTEN M22 x M16

ADAPTER NUTS

TO 89 ft-lbs

LIST OF REQUIRED MATERIALS FOR BOOM MOUNTING SYSTEM

DESIGN LOAD:

F max: 9,200 N (2067 lbf) Mb max: 12,500 Nm (9,220 ft-lbs) NOTES:

A) Components for TRUMPF EMS mounting systems are TRUMPF provided and installed BY OTHERS. If necessary, all hardware required to attach mounting plate to structure is PROVIDED BY OTHERS and INSTALLED BY OTHERS.

TRUMPF recommends installation during the early phase of construction.

B) Engineer of record must approve all fabricated structures and attachment methods.

C) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.

D) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.

E) Required opening in finished ceiling is 23" x 23" 1/2" centered. Ceiling cover (not shown) is installed by TRUMPF during final installation, and is 25.125" x 25.125". A gasket is used to seal the space between the ceiling cover and finished ceiling.

F) M16 all thread must end 3" 1/4" below the room side of the finished ceiling. All thread will be provided with protective plastic covers by TRUMPF. Following installation of the mounting system by OTHERS, the protective covers shall be reinstalled by OTHERS to protect the all thread during construction.

G) Gas risers are mounted as shown with DISS connections facing the interior of the mount. Services are then brazed to 1/2" OD copper outside the support structure, above the finished ceiling. Gas risers consist of gas specific DISS nipples brazed to 1/2" OD copper tubing. Gas riser must be installed horizontally.

H) All gas risers have a color coded label according to NFPA requirements.

I) Junction boxes for electrical connections are provided by electrical contractor, and can be mounted to the TRUMPF Universal Interface Plate. Use Raco 265 (4 11/16" x 4 11/16" x 2 1/8") or similar.

PAGE CONTENTS:

DWG. NO.:

DATE:

MOUNT:

CAT. NO:

QUANTITY ON PLAN:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

TRUMPF Medical Systems, Inc.

1046 LeGrand Blvd, Charleston, SC 29492

Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com

NOT TO SCALE

THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!

This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.

LOUIS STOKES CLEVELAND VA MEDICAL

BOOM MOUNTING SYSTEM

8" INTERSTITIAL SPACE

MOUNTING SYSTEM

110106 943805-MA

01 SAW4 12-16-2021A

UNITS: [In.]/mm

Submittal Package Revision 02: 01/19/2022 : Page 42 of 86

5"

14 1 2 "

ROUGH OPENING

10"

TRUMPF

MOUNTING PLATE

(110036)

(INSTALLED BY

OTHER)

SUB-STRUCTURE SPECIFIED BY

ENGINEER OF RECORD

(PROVIDED AND INSTALLED BY

OTHER)

THREADED ROD AND HARDWARE

APPROVED BY ENGINEER OF RECORD

(PROVIDED AND INSTALLED BY OTHER)

VIEW 'B'

LIGHT MOUNTING PLATE (110036)

14"

14"

10 1 4 "

10 1 4 "

(TYP 4X)

.81" THRU

VIEW 'A'

SIDE VIEW OF MOUNTING SYSTEM

(ALL COMPONENTS INSTALLED BY CONTRACTOR)

FINISHED

CEILING

DESIGN LOAD:

F max: 1,780 N (400 lbf) Mb max: 2,033 Nm (1500 ft-lbs)

VIEW 'C'

ISOMETRIC VIEW OF MOUNTING SYSTEM

(COMPLETED PRE-INSTALLATION)

CONDUIT TO

TRANSFORMER

ENCLOSURE

PAGE CONTENTS:

DWG. NO.:

DATE:

MOUNT:

CAT. NO:

QUANTITY ON PLAN:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

TRUMPF Medical Systems, Inc.

1046 LeGrand Blvd, Charleston, SC 29492

Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com

NOT TO SCALE

THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!

This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.

LOUIS STOKES CLEVELAND VA MEDICAL

LIGHT MOUNTING SYSTEM

FLAT CEILING COVER

MOUNTING SYSTEM

MS-1 943805-MB

01 SAW4 12-16-2021B

UNITS: [In.]/mm

Submittal Package Revision 02: 01/19/2022 : Page 43 of 86

STRUCTURE BY OTHER

MOUNTING

HARDWARE BY

OTHER

CEILING COVER

(INSTALLED BY

TRUMPF)

LIGHT TUBE

(INSTALLED BY

TRUMPF)

11.875"

.7

5"

8"

1.125" HOLES

9.25"

13 1 16 "

LIST OF PRE-INSTALL MATERIALS FOR LIGHT MOUNTING SYSTEMS

NOTES:

A) TRUMPF recommends installation during the early phase of construction. Light mounting plate is provided by TRUMPF but installed by others.

B) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.

C) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.

D) Engineer of record must approve all fabricated structures and attachment methods.

E) TRUMPF recommends 3/4" ASTM-A307 hardware for attachment of the TRUMPF light mounting plate.

F) Required opening in finished ceiling is 14.5" x 14.5" 1/4" centered. Ceiling cover is installed by TRUMPF during final installation and is 15.75" x 15.75".

G) Transformer enclosures provided by TRUMPF and installed by other in an accessible location. Access panels are required if located above hard ceiling.

H) Refer to Preinstall guide (110720) for required conduit runs for transformer enclosures, wall controls, and camera control.

DESIGN LOAD:

F max: 1,780 N (400 lbf) Mb max: 2,033 Nm (1500 ft-lbs)

FOR 3/4" TRADE

SIZE CONDUIT

TERMINATION

VIEW 'D'

TRANSFORMER ENCLOSURE

(INSTALLED BY OTHER)

VIEW 'E'

ISOMETRIC VIEW OF MOUNTING SYSTEM

PAGE CONTENTS:

DWG. NO.:

DATE:

MOUNT:

CAT. NO:

QUANTITY ON PLAN:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

TRUMPF Medical Systems, Inc.

1046 LeGrand Blvd, Charleston, SC 29492

Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com

NOT TO SCALE

THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!

This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.

LOUIS STOKES CLEVELAND VA MEDICAL

LIGHT MOUNTING SYSTEM

FLAT CEILING COVER

MOUNTING SYSTEM

MS-1 943805-MB

01 SAW4 12-16-2021B

UNITS: [In.]/mm

Submittal Package Revision 02: 01/19/2022 : Page 44 of 86

20.35"

1.02" TYP

24"

23" +/- 1/2"

FINISHED CEILING OPENING

8"

O

F

TR

U M

PF

M

O U

N

TI

N G

P

LA

T

TO

F

IN

IS

H

ED

C

EI

LN

IG

3" +/- 1/4"

MOUNTING PLATE WELDED, BOLTED, OR

ANCHORED TO STRUCTURE BY OTHER.

ATTACHMENT METHODS MUST BE

PROVIDED AND INSTALLED BY OTHER AND

APPROVED BY ENGINEER OF RECORD

FINISHED

CEILING

DESIGN LOAD:

F max: 9,200 N (2067 lbf) Mb max: 12,500 Nm (9,220 ft-lbs)

NOTES:

1) Components for TRUMPF EMS mounting systems are TRUMPF provided and installed BY OTHERS. If necessary, all hardware required to attach mounting plate to structure is PROVIDED BY OTHERS and INSTALLED BY OTHERS.

TRUMPF recommends installation during the early phase of construction.

2) Engineer of record must approve all fabricated structures and attachment methods.

3) Maximum horizontal deflection of anchor plate is 0.1" with a horizontal load of 100 lbs.

4) Maximum angular deflection of 1/3 when structure is subjected to model specific load and moment.

5) Required opening in finished ceiling is 23" x 23" 1/2" centered. Ceiling cover (not shown) is installed by TRUMPF during final installation, and is 25.125" x 25.125". A gasket is used to seal the space between the ceiling cover and finished ceiling.

6) M16 all thread must end 3" 1/4" below the room side of the finished ceiling. All thread will be provided with protective plastic covers by TRUMPF. Following installation of the mounting systems by OTHERS, the protective plastic covers shall be reinstalled by OTHERS to protect the all thread during construction.

7) Gas risers are mounted as shown with DISS connections facing the interior of the mount . Services are then brazed to 1/2" OD copper outside the support structure, above the finished ceiling. Gas risers consist of gas specific DISS nipples brazed to 1/2" OD copper tubing. Gas risers must be installed horizontally.

8) All gas risers have a color coded label according to NFPA requirements.

9) Junction boxes for electrical connections are provided by electrical contractor, and can be mounted to the TRUMPF Interface Plate. Use Raco 265 (4 11/16" x 4 11/16" x 2 1/8") or similar.

VIEW 'A'

MOUNTING PLATE DIMESIONS

VIEW 'B'

MOUNTING SYSTEM ELEVATIONS

PAGE CONTENTS:

DWG. NO.:

DATE:

MOUNT:

CAT. NO:

QUANTITY ON PLAN:

RELEASED FOR PRODUCTION

(SIGNATURE CONFIRMS ACKNOWLEDGEMT OF ALL DRAWING NOTES)

SIGNATURE:

DATE: INSTALLATION DATE: REV.:

TRUMPF Medical Systems, Inc.

1046 LeGrand Blvd, Charleston, SC 29492

Phone: 1-888-474-9359 Fax: 1-843-534-0206 www.us.trumpf-med.com

NOT TO SCALE

THIS DRAWING IS NOT TO BE USED FOR CONSTRUCTION!

This drawing contains proprietary information and is property of TRUMPF Medical Systems, Inc. It must not be reproduced in whole or in part without written permission from TRUMPF Medical Systems, Inc., and is subject to return upon request.

LOUIS STOKES CLEVELAND VA MEDICAL

BOOM…